Electrochemistry
Online ISSN : 2186-2451
Print ISSN : 1344-3542
ISSN-L : 1344-3542
Notes
Preparation of Immunosensors Using a Microfluidic Device with an Interdigitated Array Electrode Modified with Antibodies
Yu HIRANOTomoyuki YASUKAWAYoshihiro SAWAYASHIKIHitoshi SHIKUFumio MIZUTANITomokazu MATSUE
Author information
JOURNAL FREE ACCESS

2010 Volume 78 Issue 2 Pages 175-177

Details
Abstract

We prepared a microfluidic device with an interdigitated array electrode modified with antibodies to develop immunosensors. A comb-type array (W1) was used for forming immunocomplexes and another (W2) for detecting mediators generated by the enzyme reaction. We used electropolymerization and avidin-biotin complexes for an addressable immobilization of antibodies on W1. Since the microfluidics significantly accelerated the formation of the immunocomplexes, a period as short as 10 min was sufficient to detect the responses. The current observed at bare array (W2) increased with increasing analyte (mouse IgG) concentration in the range of 1.0–100 ng/mL. Therefore, present procedure is suitable for rapid immunosensing in a simple device.

Content from these authors
© 2010 The Electrochemical Society of Japan
Previous article Next article
feedback
Top