Kim and Hwang: Intraoperative, rapid administration of high-dose fentanyl by a dysfunctioning plunger lock of a syringe pump; the siphon effect
Syringe pumps are designed to deliver drugs in a safe manner, however, these devices can be potentially dangerous if operated improperly. A 50-year-old male was scheduled to undergo an elective liver transplantation. After induction of anesthesia, the Baxter AS40A pump was loaded with a 50 ml Becton Dickinson syringe containing 2,500 µg of fentanyl. Continuous infusion at 5 ml/hr was started without knowing that part of the plunger lock, i.e. the upper plunger clamp assembly, was damaged (Fig. 1), and, therefore, the plunger was incompletely clamped without notice. Ten minutes later the patient experienced marked sinus bradycardia and hypotension for which a rescue drug of 10 µg of epinephrine was injected twice. Surprisingly, the syringe loaded on AS40A was found to be empty and the missing 50 ml of fentanyl was concluded to have been infused rapidly into the patient. However, even at this point there was no alarm signal or error message to indicate a dysfunctional plunger clamp.
Of many possible reasons, a siphon effect was strongly suspected to be the cause of this mishap, which refers to the continuous flow of liquids through enclosed conduit by the force of gravity once it had been initiated, because we can acknowledge that a complete spillage of 50 ml syringe volume occurred within five minutes when the plunger lock was unclamped and manually started. Although high-dose fentanyl administration has been seen to be hemodynamically stable [1], such as in this case, bolus administration of another drug such as dopamine or norepinephrine, which is usually delivered using a Baxter infusion pump, may have catastrophic effects. Although a newer version of the Baxter AS40A pump has an alarm sensor for the syringe plunger lock, understanding the specifics of commonly used, older equipment in the operating room is crucial [2], and meticulous attention to a possibly improperly secured plunger head is always warranted in order to avoid potentially adverse outcomes.

References

1. Coda BA. Edited by Barash PG, Cullen BF, Stoelting RKOpiod. Clinical Anesthesia. 2006, 5th ed. : Philadelphia, Lippincott Williams & Wilkins. pp 353-383.

2. Christodoulou C. Incident with a Baxter AS40A syringe pump: vigilance is warranted. Can J Anaesth 2004; 51: 185-186. PMID: 14766699.
crossref pmid
Fig. 1
A Baxter AS40A pump with a normal (A) and damaged (B) plunger clamp assembly which dose not firmly hold the plunger.
kjae-62-298-g001.jpg


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next