Korean J Anesthesiol Search

CLOSE


Korean Journal of Anesthesiology 2009;57(2):217-220.
DOI: https://doi.org/10.4097/kjae.2009.57.2.217   
Radial nerve injury due to automatic blood pressure monitoring: A case report.
Young Soo Kim, Kyung Hwa Kwak, Jung Gil Hong
Department of Anesthesiology and Pain Medicine, School of Medicine, Kyungpook National University, Daegu, Korea. kwakkh@knu.ac.kr
Abstract
We experienced an unusual complication of acute radial nerve palsy presenting as wrist drop after application of automated cycled blood pressure monitoring for 3 hours. A 19-year-old ASA physical status 1 female was scheduled to undergo nail removal, had been operated internal transport over the nail lengthening for fibular hemimelia. Blood pressure cuff was affixed to her right upper arm and worked automatically every 5 minute during surgery. One day after operation she complained of pain over the lower lateral aspect of the right upper arm and examination revealed zero power of the wrist and finger extensor muscles. Electromyelography (EMG) and nerve conduction velocity (NCV) revealed right radial neuropathy. She was discharged 20 days after operation with improvement of the right upper arm pain. After three months of physical therapy, the muscle power of wrist extensors reverted to completely normal and the muscle power of the finger extensors improved to fair.
Key Words: Blood pressure; Complications; Cuff; Nerve injury; Radial nerve


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI