Korean Circ J. 2005 Sep;35(9):672-676. English.
Published online Sep 30, 2005.
Copyright © 2005 The Korean Society of Circulation
Original Article

Primary Sirolimus-Eluting Stent Implantation for Patients with Acute ST-Segment Elevation Myocardial Infarction

Tae Hyun Yang, Myeong Ki Hong, Kyoung Ha Park, Mi Jeong Kim, Bong Ryong Choi, Sung Ju Oh, Bong Ki Lee, Young Hak Kim, Ki Hoon Han, Cheol Whan Lee, Duk Hyun Kang, Jae Kwan Song, Jae Joong Kim, Seong Wook Park and Seung Jung Park
    • Department of Medicine, Inje University College of Medicine, Busan, Korea.
    • Department of Medicine, University of Ulsan College of Medicine, Asan Medical Center, Seoul, Korea.
    • Department of Medicine, Seoul Adventist Hospital, Seoul, Korea.

Abstract

BACKGROUND AND OBJECTIVES: Sirolimus-eluting stents (SES) have been shown to significantly inhibit neointimal hyperplasia, resulting in reduced restenosis compared with bare metal stents (BMS). However, the efficacy and safety of SES implantation for patients with acute ST-segment elevation myocardial infarction (STEMI) remain unclear.

SUBJECTS AND METHODS: Primary stenting was performed using SES in 74 patients (mean age: 58.0±12.7 years, 59 males) and BMS in 88 patients (mean age: 59.3±10.7 years, 63 males) between April 2003 and July 2004. We retrospectively compared the incidence of 6-month angiographic restenosis and the major adverse cardiac events (MACE) defined as cardiac death, non-fatal myocardial infarction and target lesion revascularization (TLR), between the SES group and the BMS group.

RESULTS: The SES group had smaller vessels (3.04±0.47 mm vs. 3.24±0.56 mm, respectively, p=0.02) and a longer stent length (33.7±14.3 mm vs. 25.0±9.6 mm, p=0.00). The procedural success rate (87.8% vs. 92.0%, respectively, p=0.37) and the peak creatine kinase-MB (239±196 ng/mL vs. 274±188 ng/mL, p=0.26) were similar. The 6-month angiographic restenosis rate (0.0% vs. 30.4%, respectively, p=0.00) and late loss (-0.03±0.55 mm vs. 1.28±0.58 mm, p=0.00) were significantly lower in the SES group compared with the BMS group. Stent thrombosis developed in only 1 case of the SES group (1.4% vs. 0.0%, respectively, p=0.45). At 6 months, SES implantation significantly reduced the incidence of MACE (6.9% vs. 19.5%, respectively, p=0.04), because of a reduction in the incidence of TLR (1.4% vs. 11.5%, p=0.01). Likewise, the MACE-free survival rate was significantly higher in the SES group (93.06% vs. 80.46%, respectively, p=0.03).

CONCLUSION: Compared with the BMS, the SES was effective in reducing the incidence of 6-month angiographic restenosis and MACE without any increased risk of stent thrombosis in the patients with STEMI who received primary stenting.

Keywords
Sirolimus; Stents; Myocardial infarction; Thrombosis


Metrics
Share
PERMALINK