Axisymmetric Contact Stress analysis of an Artificial Hip Joint of the Conical Fitting Type

Yong San Yoon, Young Yong Kim, Kye Lim Lee, Seung Key Kim and Young Hoon Kang

    Abstract

    The purpose of this study is to investigate the effect of the ball-cone contact interface variation on the ceramic ball stress which is the dominant factor of the design of the moldular artificial hip joint. A contact stress analysis problem is formulated using the optimization principle and finite element analysis technique on the ball-cone system of three dimensional axi-symmetrical model, Numerical procedure is employed to get the solution for varying angles of cones and angular mismatches between the ball and cone interfaces. There is a fair amount of friction in the interface and included in the analysis model.

    Keywords
    Contact stress; Finite element analysis