Original Article

The effect of heat to remove cement on implant titanium abutment and screw

Hyo-Gyoung Yi, Ki-Sung Gil, Jung-Jin Lee, Seung-Geun Ahn and Jae-Min Seo
    • Department of Prosthodontics, School of Dentistry and Institute of Oral Bio-Science, Chonbuk National University, Jeonju, Republic of Korea.
Received January 09, 2018; Revised May 11, 2018; Accepted May 30, 2018.

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

Abstract

Purpose

The purpose of this study was to investigate the effect of heat applied to disintegrate cement on the removal torque value and fracture strength of titanium abutment and abutment screw.

Materials and methods

Implants, titanium abutments and abutment screws were prepared for each 20 piece. Implant abutments and screws were classified as the control group in which no heat was applied and the experimental group was heated in a vacuum furnace to 450℃ for 8 minutes and cooled in air. The abutments and screws were connected to the implants with 30 Ncm tightening torque at interval 10 minutes and the removal torque value was measured 15 minutes later. And the fracture strength of abutment screw was measured using universal testing machine.

Results

The mean removal torque value was 27.84 ± 1.07 Ncm in the control group and 26.55 ± 1.56 Ncm in the experimental group and showed statistically significant difference (P < .05). The mean fracture strength was 731.47 ± 39.46 N in the control group and 768.58 ± 46.73 N in the experimental group and showed statistically no significant difference (P > .05).

Conclusion

The heat applied for cement disintegration significantly reduced the removal torque value of the abutment screw and did not significantly affect fracture strength of the abutment screw. Therefore, in the case of applying heat to disintegrate cement it is necessary to separate the abutment screw or pay attention to the reuse of the heated screw. However further studies are needed to evaluate the clinical reuse of the heated screw.

Keywords
Implant abutment; Abutment screw; Removal torque value; Fracture strength

Figures

Fig. 1
Materials used in the experiment. (A) Implant, (B) Abutment, (C) Abutment screw.

Fig. 2
Implant embedding and fixation with specimen holder. (A) Autopolymerizing acrylic resin replica, (B, C) Putty die, (D) Implant embedded with acrylic resin, (E) Embedded implant in specimen holder. 3 mm below the platform was not embedded to reflect vertical bone resorption.

Fig. 3
Vacuum furnace.

Fig. 4
Experimental design for removal torque value test. (A) After implant was fixed with specimen holder, abutment screwed on the implant. Abutment screw was tightened with an insertion torque of 30 Ncm and retightened after 10 minutes, (B) Digital torque meter was used to reach reproducible and accurate force. Removal torque value was measured after 15 minutes using the digital torque meter.

Fig. 5
Experimental design for static compressive loading test. (A) Universal testing machine, (B) Implant-screw-abutment assembly was fixed in specimen holder. The specimen was positioned so that the load was applied at an angle of 30 degrees to the long axis of the assembly. The tin foil was placed between abutment and loading piston for even force.

Fig. 6
Removal torque value of abutment screw. Histogram showing mean removal torque value of abutment screws. The removal torque value was significantly different between the two groups (P < .05).

Fig. 7
Fractured abutment screw at the shank area.

Fig. 8
Fracture load of abutment screw. Histogram showing mean fracture load of abutment screws. The fracture load was not significantly different between the two groups (P > .05).

Tables

Table 1
Mechanical properties of the base metal

Table 2
Descriptive statistics of removal torque value of abutment screw

Table 3
Descriptive statistics of fracture strength of abutment screw

References

    1. Misch CE. In: Contemporary implant dentistry. 3rd ed. St. Louis: Mosby Elsevier; 2008.
    1. Chung CH, Son MK. The classification and comparison of implant prosthesis according to types of retention. Part I: screw retained prosthesis vs cement retained prosthesis. Implantology 2010;14:138–151.
    1. Kim JH, Yun BH, Jang JE, Huh JB, Jeong CM. Retrievable SCP (screw-cement prosthesis) implant-supported fixed partial dentures in a fully edentulous patient: a case report. J Korean Acad Prosthodont 2012;50:318–323.
    1. Chung CH, Son MK. The classification and comparison of implant prosthesis according to types of retention. Part II: Screw-cement retained prosthesis. Implantology 2011;15:58–70.
    1. Taylor TD. Prosthodontic problems and limitations associated with osseointegration. J Prosthet Dent 1998;79:74–78.
    1. Rangert B, Jemt T, Jörneus L. Forces and moments on Branemark implants. Int J Oral Maxillofac Implants 1989;4:241–247.
    1. Pauletto N, Lahiffe BJ, Walton JN. Complications associated with excess cement around crowns on osseointegrated implants: a clinical report. Int J Oral Maxillofac Implants 1999;14:865–868.
    1. Arora A, Upadhayaya V, Mittal S, Goyal I. Techniques for retrievability of cement retained implant prosthesis. J Dent Implant 2014;4:161–164.
    1. Aparicio C. A new method for achieving passive fit of an interim restoration supported by Brånemark implants: a technical note. Int J Oral Maxillofac Implants 1995;10:614–618.
    1. Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang NP. A systematic review of the 5-year survival and complication rates of implant-supported single crowns. Clin Oral Implants Res 2008;19:119–130.
    1. Shin SY. Prosthodontic problems and complications associated with osseointegration. J Dent Rehabil Appl Sci 2015;31:349–357.
    1. Hong JY, Chae GJ, Jung UW, Kim CS, Cho KS, Chai JK, Kim CK, Choi SH. Implant-related complications and treatment of the ailing implants. Implantology 2007;11:44–54.
    1. Rosenstiel SF, Land MF, Fujimoto J. In: Contemporary fixed prosthodontics. 5th ed. St. Louis, USA: Elsevier; 2016.
    1. Krishnan V, Tony Thomas C, Sabu I. Management of abutment screw loosening: review of literature and report of a case. J Indian Prosthodont Soc 2014;14:208–214.
    1. Linkevicius T, Vindasiute E, Puisys A, Linkeviciene L, Svediene O. Influence of the temperature on the cement disintegration in cement-retained implant restorations. Stomatologija 2012;14:114–117.
    1. Veiga C, Davim JP, Loureiro AJR. Properties and applications of titanium alloys: A brief review. Rev Adv Mater Sci 2012;32:133–148.
    1. ISO 14801:2016. Dentistry implants - dynamic fatigue test for endosseous dental implants. Geneva (Switzerland): International Organization for Standardization; 2016.
    1. Leelanarathiwat K, Asvanund P, Anunmana C. Removal torque of screw-and cement-retained cantilever fixed prosthesis on angled abutment after cyclic loading. Mahidol Dent J 2016;36:269–277.
    1. Gracis S, Michalakis K, Vigolo P, Vult von Steyern P, Zwahlen M, Sailer I. Internal vs. external connections for abutments/reconstructions: a systematic review. Clin Oral Implants Res 2012;23:202–216.
    1. Theoharidou A, Petridis HP, Tzannas K, Garefis P. Abutment screw loosening in single-implant restorations: a systematic review. Int J Oral Maxillofac Implants 2008;23:681–690.
    1. Im SM, Kim DG, Park CJ, Cha MS, Cho LR. Biomechanical considerations for the screw of implant prosthesis: A literature review. J Korean Acad Prosthodont 2010;48:61–68.
    1. Al Jabbari YS, Fournelle R, Ziebert G, Toth J, Iacopino AM. Mechanical behavior and failure analysis of prosthetic retaining screws after long-term use in vivo. Part 4: Failure analysis of 10 fractured retaining screws retrieved from three patients. J Prosthodont 2008;17:201–210.
    1. Gupta S, Gupta H, Tandan A. Technical complications of implant-causes and management: A comprehensive review. Natl J Maxillofac Surg 2015;6:3–8.
    1. Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529–536.
    1. Winkler S, Ring K, Ring JD, Boberick KG. Implant screw mechanics and the settling effect: overview. J Oral Implantol 2003;29:242–245.
    1. Kyung KY, Cha HS, Lee JH. The effect of a titanium socket with a zirconia abutment on screw loosening after thermocycling in an internally connected implant: a preliminary study. J Dent Rehabil Appl Sci 2017;33:114–118.
    1. Cantwell A, Hobkirk JA. Preload loss in gold prosthesis-retaining screws as a function of time. Int J Oral Maxillofac Implants 2004;19:124–132.
    1. Huh YH, Cho LR, Kim DG, Park CJ. Comparison of implant torque controllers using detorque value. J Dent Rehabil Appl Sci 2010;26:419–432.
    1. Lang LA, Kang B, Wang RF, Lang BR. Finite element analysis to determine implant preload. J Prosthet Dent 2003;90:539–546.
    1. Stüker RA, Teixeira ER, Beck JC, da Costa NP. Preload and torque removal evaluation of three different abutment screws for single standing implant restorations. J Appl Oral Sci 2008;16:55–58.
    1. Hong SY, Markus I, Jeong WC. New cooling approach and tool life improvement in cryogenic machining of titanium alloy Ti-6Al-4V. Int J Mach Tools Manuf 2001;41:2245–2260.
    1. Zherebtsov S, Salishchev G, Galeyev R, Maekawa K. Mechanical properties of Ti-6Al-4V titanium alloy with submicrocrystalline structure produced by severe plastic deformation. Mater Trans 2005;46:2020–2025.
    1. Ming Q, Yongzhen Z, Jun Z, Jianheng Y. Dry friction characteristics of Ti-6Al-4V alloy under high sliding velocity. J Wuhan Univ Technol Mat Sci 2007;22:582.
    1. Revankar GD, Shetty R, Rao SS, Gaitonde VN. Wear resistance enhancement of titanium alloy (Ti-6Al-4V) by ball burnishing process. J Mater Res Technol 2017;6:13–32.
    1. Swarnakar AK, van der Biest O, Baufeld B. Young's modulus and damping in dependence on temperature of Ti-6Al-4V components fabricated by shaped metal deposition. J Mater Sci 2011;46:3802–3811.
    1. da Rocha SS, Adabo GL, Henriques GE, Nóbilo MA. Vickers hardness of cast commercially pure titanium and Ti-6Al-4V alloy submitted to heat treatments. Braz Dent J 2006;17:126–129.
    1. Meyers MA, Benavides HAC, Bruhl SP, Colorado HA, Dalgaard E, Elias CN, Figueiredo RB, Garcia Rincon O, Kawasaki M, Langdeon TG, Mangalaraja RV, Marroquin MCG, da Cunha Rocha A, Schoenung J, Costa e Silva A, Wells M, Yang W, editors. The Minerals, Metals & Materials Series; the 3rd Pan American Materials Congress; Springer; 2017. pp. 381-391.
    1. Liu G, Zhu D, Shang Jian. Enhanced fatigue crack growth resistance at elevated temperature in TiC/Ti-6Al-4V composite: Microcrack-induced crack closure. Metall Mater Trans A 1995;26:159–166.
    1. Lee KA, Kim YK, Yu JH, Park SH, Kim MC. Effect of heat treatment on microstructure and impact toughness of Ti-6Al-4V manufactured by selective laser melting process. Arch Metall Mater 2017;62:1341–1346.
    1. Tan BF, Tan KB, Nicholls JI. Critical bending moment of implant-abutment screw joint interfaces: effect of torque levels and implant diameter. Int J Oral Maxillofac Implants 2004;19:648–658.

Metrics
Share
Figures

1 / 8

Tables

1 / 3

PERMALINK