A Study of N-Isopropyl Acrylamide (NIPAM)-Based Polymer Gel Dosimeter by Using Raman Spectroscopy

Article Preview

Abstract:

Polymer gel is a kind of radiation dosimetry system that has been applied widely in radiotherapy treatment. In this study, the samples of NIPAM-based polymer gel which posses a good potential as 3D radiation dosimetry are synthesized. The samples of polymer gel were irradiated up to 20 Gy by using gamma cell instrument with 60Co sources at a constant dose rate. Upon irradiation, then the chemical changes in polymer gel were characterized by using Raman spectroscopy. The polymerization was referred to an increment in Raman intensity at 815 cm-1, assigned for C-C stretching mode of NIPAM polymer gel, as the dose increased. The consumptions of the co-monomers were referred to a decrement in Raman intensities at 1025 cm-1 and 2353 cm-1 assigned for C=C stretching modes of NIPAM and BIS respectively, as the dose increased. Result shows that the amount of carbon single bonds increases while the amount of carbon covalent bonds decreases following irradiation.

You might also be interested in these eBooks

Info:

Periodical:

Pages:

103-107

Citation:

Online since:

June 2015

Export:

Price:

* - Corresponding Author

[1] Senden, R., et al., Polymer gel dosimeters with reduced toxicity: a preliminary investigation of the NMR and optical dose–response using different monomers. Physics in medicine and biology, 2006. 51(14): p.3301.

DOI: 10.1088/0031-9155/51/14/001

Google Scholar

[2] De Deene, Y., et al., The fundamental radiation properties of normoxic polymer gel dosimeters: a comparison between a methacrylic acid based gel and acrylamide based gels. Physics in medicine and biology, 2006. 51(3): p.653.

DOI: 10.1088/0031-9155/51/3/012

Google Scholar

[3] Mesbahi, A., V. Jafarzadeh, and N. Gharehaghaji, Optical and NMR dose response of N-isopropylacrylamide normoxic polymer gel for radiation therapy dosimetry. Reports of Practical Oncology & Radiotherapy, 2012. 17(3): pp.146-150.

DOI: 10.1016/j.rpor.2012.03.009

Google Scholar

[4] Jirasek, A., et al., Characterization of monomer/crosslinker consumption and polymer formation observed in FT-Raman spectra of irradiated polyacrylamide gels. Physics in medicine and biology, 2001. 46(1): p.151.

DOI: 10.1088/0031-9155/46/1/311

Google Scholar

[5] Fong, P.M., et al., Polymer gels for magnetic resonance imaging of radiation dose distributions at normal room atmosphere. Physics in medicine and biology, 2001. 46(12): p.3105.

DOI: 10.1088/0031-9155/46/12/303

Google Scholar

[6] De Deene, Y., et al., A basic study of some normoxic polymer gel dosimeters. Physics in medicine and biology, 2002a. 47(19): p.3441.

DOI: 10.1088/0031-9155/47/19/301

Google Scholar

[7] Hsieh, B., et al., A study on dose response of NIPAM-based dosimeter used in radiotherapy. Journal of Radioanalytical and Nuclear Chemistry, 2011. 290(1): pp.141-148.

DOI: 10.1007/s10967-011-1161-4

Google Scholar

[8] Baldock, C., et al., Fourier transform Raman spectroscopy of polyacrylamide gels (PAGs) for radiation dosimetry. Physics in medicine and biology, 1998. 43(12): p.3617.

DOI: 10.1088/0031-9155/43/12/017

Google Scholar