Review

Conduction System Pacing Today and Tomorrow

1
Department of Cardiology, Inselspital, Bern University Hospital, University of Bern, 3010 Bern, Switzerland
2
ARTORG Center for Biomedical Engineering Research, University of Bern, 3008 Bern, Switzerland
3
Swiss Institute for Translational and Entrepreneurial Medicine, 3010 Bern, Switzerland
*
Author to whom correspondence should be addressed.
J. Clin. Med. 2022, 11(24), 7258; https://doi.org/10.3390/jcm11247258
Submission received: 22 October 2022 / Revised: 29 November 2022 / Accepted: 5 December 2022 / Published: 7 December 2022
(This article belongs to the Special Issue Advances in Cardiac Pacing and Cardiac Resynchronization Therapy)

Abstract

:
Conduction system pacing (CSP) encompassing His bundle (HBP) and left bundle branch area pacing (LBBAP) is gaining increasing attention in the electrophysiology community. These relatively novel physiological pacing modalities have the potential to outperform conventional pacing approaches with respect to clinical endpoints, although data are currently still limited. While HBP represents the most physiological form of cardiac stimulation, success rates, bundle branch correction, and electrical lead performance over time remain a concern. LBBAP systems may overcome these limitations. In this review article, we provide a comprehensive overview of the current evidence, implantation technique, device programming, and follow-up considerations concerning CSP systems. Moreover, we discuss ongoing technical developments and future perspectives of CSP.

1. Introduction

Implantable cardiac pacemakers (PMs) have been the mainstay for the treatment of bradyarrhythmia for decades. While these devices undoubtedly represent a unique success story since their first implantation in 1958, PMs have evolved significantly during the past years. With the introduction of leadless devices and conduction system pacing, the major limitations of conventional systems such as transvenous leads or unphysiological myocardial stimulation have been addressed. Conduction system pacing systems in particular have introduced a large variety of pacing options, which allow for tailoring antibradycardia stimulation individually to the patient’s needs. On the other hand, implantation complexity has increased considerably, and novel follow-up challenges have emerged. The goal of this review is to provide a state-of-the-art overview of conduction system pacing techniques, specific challenges, and clinically important implantation and follow-up tips.

2. Conduction System Pacing–Concept and Evidence

Conduction system pacing (CSP) encompasses His bundle pacing (HBP) and left bundle branch area pacing (LBBAP). The general rationale of CSP is to maximally benefit from the heart’s intrinsic conduction system to stimulate the myocardium in the most physiologic and efficient way possible.
HBP was already performed more than 20 years ago in humans [1]. Shortly thereafter, this solution was followed by early attempts of left ventricular septal pacing using a dedicated novel lead design [2]. A refined approach for left ventricular bundle branch pacing was proposed by Huang et al. in 2017 [3]. Given the relatively recent introduction of these concepts, the available evidence for CSP is still limited compared to the exhaustive evidence underlining the benefits of cardiac resynchronization therapy (CRT), which, similar to CSP, aims at improving synchronous left ventricular activation. Subsequently, we provide a short overview of the key studies published so far that have investigated the use of HBP and LBBAP systems.

2.1. His Bundle Pacing

Soon after the first HBP system implantations, acute beneficial hemodynamic and echocardiographic effects compared to conventional right ventricular (RV) pacing were demonstrated. Ventricular activation time was shortened during selective and non-selective HBP compared to RV pacing [4]. HBP also decreased interventricular dyssynchrony and reduced mitral valve regurgitation [5]. After the implantation of HBP in lieu of CRT, tricuspid valve regurgitation was reduced as well [6]. These mechanistic effects seem to translate into clinical outcomes: In nonrandomized registry studies, HBP was associated with a reduction of the combined endpoint of mortality and heart failure hospitalizations during long-term follow-up compared to RV pacing [7,8].
In CRT candidates, HBP leads to an improved NYHA functional class by QRS shortening and improvement of the LV ejection fraction [9]. Two randomized trials (His-SYNC and His-Alternative) have also compared HBP to CRT in an on-treatment fashion. HBP was equivalent or even superior to CRT with respect to QRS narrowing compared to baseline and echocardiographic response rates [10,11]. However, a crossover was required in up to almost half of the patients in the HBP group due to procedural failure. In a very recent randomized crossover analysis, HBP provided a significant improvement in the LV ejection fraction compared to CRT after AV nodal ablation [12]. These studies strongly support the only class IIa recommendation for CSP in the latest European pacing guidelines: In case coronary sinus lead implantation is unsuccessful, HBP lead implantation should be considered as an alternative [13].
HBP is the most physiologic pacing strategy, although the technique suffers from major limitations. The significant learning curve during the first 50–100 lead implantations needs to be overcome first [14]. Reported success rates vary widely from 50%-92% depending on the indication and operator experience [9,15]. Success rates in the presence of bundle branch blocks or a complete AV block may hardly be higher than 50% even in experienced centers [15]. Furthermore, even after successful implantation, power consumption of the device is higher compared to conventional RV pacing and may further increase over time [7]. This shortens the battery life and may even require HBP lead revision or lead deactivation during follow-up.

2.2. Left Bundle Branch Area Pacing

LBBAP comprises true left bundle branch pacing (LBBP), left bundle fascicular pacing (LBFP), and left ventricular septal pacing (LVSP). Definitions and distinctions between these modalities, however, are not uniformly used and differ amongst studies. Therefore, caution is warranted when comparing different trials.
Compared to HBP, data on LBBAP are even more scarce and largely based on (smaller) registry studies. In general, capture thresholds are low and remain stable over time, and significant tricuspid valve regurgitation is rare [16]. Patients with a wide QRS complex benefit from the improvement of the ejection fraction [16]. LBBAP shortens the QRS complex significantly compared to RV pacing [17]. Clinically, this results in lower rates of heart failure hospitalization and even lower all-cause mortality according to registry data [17]. Despite pacing from a more distal site than HBP, the paced QRS duration of LBBAP is not significantly longer and—depending on the indication—in the range of 120 ms (Figure 1) [18]. Full correction of a left bundle branch block can more often and more easily be achieved compared to HBP.
Patients with a conventional CRT indication may benefit from LBBAP as well. In a large observational study that included 477 CRT and CSP patients (219 CRT, 171 LBBAP, and 87 HBP systems), the paced QRS duration was shorter in LBBAP systems (133 vs. 153 ms). Echocardiographic response rates were greater in patients receiving a CSP system compared to patients with conventional CRT. In patients with LBBAP, LVEF improved from 26 to 41% compared to 26 to 34% in the conventional CRT group [19]. This also translates into favorable clinical outcomes: The risk of death or heart failure hospitalization was reduced significantly. In a small propensity-score-matched analysis, LBBAP outperformed conventional CRT with respect to the functional NYHA class [20]. These promising findings are supported by the prospective randomized LBBP-RESYNC trial that included 40 patients with non-ischemic cardiomyopathy, a left bundle branch block, and CRT indication. LBBAP demonstrated greater improvement in the LV ejection fraction compared to conventional CRT even in the intention-to-treat analysis [21]. LBBAP is also promising in patients with right bundle branch block, heart failure, and mildly reduced LV dysfunction as it narrows QRS duration and improves LV function [22]. Thus, at least in the case of failed coronary sinus lead implantation in CRT candidates, LBBAP may—similarly to HBP—serve as a bail out with consecutive improvement of the LV ejection fraction (~10% increase compared to baseline) [23]. In an international multicenter study, LBBAP was performed after failed coronary sinus lead implantation and resulted in a QRS narrowing of almost 30 ms [23]. Unlike HBP, LBBAP offers excellent capture thresholds and R-wave amplitudes in this situation [23]. This makes LBBAP leads suitable for sensing in implantable cardioverter-defibrillators (ICDs), as was demonstrated in the CROSS-LEFT pilot study [24]. However, despite the larger target area for lead implantation, the challenges and learning curve of transseptal LBBAP lead implantation should not be underestimated. Up to 150 lead implantations are required to achieve optimal implantation results. The implantation duration and fluoroscopy time still remain slightly longer compared to conventional RV lead implantation [25]. Finally, a lead position deep in the interventricular septum raises concerns regarding the long-term extractability of LBBAP leads [26].
Success rates of LBBAP lead implantation attempts are >90% for conventional bradycardia indications and >80% for heart failure indications according to a European multicenter registry (MELOS) [27]. However, true full LBBP capture (not including fascicular pacing or LV septal pacing) may be achieved in <50% of cases [27,28]. Reported complication rates are approximately 8%, whereof 4% represent intraoperative perforations into the LV, which usually do not result in sequela after the intervention [27]. Since LBBAP is a very young technique, the latest European and American pacing guidelines refrained from any recommendations regarding its use at this point in time.

3. CSP Implantation Technique

CSP lead implantation differs from traditional lead implantation regarding several key aspects and is more challenging. High-resolution fluoroscopy, a 12-lead ECG, dedicated implantation material, and trained implanters are required for CSP lead implantation. The introduction of three-dimensional (3D) guiding sheaths allows for better reach of the target area and has boosted interest in CSP in the past few years.

3.1. Implantation Guiding Sheaths

Dedicated 3D guiding sheaths for CSP are available in various dimensions from Medtronic (Dublin, Ireland), Biotronik (Berlin, Germany), Boston Scientific (Marlborough, MA, USA), and Abbott (Chicago, IL, USA). They all require a 7–9F introducer sheath and provide a side-port for contrast delivery to perform septography. The sheaths facilitate reaching a suitable septal target site and provide additional backup, which may be of particular importance for LBBAP lead implantation, as the lead needs to penetrate deep into the interventricular septum. Some sheaths are deflectable or even equipped with mapping electrodes, which may be helpful for mapping the His bundle. The guiding sheaths are inserted using a J-wire at the beginning of the implantation and are slit at the end of the procedure.

3.2. CSP Leads

A lumen-less lead (3830 SelectSecure, Medtronic) with a 4.1F lead body and a 1.8 mm long fixed helix has initially been used for HBP in particular. However, for LBBAP, conventional stylet-driven leads (Solia S, Biotronik; Fineline II and Ingevity+, Boston Scientific; Tendril 2088TC, Abbott) have been used for CSP as well [29,30]. The lead implantation techniques vary slightly amongst the different electrode designs [29], and learning curves for the individual leads must be assumed. No systematic performance differences between lumen-less and stylet-driven leads regarding acute implantation success rates, screw-in attempts, and fluoroscopy time have been identified so far [30]. However, concerns regarding a higher rate of lead failures of extendable helix leads compared to fixed-helix leads have been raised recently [26].

3.3. Additional Material

Sterile cables with crocodile clamps allow for connecting the CSP lead to the electrophysiology system or a pacing system analyzer (PSA). This is essential for (pace) mapping, repetitive impedance measurements, and standard lead measurements.

3.4. HBP Lead Implantation Technique

The target site for HBP is fibers of the specific conduction system in the right atrium and the penetrating bundle, which is approximately 2–8 mm long and 1 mm thick. The HBP lead is advanced through the guiding sheath while mapping the target area in a unipolar fashion until a His bundle potential can be recorded. The lead is fixated by (usually) 4–8 clockwise turns of the whole lead. Selective or non-selective pacing (with correction of a left bundle branch block in the case of HBP in lieu of CRT) is ensured by uni- and bipolar pacing from the HBP lead. If correction of the QRS cannot be achieved, a His-optimized CRT (HOT-CRT) system with the additional implantation of a CS lead may be a more effective option, resulting in complex programming of the device. Distinguishing non-selective HBP from RV septal pacing may be challenging, thus, programmed stimulation and cutoff values for QRS duration, R-wave peak time in V6, and morphological criteria may help to identify the true His bundle capture [31,32]. Pacing thresholds should be <2.5 V@1 ms in non-PM-dependent patients and lower in dependent patients [33] (<1.5 V@1 ms in the author’s opinion). The implantation of a permanent RV backup lead is recommended in patients undergoing a pace-and-ablate strategy due to a potential threshold rise after subsequent AV nodal ablation [13] or PM-dependent patients due to the risk of dislocation of HBP leads.

3.5. LBBAP Lead Implantation Technique

The target implantation area for an LBBAP lead is significantly larger [34] but the lead needs to be advanced through the interventricular septum until it reaches the LV subendocardium [35]. To identify a suitable starting point for the lead-drilling process on the right ventricular septal side, fluoroscopy, and pace mapping are used [36]. Contrast-enhanced imaging strategies may further facilitate lead implantation [37]. Subsequently, the lead is forced gradually through the septum by clockwise turns (Figure 2). Lead implanters should be aware of fixation issues such as lead entanglement (prevents septal penetration and destroys the lead) and drilling (favors lead dislocation). During the screw-in of the lead, continuous unipolar pacing is performed to monitor the fixation progress through the septum by recognizing ECG changes (Figure 3). Unipolar impedances <420Ω and the loss of injury current are specific indicators for undesired septal perforation [38]. Sensing and pacing thresholds are usually similar or even better compared with conventional RV lead implantation [27]. The 12-lead surface ECG is essential for the verification of conduction system recruitment. A Qr or qR complex in V1 is desired. Moreover, an R-wave peak time in V6 < 80 ms [39], a V1–V6 interpeak interval > 40 ms [40] (Figure 1), and a left bundle branch potential during an intrinsic rhythm (Figure 4) are highly specific for successful implantation of the lead in the left bundle.

4. Considerations on Device Programming and Follow-Up

CSP device programming and follow-up are challenging for several reasons. Up to now, there have been no designated devices in the market addressing CSP programming requirements. The CSP lead can be connected to the atrial, RV, or LV ports of the PM, which increases the overall complexity of CSP programming considerably. In addition, device follow-ups are more time-consuming and require a 12-lead ECG for threshold testing. The CSP lead may pace multiple structures simultaneously (e.g., conduction system with or without correction of a bundle branch block as well as myocardium). The proper identification of the individual thresholds of these structures is crucial for tailored programming of CSP systems [41].
HBP systems may suffer from poor CSP lead performance (low-amplitude ventricular signals, oversensing of atrial and His signals, and high thresholds), which makes programming His bundle PMs often cumbersome and challenging [42]. Moreover, non-selective and selective stimulation with or without correction of a bundle branch block needs to be distinguished from RV septal pacing, and the pacing output needs to be adjusted accordingly. Full correction of bundle branch blocks is desired if an HBP system is implanted in lieu of a CRT.
In contrast, programming LBBAP systems is relatively simple. The LBBAP lead is usually the only ventricular lead since a backup lead is rarely used and an additional coronary sinus lead does not provide better LV activation in the case of true LBB stimulation. For resynchronization, as in HOT-CRT, a CS-lead may be added if full recruitment of the left bundle cannot be achieved, resulting in a left-optimized CRT (LOT-CRT).
In Table 1, we provide an overview of the most important programming recommendations for LBBAP systems with a CSP lead in a ventricular port of the device.

5. Future Perspectives

5.1. Clinical Evidence

Currently, published studies investigating CSP are often monocentric, retrospective, and descriptive in nature. Randomized studies investigating CSP approaches are scarce, but numerous trials have been initiated. Amongst these, only some studies are briefly introduced here. The multicentric PROTECT-HF study aims to randomize 2600 patients with bradycardia pacing indication and a mildly reduced LV ejection fraction to either RV pacing or CSP (HBP or LBBAP) with the primary composite endpoint of death and heart failure hospitalizations. The smaller LEAP-BLOCK study is a randomized controlled trial designed to investigate whether LBBAP reduces the risk of RV-pacing-induced cardiac dysfunction compared to conventional pacing in patients with an AV block and normal LV function.
Randomized trials comparing CSP and CRT systems have also been initiated but encompass only slightly more than 100 patients each. The CONSYST-CRT trial compares both modalities in patients with a class I or IIa CRT indication in a randomized non-inferiority fashion. Resynchronization will be provided by a conventional coronary sinus or a CSP lead. A composite endpoint of mortality, cardiac transplantation, heart failure hospitalization, and insignificant improvement of the LV ejection fraction after implantation will be analyzed. The LEFT-BUNDLE-CRT trial is a similar analysis with 176 patients qualifying for CRT-P/CRT-D implantation. The primary non-inferiority outcome measure will be a positive response to resynchronization defined by a clinical score or LV end-systolic volume.
While all these studies will be important to assess the future prospects of CSP, key limitations have to be considered at this early stage of CSP system evolution. The limited implantation success rate and the long learning curve for implanters will inevitably result in a significant amount of crossovers in randomized trials hampering intention-to-treat analyses. Moreover, non-uniform success criteria used for LBBAP will impair comparability amongst studies.

5.2. Technical Improvements

The introduction of dedicated delivery systems for CSP leads has enabled operators to routinely implant CSP leads. However, the available systems are still far from perfect, and manufacturers are constantly improving CSP hardware.
Delivery sheaths have improved significantly with the introduction of different lengths, curves, and braided tubes for variable stiffness along the sheath body. Nonetheless, the manufacturer’s sheath portfolios still lack valuable tools (e.g., a longer sheath for delivery of stylet-driven leads for larger patients).
Future PM generators will offer the possibility to select the conduction system as a pacing site. Nominal programming values will be suggested accordingly, facilitating device programming. Moreover, improved dedicated CSP lead designs are under development. Lead prototypes include various designs such as fixed-screw leads offering a longer screw and a tempered tip design for better septal penetration.
Finally, merging leadless technology and conduction system pacing approaches may be a highly desirable form of pacing in the future. Leadless systems have evolved rapidly during the past years and novel VDD systems provide relatively reliable AV synchrony [43]. True dual-chamber systems using conductive communication [44] are undergoing first-in-man trials. The stimulation electrode of the ventricular device in such systems might be positioned in the conduction system via a transseptal route. Such PMs may omit pocket and important lead-related complications while delivering a more physiological form of pacing than contemporary leadless PMs. A prototype of a leadless CSP PM undergoing pre-clinical in vivo implantation is shown in Figure 5.

6. Conclusions

Conduction system pacing is becoming increasingly popular. Novel hardware solutions provided by manufacturers and dedicated training courses for implanters facilitate the adoption of this new technique. However, despite the general enthusiasm CSP faces nowadays, structured evidence from randomized trials is still limited. Data are very encouraging but whether or not CSP will be the new Holy Grail for antibradycardia pacing remains to be investigated further. Thus, despite an increasingly liberal implantation practice, patient-specific counseling and case-by-case decisions seem advisable.

Author Contributions

Conceptualization, A.H. and F.N.; methodology, S.C. and A.K.; investigation, A.H., F.N., S.C. and A.K.; writing—original draft preparation, A.H. and F.N.; writing—review and editing, J.S., S.H.B., A.M., G.T., A.R., H.T., L.R. and T.R.; visualization, C.G. All authors have read and agreed to the published version of the manuscript.

Funding

This research received no external funding.

Institutional Review Board Statement

The animal study protocol was approved by the Cantonal Ethics Committee.

Informed Consent Statement

Not applicable.

Data Availability Statement

Not applicable.

Conflicts of Interest

A.H. has received travel/educational grants from Medtronic and Philips/Spectranetics. He is a consultant/advisor for DiNAQOR and Biotronik and a proctor for Medtronic. He is co-founder and head of Act-Inno, a cardiovascular device testing company. He has received research grants from the Swiss National Science Foundation, the Swiss Heart Foundation, the Swiss Heart Rhythm Foundation, the Swiss Pacemaker Foundation, the Hasler Foundation, the Velux Foundation, and Novartis. F.N. has received travel/educational grants from Medtronic, Abbott, and Boston Scientific and Philips/Spectranetics. The spouse of J.S. is an employee and stock owner of Boston Scientific. C.G. received funding from the Swiss National Science Foundation, InnoSuisse, Center for Artificial Intelligence University Bern and the GAMBIT foundation, outside of the submitted work. L.R. has received speaker/consulting honoraria from Abbott and Medtronic and research grants from Medtronic. T.R. has received research grants from the Goldschmidt-Jacobson Foundation, the Swiss National Science Foundation, the Swiss Heart Foundation, and the sitem insel support fund, all for work outside the submitted study. He has received speaker/consulting honoraria or travel support from Abbott/SJM, Bayer, Biosense-Webster, Biotronik, Boston-Scientific, Daiichi Sankyo, Farapulse, Medtronic, and Pfizer-BMS, all for work outside the submitted study. He has received support for his institution’s fellowship program from Abbott/SJM, Biosense-Webster, Biotronik, Boston-Scientific, and Medtronic for work outside the submitted study. The other authors have nothing to disclose relevant to this manuscript.

References

  1. Deshmukh, P.; Casavant, D.A.; Romanyshyn, M.; Anderson, K. Permanent, direct His-bundle pacing: A novel approach to cardiac pacing in patients with normal His-Purkinje activation. Circulation 2000, 101, 869–877. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  2. Grosfeld, M.J.W.; Res, J.C.J.; Vos, D.H.S.; De Boer, T.J.M.; Bos, H.J. Testing a new mechanism for left interventricular septal pacing: The transseptal route A feasibility and safety study. Europace 2002, 4, 439–444. [Google Scholar] [CrossRef] [PubMed]
  3. Huang, W.; Su, L.; Wu, S.; Xu, L.; Xiao, F.; Zhou, X.; Ellenbogen, K.A. A Novel Pacing Strategy With Low and Stable Output: Pacing the Left Bundle Branch Immediately Beyond the Conduction Block. Can. J. Cardiol. 2017, 33, 1736.e1–1736.e3. [Google Scholar] [CrossRef] [PubMed]
  4. Zhang, J.; Guo, J.; Hou, X.; Wang, Y.; Qian, Z.; Li, K.; Ge, P.; Zou, J. Comparison of the effects of selective and non-selective His bundle pacing on cardiac electrical and mechanical synchrony. Europace 2018, 20, 1010–1017. [Google Scholar] [CrossRef]
  5. Catanzariti, D.; Maines, M.; Cemin, C.; Broso, G.; Marotta, T.; Vergara, G. Permanent direct his bundle pacing does not induce ventricular dyssynchrony unlike conventional right ventricular apical pacing. An intrapatient acute comparison study. Journal of interventional cardiac electrophysiology: An international journal of arrhythmias and pacing. J. Interv. Card Electrophysiol. 2006, 16, 81–92. [Google Scholar]
  6. Zaidi, S.M.J.; Sohail, H.; Satti, D.I.; Sami, A.; Anwar, M.; Malik, J.; Mustafa, B.; Mustafa, M.; Mehmoodi, A. Tricuspid regurgitation in His bundle pacing: A systematic review. Ann. Noninvasive Electrocardiol. 2022, 27, e12986. [Google Scholar] [CrossRef]
  7. Vijayaraman, P.; Naperkowski, A.; Subzposh, F.A.; Abdelrahman, M.; Sharma, P.S.; Oren, J.W.; Dandamudi, G.; Ellenbogen, K.A. Permanent His-bundle pacing: Long-term lead performance and clinical outcomes. Heart Rhythm 2018, 15, 696–702. [Google Scholar] [CrossRef]
  8. Abdelrahman, M.; Subzposh, F.A.; Beer, D.; Durr, B.; Naperkowski, A.; Sun, H.; Oren, J.W.; Dandamudi, G.; Vijayaraman, P. Clinical Outcomes of His Bundle Pacing Compared to Right Ventricular Pacing. J. Am. Coll. Cardiol. 2018, 71, 2319–2330. [Google Scholar] [CrossRef]
  9. Sharma, P.S.; Dandamudi, G.; Herweg, B.; Wilson, D.; Singh, R.; Naperkowski, A.; Koneru, J.N.; Ellenbogen, K.A.; Vijayaraman, P. Permanent His-bundle pacing as an alternative to biventricular pacing for cardiac resynchronization therapy: A multicenter experience. Heart Rhythm 2018, 15, 413–420. [Google Scholar] [CrossRef] [PubMed]
  10. Upadhyay, G.A.; Vijayaraman, P.; Nayak, H.M.; Verma, N.; Dandamudi, G.; Sharma, P.S.; Saleem, M.; Mandrola, J.; Genovese, D.; Oren, J.W.; et al. On-treatment comparison between corrective His bundle pacing and biventricular pacing for cardiac resynchronization: A secondary analysis of the His-SYNC Pilot Trial. Heart Rhythm 2019, 16, 1797–1807. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  11. Vinther, M.; Risum, N.; Svendsen, J.H.; Møgelvang, R.; Philbert, B.T. A Randomized Trial of His Pacing Versus Biventricular Pacing in Symptomatic HF Patients With Left Bundle Branch Block (His-Alternative). JACC Clin. Electrophysiol. 2021, 7, 1422–1432. [Google Scholar] [CrossRef] [PubMed]
  12. Huang, W.; Wang, S.; Su, L.; Fu, G.; Su, Y.; Chen, K.; Zou, J.; Han, H.; Wu, S.; Sheng, X.; et al. His-bundle pacing vs biventricular pacing following atrioventricular nodal ablation in patients with atrial fibrillation and reduced ejection fraction: A multicenter, randomized, crossover study—The Alternative-AF trial. Heart Rhythm 2022, 19, 1948–1955. [Google Scholar] [CrossRef] [PubMed]
  13. Glikson, M.; Nielsen, J.C.; Kronborg, M.B.; Michowitz, Y.; Auricchio, A.; Barbash, I.M.; Barrabés, J.A.; Boriani, G.; Braunschweig, F.; Brignole, M.; et al. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy. Eur. Heart J. 2021, 42, 3427–3520. [Google Scholar] [CrossRef]
  14. Keene, D.; Arnold, A.D.; Jastrzębski, M.; Burri, H.; Zweibel, S.; Crespo, E.; Chandrasekaran, B.; Bassi, S.; Joghetaei, N.; Swift, M.; et al. His bundle pacing, learning curve, procedure characteristics, safety, and feasibility: Insights from a large international observational study. J. Cardiovasc. Electrophysiol. 2019, 30, 1984–1993. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  15. Bhatt, A.G.; Musat, D.L.; Milstein, N.; Pimienta, J.; Flynn, L.; Sichrovsky, T.; Preminger, M.W.; Mittal, S. The Efficacy of His Bundle Pacing: Lessons Learned From Implementation for the First Time at an Experienced Electrophysiology Center. JACC Clin. Electrophysiol. 2018, 4, 1397–1406. [Google Scholar] [CrossRef]
  16. Su, L.; Wang, S.; Wu, S.; Xu, L.; Huang, Z.; Chen, X.; Zheng, R.; Jiang, L.; Ellenbogen, K.A.; Whinnett, Z.I.; et al. Long-Term Safety and Feasibility of Left Bundle Branch Pacing in a Large Single-Center Study. Circ. Arrhythm. Electrophysiol. 2021, 14, e009261. [Google Scholar] [CrossRef]
  17. Sharma, P.S.; Patel, N.R.; Ravi, V.; Zalavadia, D.V.; Dommaraju, S.; Garg, V.; Larsen, T.R.; Naperkowski, A.M.; Wasserlauf, J.; Krishnan, K.; et al. Clinical outcomes of left bundle branch area pacing compared to right ventricular pacing: Results from the Geisinger-Rush Conduction System Pacing Registry. Heart Rhythm 2022, 19, 3–11. [Google Scholar] [CrossRef]
  18. Vijayaraman, P.; Rajakumar, C.; Naperkowski, A.M.; Subzposh, F.A. Clinical outcomes of left bundle branch area pacing compared to His bundle pacing. J. Cardiovasc. Electrophysiol. 2022, 33, 1234–1243. [Google Scholar] [CrossRef]
  19. Vijayaraman, P.; Zalavadia, D.; Haseeb, A.; Dye, C.; Madan, N.; Skeete, J.R.; Vipparthy, S.C.; Young, W.; Ravi, V.; Rajakumar, C.; et al. Clinical outcomes of conduction system pacing compared to biventricular pacing in patients requiring cardiac resynchronization therapy. Heart Rhythm 2022, 19, 1263–1271. [Google Scholar] [CrossRef]
  20. Wang, Y.; Gu, K.; Qian, Z.; Hou, X.; Chen, X.; Qiu, Y.; Jiang, Z.; Zhang, X.; Wu, H.; Chen, M.; et al. The efficacy of left bundle branch area pacing compared with biventricular pacing in patients with heart failure: A matched case–control study. J. Cardiovasc. Electrophysiol. 2020, 31, 2068–2077. [Google Scholar] [CrossRef]
  21. Wang, Y.; Zhu, H.; Hou, X.; Wang, Z.; Zou, F.; Qian, Z.; Wei, Y.; Wang, X.; Zhang, L.; Li, X.; et al. Randomized Trial of Left Bundle Branch vs Biventricular Pacing for Cardiac Resynchronization Therapy. J. Am. Coll. Cardiol. 2022, 80, 1205–1216. [Google Scholar] [CrossRef] [PubMed]
  22. Vijayaraman, P.; Cano, O.; Ponnusamy, S.S.; Molina-Lerma, M.; Chan, J.Y.; Padala, S.K.; Sharma, P.S.; Whinnett, Z.I.; Herweg, B.; Upadhyay, G.A.; et al. Left bundle branch area pacing in patients with heart failure and right bundle branch block: Results from International LBBAP Collaborative-Study Group. Heart Rhythm O2 2022, 3, 358–367. [Google Scholar] [CrossRef] [PubMed]
  23. Vijayaraman, P.; Herweg, B.; Verma, A.; Sharma, P.S.; Batul, S.A.; Ponnusamy, S.S.; Schaller, R.D.; Cano, O.; Molina-Lerma, M.; Curila, K.; et al. Rescue left bundle branch area pacing in coronary venous lead failure or nonresponse to biventricular pacing: Results from International LBBAP Collaborative Study Group. Heart Rhythm 2022, 19, 1272–1280. [Google Scholar] [CrossRef] [PubMed]
  24. Clementy, N.; Bodin, A.; Ah-Fat, V.; Babuty, D.; Bisson, A. Dual-chamber ICD for left bundle branch area pacing: The cardiac resynchronization and arrhythmia sensing via the left bundle (cross-left) pilot study. Journal of interventional cardiac electrophysiology: An international journal of arrhythmias and pacing. J. Interv. Card Electrophysiol. 2022; online ahead of print. [Google Scholar] [CrossRef]
  25. Wang, Z.; Zhu, H.; Li, X.; Yao, Y.; Liu, Z.; Fan, X. Comparison of Procedure and Fluoroscopy Time Between Left Bundle Branch Area Pacing and Right Ventricular Pacing for Bradycardia: The Learning Curve for the Novel Pacing Strategy. Front. Cardiovasc. Med. 2021, 8, 695531. [Google Scholar] [CrossRef] [PubMed]
  26. Tan, E.S.J.; Lee, J.; Boey, E.; Soh, R.; Sim, M.G.; Yeo, W.; Seow, S.; Kojodjojo, P. Use of extendable helix leads for conduction system pacing: Differences in lead handling and performance lead design impacts conduction system pacing. J. Cardiovasc. Electrophysiol. 2022, 33, 1550–1557. [Google Scholar] [CrossRef]
  27. Jastrzębski, M.; Kiełbasa, G.; Cano, O.; Curila, K.; Heckman, L.; De Pooter, J.; Chovanec, M.; Rademakers, L.; Huybrechts, W.; Grieco, D.; et al. Left bundle branch area pacing outcomes: The multicentre European MELOS study. Eur. Heart J. 2022, 43, 4161–4173. [Google Scholar] [CrossRef]
  28. Shimeno, K.; Tamura, S.; Hayashi, Y.; Abe, Y.; Naruko, T.; Fukuda, D. Achievement rate and learning curve of left bundle branch capture in left bundle branch area pacing procedure performed to demonstrate output-dependent QRS transition. J. Cardiovasc. Electrophysiol. 2022, 33, 2183–2191. [Google Scholar] [CrossRef]
  29. De Pooter, J.; Wauters, A.; Van Heuverswyn, F.; Le Polain de Waroux, J.B. A Guide to Left Bundle Branch Area Pacing Using Stylet-Driven Pacing Leads. Front. Cardiovasc. Med. 2022, 9, 844152. [Google Scholar] [CrossRef]
  30. De Pooter, J.; Calle, S.; Timmermans, F.; Van Heuverswyn, F. Left bundle branch area pacing using stylet-driven pacing leads with a new delivery sheath: A comparison with lumen-less leads. J. Cardiovasc. Electrophysiol. 2021, 32, 439–448. [Google Scholar] [CrossRef]
  31. Burri, H.; Jastrzebski, M.; Vijayaraman, P. Electrocardiographic Analysis for His Bundle Pacing at Implantation and Follow-Up. JACC Clin. Electrophysiol. 2020, 6, 883–900. [Google Scholar] [CrossRef] [PubMed]
  32. Jastrzębski, M.; Moskal, P.; Bednarek, A.; Kiełbasa, G.; Vijayaraman, P.; Czarnecka, D. Programmed His Bundle Pacing. Circ. Arrhythmia Electrophysiol. 2019, 12, e007052. [Google Scholar] [CrossRef] [PubMed]
  33. Vijayaraman, P.; Dandamudi, G.; Zanon, F.; Sharma, P.S.; Tung, R.; Huang, W.; Koneru, J.; Tada, H.; Ellenbogen, K.A.; Lustgarten, D.L. Permanent His bundle pacing: Recommendations from a Multicenter His Bundle Pacing Collaborative Working Group for standardization of definitions, implant measurements, and follow-up. Heart Rhythm 2018, 15, 460–468. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  34. Lu, W.; Lin, J.; Chen, K.; Dai, Y.; Chen, R.; Hu, Q.; Li, Y.; Cheng, C.; Zhou, Y.; Zhang, S. Quantitative distance and electrocardiographic parameters for lead-implanted site selection to enhance the success likelihood of left bundle branch pacing. Clin. Res. Cardiol. 2021, 111, 1219–1230. [Google Scholar] [CrossRef]
  35. Cabrera, J.-Á.; Porta-Sánchez, A.; Tung, R.; Sánchez-Quintana, D. Tracking Down the Anatomy of the Left Bundle Branch to Optimize Left Bundle Branch Pacing. JACC Case Rep. 2020, 2, 750–755. [Google Scholar] [CrossRef] [PubMed]
  36. Huang, W.; Chen, X.; Su, L.; Wu, S.; Xia, X.; Vijayaraman, P. A beginner’s guide to permanent left bundle branch pacing. Heart Rhythm 2019, 16, 1791–1796. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  37. Liu, X.; Niu, H.-X.; Gu, M.; Chen, X.; Hu, Y.; Cai, M.; Zhang, N.; Zhao, J.; Zhou, X.; Gold, M.R.; et al. Contrast-enhanced image-guided lead deployment for left bundle branch pacing. Heart Rhythm 2021, 18, 1318–1325. [Google Scholar] [CrossRef] [PubMed]
  38. Shali, S.; Wu, W.; Bai, J.; Wang, W.; Qin, S.; Wang, J.; Liang, Y.; Chen, H.; Su, Y.; Chen, X.; et al. Current of injury is an indicator of lead depth and performance during left bundle branch pacing lead implantation. Heart Rhythm 2022, 19, 1281–1288. [Google Scholar] [CrossRef]
  39. Jastrzębski, M.; Kiełbasa, G.; Curila, K.; Moskal, P.; Bednarek, A.; Rajzer, M.; Vijayaraman, P. Physiology-based electrocardiographic criteria for left bundle branch capture. Heart Rhythm 2021, 18, 935–943. [Google Scholar] [CrossRef]
  40. Jastrzębski, M.; Burri, H.; Kiełbasa, G.; Curila, K.; Moskal, P.; Bednarek, A.; Rajzer, M.; Vijayaraman, P. The V6-V1 interpeak interval: A novel criterion for the diagnosis of left bundle branch capture. Europace 2022, 24, 40–47. [Google Scholar] [CrossRef]
  41. Bakelants, E.; Burri, H. Troubleshooting Programming of Conduction System Pacing. Arrhythmia Electrophysiol. Rev. 2021, 10, 85–90. [Google Scholar] [CrossRef]
  42. Burri, H.; Keene, D.; Whinnett, Z.; Zanon, F.; Vijayaraman, P. Device Programming for His Bundle Pacing. Circ. Arrhythmia Electrophysiol. 2019, 12, e006816. [Google Scholar] [CrossRef] [PubMed]
  43. Neugebauer, F.; Noti, F.; van Gool, S.; Roten, L.; Baldinger, S.H.; Seiler, J.; Madaffari, A.; Servatius, H.; Ryser, A.; Tanner, H.; et al. Leadless atrioventricular synchronous pacing in an outpatient setting: Early lessons learned on factors affecting atrioventricular synchrony. Heart Rhythm 2021, 19, 748–756. [Google Scholar] [CrossRef] [PubMed]
  44. Bereuter, L.; Gysin, M.; Kueffer, T.; Kucera, M.; Niederhauser, T.; Fuhrer, J.; Heinisch, P.; Zurbuchen, A.; Obrist, D.; Tanner, H.; et al. Leadless Dual-Chamber Pacing: A Novel Communication Method for Wireless Pacemaker Synchronization. JACC Basic Transl. Sci. 2018, 3, 813–823. [Google Scholar] [CrossRef] [PubMed]
Figure 1. Paced QRS morphology in a patient with a dual-chamber PM and an LBBAP lead. V1 exhibits an R’, R-wave peak time in V6 is ~70 ms, the V1–V6 interpeak interval is ~35 ms and paced QRS duration is ~120 ms.
Figure 1. Paced QRS morphology in a patient with a dual-chamber PM and an LBBAP lead. V1 exhibits an R’, R-wave peak time in V6 is ~70 ms, the V1–V6 interpeak interval is ~35 ms and paced QRS duration is ~120 ms.
Jcm 11 07258 g001
Figure 2. Computed tomography (CT) images of a patient with AV block after transcatheter aortic valve replacement and consecutive dual-chamber PM implantation with a ventricular lead in LBBAP position. (A) The deep septal implantation site of the CSP lead in a short-axis projection. (B) The basal implantation site in the proximal left bundle in a reconstruction. The resulting ECG is shown in Figure 1.
Figure 2. Computed tomography (CT) images of a patient with AV block after transcatheter aortic valve replacement and consecutive dual-chamber PM implantation with a ventricular lead in LBBAP position. (A) The deep septal implantation site of the CSP lead in a short-axis projection. (B) The basal implantation site in the proximal left bundle in a reconstruction. The resulting ECG is shown in Figure 1.
Jcm 11 07258 g002
Figure 3. ECG leads V1 and V6 during LBBAP lead implantation. Continuous unipolar pacing is performed from the lead’s tip. ① A wide QRS reflecting right ventricular septal pacing with a notch near the nadir of the QRS (one of the criteria for a suitable starting position for lead drilling). ② During the drilling process, this notch starts to move towards the end of the QRS in V1 and the QRS duration is shortened (deep septal pacing). ③ Finally, an r’ appears in V1 and the QRS suddenly shortens again significantly, indicating CSP capture.
Figure 3. ECG leads V1 and V6 during LBBAP lead implantation. Continuous unipolar pacing is performed from the lead’s tip. ① A wide QRS reflecting right ventricular septal pacing with a notch near the nadir of the QRS (one of the criteria for a suitable starting position for lead drilling). ② During the drilling process, this notch starts to move towards the end of the QRS in V1 and the QRS duration is shortened (deep septal pacing). ③ Finally, an r’ appears in V1 and the QRS suddenly shortens again significantly, indicating CSP capture.
Jcm 11 07258 g003
Figure 4. 12-lead ECG from a patient, who underwent LBBAP lead implantation. The unipolar electrogram (CSPuni) shows a left bundle branch potential during intrinsic rhythm, which is indicative of successful positioning of the lead in the specific conduction system (arrow).
Figure 4. 12-lead ECG from a patient, who underwent LBBAP lead implantation. The unipolar electrogram (CSPuni) shows a left bundle branch potential during intrinsic rhythm, which is indicative of successful positioning of the lead in the specific conduction system (arrow).
Jcm 11 07258 g004
Figure 5. Implantation of a leadless CSP PM in a domestic pig. The steerable implantation catheter (asterisk) allows positioning the stimulation electrode towards the interventricular septum (solid arrow). The main body of the leadless PM is implanted in an apical position (dotted arrow). The corresponding ECG with a simulated QRS duration of <100 ms is shown on the right.
Figure 5. Implantation of a leadless CSP PM in a domestic pig. The steerable implantation catheter (asterisk) allows positioning the stimulation electrode towards the interventricular septum (solid arrow). The main body of the leadless PM is implanted in an apical position (dotted arrow). The corresponding ECG with a simulated QRS duration of <100 ms is shown on the right.
Jcm 11 07258 g005
Table 1. LBBAP programming recommendations for systems with a CSP lead in a ventricular port.
Table 1. LBBAP programming recommendations for systems with a CSP lead in a ventricular port.
ParameterRecommendationReasoning
Sensing vectorBipolarBipolar sensing is preferred to avoid oversensing of pectoral myopotentials.
SensitivityAs a conventional leadSensing values are usually excellent and rarely cause concerns.
Output polarityBipolar (extended bipolar/unipolar may be considered for better visibility of the PM stimulus)Bipolar pacing is usually more power-efficient but simultaneous cathodal and anodal capture may mimic lead microdislocation. Be aware of undesired anodal RV capture in case of an extended bipolar configuration.
Output voltageAs a conventional lead (2× threshold, 0.4 ms pulse duration)Capture thresholds are usually excellent and remain stable over time.
AV delayLBBAP lead in RV port: subtraction of LBB-V interval from desired AV delay
LBBAP lead in LV port: program VV delay first, adjust AV delay thereafter in a similar fashion
The resulting AV delay corresponds to the programmed AV delay plus the conduction time from the left bundle to the ventricle (LBB-V interval, ~15–25 ms).
VV delayLBBAP lead in LV port:
option 1: program LV first (~60–80 ms, in case of LV septal pacing, fusion may shorten QRS)
option 2: deactivate RV pacing in non PM-dependent patients
option 3: program subthreshold stimulation in the RV
Promote ventricular activation via the conduction system in case of proper left bundle activation. In case of deep septal lead position, LV-RV-fusion (similar to CRT) may further decrease QRS duration. Subthreshold stimulation may cause confusion and provides no ventricular backup.
Automatic capture controlMonitorMay be activated in the chronic stage when correct functioning has been verified.
Automatic sensingOnSensing values are usually excellent and rarely cause concerns.
Ventricular safety pacingOnTurn off if CSP lead is connected to the atrial port (after crosstalk test).
Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Share and Cite

MDPI and ACS Style

Haeberlin, A.; Canello, S.; Kummer, A.; Seiler, J.; Baldinger, S.H.; Madaffari, A.; Thalmann, G.; Ryser, A.; Gräni, C.; Tanner, H.; et al. Conduction System Pacing Today and Tomorrow. J. Clin. Med. 2022, 11, 7258. https://doi.org/10.3390/jcm11247258

AMA Style

Haeberlin A, Canello S, Kummer A, Seiler J, Baldinger SH, Madaffari A, Thalmann G, Ryser A, Gräni C, Tanner H, et al. Conduction System Pacing Today and Tomorrow. Journal of Clinical Medicine. 2022; 11(24):7258. https://doi.org/10.3390/jcm11247258

Chicago/Turabian Style

Haeberlin, Andreas, Siro Canello, Andreas Kummer, Jens Seiler, Samuel H. Baldinger, Antonio Madaffari, Gregor Thalmann, Adrian Ryser, Christoph Gräni, Hildegard Tanner, and et al. 2022. "Conduction System Pacing Today and Tomorrow" Journal of Clinical Medicine 11, no. 24: 7258. https://doi.org/10.3390/jcm11247258

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop