Yonsei Med J. 1994 Jun;35(2):209-217. English.
Published online Feb 20, 2002.
Copyright © 1994 The Yonsei University College of Medicine
Original Article

Ten years of experience with various penile prosthesis in Korean

Hyung Ki Choi, In Rae Cho and Zhong Cheng Xin
    • Department of Urology, Yonsei University College of Medicine, Seoul, Korea.

Abstract

Currently there are more than 10 types of penile prosthesis available, ranging from the very simple to the very sophisticated. We review our experiences with various penile prosthesis, with particular regard to the complication rate. From Dec. 1983 to Jul. 1993, we implanted 295 penile prosthesis of eight different types. The average age of patients was 44 years. Every patient was evaluated with various multidisplinary diagnostic approaches. The etiologies of impotence were vasculogenic 29%, diabetogenic 22%, spinal cord injury 16%, pelvic bone injury 11%, etc. The types of implanted prosthesis were AMS malleable 143, Jonas 42, Dynaflex 36, Hydroflex 8, Uni-Flate 1000 2, AMS 700 CXM 58, Ultrex 3, Mentor alpha-1 3 and the mean follow-up period was 34 months. The diameters of implanted prosthesis were from 9.5 mm to 13 mm, mostly 9.5 mm (52.9%). The length of implanted prosthesis were from 10 cm to 20 cm, mostly 16 approximately 18 cm (68.8%). Cases with uneven diameters or lengths were 20 (6.8%). The int aoperative complications were 1 corporeal rupture and 1 bladder rupture, and the postoperative complications were 2 prosthesis infections, 2 mechanical failures, and 1 prosthesis infection with mechanical failure. In those 4 patients reimplantations were successful. More than 99% (290/291) patients still have functioning prosthesis. Every prosthesis has their advantages and disadvantages. Factors to be analysed in the selection of proper prosthesis should include patients economic status, education, personality, social activity, hand dexterity, and penile size.

Keywords
Penile prosthesis; complication; AMS 700 CXM


Metrics
Share
PERMALINK