Abstract
A flow injection analysis method is proposed for the spectrophotometry determination of trace amounts of iodide based on its catalytic effect on the oxidation of 4,4′-bis(dimethylamino)diphenylmethane (tetrabase) by Chloramine T in a weakly acidic solution. Iodide in water samples can be determined at a sampling rate of 85 h-1 with 300 μl sample injections. The detection limit is 0.1 μgl-1 and the calibration graph for iodide is linear up to 2 μgl-1, showing a slightly negative deviation in the higher concentration range. The relative standard deviations are 2.8 and 0.8% for 0.40 and 4.0 μgl-1 iodide, respectively. Although free iodine shows a somewhat lower catalytic effect and iodate no catalytic effect up to at least 50 μg l-1 as iodine, these species show the same effect as iodide upon being reduced with thioacetamide in an acidic solution. The procedure has been applied to the determination of iodide in natural water samples as well as iodine impurity in both distilled and redistilled water samples. It has also been used to examine iodine contamination in ambient air.
Similar content being viewed by others
References
T. Deguchi, A. Tanaka, I. Sanemasa and H. Nagai, Bunseki Kagaku, 32, 23 (1983).
E. Jungreis and I. Gedalia, Mikrochim. Acta [Wien], 1960, 10.
E. P. Lankmayr, B. Maichin and G. Knapp, J. Chromatogr., 224, 239 (1981).
W. Buchberger and K. Winsauer, Mikrochim, Acta [Wien], 1985, 347.
W. Buchberger, J. Chromatogr., 493, 129 (1988).
N. Yonehara, Bull Chem. Soc. Jpn., 37, 1101 (1964).
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Yonehara, N., Kozono, S. & Sakamoto, H. Flow Injection-Spectrophotometric Determination of Trace Amounts of Iodide by Its Catalytic Effect on the 4,4-Bis(dimethylamino)-diphenylmethane-Chloramine T Reaction. ANAL. SCI. 7, 229–234 (1991). https://doi.org/10.2116/analsci.7.229
Received:
Accepted:
Published:
Issue Date: