Abstract
A combination of palladium and sodium azide is proposed as a mixed modifier for cadmium determination in urine by graphite-furnace atomic absorption spectrometry using calibration against matrix-free standard solutions. It permits the use of a pretreatment temperature of 900–1000°C and markedly reduces the background absorption caused by inorganic salt matrices. The urine was simply diluted 1+1 with the addition of 1% HCl, Pd and NaN3. The recovery of cadmium obtained from urine was higher than 95%. The relative standard deviation at a cadmium level of 2.10–4.56 µg l-1 was smaller than 4%. The detection limit (3 s, n=20) is 0.16 µg l−1 of cadmium.
Similar content being viewed by others
References
R. D. Ediger, At. Absorpt. Newsl., 14, 127 (1975).
K. G. Feitsma, J. P. Franke and P. A. Zeeuw, Analyst [London], 109, 789 (1991).
K. S. Subramanian, J. C. Meranger and J. E. MacKeen, Anal. Chem., 55, 1064 (1983).
J. S. Verbeke, Q. Yang, W. Penninck and F. Vandervoort, J. Anal. At. Spectrom., 5, 393 (1990).
C. R. Lan, Analyst [London], 118, 189 (1993).
R. Kobayashi, K. Imaizumi, Anal. Sci., 7 (supplement), 1197 (1991).
Y. Morishiye, K. Hirokowa, K. Yasuda, Fresenius’ J. Anal. Chem., 350, 410 (1994).
E. Bulska, Z. Grobenski, G. Schlemmer, J. Anal. At. Spectrom., 5, 203 (1990).
X. Yin, G. Schlemmer, B. Welz, Anal. Chem., 59, 1462 (1987).
G. Schlemmer, B. Welz, Spectrochim. Acta, 41B, 1157 (1986).
X. Q. Shan, Z. M. Ni, Acta Chim. Sin., 37, 261 (1979).
P. B. Barrera, J. M. Pineiro, A. M. Pineiro, A. B. Barrera, Fresenius’ J. Anal. Chem., 174 (1996).
A. Brzezinska, J. C. Van Loon, Fresenius’ Z. Anal. Chem., 331, 707 (1989).
A. Volynsky, S. Tikhomirov, A. Elagin, Analyst [London], 116, 145 (1991).
E. Bulska, W. Jedral, J. Anal. At. Spectrom., 10, 49 (1995).
L. M. Voth-Beach and D. E. Shrader, J. Anal. At. Spectrom., 2, 45 (1987).
M. Yizai, L. Zhikun, W. Xiouhui, W. Jiazhen and L. Yangguan, J. Anal. At. Spectrom., 9, 679 (1994).
M. Hansen, “Constitution of Binary Alloys”, McGarw-Hill, New York, 1958.
Y. Terui, K. Yasuda, K. Hirakowa, Anal. Sci., 7, 397 (1991).
B. R. Culver, T. Surles, Anal. Chem., 47, 920 (1975).
L. Elçi, U. Şahin, S. Öztasş, Talanta, 44, 1017 (1997).
L. Elçi, U. Şahin, A. Ülgen, Trace Elements and Electrolytes, 2, 106 (1998).
D. A. Skoog, D. M. West and F. J. Holler, “Fundamentals of Analytical Chemistry”, 6th ed., Saunders College Publishing, New York, 1991.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Saraçoğlu, S., Elçi, L. Effectiveness of Palladium–Sodium Azide Modifier for the Direct Determination of Urinary Cadmium by Graphite-Furnace Atomic Absorption Spectrometry. ANAL. SCI. 15, 569–573 (1999). https://doi.org/10.2116/analsci.15.569
Received:
Accepted:
Published:
Issue Date: