Soil & Water Res., 2020, 15(2):75-84 | DOI: 10.17221/227/2018-SWR

The laboratory calibration of a soil moisture capacitance probe in sandy soilsOriginal Paper

Marina Campora1, Anna Palla*,2, Ilaria Gnecco2, Rossella Bovolenta2, Roberto Passalacqua2
1 Genoa, Italy
2 Department of Civil, Chemical and Environmental Engineering, University of Genoa, Genoa, Italy

Determining and mitigating landslide risk is a technical-scientific objective, particularly for the protection and proper territorial management and planning. The slope stability depends on the pore pressure distribution, which is influenced by the saturation front propagation through the unsaturated zone, whose monitoring is useful to understand any possible instabilities. Such monitoring may be undertaken by sensors based on the measurement of the relative dielectric permittivity. Reliable relationships between the measurement and the soil moisture are necessary. The main objective of this study is to assess a laboratory calibration protocol for a specific capacitance sensor (Drill & Drop, Sentek Sensor Technologies). Two monogranular sands have been selected for the calibration purpose. The laboratory tests were performed under three relative density values (DR equal to 40%, 60% and 80%) for seven volumetric water content values (θv ranging from 0.00% to 36.26%). Based on the experimental measurements, the soil-specific calibration curves were determined at an assigned relative density value; in particular, a simple power law is adopted to describe the probe's reading as a function of the volumetric water content. The results point out that the relative density values slightly affect the tests, thus, the soil-specific calibration curves are derived based on a simple regression analysis fitting the whole set of the laboratory tests validated for each sand. The calculated coefficient of determination (R2 = 0.96/0.99) and root mean square error (RMSE = 1.4%/2.8%) values confirm the goodness of fit. In order to propose more general fitting curves, suitable for both the investigated sands, multiple linear regressions are performed by considering θv and the mean grain size, D50 as independent variables; again, the R2 and RMSE values equal to 0.97 and 2.41%, respectively, confirm the suitability of the calibration curve. Finally, the laboratory calibration curves are compared with the manufacturer-supplied curves, thus, enhancing the need for the soil-specific calibration.

Keywords: particle size distribution; regression analysis; relative density; soil water monitoring; volumetric water content

Published: June 30, 2020  Show citation

ACS AIP APA ASA Harvard Chicago IEEE ISO690 MLA NLM Turabian Vancouver
Campora M, Palla A, Gnecco I, Bovolenta R, Passalacqua R. The laboratory calibration of a soil moisture capacitance probe in sandy soils. Soil & Water Res.. 2020;15(2):75-84. doi: 10.17221/227/2018-SWR.
Download citation

References

  1. Bogena H.R., Huisman J.A., Oberdörster C., Vereecken H. (2007): Evaluation of a low-cost soil water content sensor for wireless network applications. Journal of Hydrology, 344: 32-42. Go to original source...
  2. Bordoni M., Meisina C., Valentino R., Lu N., Bittelli M., Chersich S. (2015): Hydrological factors affecting rainfall-induced shallow landslides: From the field monitoring to a simplified slope stability analysis. Engineering Geology, 193: 19-37. Go to original source...
  3. Bovolenta R., Passalacqua R., Federici B., Sguerso D. (2016): LAMP - Landslide Monitoring and Predicting for the analysis of landslide susceptibility triggered by rainfall events. In: Proc. 12th Int. Symp. Landslides and Engineered Slopes. Experience, Theory and Practice, Napoli, June 12-19, 2016: 517-522. Go to original source...
  4. Campbell C.S., Campbell G.S., Cobos D.R., Bissey L.L. (2009): Calibration and Evaluation of an Improved Lowcost Soil Moisture Sensor. Application Note. Pullman, Decagon Services, Inc.
  5. De Carteret R., Buzzi O., Fityus S. (2013): Installation, calibration and application of dielectric sensors in laboratory studies of compacted unsaturated granular materials. Geotechnical Testing Journal, 36: 1-12. Go to original source...
  6. Fares A., Awal R., Bayabil H.K. (2016): Soil water content sensor response to organic matter content under laboratory conditions. Sensors, 16: 1239-1249. Go to original source... Go to PubMed...
  7. Federici B., Bovolenta R., Passalacqua R. (2015): From rainfall to slope instability: an automatic GIS procedure for susceptibility analyses over wide areas. Geomatics, Natural Hazards & Risk, 6 (Special Issue): 454-472. Go to original source...
  8. Gaskin G.J., Miller J.D. (1995): Measurement of soil water content using a simplified impedance measuring technique. Journal of Agricultural Engineering Research, 63: 153-160. Go to original source...
  9. Godt J.W., Baum R.L., Lu N. (2009): Landsliding in partially saturated materials. Geophysical Research Letters, 36: L02403. Go to original source...
  10. Iezzoni H.M., McCartney J.S. (2016): Calibration of capacitance sensors for compacted silt in non-isothermal applications. Geotechnical Testing Journal, 39: 169-180. Go to original source...
  11. Passalacqua R., Bovolenta R., Federici B. (2015): An Integrated Hydrological-Geotechnical Model in GIS for the Analysis and Prediction of Large-Scale Landslides Triggered by Rainfall Events. In: Lollino G. et al. (eds): Engineering Geology for Society and Territory - Volume 2. Cham, Springer: 1799-1803. Go to original source...
  12. Provenzano G., Rallo G., Ghazouani H. (2016): Assessing field and laboratory calibration protocols for the Diviner 2000 probe in a range of soils with different textures. Journal of Irrigation and Drainage Engineering, 142: 1-12. Go to original source...
  13. Sentek Technologies Inc. (2011): Calibration manual for Sentek soil moisture sensors. Version 2.0. Sentek Sensor Technologies, Stepney, SA, Australia.
  14. Sentek Technologies Inc. (2014): Sentek Drill & Drop probe manual. Sentek Sensor Technologies, Stepney, SA, Australia.
  15. Tarantino A., Pozzato A. (2008): Theoretical analysis of the effect of temperature, cable length and double-impedance probe head on TDR water content measurement. In: Unsaturated Soils: Advances in Geo-Engineering. Proc. 1st European Conf. Unsaturated Soils, E-UNSAT: 165-171. Go to original source...
  16. Tarantino A., Ridley A.M., Toll D.G. (2008): Field measurement of suction, water content, and water permeability. Geotechnical and Geological Engineering, 26: 751-782. Go to original source...
  17. Wang J.R., Schmugge T.J. (1980): An empirical model for the complex dielectric permittivity of soils as a function of water content. IEEE Transactions on Geoscience and Remote Sensing, GE-18: 288-295. Go to original source...

This is an open access article distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International (CC BY NC 4.0), which permits non-comercial use, distribution, and reproduction in any medium, provided the original publication is properly cited. No use, distribution or reproduction is permitted which does not comply with these terms.