Skip to main content
Advertisement
Browse Subject Areas
?

Click through the PLOS taxonomy to find articles in your field.

For more information about PLOS Subject Areas, click here.

  • Loading metrics

Frequency and clinical impact of retained implantable cardioverter defibrillator lead materials in heart transplant recipients

  • Jun Kim ,

    Contributed equally to this work with: Jun Kim, Jongmin Hwang

    Affiliation Department of Cardiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea

  • Jongmin Hwang ,

    Contributed equally to this work with: Jun Kim, Jongmin Hwang

    Affiliation Department of Cardiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea

  • Jin Hee Choi,

    Affiliation Department of Cardiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea

  • Hyo-In Choi,

    Affiliation Department of Cardiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea

  • Min-Seok Kim,

    Affiliation Department of Cardiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea

  • Sung-Ho Jung,

    Affiliation Thoracic and Cardiovascular Surgery, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea

  • Gi-Byoung Nam,

    Affiliation Department of Cardiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea

  • Kee-Joon Choi,

    Affiliation Department of Cardiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea

  • Jae Won Lee,

    Affiliation Thoracic and Cardiovascular Surgery, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea

  • You-Ho Kim,

    Affiliation Department of Cardiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea

  • Jae-Joong Kim

    jjkim@amc.seoul.kr

    Affiliation Department of Cardiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea

Abstract

End-stage heart failure patients with implantable cardioverter-defibrillator (ICD) with/without cardiac resynchronization therapy (CRT-D) often require heart transplantation (HTPL) as a last-resort treatment. We aimed to assess the frequency and clinical impact of retained ICD lead materials in HTPL patients. In this retrospective single center study, we examined the clinical records and chest radiographs of patients with ICD and CRT-D who underwent HTPL between January 1992 and July 2014. Of 40 patients with ICD and CRT-D at HTPL, 19 (47.5%) patients had retained ICD lead materials within the central venous system. Retained ICD lead materials following HTPL were more frequently noted in patients with longer implantation durations until HTPL. None of the patients underwent extraction procedures after HTPL. All patients were asymptomatic and did not exhibit significant complications or death related to the retained ICD lead materials. Seven (7/40, 17.5%) patients without any retained ICD lead materials underwent magnetic resonance imaging (MRI) during the follow-up period (median, 29.5 months); none of the patients with retained lead materials were given MRI. Considering the common use of MRI in HTPL patients, further studies on the prophylactic extraction of retained ICD lead materials and safety of MRI in these patients are needed.

Introduction

Implantable cardiac devices are used in advanced heart failure patients to improve the symptom, and prognosis. Implantable cardioverter-defibrillators (ICD) can prevent sudden cardiac death attributed to ventricular tachyarrhythmia, and cardiac resynchronization therapy (CRT), usually with defibrillation capability (CRT-D), can reduce symptoms and hospitalizations due to ventricular dyssynchrony in these patients [13]. Despite the beneficial effects of this therapy, end-stage heart failure patients often require heart transplantation (HTPL) because it is the regarded as the last-resort treatment option in these patients, and is associated with excellent long-term survival [4, 5]. The implanted ICD/CRT-D is usually removed at the time of HTPL operation. The generator is removed from the pocket and the portion of the leads dwelling within the atrium, ventricle, and lower part of the superior vena cava (SVC) are removed with the excision of the recipient’s heart. However, the leads within the central venous system can be difficult to remove without specialized extraction tools or techniques. This is particularly the case because ICD/CRT-D leads contain defibrillating coils. In this situation, tissue ingrowth and adhesion around coils act as binding points to the walls of large veins and hence complete endothelialization of lead materials into the vessel wall occurs more frequently [6]. As heart transplantation is never an elective procedure, it is not feasible to provide an additional team for lead removal, which might prolong the operation time. Evaluating whether lead materials have been completely removed requires fluoroscopic imaging of the procedural field, but surgical rooms are usually not equipped with an X-ray machine [7]. Therefore, complete removal of previously-implanted ICD/CRT-D during HTPL operation presents quite a clinical challenge, and, as a result, parts of the lead materials especially within the central venous system are often retained after the surgery [8].

Recently, indications for ICD/CRT-D have expanded, so the number of patients undergoing HTPL with implanted device is expected to increase. But the published literature regarding the frequency and clinical implications of retained ICD/CRT-D lead materials following HTPL is still lacking. The purpose of this study is to investigate the frequency of retained ICD/CRT-D lead materials in the central venous system after HTPL operation and to assess its clinical sequelae. In addition, we evaluated the proportion of patients who underwent magnetic resonance imaging (MRI) examination, which is one of the most important clinical issues in these patients due to the possible hazardous complications by retained lead materials [9].

Methods

Study patients

From November 1992 to December 2014, 474 consecutive patients underwent orthotopic HTPL at Asan Medical Center, Seoul, South Korea. Of these, 40 patients with ICD or CRT-D in situ at the time of HTPL were included in this study. Baseline demographic, historical, and procedural data were obtained from electronic medical records.

HTPL operation, postoperative management, and follow-up

Prior to January 1999, standard anastomosis was performed. However, since then, all HTPL procedures have involved bicaval anastomosis. The details of the HTPL surgical procedures have been described previously [10]. At the time of HTPL operation, only manual traction from the implant site was made to remove the leads, with no further attempt to extract using extraction tools such as a locking stylet, mechanical sheaths, or snares. After the operation, all recipients were managed according to protocols that included immunosuppression, infection prophylaxis, endomyocardial biopsy, and coronary angiography; the details of these procedures have been described previously [5]. All HTPL patients received regular clinical follow-up at Asan Medical Center.

Determination of retained lead materials

Chest radiographs were obtained in all patients after HTPL, and were reviewed independently by 2 cardiac electrophysiologists for the presence of retained ICD/CRT-D lead materials. Retained materials were classified based on the extent of remnants in the central venous system, and were classified into those localized to the SVC, those exist from SVC to left or right brachiocephalic vein, and those exist from SVC to left or right subclavian vein.

Assessment of clinical implications

We retrospectively reviewed the data of study patients with focus on the following topics to assess the clinical implications: (1) any subsequent procedure performed to remove the retained lead not removed during HTPL, (2) any infection related to the retained lead, (3) evidence of embolization or erosion of the retained lead materials, and (4) mortality. In addition, the frequency and sites of the MRI examination after HTPL regardless of whether it was performed at our institution or not, were also investigated.

Statistical analysis

SPSS version 18.0 (SPSS Inc., Chicago, IL) was used for all statistical analyses. Continuous variables are presented as means (±SD) or medians (quartiles, Q1–Q3) and categorical variables as numbers and percentages. Continuous variables were compared using the t-test or Mann–Whitney U test, and categorical variables were compared using χ2 statistics or Fisher's exact tests. A p value of <0.05 was considered statistically significant.

Ethics statement

The study protocol was reviewed and approved by the institutional review board of Seoul Asan Medical center (IRB No. 2015–0068), which waived the requirement for informed consent based on the retrospective nature of the study. None of the transplant donors were from a vulnerable population and all donors or next of kin provided written informed consent that was freely given.

Result

Clinical characteristics of the study population

Of 474 patients who underwent heart transplantation (HTPL) at our institution, 40 patients had ICD or CRT-D in situ at the time of HTPL. The mean age was 50±9.63 years, and 30 patients (75%) were male. The underlying cardiac diseases that led to the HTPL included dilated cardiomyopathy in 27 patients (67.5%), ischemic cardiomyopathy in 7 patients (17.5%), and other causes in 6 patients (15%). Among 40 patients, 35 patients (87.5%) had an ICD, whereas 5 patients (12.5%) had a CRT-D. The median duration between ICD/CRT-D implantation and HTPL was 30 months (17–62 months), and the median number of leads was 2 (1–4).

Nineteen out of 40 patients (47.5%) had retained ICD/CRT-D lead materials within the central venous system after HTPL, as confirmed by postoperative chest radiography. The clinical characteristics of those with and without retained lead materials are shown in Table 1 (Individual-level data are presented in the S1 Table). Patients with retained lead materials had a longer interval from device implantation to HTPL than those without (p = 0.01).

thumbnail
Table 1. Baseline characteristics of patients at the time of heart transplantation.

https://doi.org/10.1371/journal.pone.0176925.t001

Retained ICD/CRT-D lead materials

Table 2 describes the nature of the retained ICD/CRT-D lead materials and their extent (Individual-level data are presented in the S1 Table). All of the lead materials retained in the central venous system involved defibrillator coil, with or without lead. The lead materials were retained in SVC only in 7 patients (7/19, 36.8%), retained from SVC to left or right brachiocephalic vein in another 7 patients (7/19, 36.8%), and retained from SVC to left or right subclavian vein in 5 patients (5/19, 26.3%).

thumbnail
Table 2. Characteristics of the retained ICD/CRT-D lead materials following heart transplantation.

https://doi.org/10.1371/journal.pone.0176925.t002

Clinical sequelae of retained ICD/CRT-D lead materials

During the follow up period, no subsequent procedure was performed to remove the retained lead materials not removed during HTPL. Infection related to the retained lead and embolization or erosion of the retained lead materials did not occur. The overall mortality rate was 7.5% (3/40) during the study period. In one patient with retained lead materials, the cause of death was pneumonia-related sepsis. In two patients without retained lead materials, lung cancer and sepsis due to opportunistic infection were the cause of death, respectively.

MRI examination after HTPL

Of 21 patients without retained ICD/CRT-D lead materials, 7 patients were examined by MRI. The sites of the MRI scan were as follows: brain in 12, lumbar spine in 1, and liver in 1 patient. None of the patients with retained ICD/CRT-D lead materials underwent MRI during the follow-up period. Of 474 patients who underwent HTPL at our institution from January 1992 to July 2014, 168 underwent 308 MRI procedures during the follow-up period. The sites of the examinations are described in Table 3 (Individual-level data are presented in the S2 Table).

Discussion

In the present study, we observed that 47.5% (19/40) of patients who had ICD/CRT-D at the time of HTPL had retained ICD/CRT-D lead materials within their central venous system. No subsequent procedures were undertaken to remove retained lead materials, and there were no retained lead materials related complications such as lead fragment embolization, migration, erosion, and infection. All patients who had retained lead materials were asymptomatic and did not display adverse clinical sequelae. None of these patients were permitted to undergo MRI.

Despite advances in techniques and tools for transvenous lead extraction, the extraction of defibrillator leads including coils remains challenging, and is associated with minimal, but significant, mortality [11, 12]. The coils of ICD leads induce extensive growth of scar tissue, which surrounds and entraps the lead; hence, complex extraction procedures are required [13, 14]. A previous study showed that areas of adherence of the ICD lead were detected in the subclavian vein (78%), brachiocephalic vein (65%), SVC (66%), and heart (73%) [8]. Dwell time, passive fixation, and dual-coil lead design were independently associated with adherence [8]. Dual coil ICD is a risk factor for extraction difficulty, and is associated with SVC rupture [15]. In the present study, patients with retained ICD/CRT-D lead materials showed longer interval from device implantation to HTPL than those without. This finding is consistent with that in previous studies, wherein long dwell time was found to be associated with the need for advanced tools during ICD extraction [1618] and lead extraction failure during HTPL [6].

Non-functioning leads themselves may be associated with a small risk of complications such as infection, venous occlusion, lead migration, and skin erosion [1922]. In the present study, there were no long-term sequelae related to retained ICD/CRT-D lead materials, such as lead-related infection, thrombosis, vascular obstruction, and embolization. The absence of vascular obstruction can be attributed to the small number of ICD/CRT-D lead(s) (1 to 2) in each patient, because the risk of vascular obstruction is positively related to the number of retained leads. Moreover, the absence of ICD/CRT-D lead material embolization can be attributed to severe fibrosis around the SVC coil. These data are consistent with the results of previous studies on patients with and without HTPL [6, 23]. Considering that the presence of abandoned ICD/CRT-D leads is not associated with risks to patients, abandoning the retained lead materials may represent a reasonable strategy, particularly in institutions where laser sheaths are not available [23, 24]. However, recent data have suggested the feasibility and safety of lead extraction after HTPL using laser sheath and snares [7].

HTPL has evolved into a treatment modality for end-stage heart failure. Although this method is associated with excellent long-term survival, graft vasculopathy, infection, and malignancy are the leading causes of death [4]. Moreover, MRI is an indispensable clinical diagnostic tool for the diagnosis of cardiac and non-cardiac diseases due to the excellent spatial resolution and multiplanar 3-dimensional analysis, as well as the absence of risk of exposure to ionizing radiation and potentially nephrotoxic iodinated contrast agents [2528]. In our heart transplant cohort, 35.4% (168/474) of patients required MRI for various reasons. The most common examination sites included the brain (72.4%), spine (11.3%), and extremities (6.8%). During the follow-up duration of 30 months, there were no difference in clinical outcome between patients with and without retained ICD/CRT-D lead materials; however, it should be noted that a greater need for MRI in heart transplant recipients increases management issues related to these retained lead materials. A remnant nonfunctional lead is a contraindication for MRI due to the potential hazard for patients [29, 30]; this supports practicing routine extraction of retained lead materials in HTPL recipients. Although previous case series have suggested the safety of MRI in patients with retained leads [6, 9], experimental data have indicated the potential harm of MRI in patients with abandoned pacemaker leads [31, 32]. Nevertheless, there is a paucity of data on the hazards of MRI, particularly in cases with retained ICD/CRT-D lead materials.

Limitations

Our current study is partially and inherently limited in that it is a retrospective investigation. However, to our knowledge, this is the largest study on heart transplant recipients with retained ICD/CRT-D lead materials in Asia; for this reason, we believe that our study holds considerable value for clinicians. Second, the use of advanced extraction tools might continue to diminish this clinical dilemma of retained lead. However, considering the advanced disease stage of the recipients, the emergency nature of HTPL, and the potential risk in lead extraction procedure, it is possible that lead extraction with advanced extraction tools should not be performed simultaneously or after HTPL. Third, the overutilization of MRI in these cases has been suggested. Our current study reflects the real-world practice of a dedicated heart transplant care team. Considering the various post-HTPL complications, MRI examination is essential in caretaking of HTPL patients, especially to exclude central nervous system infection and to evaluate spine or joints disorders. However, due to the retrospective nature of our analysis, we could not assess the appropriateness of the MRI examination in all the patients.

Conclusion

Advanced heart failure patients require implantable cardiac devices such as ICD/CRT-D to improve their disease status and prognosis. Some of those patients, however, eventually require HTPL at the end stage of heart failure as a last-resort treatment. These implanted devices are usually removed at the time of HTPL operation. However, the defibrillating lead coils within the central venous system are difficult to be completely removed due to adhesion to host tissue. Therefore, ICD/CRT-D lead materials are frequently retained in the central venous system of HTPL patients. Although there were no clinical sequelae related to retained ICD/CRT-D lead materials in our study patients, the greater need for MRI in HTPL patients warrant further studies geared toward recommending routine extraction of retained lead materials for the safety of MRI procedure in this group of patients.

Supporting information

S1 Table. Individual-level data underlying results presented in the Tables 1 and 2.

https://doi.org/10.1371/journal.pone.0176925.s001

(XLSX)

S2 Table. Individual-level data underlying results presented in the Table 3.

https://doi.org/10.1371/journal.pone.0176925.s002

(XLSX)

Acknowledgments

The authors are grateful to the biostatistician Seunghee Baek, Ph.D, working at Department of Clinical Epidemiology and Biostatistics in Asan Medical Center for statistical assistance.

Author Contributions

  1. Conceptualization: JK JJK.
  2. Data curation: JK JH JHC HIC MSK.
  3. Formal analysis: JK JH JHC.
  4. Investigation: JK JH JHC JJK MSK SHJ JWL.
  5. Methodology: JK JH JHC JJK.
  6. Project administration: JJK JWL GBN KJC YHK.
  7. Resources: JK JJK SHJ JWL GBN KJC YHK.
  8. Supervision: JK JJK JWL GBN KJC YHK.
  9. Validation: JK JJK SHJ.
  10. Visualization: JK JH JHC HIC MSK.
  11. Writing – original draft: JK JH JJK.
  12. Writing – review & editing: JK JH JJK.

References

  1. 1. Investigators TAvIDA. A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. The Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. N Engl J Med. 1997; 337: 1576–1583. pmid:9411221
  2. 2. Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002; 346: 877–883. pmid:11907286
  3. 3. Bardy GH, Lee KL, Mark DB, Poole JE, Packer DL, Boineau R, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med. 2005; 352: 225–237. pmid:15659722
  4. 4. Stehlik J, Edwards LB, Kucheryavaya AY, Benden C, Christie JD, Dipchand AI, et al. The Registry of the International Society for Heart and Lung Transplantation: 29th official adult heart transplant report—2012. J Heart Lung Transplant. 2012; 31: 1052–1064. pmid:22975095
  5. 5. Jung SH, Kim JJ, Choo SJ, Yun TJ, Chung CH, Lee JW. Long-term mortality in adult orthotopic heart transplant recipients. J Korean Med Sci. 2011; 26: 599–603. pmid:21532848
  6. 6. Martin A, Voss J, Shannon D, Ruygrok P, Lever N. Frequency and sequelae of retained implanted cardiac device material post heart transplantation. Pacing Clin Electrophysiol. 2014; 37: 242–248. pmid:24428516
  7. 7. Kusmierski K, Przybylski A, Oreziak A, Sobieszczanska-Malek M, Kolsut P, Rozanski J. Post heart transplant extraction of the abandoned fragments of pacing and defibrillation leads: proposed management algorithm. Kardiol Pol. 2013; 71: 159–163. pmid:23575709
  8. 8. Segreti L, Di Cori A, Soldati E, Zucchelli G, Viani S, Paperini L, et al. Major predictors of fibrous adherences in transvenous implantable cardioverter-defibrillator lead extraction. Heart Rhythm. 2014; 11: 2196–2201. pmid:25111324
  9. 9. Higgins JV, Gard JJ, Sheldon SH, Espinosa RE, Wood CP, Felmlee JP, et al. Safety and outcomes of magnetic resonance imaging in patients with abandoned pacemaker and defibrillator leads. Pacing Clin Electrophysiol. 2014; 37: 1284–1290. pmid:24809591
  10. 10. Choo SJ, Kim JJ, Kim SP, Lee JW, Wan RS, Park NH, et al. Heart transplantation: a retrospective analysis of the long-term results. Yonsei Med J. 2004; 45: 1173–1180. pmid:15627315
  11. 11. Brunner MP, Cronin EM, Duarte VE, Yu C, Tarakji KG, Martin DO, et al. Clinical predictors of adverse patient outcomes in an experience of more than 5000 chronic endovascular pacemaker and defibrillator lead extractions. Heart Rhythm. 2014; 11: 799–805. pmid:24444444
  12. 12. Gomes S, Cranney G, Bennett M, Li A, Giles R. Twenty-year experience of transvenous lead extraction at a single centre. Europace. 2014; 16: 1350–1355. pmid:24554523
  13. 13. Kennergren C, Bjurman C, Wiklund R, Gabel J. A single-centre experience of over one thousand lead extractions. Europace. 2009; 11: 612–617. pmid:19329797
  14. 14. Epstein AE, Kay GN, Plumb VJ, Dailey SM, Anderson PG. Gross and microscopic pathological changes associated with nonthoracotomy implantable defibrillator leads. Circulation. 1998; 98: 1517–1524. pmid:9769305
  15. 15. Epstein LM, Love CJ, Wilkoff BL, Chung MK, Hackler JW, Bongiorni MG, et al. Superior vena cava defibrillator coils make transvenous lead extraction more challenging and riskier. J Am Coll Cardiol. 2013; 61: 987–989. pmid:23449433
  16. 16. Bracke F, Meijer A, Van Gelder B. Extraction of pacemaker and implantable cardioverter defibrillator leads: patient and lead characteristics in relation to the requirement of extraction tools. Pacing Clin Electrophysiol. 2002; 25: 1037–1040. pmid:12164443
  17. 17. Jones SOt, Eckart RE, Albert CM, Epstein LM. Large, single-center, single-operator experience with transvenous lead extraction: outcomes and changing indications. Heart Rhythm. 2008; 5: 520–525. pmid:18325849
  18. 18. Roux JF, Page P, Dubuc M, Thibault B, Guerra PG, Macle L, et al. Laser lead extraction: predictors of success and complications. Pacing Clin Electrophysiol. 2007; 30: 214–220. pmid:17338718
  19. 19. Furman S, Behrens M, Andrews C, Klementowicz P. Retained pacemaker leads. J Thorac Cardiovasc Surg. 1987; 94: 770–772. pmid:3669704
  20. 20. Zerbe F, Ponizynski A, Dyszkiewicz W, Ziemianski A, Dziegielewski T, Krug H. Functionless retained pacing leads in the cardiovascular system. A complication of pacemaker treatment. Br Heart J. 1985; 54: 76–79. pmid:3893487
  21. 21. Suga C, Hayes DL, Hyberger LK, Lloyd MA. Is there an adverse outcome from abandoned pacing leads? J Interv Card Electrophysiol. 2000; 4: 493–499. pmid:11046188
  22. 22. Bohm A, Pinter A, Duray G, Lehoczky D, Dudas G, Tomcsanyi I, et al. Complications due to abandoned noninfected pacemaker leads. Pacing Clin Electrophysiol. 2001; 24: 1721–1724. pmid:11817804
  23. 23. Amelot M, Foucault A, Scanu P, Gomes S, Champ-Rigot L, Pellissier A, et al. Comparison of outcomes in patients with abandoned versus extracted implantable cardioverter defibrillator leads. Arch Cardiovasc Dis. 2011; 104: 572–577. pmid:22117909
  24. 24. Bode F, Himmel F, Reppel M, Mortensen K, Schunkert H, Wiegand UK. Should all dysfunctional high-voltage leads be extracted? Results of a single-centre long-term registry. Europace. 2012; 14: 1764–1770. pmid:22753865
  25. 25. Borhani AA, Hosseinzadeh K, Almusa O, Furlan A, Nalesnik M. Imaging of posttransplantation lymphoproliferative disorder after solid organ transplantation. Radiographics. 2009; 29: 981–1000; discussion 1000–1002. pmid:19605652
  26. 26. Chitsaz S, Bagheri J, Mandegar MH, Rayatzadeh H, Razavi J, Azadi L. Extensive sino-orbital zygomycosis after heart transplantation: a case report. Transplant Proc. 2009; 41: 2927–2929. pmid:19765477
  27. 27. Drachman BM, DeNofrio D, Acker MA, Galetta S, Loh E. Cortical blindness secondary to cyclosporine after orthotopic heart transplantation: a case report and review of the literature. J Heart Lung Transplant. 1996; 15: 1158–1164. pmid:8956125
  28. 28. Taillandier J, Alemanni M, Cerrina J, Le Roy Ladurie F, Dartevelle P. Aspergillus osteomyelitis after heart-lung transplantation. J Heart Lung Transplant. 1997; 16: 436–438. pmid:9154954
  29. 29. Levine GN, Gomes AS, Arai AE, Bluemke DA, Flamm SD, Kanal E, et al. Safety of magnetic resonance imaging in patients with cardiovascular devices: an American Heart Association scientific statement from the Committee on Diagnostic and Interventional Cardiac Catheterization, Council on Clinical Cardiology, and the Council on Cardiovascular Radiology and Intervention: endorsed by the American College of Cardiology Foundation, the North American Society for Cardiac Imaging, and the Society for Cardiovascular Magnetic Resonance. Circulation. 2007; 116: 2878–2891. pmid:18025533
  30. 30. Nazarian S, Hansford R, Roguin A, Goldsher D, Zviman MM, Lardo AC, et al. A prospective evaluation of a protocol for magnetic resonance imaging of patients with implanted cardiac devices. Ann Intern Med. 2011; 155: 415–424. pmid:21969340
  31. 31. Langman DA, Goldberg IB, Finn JP, Ennis DB. Pacemaker lead tip heating in abandoned and pacemaker-attached leads at 1.5 Tesla MRI. J Magn Reson Imaging. 2011; 33: 426–431. pmid:21274985
  32. 32. Mattei E, Gentili G, Censi F, Triventi M, Calcagnini G. Impact of capped and uncapped abandoned leads on the heating of an MR-conditional pacemaker implant. Magn Reson Med. 2015; 73: 390–400. pmid:24436030