Simultaneous Quantitative MRI Mapping of T1, T2* and Magnetic Susceptibility with Multi-Echo MP2RAGE

  • Riccardo Metere ,

    metere@cbs.mpg.de (RM); moeller@cbs.mpg.de (HEM)

    Affiliation Max Planck Institute for Human Cognitive and Brain Sciences, Leipzig, Germany

  • Tobias Kober,

    Affiliations Advanced Clinical Imaging Technology, Siemens Healthcare HC CMEA SUI DI BM PI, Lausanne, Switzerland, Department of Radiology, University Hospital (CHUV), Lausanne, Switzerland, LTS5, École Polytechnique Fédérale de Lausanne, Lausanne, Switzerland

  • Harald E. Möller ,

    metere@cbs.mpg.de (RM); moeller@cbs.mpg.de (HEM)

    ‡ HEM and AS are joint senior authors on this work.

    Affiliation Max Planck Institute for Human Cognitive and Brain Sciences, Leipzig, Germany

  • Andreas Schäfer

    ‡ HEM and AS are joint senior authors on this work.

    Affiliations Max Planck Institute for Human Cognitive and Brain Sciences, Leipzig, Germany, Diagnostic Imaging—Magnetic Resonance—Research & Development, Siemens Healthcare GmbH, Erlangen, Germany

Abstract

The knowledge of relaxation times is essential for understanding the biophysical mechanisms underlying contrast in magnetic resonance imaging. Quantitative experiments, while offering major advantages in terms of reproducibility, may benefit from simultaneous acquisitions. In this work, we demonstrate the possibility of simultaneously recording relaxation-time and susceptibility maps with a prototype Multi-Echo (ME) Magnetization-Prepared 2 RApid Gradient Echoes (MP2RAGE) sequence. T1 maps can be obtained using the MP2RAGE sequence, which is relatively insensitive to inhomogeneities of the radio-frequency transmit field, . As an extension, multiple gradient echoes can be acquired in each of the MP2RAGE readout blocks, which permits the calculation of and susceptibility maps. We used computer simulations to explore the effects of the parameters on the precision and accuracy of the mapping. In vivo parameter maps up to 0.6 mm nominal resolution were acquired at 7 T in 19 healthy volunteers. Voxel-by-voxel correlations and the test-retest reproducibility were used to assess the reliability of the results. When using optimized paramenters, T1 maps obtained with ME-MP2RAGE and standard MP2RAGE showed excellent agreement for the whole range of values found in brain tissues. Simultaneously obtained and susceptibility maps were of comparable quality as Fast Low-Angle SHot (FLASH) results. The acquisition times were more favorable for the ME-MP2RAGE (≈ 19 min) sequence as opposed to the sum of MP2RAGE (≈ 12 min) and FLASH (≈ 10 min) acquisitions. Without relevant sacrifice in accuracy, precision or flexibility, the multi-echo version may yield advantages in terms of reduced acquisition time and intrinsic co-registration, provided that an appropriate optimization of the acquisition parameters is performed.

1 Introduction

Quantitative mapping of Magnetic Resonance (MR) relaxation times has become a useful tool in brain research as well as for clinical applications due to the possibility of directly comparing results across subjects and sites. The knowledge of relaxation times is further essential for understanding the biophysical mechanisms underlying image contrast. Recently, the correlation between relaxation times and brain tissue composition has been highlighted [1, 2]. For example, it has been shown that the effective transverse relaxation time, , is influenced by iron and myelin content, whereas the longitudinal relaxation time, T1, depends on myelin but only little on iron content [2, 3]. Three-Dimensional (3D) T1 maps are also frequently used for differentiating brain tissue types, especially for White Matter (WM), Gray Matter (GM), and Cerebro-Spinal Fluid (CSF) segmentation [36].

T1 maps can be obtained, for example, with ‘Fast Low-Angle SHot’ (FLASH) [7, 8] employing different flip angles [9] or with inversion-recovery sequences [10]. If the recovery curve is sampled with sufficient density, separation of multiple water compartments in the brain based on their T1 values has been achieved by relaxographic techniques [11, 12]. Recently, the ‘Magnetization-Prepared 2 RApid Gradient Echoes’ (MP2RAGE) sequence has been proposed for an efficient 3D mapping of T1 [5, 13]. It consists of two Gradient-Recalled Echo (GRE) image volumes acquired within a single repetition at two different inversion times, TI,1 and TI,2. Assuming that the recovery can be characterized by a single exponential, accurate T1 estimates are obtained as long as the acquisition parameters are properly chosen, allowing for sufficient insensitivity to inhomogeneities of the Radio-Frequency (RF) transmit magnetic field amplitude, .

Mapping of requires a gradient acquisition scheme with multiple echoes sampled at different times, TE,i. Usually, the FLASH sequence with multi-echo (ME) readout is employed for -related relaxometry, because of its relatively high robustness against inhomogeneities of the main magnetic field amplitude, B0.

In recent years, Quantitative Susceptibility Mapping (QSM) has been developed to measure another intrinsic property of tissues, the bulk magnetic susceptibility, χ [1419]. It employs maps of the spatial variation of B0 extracted from the signal phase of a GRE [20]. Molecular anisotropy of the susceptibility and orientation of micro-structural components have been shown to influence the phase contrast as well [18, 2127]. Furthermore, QSM has led to the possibility of differentiating between diamagnetic myelin and paramagnetic iron deposition [17, 2831] that is already exploited in brain research and clinical applications.

As a typical drawback, many techniques for quantitative mapping of tissue parameters require longer scan times than conventional T1-, - or susceptibility-weighted MR Imaging (MRI). This imposes limitations in clinical studies where the available scan time is often restricted. Hence, more versatile sequence implementations permitting the simultaneous acquisition of image data with different contrasts are of interest. Previous work has shown that collecting multiple echoes in conventional ‘Magnetization-Prepared RApid Gradient Echo’ (MP-RAGE) imaging [32] may be helpful in brain morphometry studies [33, 34]. In this work, we introduce an ME version of the MP2RAGE sequence that permits simultaneous 3D mapping of T1, , and χ. Computer simulations and experimental verification in healthy human brains at 7T demonstrate that all parameters can be obtained with similar accuracy as achieved using separately acquired MP2RAGE and ME-FLASH scans.

2 Methods

In order to investigate whether the ME-MP2RAGE sequence permits robust mapping of T1, and χ, we used a two–step approach: (i) the effects of the acquisition parameters on the accuracy and precision of the maps were investigated by simulations and experiments, and (ii) after finding a suitable set of parameters, the reproducibility of the maps was evaluated in comparison to standard MP2RAGE or ME-FLASH results.

2.1 Description of the Pulse Sequence

The ME-MP2RAGE pulse sequence, shown in Fig 1, is obtained from MP2RAGE by acquiring multiple gradient echoes for each phase-encoding step. The signal expression for MP2RAGE derived in [5] is also valid for ME-MP2RAGE. Briefly, the signals (at any echo time, TE,i) can be written as: (1) and (2) where M0 is the equilibrium magnetization, E1 ≡ exp(−TR,GRE/T1), EA ≡ exp(−TA/T1), EC ≡ exp(−TC/T1), and n is the number of k-space lines in one GRE block. The intervals TR,GRE, TA, and TC, as well as the flip angles α1 and α2 are defined in Fig 1; η ≡ (1/2)[1 − Mz(0+)/Mz(0)] is the inversion efficiency of the adiabatic pulse, where Mz(0) and Mz(0+) are the longitudinal magnetizations immediately before and after the application of the pulse, respectively [35]. is the steady-state longitudinal magnetization, which is obtained by solving the Bloch equations with appropriate boundary conditions (see Eq A1.4 in [5]). Finally, all scanner-dependent parameters required for converting magnetization into signal voltage are lumped together in the scaling constant ξ.

thumbnail
Fig 1. Schematic diagram of the ME-MP2RAGE sequence.

After an adiabatic inversion pulse (INV), two GRE readout blocks are collected with excitation pulse flip angles, α1 and α2. Both GRE blocks consist of n acquisitions of k-space lines, each of duration TR,GRE, stepping linearly through the second phase-encoding direction (as in standard MP2RAGE). To each TR,seq corresponds a k-space line acquisition for the first phase-encoding direction. The center of k-space is acquired at times TI,1 and TI,2. Examples of the additional mono-polar gradient lobes of the ME readouts are indicated by blue color. The sequence repetition time, TR,seq, is defined as the time between two successive inversion pulses. The total acquisition time is thus defined by TR,seq multiplied by the number of steps in the second phase-encoding direction. TA, TB, and TC denote, respectively, the durations from the initial inversion pulse to the onset of the first GRE block, between the two GRE blocks, and from the end of the second GRE block to the next inversion pulse. Note that the first and the second phase-encoding direction may be interchanged.

https://doi.org/10.1371/journal.pone.0169265.g001

The (complex) signals from the two inversion-contrast image volumes are combined by computing (3) where S* denotes the complex conjugate of S and returns the real part of x. The expression for ρ can be found analytically, and it is used to invert the equation for T1 numerically for a specific range of T1 values to calculate the maps. Several parameter combinations can yield equivalent T1 maps, provided that the appropriate expression for ρ (or a suitable approximation) is used, and specific requirements on such parameters are met, as detailed in [5].

The same equations can also be used to show that the (absolute) signal level of the image volume recorded at TI,2, being acquired later on the recovery curve, is generally larger than the one recorded at TI,1, in agreement with experimental observations. Hence, the second inversion image volume was employed for obtaining and χ estimates throughout this work. Instead, the T1 estimates were always obtained from the inversion contrast recorded with the first echo.

2.2 Simulation Studies

In order to evaluate the sensitivity of the ME-MP2RAGE signal to variations, we used simulations based on the Bloch equations. In particular, we plotted the signal expression ρ as a function of T1 with additional consideration of ±20% and ±40% variations in (through the flip angles). The convergence of the varied curves to the original signal expression indicates insensitivity to inhomogeneity of the field.

Because of the well-known inhomogeneity at 7 T, the nominal flip angle, αnom, may differ significantly from the effective flip angle, αeff, generated experimentally in a tissue. For more realistic comparisons, the flip angle values of the simulations and the experiments were matched using the efficiency factor ηααnom / αeff, where αeff was determined experimentally using maps.

The (ME-)MP2RAGE sequence yields T1 maps through an inversion-recovery acquisition where a mono-exponential recovery constant is estimated from two support points, TI,(1,2), by inverting the signal expression for T1. Similarly, maps are obtained via a fitting procedure, where a mono-exponential decay constant is interpolated on ME measurements.

It is possible to investigate, through Monte-Carlo-like simulations, with the addition of noise [36], how the Signal-to-Noise Ratio (SNR) and the number and distribution of the support points affect the quality of the relaxation constant estimates. In particular, nT = 100 different relaxation time values from the intervals T1/ ms ∈ [500, 3500] and were assumed, and up to N = 20000 simulations of the obtained signal levels were performed for each value with random variation of the noise contribution to compute standard deviations (SD), σT, of the relaxation-time estimates. Based on typical experimental results, the SNR was set to 25, 50, or 100, with TI,1/ ms ∈ [500, 1500] and TI,2/ ms ∈ [1500, 3500] (constrained to TI,2TI,1 ≥ 1000 ms) in simulations of T1 mapping, and with TE,i/ ms ∈ [2, 30] and nE ∈ [3, 5] in simulations of mapping (nE is the number of echoes). The mean and SD of the σT values for the full range of relaxation times, μσ and σσ, were then used to assess the robustness of the estimation.

Unless otherwise noted, simulations, data analyses (see below), and the visualization of the results were performed using the “Scientific Python” ecosystem [3739].

2.3 Acquisition Parameter Considerations

This work targets at acquiring multiple quantitative maps of the full brain at sub-millimeter resolution with an acquisition time similar to current clinical practice, which implies that only a restricted set of acquisition parameter values are desirable or accessible. In order to achieve an advantage from the simultaneous mapping approach, the following conditions should be met: (i) the T1 map obtained with ME-MP2RAGE should be as accurate as a corresponding map acquired with MP2RAGE, and (ii) the acquisition time required for ME-MP2RAGE should be shorter than the sum of the acquisition times of corresponding MP2RAGE and ME-FLASH scans.

We specifically considered the following influences from the most relevant acquisition parameters: (a) lower values of α1,2 translate into improved insensitivity but also lower SNR, which directly affects the accuracy and precision; to maximize SNR, α2 was set to the Ernst angle, assuming T1 ≈ 1.6 s as the mean of the expected values for WM (≈ 1.2 s) and GM (≈ 2.0 s) [10]; (b) the number of k-space lines per GRE block affects the block duration and limits the range of possible inversion times; n increases with matrix size and Field Of View (FOV) and may be decreased through Parallel Acquisition Techniques (PAT); (c) the choice of TI,(1,2) has an effect on the sensitivity and the accessible value range for T1 estimates as well as its precision; (d) the repetition time of the sequence, TR,seq, defines the total acquisition time and also has an (indirect) impact on the sensitivity to variation; (e) a short first TE is desirable to maximize the SNR for T1 mapping; however, a relatively long (later) TE in the order of 15 ms is useful for χ mapping (at the cost of longer TR,GRE with concomitant constrains for matrix size and FOV) as it increases the Contrast-to-Noise Ratio (CNR) of the phase images and avoids possible anisotropy bias in the WM caused by non-linear phase evolution at short TE [25, 26]; finally, acquiring more echoes improves mapping but might require a larger bandwidth, Δν.

The relatively large number of parameters (and, hence, optimization criteria) and their complex relations makes it impossible to define an unbiased cost function that simultaneously maximizes the accuracy and precision of all maps.

For this reason, a manual optimization approach was adopted: initially, the FOV, matrix, and PAT parameters were defined to determine n; then, we accommodated the maximum number of echoes within a fixed TR,GRE, long enough for reliable and χ mapping; lastly, we adjusted all other parameters to achieve sufficient insensitivity, and T1 mapping accuracy with the minimum possible TR,seq.

For the MP2RAGE acquisitions, a similar optimization approach was used, since the values suggested by [5] were not optimal for the desired resolution and FOV. In this case, the shorter TR,GRE (due to nE = 1) allowed to increase n, which was exploited to decrease the acquisition time. This was achieved both by shortening TR,seq and by switching the first and second phase-encoding directions, so that the main (TR,seq) loop is performed along the GRAPPA-accelerated direction. However, in this case, the obtained acquisition parameters cause a systematically lower and potentially ambiguous T1 estimate in the CSF value range, which is not relevant in most applications. Note that this feature is intrinsic to the MP2RAGE-based T1 estimation procedure (see also: Fig 2) and the acquisition parameters should be adjusted to avoid this issue in the interval of interest.

thumbnail
Fig 2. Estimated T1 values as a function of (a) the ME-MP2RAGE and (b) the MP2RAGE signal intensity parameter ρ.

The green lines correspond to the effective acquisition parameters while red/blue lighter/darker lines indicate, respectively, ±20% and ±40% offsets of . The flip angle values are adjusted for the accuracy factor ηα. Note that the estimated T1 is limited to an upper value of approximately 3 s for the acquisition parameters that were used for MP2RAGE, which would result in T1 values exceeding this limit (e.g., in CSF) to be underestimated.

https://doi.org/10.1371/journal.pone.0169265.g002

Exploratory measurements (summarized in the Supporting Information) with a large coverage of different acquisition parameters were used to infer their effects, in relation to the corresponding simulations. The manually chosen parameters were then tested on a cohort of subjects to provide additional experimental evidence (allowing for group statistics).

2.4 MRI Experiments

Quantitative 3D MRI of the brain was performed in 19 healthy subjects (9 females, 23–31 years) on a Magnetom 7T scanner (Siemens Healthcare, Erlangen, Germany).

This study was approved by the Ethics Committee at the Medical Faculty of the University of Leipzig (application n.: 273-14-25082014). All participants had given informed written consent prior to the examination.

The ME-MP2RAGE pulse sequence was provided by Siemens. The acquisitions were performed using a monopolar readout (to avoid potential effects related to a constant offset in the gradients system), Generalized auto-calibRating Partially Parallel Acquisitions (GRAPPA) along the first phase-encoding (PE) direction, and 6/8 partial Fourier factors (both for the first and the second PE direction). The in vivo experiments were divided into two studies.

In Study 1, exploratory measurements were performed with a circularly-polarized transmit/24-channel receive head coil and a 32-channel receive version of the same coil design (Nova Medical, Wilmington, MA, USA) to evaluate the following range of acquisition parameters: FOV between 224×224×145.6 mm3 and 192×168×112 mm3 with nominal isotropic resolutions in the 0.6–0.9 mm range; GRAPPA acceleration factors, f = 2–3; TR,seq = 4700–8000 ms; TI,(1,2) = 750–1100, 2750–3500 ms; α1,2 = 2–5, 3–10 deg; TE = 2.32–23.05 ms; nE = 3–6; and Δν = 280–500 Hz/px yielding total acquisition times, Tacq = 5:55–22:25 min. MP2RAGE and ME-FLASH data were also recorded for the validation of the results with acquisition parameters similar to previously published values [46] or in-house standard settings. The FOV, nominal resolution, and GRAPPA acceleration were matched to the specific ME-MP2RAGE acquisition; further parameters were: for MP2RAGE TR,seq = 5000–8000 ms; α1,2 = 4–5, 3–10 deg; TI,(1,2) = 800, 2400 ms; TE = 2.15 ms; Δν = 280 Hz/px; Tacq = 9:42 min; and for ME-FLASH TR = 31–35 ms; α = 10–11 ms; TE = 2.94–29.59 ms; nE = 5–6; Δν = 280–500 Hz/px; Tacq = 6:04–12:40 min.

From the exploratory results, a preferred set of ME-MP2RAGE parameters (see Table 1) was derived to test, in Study 2, the intra-subject reproducibility with a FOV of 192 × 144 × 134.4 mm3 (axial orientation, first PE direction from right to left, second PE with 14% oversampling) and an acquisition matrix of 320 × 280 × 224 (i.e., 0.6 mm isotropic nominal resolution).

thumbnail
Table 1. Details of the manually optimized acquisition parameters that were used for the simulations in Fig 2 and for the acquisitions of Study 2 with a nominal spatial resolution of 0.6 mm (isotropic).

https://doi.org/10.1371/journal.pone.0169265.t001

These scans were recorded with the 32-channel coil on 7 of the 19 subjects. To obtain reference data, further measurements were made in the same sessions with MP2RAGE and ME-FLASH (parameters included in Table 1) using the same FOV, matrix size, orientation, GRAPPA acceleration, and partial Fourier settings.

A typical session, thus, consisted of two MP2RAGE, two ME-FLASH and two ME-MP2RAGE acquisitions acquired in random order without repositioning.

In some sessions, additional maps were acquired with 3 mm isotropic nominal resolution employing an in-house modification (based on complex instead of magnitude images) of the Actual Flip-angle Imaging (AFI) technique [40]. Due to time restrictions, these maps were acquired in place of an ME-FLASH or MP2RAGE scan; thus, in such cases, ME-FLASH or MP2RAGE reproducibility results are not available.

2.5 Image Processing

T1 maps were reconstructed on-line by algorithms integrated in the Image Calculation Environment (ICE) provided by the vendor. maps were reconstructed off-line by log-linear least-squares fitting to the signal magnitudes, using the standard polynomial fit approach based on singular value decomposition. The QSM analysis was implemented in C++ using the ODIN library [41]. Susceptibility mapping was based on the phase of the echo acquired at the longest TE for ME-MP2RAGE (TE = 14.77 ms), and a similar value (TE = 15.00 ms) was used for the ME-FLASH reference scans in order to improve comparability. The phase images were unwrapped using a Fourier method [42], which offers the advantage of describing singularities as continuous functions, and high-pass filtered using the Sophisticated Harmonic Artifact Reduction for Phase data (SHARP) approach [43]. Filtered phase images were converted into units of ppb (division by 10−9 γ TE B0; γ is the gyromagnetic ratio of the proton), and quantitative maps were calculated with the Superfast Dipole Inversion (SDI) method [44].

Image segmentation and registration of brain tissues were accomplished using the FMRIB Software Library (FSL), Ver. 5.0 [45]. For each subject, a brain mask was generated (from the GRE readout at TI,2 for MP2RAGE or ME-MP2RAGE) using the Brain Extraction Tool (BET2) (with the ‘robust brain center estimation flag’ [-R] active), a Gaussian filter on BET’s result (for smoother edges) with SD σ = 0.5 px, and thresholding for values above 5% of the maximum. The extracted masks were furthermore shrunk (to avoid border effects) by the application of a binary erosion filter with a “spherical” kernel of diameter 1 px, iterated five times to remove the outer layer of CSF. Maps were then linearly co-registered using the FMRIB Linear Image Registration Tool (FLIRT) with rigid-body transformations, the correlation ratio metric and the previously obtained mask as weights. Once the volumes were registered, the non-brain tissues were masked out.

2.6 Statistical Analyses

Results obtained with ME-MP2RAGE were compared against MP2RAGE as reference for T1 maps and against ME-FLASH as reference for and χ maps. The equivalence of the maps was evaluated voxel-by-voxel with two-dimensional (2D) correlation histograms and difference images. For the quantification of correlations and mutual consistency of the maps, the following parameters were investigated:

  1. The squared Pearson’s correlation coefficient, r2, as an indication of how well the test measurements reproduce the reference measurements (neglecting the noise of the reference measurements).
  2. The means and SDs of the image volumes’ difference, as an estimate of respectively accuracy and precision, (4) as well as their absolute difference, (5) where xi and yi are defined by the map intensity of a voxel in the test data and in the reference data, respectively, and N is the number of voxels after the masking step. Note that μ|D| and σ|D| are equivalent (except for a factor ) to the average and SD of the Euclidean distance from the identity line in the 2D histogram, and reflect the combined effects of deviation from the identity line and (random) spreading.

While r2 is often used in correlation studies, it only reflects potential deviations from a linear relationship, but it is not sensitive to the slope of the linear component of the relationship. It is thus of limited relevance if the measured values should be, ideally, identical as in the current case. As r2 is dimensionless, it more directly permits inter-modality comparisons. The accuracy and the precision of the measurement can be estimated by μD and σD, respectively (in units of the quantity being measured). Additionally, if the accuracy is good, that is when μD ≈ 0, then μ|D| (along with σ|D| as its error) may be used as an estimate of the overall reproducibility. A sensible reference for these parameters is the size of the value range of the obtained maps. Note that these parameters are quite sensitive to outliers and may vary considerably depending on the efficacy of some of the procedures used in this work (e.g., the brain tissue masking, the fits used to calculate the maps, the registration, the considered value range, etc.).

To investigate the effects of potential errors arising from the registration procedure, we performed the voxel-by-voxel analysis described earlier on a given map in comparison to the same map after application of a linear rigid transformation. More specifically, translations in axial direction were examined for voxel offset values between 0.0 and 1.0 px; rotations about the axial direction and centered in the geometrical center of the FOV were examined for rotation angles between 0.0 and 1.0°; and rotations followed by translations were examined within the same value ranges. A complementary estimate of the effects of the registration step was obtained by applying the usual voxel-by-voxel analysis to “self-registered” maps. More precisely, reference maps were compared with the same maps registered back onto themselves after the application of a significant rigid transformation (in this study, a 10° rotation). These analyses were performed for T1 maps obtained with MP2RAGE and for and χ maps obtained with ME-FLASH.

Because of the relatively complex reconstruction technique required for QSM, it is not possible to use an equivalent simulation approach as for relaxometry to study the impact of the SNR. Nevertheless, it is possible to investigate the effects of SNR on the final QSM result by using an approach similar to what was done to investigate the effects of mis-registration. Particularly, we reconstructed several QSM maps obtained from the phase of the ME-FLASH complex images with the addition of Gaussian noise (with increasing SD in a range of 0–30 arb.units, corresponding to up to ≈ 100% decrease in SNR). The effect of the SNR on the χ maps was evaluated using the same voxel-by-voxel comparison analyses described earlier.

Additionally, we performed an analysis on a Region-Of-Interest (ROI) basis, where we calculated the group average parametric values across the different brain areas: frontal lobe (Fro.), temporal lobe (Tem.), parietal lone (Par.), occipital lobe (Occ.), insula (Ins.), putamen (Put.), caudate (Caud.), thalamus (Tha.), cerebellum (Cereb.) as defined by the MNI Structural Atlas. Additionally, white matter (WM) from the Harvard-Oxford Subcortical Structural Atlas was considered. Volumes from the second inversion images of the ME-MP2RAGE data were registered non-linearly to the MNI152 1 mm nominal resolution template (using default options) and such transformations were then used to bring the maps to the same space. Then, for each area (defined by the probabilistic masks thresholded for above 80%, which may incidentally leave out some portions of the WM), all the subjects from Study 2 were group-averaged, and the group SD was also calculated.

3 Results

3.1 Simulations

The results obtained by relaxometry simulations (an overview of which is given in S3 and S4 Tables with additional details reported in S1 and S2 Figs) indicate that, for a given SNR, the number of echoes and their distribution is crucial for mapping, while the specific choice of inversion times (for the range examined in our study) is less critical for T1 mapping. As a rule of thumb, choices of inversion times according to TI,1/2 < T1 < 2TI,2 and echo times according to (where max indicates the longest TE) yield robust estimates of T1 and , respectively, unless the SNR is insufficient. Detailed results for the specific settings of TI and TE that were selected for Study 2 are shown in Fig 3. They indicate that the accuracy and precision (represented graphically by the distance from the identity line and by the size of the error bars, respectively) depend on the exact value of the relaxation times, and at this level of SNR are comparable between the test (ME-MP2RAGE) and the reference (MP2RAGE for T1, ME-FLASH for ) sequences. Additional information on the SNR dependence can be found in the support material indicated above.

thumbnail
Fig 3. Simulations for the expected accuracy and precision ofT1 and maps at SNR = 50.

Rows refer to T1 (a, b) and (c, d) simulations, while columns indicate the choice of the simulation parameters reflecting the acquisition: ME-MP2RAGE (a, c), MP2RAGE (b) or ME-FLASH (d). Each panel contain a plot of the estimated relaxation time as a function of the exact value, with the error bars indicating the standard deviations, σT, for N = 20000 simulations. The distance from the identity line indicates the expected accuracy, while the size of the error bars reflect the expected precision. For T1 / exponential recovery simulations, the range 0.5–3.5 s was probed, and the exact acquisition parameters simulated were TI,(1,2) = 800, 2400 ms for ME-MP2RAGE, and TI,(1,2) = 750, 2900 ms for MP2RAGE. For T2 / exponential decay simulations, the range 2–60 ms was probed, and the exact acquisition parameters simulated were nE = 4, TE,1 = 2.5 ms, ΔTE ≈ 4.2 ms for ME-MP2RAGE, and nE = 5, TE,1 = 3.0 ms, ΔTE = 6.0 ms for ME-FLASH.

https://doi.org/10.1371/journal.pone.0169265.g003

Fig 4 shows combined histograms of the flip-angle accuracy factor measured in the brain with the 32-channel coil. The mean plus/minus SD were 0.8 ± 0.2. Hence, flip angle values used in the simulations were divided by 0.8 and rounded to the nearest integer to obtain corresponding settings for αnom in the experimental protocols.

thumbnail
Fig 4. inhomogeneity displayed as histograms of the flip-angle accuracy factor inside the brain of six healthy human volunteers.

Voxels outside the head or not containing brain tissue were masked out. The dashed lines indicate the mean (thick line) plus/minus the standard deviations (thin lines) of ηα across subjects.

https://doi.org/10.1371/journal.pone.0169265.g004

The variation of simulated intensity parameters ρ of the (ME-)MP2RAGE signal in dependence of T1 is presented in Fig 2. The underlying acquisition parameters are those from Table 1. These graphs were used to visually inspect the sensitivity and the acquisition parameters were selected accordingly. The level of the steady-state magnetization is the major factor affecting sensitivity, while the accessible T1 range is mostly influenced by the specific choice of α1,2 and TI,(1,2). Therefore, the number of k-space lines acquired in each GRE block, n, and TR,GRE, which relate directly to the image resolution and SNR, have a stronger impact on the sensitivity when the inter-acquisition times TA, TB, and TC are shorter. Therefore, a relatively long TR,seq is often needed for higher accuracy and robustness.

3.2 MRI Experiments

The results from the exploratory Study 1 (summarized in the S5S7 Tables) were used to investigate the impact of the acquisition scheme on the mapping. Further results obtained with a nominal resolution of 0.6 mm will be discussed in greater detail below. For higher resolutions, the statistical analyses did not indicate a strong impact from the particular imaging sequence. At lower resolutions, T1 maps obtained with ME-MP2RAGE and MP2RAGE were also essentially identical, whereas and χ maps obtained with ME-FLASH seemed slightly more accurate than those obtained with ME-MP2RAGE. This counter-intuitive observation is explained by the different PE schemes used in ME-MP2RAGE acquisitions with lower (e.g., 0.9 mm with first PE direction from head to foot) or higher (e.g., 0.6 mm with first PE direction from left to right) nominal resolutions, which yielded stronger restrictions in the choice of TE,max (and, hence, less accurate fits) for the lower-resolution scans.

Based on the combined results from the simulations and Study 1, the parameters in Table 1 were defined for an in-depth evaluation of data acquired with a high nominal resolution of 0.6 mm. An example of the data obtained with the ME-MP2RAGE sequence is shown briefly in Fig 5.

thumbnail
Fig 5. Example of the images obtained from an ME-MP2RAGE acquisition.

Each row represents a different echo time. The columns show in order: first inversion magnitude (1st) and phase (2nd); second inversion magnitude (3rd) and phase (4th). Magnitude images are shown in arb.units, while phase image are in radians, both using a gray scale. Note that: (i) the phase images for the first inversion point show an abrupt change in their value corresponding to the zero crossing of the signal in the T1 recovery curve; (ii) the phase images for the second inversion point present some coil combination pole artifacts resulting in a corresponding degradation of the QSM maps at these locations.

https://doi.org/10.1371/journal.pone.0169265.g005

A comparison of T1 maps acquired with ME-MP2RAGE and with MP2RAGE is shown in Fig 6. Similar comparisons of and χ maps derived with ME-MP2RAGE and with ME-FLASH are shown in Figs 7 and 8, respectively. These results are based on single subject acquisitions. Note that the χ maps are displayed in gray scale to be consistent with the QSM literature, but a diverging color map with linear luminance could be a better option for representing signed data (see also S8 Fig). The group statistics results are reported in Table 2. Individual subject results are reported in S8S10 Tables.

thumbnail
Fig 6. Single subject acquisition of T1 maps acquired with (a) ME-MP2RAGE, (c) MP2RAGE and (e) their difference, and test-retest reproducibility evaluation with voxel-by-voxel correlation 2D histograms for (b) ME-MP2RAGE, (d) MP2RAGE, and (f) ME-MP2RAGE versus MP2RAGE.

Note how the CSF voxel are underestimated by the MP2RAGE as a result of the specific choice of the acquisition parameters.

https://doi.org/10.1371/journal.pone.0169265.g006

thumbnail
Fig 7. Single subject acquisition of maps acquired with (a) ME-MP2RAGE, (c) ME-FLASH and (e) their difference, and test-retest reproducibility evaluation with voxel-by-voxel correlation 2D histograms for (b) ME-MP2RAGE, (d) ME-FLASH, and (f) ME-MP2RAGE versus ME-FLASH.

https://doi.org/10.1371/journal.pone.0169265.g007

thumbnail
Fig 8. Single subject acquisition of χ maps acquired with (a) ME-MP2RAGE, (c) ME-FLASH and (e) their difference, and test-retest reproducibility evaluation with voxel-by-voxel correlation 2D histograms for (b) ME-MP2RAGE, (d) ME-FLASH, and (f) ME-MP2RAGE versus ME-FLASH.

https://doi.org/10.1371/journal.pone.0169265.g008

thumbnail
Table 2. Summary of the group averages μg and SDs σg results of the correlation parameters for the T1, and χ maps acquired during Study 2.

https://doi.org/10.1371/journal.pone.0169265.t002

The 2D histograms of voxel-by-voxel correlations of T1 maps underline that the acquisition scheme produces very consistent results, both for within- as well as across-sequence comparisons. This is also supported by the mean and SDs of the image volumes’ differences, μD and σD, being close to zero (relative to the T1 value range). However, some deviation is observed for very long T1 values corresponding to CSF voxels, as anticipated, because the selected MP2RAGE acquisition parameters were expected to systematically underestimate these values (see also Figs 2 and 6 and §2.3)

For the and χ maps, the 2D histograms and voxel-by-voxel correlations indicate that both ME-MP2RAGE and ME-FLASH do not achieve a similar degree of global reproducibility as observed for the T1 maps. While both and χ maps are essentially compatible across sequence types with μD values close to zero, the σD values are not negligible and, at least in the case of , of almost the same order as the value range. Also, the bias introduced in the QSM results by the pole artifacts in the phase images seems to be independent of the acquisition scheme (see also S9 Fig). However, this does not seem to indicate inaccuracy of a specific acquisition scheme as the reproducibility does not substantially change for the within-sequence comparison. These results were consistently observed for all subjects.

The inter-acquisition averages of μD, μ|D|, σD, σ|D| for the test ME-MP2RAGE acquisition, for the reference MP2RAGE and ME-FLASH acquisitions, and for the cross-comparisons are always within twice their respective SDs. Additionally, the r2 parameters for different acquisition schemes of the same mapping are always within a SD of each other. This suggests that the accuracy, precision and reproducibility of the mappings are only marginally dependent on the tested acquisition schemes.

The results from the ROI-based analysis for all maps are presented in Table 3.

thumbnail
Table 3. Group averages, μg, and SDs, σg, for the ROI-based analysis of ME-MP2RAGE acquisitions.

Abbreviations: WM = White Matter; Fro. = Frontal Lobe; Tem. = Temporal Lobe; Par. = Parietal Lobe; Occ. = Occipital Lobe; Ins. = Insula; Put. = Putamen; Caud. = Caudate; Tha. = Thalamus; Cereb. = Cerebellum. Susceptibility is indicated here by Δχ as a reminder of the values being referenced to an arbitrary offset, implying that the group average and SD values are expected to be biased by that and, possibly, by the pole artifacts of the phase images.

https://doi.org/10.1371/journal.pone.0169265.t003

3.3 Effects from Mis-registration and Noise

The effects of registration are reported in Fig 9, where we show the 2D correlation histograms both on artificially mis-registratered and on self-registered T1, , and χ maps, respectively. The original maps were obtained with MP2RAGE for T1 and with ME-FLASH for and χ in these cases. Additional results for correlation histograms and the statistical comparisons related to the mis-registration are provided in Simultaneous S3S5 Figs and S11S13 Tables, respectively.

thumbnail
Fig 9. 2D correlation histograms showing the effects of registration.

Columns refer to T1 (a, c), (b, d) and χ (e, f) maps, while rows show either mis-registration (a, b, c) or self-registration (d, e, f) effects, respectively. The mis-registration and the self-registration effects are here illustrated by comparing each map with itself: after the application of a small transformation without registration (mis-registration) or after the application of a large transformation followed by a registration step (self-registration). The x-axis indicate the unmodified map, while the y-axis indicate the transformed or self-registered map. The small tranformation considered for the misregitration is a 0.5px translation followed by a 0.5° rotation. The large transformation considered for the self-registration is a 10° rotation.

https://doi.org/10.1371/journal.pone.0169265.g009

The 2D histograms obtained from these analyses are comparable to those obtained from the comparison of different acquisitions, even for relatively small transformation parameters. These essentially capture the global effects of the considered transformations. Moreover, the correlation coefficients indicate different sensitivities for the different map types. Specifically, a relatively linear behavior of r2 for the explored range of parameters is found for T1 and χ maps, while there is a remarkably non-linear drop in r2 for the same range of transformations in the case of (e.g., r2 dropped from 0.952 to 0.918 for T1 and from 0.913 to 0.850 for χ but from 0.615 to 0.401 for in analyses of subtle translations of 0.375 px and 0.500 px). The comparisons of the “self-registered” maps indicate that the effect of the registration step on the correlation analyses is at least of the same order of magnitude of a 0.2° rotation, which is alone sufficient to explain an r2 of ≈ 0.99, 0.91 and 0.97 for T1, and χ respectively, the major effect again being observed with maps. Note that self-registration results indicate only a lower limit on the effects of the registration step because of the negative effects of noise. Additionally, neither r2 nor the other correlation parameters are linear with respect to SNR, mis-registration or their combination; therefore these results cannot be used to quantify the single effects.

The explored noise level range correspond to up to about half of the SNR of the original images. The values of the obtained coefficients indicate that such SNR reductions degrade the voxel-by-voxel correlations with an effect size that is comparable to that from a translation of approximately 0.5 px, a rotation by approximately 0.9°, or a 0.4° rotation followed by a 0.4 px translation. An example of the effects of an increased noise level on χ maps is reported in S6 Fig. A summary of the correlation analysis is provided in S14 Table.

4 Discussion

4.1 Acquisition Parameters

In this work, simulations and exploratory MRI experiments were employed to optimize an ME-MP2RAGE sequence modification for high-resolution simultaneous mapping of T1, , and χ. The proposed set of acquisition parameters is specifically tailored to the tested resolutions and FOV. The software tools developed for this work can effectively be used for optimizing the acquisition parameters for different use cases. However, their limitations should be properly considered.

Particularly, the solutions of the Bloch equations for evaluating the sensitivity are ignoring the experimentally available SNR. This is of particular relevance when considering that the T1 maps are more insensitive to inhomogeneities at lower (relative) , which is typically achieved by lowering the flip angles α1 and α2 or increasing the TR,seq. However, this might have a negative impact on the acquisition time or the SNR (depending on TR,GRE, which in turn determines the Ernst angle).

The fitting procedures that were used to infer the influence of SNR and sampling points on the accuracy and precision of the estimated relaxation time constants (i.e., T1 from TI,(1,2) and from different TE,i) are based on the implicit assumption that the signal is well modeled by a mono-exponential function. However, recent works suggest that more complex curves (e.g., multi-exponential) might be required for modeling both longitudinal [12] and transverse [25, 46, 47] magnetization behaviors. More elaborate acquisition schemes with many more support points would be required to further investigate this aspect, which is neglected at this level.

Additionally, while the physical experiment is fundamentally identical, the methods for obtaining T1 are slightly different for the simulation and the measurements. While the simulations rely on the fitting of magnitudes images, the measurements use a combination of the complex images to obtain the MP2RAGE signal which is then converted to T1 via a look-up table (see also: Fig 2). In fact, the MP2RAGE approach is more accurate [5], essentially because the phase images provide additional information, particularly for the first inversion image, where the magnitude-only images have very low intensity due to the proximity to the zero-crossing in the inversion recovery curve. Therefore, the simulation constitutes a lower limit (except for SNR considerations) to the accuracy of the T1 estimates.

The proposed procedure for the choice of the acquisition parameters requires a manual optimization, which inherently implies a certain degree of arbitrariness. While a more formal approach would be possible, there are several aspects that should be considered. Firstly, the arbitrariness is inherent to the problem, because a particular desired feature (e.g., resolution, accuracy, speed) is obtained at the expenses of the others. For example, it would be possible to write a cost function for insensitivity and a cost function for the coverage of a specific range of T1 values. However, there is no unbiased way of combining the two without introducing some degree of arbitrariness, and further weights would need to be integrated into the model in order to cover the most relevant aspects of the acquisition. Such weights might not have immediate physical links with the desired features, making it difficult to find a rationale for their definition. Secondly, given the structure of the manifold parameters to be explored, there exist several suboptimal combinations as already suggested by [5], and an optimization algorithm might not find the best solution. Although a metric could be defined to describe the marginal gain associated with slight changes in parameters (e.g., by taking the partial derivatives of “first order” cost functions with respect to the parameters optimized by the other cost functions), this would lead to a much more complex mathematical model and would introduce additional arbitrary weights. Lastly, given the limits to specify sequence parameters on the protocol level (e.g., the flip angle can only be set to integer values in our current implementation), an automated optimization procedure is likely to produce a set of parameters that is not accessible with the same exactness in an experimental situation. Ultimately, this does not justify the additional complexity of such an approach. For these reasons, we performed the simulations by choosing the parameters with a rationale based on provisions of the desired resolution and scan time. These results were then extensively tested experimentally, thus providing additional information on the impact of aspects that were neglected or only partially addressed by the simulations.

4.2 Simultaneous Parameter Mapping

Focusing on the voxel-by-voxel comparisons of results obtained with different or the same acquisition modalities, a major point to discuss is the validity of the method for assessing the impact of the acquisition technique on the map accuracy. Since there was no direct access to the “ground truth”, we assumed that a more-established acquisition technique can be used as a “gold standard” for a specific map. Based on this, we consider two different acquisition schemes to be experimentally equivalent if they have the same degree of reproducibility (e.g., similar μD and σD) as two separate repetitions with the same sequence. This rationale relies on the assumption that the assessment of the reproducibility is unbiased. However, the registration step has an important effect on the parameters used to quantify reproducibility. This bias is always present (for non-simultaneous acquisitions), arising from the numerical interpolation required to address geometrical transformations acting on a sub-voxel scale—even in ideal cases where other spatial inaccuracies due to movements of the subject during the acquisition or noise associated with physiological processes could be ignored. While errors arising during registration are not sufficient to entirely explain the observed level of reproducibility, our simulation results are compatible with the hypothesis that even subtle mis-registrations are able to largely explain the values of the statistical coefficients used to assess the reproducibility. This is especially true for the measured (and, partially, also for χ) maps, because the inter-voxel values fluctuations are observed at a much finer spatial scale compared to the measured T1 maps. These differences are mostly localized at the boundaries of regions characterized by different average values (e.g. GM/WM, GM/CSF, etc.) and are sufficiently strong to be captured by the global coefficients and plots considered.

The lower reproducibility that is observed for the and χ (compared to T1) may not entirely be explained by registration issues, and further investigation may be required to elucidate other potentially relevant sources of error, for example physiological noise or motion. This would improve the understanding of the link between the tissue microstructure and mapping at a finer level. To reduce the effect of noise in these maps, it would be possible to introduce in the fit algorithm a regularization term incorporating additional information, for example from neighboring voxels [12, 46] or from less noisy acquisitions like the simultaneously measured T1.

Regardless of the acquisition scheme, our results also indicate a higher reproducibility of χ results compared to . However, this result would need a more robust validation. In fact, the correlation coefficients considered in this work are not appropriate for accurate quantification of this aspect. Additionally, the QSM results are obtained on a smaller volume (due to limitations of the processing pipeline) and, hence, a comparison of both contrasts might be biased by this difference. Note also that values were generally longer in the voxels excluded for QSM and might be estimated with an increased error because the acquisition parameters had been optimized for values up to ≈ 45 ms.

The SNR available for and χ estimates was inherently lower for ME-MP2RAGE as compared to ME-FLASH with the tested parameters. This is because substantially longer TR values and higher flip-angles were used for ME-FLASH. Additionally, the SNR of the second GRE block in (ME-)MP2RAGE is inversely modulated by T1 relaxation. Despite these differences in the SNR, the reproducibility coefficients are not appreciably different across acquisition schemes, suggesting that either the mapping or the comparison procedure (or both) are not sensitive enough for subtle SNR variations.

Of note, when choosing TE for both ME-FLASH and ME-MP2RAGE, we ignored potential effects related to the phase difference between water and fat. This was not a significant issue because usually the fat contribution to the signal in brain tissues is negligible, since most lipid molecules are immobilized in membranes and will have a in the μs range, which is too short to be observed with the typical echo times of a standard ME-FLASH experiment. Otherwise, we would have expected an acquisition-related bias in the maps, which would have caused, for example, the μD parameter between different acquisition to appreciably deviate from zero, and this was not observed.

The T1 values obtained from the ROI-based analysis indicate excellent agreement with [10] for WM, while GM values are consistently lower in our experiments. This might be explained by a slightly different definition of the ROIs and varying partial voluming effects, especially with CSF. values are generally in agreement with previously published results (e.g., [48, 49]), although it should be noted that results are always modulated—and therefore biased—by the B0 shimming. The susceptibility results are relative to an arbitrary reference, which was chosen to obtain an average susceptibility value very close to zero. For this reason, the group-averages in this case may be of limited significance.

In its current form, the ME-MP2RAGE pulse sequence achieves simultaneous acquisition of T1, , and χ maps as a trade-off between resolution, SNR, and scan time. Our set of parameters (Table 1) showed that a slightly reduced SNR to stay within the constrained acquisition time was still sufficient for relatively accurate simultaneous mapping. A more significant impact on the map accuracy, however, is expected for more aggressive reductions of the scan time (e.g., by further increasing the GRAPPA factor). Instead, if the resolution and FOV constraints are relaxed, it is possible to reach the point where the MP2RAGE and its ME variant have identical scan times, but the latter allows the simultaneous acquisition of T1, and χ maps without a further penalty for the achieved accuracy. This situation might be more interesting for 3 T, where typical resolutions are lower. This aspect remains to be experimentally investigated, but the already mentioned simulation tools can be effectively used to predict the reliability of T1 estimates.

Further sequence developments may allow to address some of the current limitations of ME-MP2RAGE. A finer control over timing and k-space coverage will improve the flexibility of the sequence, thus allowing for better trade-offs in resolution, scan time, and maps accuracy. This could be achieved for example by mixing the first and the second phase encoding directions in each GRE block (eventually integrating compressed sensing techniques) and/or by using a different number (and eventually timings) for the acquired echoes across the GRE blocks, but this would require both major modification to the sequence code and careful consideration on the effects on the T1 point-spread function, and ultimately a separate study to evaluate its accuracy in estimating T1 maps.

Such modifications may also lead to the opportunity of acquiring more sampling points in the inversion recovery curve. This is interesting because it might offer an effective and time saving sequence for the purpose of investigating non-mono-exponential behaviors, and hence partial volumes effects.

Another approach toward the acquisition of multiple images at different inversion times was recently proposed with the MPnRAGE sequence [50], where a radial readout scheme is employed instead of the cartesian one.

5 Conclusion

The ME-MP2RAGE scheme with an appropriate choice of acquisition parameters permits the quantification of T1, , and magnetic susceptibility χ in vivo at 7 T. The resulting maps are reasonably well comparable to results obtained from more-established techniques, such as standard MP2RAGE and ME-FLASH. The time required for recording multiple 3D maps simultaneously, even at high nominal resolution of 0.6 mm, is shorter than the time required to obtain corresponding maps from separate acquisitions. This aspect is even more favorable at lower resolution, where the timing is more flexible and the acquisition becomes more efficient with the current pulse-sequence design.

This sequence modification may allow researchers and clinicians to gain additional useful tissue information from a single experiment with improved consistency regarding subtle head motion and without a need for registration of image volumes with different contrast.

Supporting Information

S1 Table. List of Abbreviations.

A list of common abbreviations used through the text.

https://doi.org/10.1371/journal.pone.0169265.s001

(PDF)

S2 Table. List of Symbols.

A list of common symbols used through the text.

https://doi.org/10.1371/journal.pone.0169265.s002

(PDF)

S1 Fig. Dependece of T1 simulations from SNR.

Plot of T1 estimates obtained with simulations versus the exact input values (range 0.5–3.5 s) as a function of the SNR. Error bars indicate the standard deviations, σT, for N = 20000 simulations. Different columns correspond to different SNR levels of (a,d) 25, (b,e) 50, and (c,f) 100, while the two rows show the results for (a–c) ME-MP2RAGE with TI,(1,2) = 800, 2400 ms and for (d–f) MP2RAGE with TI,(1,2) = 750, 2900 ms.

https://doi.org/10.1371/journal.pone.0169265.s003

(TIFF)

S2 Fig. Dependece of simulations from SNR.

Plot of estimates obtained with simulations versus the exact input values (range 2–60 ms) as a function of the SNR. Error bars indicate the standard deviations, σT, for N = 20000 simulations. Different columns correspond to different SNR levels of (a,d) 25, (b,e) 50, and (c,f) 100, while the two rows show the results for (a–c) ME-MP2RAGE with nE = 4, TE,1 = 2.5 ms, and ΔTE ≈ 4.2 ms and for (d–f) ME-FLASH with nE = 5, TE,1 = 3.0 ms, and ΔTE = 6.0 ms.

https://doi.org/10.1371/journal.pone.0169265.s004

(TIFF)

S3 Fig. 2D histograms showing mis-registration effects in T1 maps.

The x-axis refers to the unmodified map, while the y-axis refers to the same map after application of: (a) a translation by 0.2 px, (b) a rotation by 0.2°, (c) a rotation by 0.2° followed by a translation by 0.2 px, (d) a translation by 0.5 px, (e) a rotation by 0.5°, (f) a rotation by 0.5° followed by a translation by 0.5 px. More details can be found in S11 Table.

https://doi.org/10.1371/journal.pone.0169265.s005

(TIFF)

S4 Fig. 2D histograms showing mis-registration effects in maps.

The x-axis refers to the unmodified map, while the y-axis refers to the same map after application of: (a) a translation by 0.2 px, (b) a rotation by 0.2°, (c) a rotation by 0.2° followed by a translation by 0.2 px, (d) a translation by 0.5 px, (e) a rotation by 0.5°, (f) a rotation by 0.5° followed by a translation by 0.5 px. More details can be found in S12 Table.

https://doi.org/10.1371/journal.pone.0169265.s006

(TIFF)

S5 Fig. 2D histograms showing mis-registration effects in χ maps.

The x-axis refers to the unmodified map, while the y-axis refers to the same map after application of: (a) a translation by 0.2 px, (b) a rotation by 0.2°, (c) a rotation by 0.2° followed by a translation by 0.2 px, (d) a translation by 0.5 px, (e) a rotation by 0.5°, (f) a rotation by 0.5° followed by a translation by 0.5 px. More details can be found in S13 Table.

https://doi.org/10.1371/journal.pone.0169265.s007

(TIFF)

S6 Fig. 2D histograms for χ maps showing effects from differing noise levels.

The x-axis refers to the unmodified map, while the y-axis refers to a map calculated from the same source complex image after adding Gaussian noise with a SD of (a) 5, (b) 10, (c) 15, (d) 20, (e) 25, (f) 30 arbitrary intensity units, corresponding to up to 100% increase in SNR. The original SNR was estimated as the maximum of the magnitude image divided by the standard deviation of the real image in a region where no signal was expected. More details can be found in S14 Table.

https://doi.org/10.1371/journal.pone.0169265.s008

(TIFF)

S7 Fig. 2D histograms for the self-registration effects.

Results for (a) T1, (b) and (c) χ maps are shown. The x-axis refers to the unmodified map, while the y-axis refers to the self-registered map, i.e. the same map after the application of a 10° rotation and then registered back (using FLIRT) onto the original map.

https://doi.org/10.1371/journal.pone.0169265.s009

(TIFF)

S8 Fig. Example of different color maps used for magnetic susceptibility mapping.

The same magnetic susceptibility map is displayed with the standard gray scale color map (a), a diverging color map with non-linear (b) and (almost) linear luminance (c). The divergin color map with linear luminance (see also: http://matplotlib.org/users/colormaps.html and “A Better Default Colormap for Matplotlib.” Berkeley Institute for Data Science, July 17, 2015. https://bids.berkeley.edu/resources/videos/better-default-colormap-matplotlib) may be a better option (although not used in previous literature) because it allows for a clear separation of positive and negative values, while retaining as much quantitative information as a gray color scale. This is actually interesting in view of the applications where a separation from paramagnetic and diamagnetic substances is desired. On the other hand, the arbitrary reference currently limits its efficacy and its adoption could be postponed until the community reaches a consensus on this.

https://doi.org/10.1371/journal.pone.0169265.s010

(TIFF)

S9 Fig. Example of the consistency of pole artifact across acquisitions.

Phase images from the same subject and session are shown side by side for the ME-MP2RAGE acquisition and the ME-FLASH acquisition. The position of the pole artifact is consistent between the two acquisition (within the limits of subject’s motion), thus suggesting a limited impact on test-retest reproducibility.

https://doi.org/10.1371/journal.pone.0169265.s011

(TIFF)

S3 Table. Simulation of SNR influence on T1 mapping.

Average and standard deviation of σT values in simulations of T1 mapping (assumed T1 values between 0.5 and 3.5 s) to evaluate the SNR influence.

https://doi.org/10.1371/journal.pone.0169265.s012

(PDF)

S4 Table. Simulation of SNR influence on mapping.

Average and standard deviation of σT values in simulations of mapping (assumed values between 2 and 60 ms) to evaluate the SNR influence.

https://doi.org/10.1371/journal.pone.0169265.s013

(PDF)

S5 Table. Correlation parameters obtained in Study 1 for T1 mapping.

The order of the listed acquisition parameters is: nominal isotropic resolution, TR,seq, α1,2, TI,(1,2), and TE.

https://doi.org/10.1371/journal.pone.0169265.s014

(PDF)

S6 Table. Correlation parameters obtained in Study 1 for mapping.

The order of the listed acquisition parameters is: nominal isotropic resolution, TR,seq, α1,2, TI,(1,2), and TE values for ME-MP2RAGE; nominal isotropic resolution, TR, α and TE values for ME-FLASH.

https://doi.org/10.1371/journal.pone.0169265.s015

(PDF)

S7 Table. Correlation parameters obtained in Study 1 for χ mapping.

The order of the listed acquisition parameters is: nominal isotropic resolution, TR,seq, α1,2, TI,(1,2), and TE for ME-MP2RAGE; nominal isotropic resolution, TR, α and TE for ME-FLASH.

https://doi.org/10.1371/journal.pone.0169265.s016

(PDF)

S8 Table. Statistical coefficients for T1 mapping from the different acquisitions.

Variations of the correlation coefficients, and means and SDs of image volume differences (as defined in Eqs 4 and 5) obtained in Study 2 for T1 maps across different subjects with the acquisition parameters from Table 1. The last part of the table are the group averages μg and SDs σg according to the acquisition scheme.

https://doi.org/10.1371/journal.pone.0169265.s017

(PDF)

S9 Table. Statistical coefficients for mapping from the different acquisitions.

Variations of the correlation coefficients, and means and SDs of image volume differences (as defined in Eqs 4 and 5) obtained in Study 2 for maps across different subjects with the acquisition parameters from Table 1. The last part of the table are the group averages μg and SDs σg according to the acquisition scheme.

https://doi.org/10.1371/journal.pone.0169265.s018

(PDF)

S10 Table. Statistical coefficients for χ mapping from the different acquisitions.

Variations of the correlation coefficients, and means and SDs of image volume differences (as defined in Eqs 4 and 5) obtained in Study 2 for χ maps across different subjects with the acquisition parameters from Table 1. The last part of the table are the group averages μg and SDs σg according to the acquisition scheme.

https://doi.org/10.1371/journal.pone.0169265.s019

(PDF)

S11 Table. Statistical coefficients for T1 mapping from mis-registration.

Variations of the correlation coefficients, and means and SDs of image volume differences (as defined in Eqs 4 and 5) obtained for systematic geometrical transformations of T1 maps.

https://doi.org/10.1371/journal.pone.0169265.s020

(PDF)

S12 Table. Statistical coefficients for mapping from mis-registration.

Variations of the correlation coefficients, and means and SDs of image volume differences (as defined in Eqs 4 and 5) obtained for systematic geometrical transformations of maps.

https://doi.org/10.1371/journal.pone.0169265.s021

(PDF)

S13 Table. Statistical coefficients for χ mapping from mis-registration.

Variations of the correlation coefficients, and means and SDs of image volume differences (as defined in Eqs 4 and 5) obtained for systematic geometrical transformations of χ maps.

https://doi.org/10.1371/journal.pone.0169265.s022

(PDF)

S14 Table. Statistical coefficients for χ mapping from differing noise levels.

Variations of the correlation coefficients, and means and SDs of image volume differences (as defined in Eqs 4 and 5) obtained for systematic increase of the noise level for the complex images used to compute χ maps.

https://doi.org/10.1371/journal.pone.0169265.s023

(PDF)

Acknowledgments

We would like to express our deepest gratitude to Gunnar Krüger, Jöran Lepsien and Pierre-Louis Bazin for the helpful discussions and suggestions.

Author Contributions

  1. Conceptualization: AS RM TK.
  2. Data curation: RM.
  3. Formal analysis: RM.
  4. Funding acquisition: HEM.
  5. Investigation: AS RM.
  6. Methodology: AS RM HEM.
  7. Project administration: RM.
  8. Resources: AS RM HEM TK.
  9. Software: AS RM TK.
  10. Supervision: AS HEM.
  11. Validation: AS RM.
  12. Visualization: RM.
  13. Writing – original draft: RM.
  14. Writing – review & editing: AS RM HEM.

References

  1. 1. Callaghan MF, Helms G, Lutti A, Mohammadi S, Weiskopf N. A general linear relaxometry model of R1 using imaging data. Magnetic Resonance in Medicine. 2015;73(3):1309–1314. pmid:24700606
  2. 2. Stüber C, Morawski M, Schäfer A, Labadie C, Wähnert M, Leuze C, et al. Myelin and iron concentration in the human brain: A quantitative study of MRI contrast. NeuroImage. 2014;93, Part 1:95–106. pmid:24607447
  3. 3. Bazin PL, Weiss M, Dinse J, Schäfer A, Trampel R, Turner R. A computational framework for ultra-high resolution cortical segmentation at 7 Tesla. NeuroImage. 2014;93, Part 2:201–209. pmid:23623972
  4. 4. Keuken MC, Bazin PL, Crown L, Hootsmans J, Laufer A, Müller-Axt C, et al. Quantifying inter-individual anatomical variability in the subcortex using 7 T structural MRI. NeuroImage. 2014;94:40–46. pmid:24650599
  5. 5. Marques JP, Kober T, Krueger G, van der Zwaag W, Van de Moortele PF, Gruetter R. MP2RAGE, a self bias-field corrected sequence for improved segmentation and T1-mapping at high field. NeuroImage. 2010;49(2):1271–1281. pmid:19819338
  6. 6. Streitbürger DP, Pampel A, Krueger G, Lepsien J, Schroeter ML, Mueller K, et al. Impact of image acquisition on voxel-based-morphometry investigations of age-related structural brain changes. NeuroImage. 2014;87:170–182. pmid:24188812
  7. 7. Frahm J, Haase A, Matthaei D. Rapid NMR imaging of dynamic processes using the FLASH technique. Magnetic Resonance in Medicine: Official Journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 1986;3(2):321–327. pmid:3713496
  8. 8. Haase A, Frahm J, Matthaei D, Hänicke W, Merboldt KD. FLASH imaging: Rapid NMR imaging using low flip-angle pulses. Journal of Magnetic Resonance. 2011;213(2):533–541. pmid:22152368
  9. 9. Helms G, Dathe H, Dechent P. Quantitative FLASH MRI at 3T using a rational approximation of the Ernst equation. Magnetic Resonance in Medicine. 2008;59(3):667–672. pmid:18306368
  10. 10. Wright PJ, Mougin OE, Totman JJ, Peters AM, Brookes MJ, Coxon R, et al. Water proton T1 measurements in brain tissue at 7, 3, and 1.5T using IR-EPI, IR-TSE, and MPRAGE: results and optimization. Magnetic Resonance Materials in Physics, Biology and Medicine. 2008;21(1–2):121–130. pmid:18259791
  11. 11. Barta R, Kalantari S, Laule C, Vavasour IM, MacKay AL, Michal CA. Modeling T1 and T2 relaxation in bovine white matter. Journal of Magnetic Resonance. 2015;259:56–67. pmid:26295169
  12. 12. Labadie C, Lee JH, Rooney WD, Jarchow S, Aubert-Frécon M, Springer CS, et al. Myelin water mapping by spatially regularized longitudinal relaxographic imaging at high magnetic fields. Magnetic Resonance in Medicine. 2014;71(1):375–387. pmid:23468414
  13. 13. Van de Moortele PF, Auerbach EJ, Olman C, Yacoub E, Ugurbil K, Moeller S. T1 weighted brain images at 7 Tesla unbiased for Proton Density, T2* contrast and RF coil receive B1 sensitivity with simultaneous vessel visualization. NeuroImage. 2009;46(2):432–446. pmid:19233292
  14. 14. de Rochefort L, Liu T, Kressler B, Liu J, Spincemaille P, Lebon V, et al. Quantitative susceptibility map reconstruction from MR phase data using bayesian regularization: Validation and application to brain imaging. Magnetic Resonance in Medicine. 2010;63(1):194–206. pmid:19953507
  15. 15. Liu T, Spincemaille P, de Rochefort L, Kressler B, Wang Y. Calculation of susceptibility through multiple orientation sampling (COSMOS): A method for conditioning the inverse problem from measured magnetic field map to susceptibility source image in MRI. Magnetic Resonance in Medicine. 2009;61(1):196–204. pmid:19097205
  16. 16. Liu T, Liu J, de Rochefort L, Spincemaille P, Khalidov I, Ledoux JR, et al. Morphology enabled dipole inversion (MEDI) from a single-angle acquisition: Comparison with COSMOS in human brain imaging. Magnetic Resonance in Medicine. 2011;66(3):777–783. pmid:21465541
  17. 17. Schweser F, Sommer K, Deistung A, Reichenbach JR. Quantitative susceptibility mapping for investigating subtle susceptibility variations in the human brain. NeuroImage. 2012;62(3):2083–2100. pmid:22659482
  18. 18. Shmueli K, Dodd SJ, Li TQ, Duyn JH. The contribution of chemical exchange to MRI frequency shifts in brain tissue. Magnetic Resonance in Medicine. 2011;65(1):35–43. pmid:20928888
  19. 19. Wharton S, Schäfer A, Bowtell R. Susceptibility mapping in the human brain using threshold-based k-space division. Magnetic Resonance in Medicine. 2010;63(5):1292–1304. pmid:20432300
  20. 20. Schäfer A, Wharton S, Gowland P, Bowtell R. Using magnetic field simulation to study susceptibility-related phase contrast in gradient echo MRI. NeuroImage. 2009;48(1):126–137. pmid:19520176
  21. 21. Denk C, Torres EH, MacKay A, Rauscher A. The influence of white matter fibre orientation on MR signal phase and decay. NMR in Biomedicine. 2011;24(3):246–252. pmid:21404336
  22. 22. He X, Yablonskiy DA. Biophysical mechanisms of phase contrast in gradient echo MRI. Proceedings of the National Academy of Sciences. 2009;106(32):13558–13563. pmid:19628691
  23. 23. Lee J, Shmueli K, Kang BT, Yao B, Fukunaga M, van Gelderen P, et al. The contribution of myelin to magnetic susceptibility-weighted contrasts in high-field MRI of the brain. NeuroImage. 2012;59(4):3967–3975. pmid:22056461
  24. 24. Liu C. Susceptibility tensor imaging. Magnetic Resonance in Medicine. 2010;63(6):1471–1477. pmid:20512849
  25. 25. Wharton S, Bowtell R. Fiber orientation-dependent white matter contrast in gradient echo MRI. Proceedings of the National Academy of Sciences of the United States of America. 2012;109(45):18559–18564. pmid:23091011
  26. 26. Wharton S, Bowtell R. Effects of white matter microstructure on phase and susceptibility maps. Magnetic Resonance in Medicine. 2015;73(3):1258–1269. pmid:24619643
  27. 27. Zhong K, Leupold J, von Elverfeldt D, Speck O. The molecular basis for gray and white matter contrast in phase imaging. NeuroImage. 2008;40(4):1561–1566. pmid:18353683
  28. 28. Schweser F, Deistung A, Lehr BW, Reichenbach JR. Differentiation between diamagnetic and paramagnetic cerebral lesions based on magnetic susceptibility mapping. Medical Physics. 2010;37(10):5165–5178. pmid:21089750
  29. 29. Langkammer C, Liu T, Khalil M, Enzinger C, Jehna M, Fuchs S, et al. Quantitative Susceptibility Mapping in Multiple Sclerosis. Radiology. 2013;267(2):551–559. pmid:23315661
  30. 30. Chen W, Zhu W, Kovanlikaya I, Kovanlikaya A, Liu T, Wang S, et al. Intracranial calcifications and hemorrhages: characterization with quantitative susceptibility mapping. Radiology. 2014;270(2):496–505. pmid:24126366
  31. 31. Fritzsch D, Reiss-Zimmermann M, Trampel R, Turner R, Hoffmann KT, Schäfer A. Seven-Tesla Magnetic Resonance Imaging in Wilson Disease Using Quantitative Susceptibility Mapping for Measurement of Copper Accumulation:. Investigative Radiology. 2014;49(5):299–306. pmid:24220252
  32. 32. Mugler JP, Brookeman JR. Three-dimensional magnetization-prepared rapid gradient-echo imaging (3D MP RAGE). Magnetic Resonance in Medicine: Official Journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 1990;15(1):152–157. pmid:2374495
  33. 33. van der Kouwe AJW, Benner T, Salat DH, Fischl B. Brain morphometry with multiecho MPRAGE. NeuroImage. 2008;40(2):559–569. pmid:18242102
  34. 34. Fujimoto K, Polimeni JR, van der Kouwe AJW, Reuter M, Kober T, Benner T, et al. Quantitative comparison of cortical surface reconstructions from MP2RAGE and multi-echo MPRAGE data at 3 and 7 T. NeuroImage. 2014;90:60–73. pmid:24345388
  35. 35. Mildner T, Müller K, Hetzer S, Trampel R, Driesel W, Möller HE. Mapping of arterial transit time by intravascular signal selection. NMR in Biomedicine. 2014;27(5):594–609. pmid:24610794
  36. 36. Gudbjartsson H, Patz S. The Rician Distribution of Noisy MRI Data. Magnetic Resonance in Medicine: Official Journal of the Society of Magnetic Resonance in Medicine / Society of Magnetic Resonance in Medicine. 1995;34(6):910–914. pmid:8598820
  37. 37. Oliphant TE. Python for Scientific Computing. Computing in Science & Engineering. 2007;9(3):10–20.
  38. 38. Millman KJ, Aivazis M. Python for Scientists and Engineers. Computing in Science Engineering. 2011;13(2):9–12.
  39. 39. Hunter JD. Matplotlib: A 2D Graphics Environment. Computing in Science & Engineering. 2007;9(3):90–95.
  40. 40. Yarnykh VL. Actual flip-angle imaging in the pulsed steady state: A method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magnetic Resonance in Medicine. 2007;57(1):192–200. pmid:17191242
  41. 41. Jochimsen TH, von Mengershausen M. ODIN—Object-oriented Development Interface for NMR. Journal of Magnetic Resonance. 2004;170(1):67–78. pmid:15324759
  42. 42. Schofield MA, Zhu Y. Fast phase unwrapping algorithm for interferometric applications. Optics Letters. 2003;28(14):1194–1196. pmid:12885018
  43. 43. Schweser F, Deistung A, Lehr BW, Reichenbach JR. Quantitative imaging of intrinsic magnetic tissue properties using MRI signal phase: An approach to in vivo brain iron metabolism? NeuroImage. 2011;54(4):2789–2807. pmid:21040794
  44. 44. Schweser F, Deistung A, Sommer K, Reichenbach JR. Toward online reconstruction of quantitative susceptibility maps: Superfast dipole inversion. Magnetic Resonance in Medicine. 2013;69(6):1581–1593. pmid:22791625
  45. 45. Jenkinson M, Beckmann CF, Behrens TEJ, Woolrich MW, Smith SM. FSL. NeuroImage. 2012;62(2):782–790. pmid:21979382
  46. 46. Hwang D, Kim DH, Du YP. In vivo multi-slice mapping of myelin water content using T2* decay. NeuroImage. 2010;52(1):198–204. pmid:20398770
  47. 47. van Gelderen P, de Zwart JA, Lee J, Sati P, Reich DS, Duyn JH. Nonexponential T2* decay in white matter. Magnetic Resonance in Medicine. 2012;67(1):110–117. pmid:21630352
  48. 48. Govindarajan ST, Cohen-Adad J, Sormani MP, Fan AP, Louapre C, Mainero C. Reproducibility of T2* mapping in the human cerebral cortex in vivo at 7 tesla MRI. Journal of Magnetic Resonance Imaging. 2015;42(2):290–296. pmid:25407671
  49. 49. Mangeat G, Govindarajan ST, Mainero C, Cohen-Adad J. Multivariate combination of magnetization transfer, T2* and B0 orientation to study the myelo-architecture of the in vivo human cortex. NeuroImage. 2015;119:89–102. pmid:26095090
  50. 50. Kecskemeti S, Samsonov A, Hurley SA, Dean DC, Field A, Alexander AL. MPnRAGE: A technique to simultaneously acquire hundreds of differently contrasted MPRAGE images with applications to quantitative T1 mapping. Magnetic Resonance in Medicine. 2016;75(3):1040–1053. pmid:25885265