Analysis of physiological noise in quantitative cardiac magnetic resonance

  • Terrence Jao ,

    Roles Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Software, Writing – original draft, Writing – review & editing

    terrencejao@gmail.com

    Affiliation Department of Biomedical Engineering, Viterbi School of Engineering, University of Southern California, Los Angeles, California, United States of America

  • Krishna Nayak

    Roles Funding acquisition, Supervision, Writing – review & editing

    Affiliation Ming Hsieh Department of Electrical and Computer Engineering, Viterbi School of Engineering, University of Southern California, Los Angeles, California, United States of America

Abstract

Purpose

To determine the impact of imaging parameters on the temporal signal-to-noise ratio (TSNR) of quantitative cardiac magnetic resonance (MR) in humans, and to determine applicability of the physiological noise covariance (PNC) model for physiological noise (PN).

Methods

We conducted MRI experiments in four healthy volunteers, and obtained series of short-axis cardiac images acquired with snapshot balanced steady-state free precession (bSSFP) and snapshot gradient echo (GRE) using a broad range of spatial resolutions and parallel imaging acceleration factors commonly used in quantitative cardiac MR. We measured regional SNR and TSNR in these datasets and fit the measurements to the PNC model for PN, which assumes that PN scales with signal strength.

Results

The relationship between SNR and TSNR in human cardiac MR without contrast preparation was well modeled by the PNC model. SNR consistently decreased as the spatial resolution (matrix size) and acceleration factor (R) increased for both GRE and bSSFP imaging. TSNR varied linearly with SNR using GRE imaging, when SNR was low (SNR < 20), and approached an asymptotic limit using bSSFP imaging, when SNR was high (SNR > 40).

Conclusions

The PNC model can be used to guide the choice of matrix size and acceleration factor to optimize TSNR in stable contrast cardiac MR, such as T2-prepared Blood-Oxygen-Level-Dependent (BOLD) and several variants of Arterial Spin Labeled (ASL) cardiac MR.

Introduction

Quantitative cardiac magnetic resonance (MR) has shown clinical value in a wide array of applications, including T1, T2, and ECV mapping as well as first-pass and non-contrast perfusion imaging. Quantitative cardiac MR maps are formed from multiple images, which make them susceptible to errors caused by signal fluctuations of physiological origin, such as cardiac or respiratory motion. This is known as physiological noise (PN) and impacts the temporal signal-to-noise (TSNR) ratio which ultimately determines measurement variability.

Recent advances in cardiac MR have enabled clinically feasible quantitative mapping of magnetic relaxation properties and perfusion of the myocardium [1]. T1 and ECV mapping can assess diffuse myocardial fibrosis and infiltration [24] and can measure extracellular volume when acquired with contrast [5,6]. Inflammation and edema in the heart can be detected with T2 mapping [7] while T2* is sensitive to iron induced field inhomogeneity in hemochromatosis [8] and hemorrhage [9]. Quantitative myocardial perfusion imaging performed using either first-pass cardiac MR [10], arterial spin labeling (ASL) [11], or blood oxygen level dependent (BOLD) [12] has been shown to detect regions of ischemia to evaluate coronary artery disease severity. These quantitative maps are all formed by imaging the myocardium multiple times with either contrast agents or different magnetization preparations to vary the sensitivity of each image to the parameter of interest. These preparations produce changes in the signal intensity of the acquired images, which are fit to models of the underlying physiology or relaxation parameter to generate a quantitative map.

Unexpected signal changes that are not from the designed preparations will cause errors in the parameter estimates. These errors arise from thermal noise as well as physiological fluctuations from cardiac, respiratory, and hemodynamic motion. Thermal noise is well understood and is modeled as an independent additive bivariate Gaussian process. It can be easily measured from noise-only data (acquired without RF excitation) or from a manually-selected noise-only region within an image. In contrast, the physiological noise distribution is largely unknown and cannot be directly measured. Physiological noise contributes to the temporal standard deviation of an image series, and can become a dominant contributor to measurement variability. In quantitative imaging, temporal signal-to-noise ratio (TSNR) is measured as the mean signal intensity within a region-of-interest (ROI) divided by its temporal standard deviation.

Triantafyllou et. al. [13] demonstrated that in the brain image acquisition parameters, including spatial resolution, parallel imaging acceleration factor, and flip angle, altered the TSNR. They were also able to experimentally confirm that physiological noise increases proportionally with the signal strength [14]. This is problematic because a simple increase of signal strength through higher field or denser receiver arrays may not translate to improvements in TSNR if physiological noise is dominant. In this case, efforts should be made to increase spatial resolution or speed acquisition through parallel imaging. Northrup et. al. [15] observed this phenomenon in ASL cardiac MR, which exhibited similar TSNR at 1.5T and 3.0T, despite the SNR advantage at 3.0T. Do et. al. [16] used parallel imaging to speed up the acquisition of ASL cardiac MR and observed an increase in TSNR despite a decrease in SNR. TSNR can be increased by using parallel imaging to shorten the image acquisition window. This has been found to reduce physiological noise, likely due to cardiac and respiratory motion [16].

In this study, we measure the SNR and TSNR of cardiac mid short-axis image series under different spatial resolutions and parallel imaging acceleration factors and assess the ratio between physiological and thermal noise. We study snapshot bSSFP and GRE acquisitions because they are the most commonly used acquisition schemes for cardiac MR T1, T2, and ECV mapping and for myocardial perfusion imaging. While the results of this work are specific to our MRI apparatus, receiver coil, and pulse sequence, it is simple to reproduce and need only be performed once. To the best of our knowledge, this is the first study that experimentally examines how the choice of imaging parameters affects TSNR and σP for quantitative cardiac MR.

Materials and methods

Noise analysis

Our analysis of physiological noise and TSNR follows the work of Triantafyllou et. al. [13] for multi-channel coil arrays in fMRI image series in the brain. The total noise, σ in a cardiac MR image series is modeled as the independent sum of thermal noise, σt and physiological noise, σp. (1) Physiological noise is described by the Krüger and Glover model (KG) [14] to scale proportionally by a constant, λ, with the MR signal level, S, such that σp = λS. According to this model for noise, TSNR can be derived from image SNR as follows: (2)

Recently, Triantafyllou et. al. proposed a new physiological noise covariance (PNC) model that captures coil to coil noise correlations from physiological fluctuations by splitting the noise covariance matrix into a thermal and physiological component [17]. This model relaxes the assumption that PN must scale with the signal level and reduces to the KG model when the physiological noise covariance matrix scales with the signal level. The correlations of physiological noise sources that are not signal level dependent are described by an additional parameter, α, to generalize the model such that (3) In practice, both λ and α are unknown and TSNR is directly measured by dividing the mean signal intensity by its temporal standard deviation. However, an important result of this model is that TSNR reaches an asymptotic limit of 1/λ when SNR becomes large. This implies that attempts to increase TSNR by increasing SNR will fail if SNR is already high.

From measurements of SNR and TSNR, the ratio of physiological noise to thermal noise can be measured as follows: (4)

This noise ratio determines whether an image series is thermal noise or physiological noise dominant. When images are thermal noise dominant, improvements in TSNR can be achieved by increasing the signal, through higher field strength of denser surface coils. However, when images are physiological noise dominant, improvements in SNR have negligible impact on TSNR. Instead, SNR can be traded for other gains, such as in spatial resolution or parallel imaging acceleration factor up until the physiological noise is comparable to thermal noise.

Experimental methods

Images were acquired on a 3T GE Signa Excite HD with an 8-channel cardiac receiver array in 4 healthy volunteers. Written informed consent was acquired from all volunteers using a protocol specifically approved by the University of Southern California institutional review board. Images were acquired using snapshot bSSFP and GRE. All images were acquired during mid-diastole at a single mid short-axis slice. For each unique set of acquisition parameters, twenty images were acquired. Images were acquired with a 1 second breath hold with a 3 second wait between acquisitions to allow the magnetization to fully recover. A cardiac cinema/video [CINE] scout scan was acquired in each subject to determine the exact timing of mid-diastole relative to the ECG R-wave. Noise only images without RF were also acquired to calculate the thermal noise and coil noise covariance matrix in each subject.

In each subject, data were acquired at various matrix sizes (192×96, 192×128, 192×192) and accelerated (1, 1.33, 1.6, 2) by GRAPPA [18] and partial Fourier imaging. Other imaging parameters were fixed throughout the experiment, which include TR/TE of 3.6/1.8 ms, FA of 500/50 for bSSFP and GRE respectively, BW of ±125 kHz, and slice thickness of 10 mm. This led to a total of 24 different imaging settings (3 matrix sizes, 4 acceleration factors, 2 imaging sequences). Differences in the size of the volunteers led to different choices of FOV and spatial resolution. Specific imaging parameters are summarized in Table 1, and inter-subject variations in the FOV and spatial resolutions are summarized in Table 2.

thumbnail
Table 1. Acquisition parameters used for acquiring cardiac image series at various resolutions and acceleration factors.

https://doi.org/10.1371/journal.pone.0214566.t001

thumbnail
Table 2. Variations in FOV and spatial resolution in the different subjects.

Resolution is reported for matrix sizes of 192 x (96, 128, 192).

https://doi.org/10.1371/journal.pone.0214566.t002

Image reconstruction and data analysis

Image were reconstructed offline using software written in Matlab (Mathworks Inc. Natick, MA, USA). Partial Fourier image reconstruction was carried out using homodyne reconstruction with a linear ramp weighting while GRAPPA reconstruction was performed using a custom implementation. Images were coil combined using an optimal B1 weighted combination [19] and aligned with non-rigid motion correction using advanced normalization tools (ANTs) [20].

SNR maps were calculated for each image using the “pseudo-replica” method described by Robson et al. [21] For each image, correctly scaled and correlated synthetic noise is added to acquired k-space data before reconstruction. This is repeated multiple times, each with a different instantiation of synthetic noise, to generate an image stack. Under the assumption that the reconstruction process acts independently on the signal and the noise, SNR can be calculated as the ratio of the image reconstructed without noise addition to the standard deviation of the image stack. In contrast, TSNR was measured from the actual multiple replicas that were acquired. It is calculated as the mean intensity of a pixel or a manually drawn ROI of the left ventricular myocardium divided by the temporal standard deviation. The SNR and TSNR for each of the 24 individual imaging parameters were averaged across the four subjects, and fit to the PNC model of Eq 3 using a non-linear least squares algorithm provided by Matlab to estimate the physiological noise scaling parameter, λ and physiological coil covariance parameter, α. The noise ratio from Eq 4 was subsequently calculated from SNR, TSNR and α. Regional analysis of the lateral and septal myocardium used ROIs corresponding to segments 2 and 3 and segments 5 and 6 of the standard AHA model [22].

Results

Fig 1 shows a representative image, SNR, TSNR, and noise ratio map from one volunteer with an acceleration factor of 1 using a 192×192 matrix. SNR is always greater than or equal to TSNR. The noise ratio map is low in areas with little movement such as the musculature in the back and highest in areas with large fluctuations in signal intensity, such as blood vessels.

thumbnail
Fig 1. Representative image, SNR, TSNR, and σP/σT maps from a healthy volunteer.

Images were acquired with a matrix size of 192x192, R = 1, Resolution = 1.67x1.67mm2. The SNR, TSNR, and σP/σT ratio within the left ventricular myocardium ROI (white dashed line) was 51.3± 9.8, 30.5±10.2, and 1.49 ± 0.56 respectively. The noise ratio map is lowest in areas with little movement, such as the muscles of the back and highest in areas with large fluctuations in signal intensity such as blood vessels. In a static phantom and stable imaging environment, TSNR = SNR. In vivo, TSNR ≤ SNR due to additional sources of σP. The σP/σT ratio determines whether σP or σT is the dominant source of noise. When σP is dominant (σP/σT >> 1), SNR gains will not improve TSNR.

https://doi.org/10.1371/journal.pone.0214566.g001

Fig 2A shows TSNR as a function of SNR in the left ventricular myocardium. Each point represents an average across subjects while the error bars represent the standard deviation within the ROI averaged across subjects. Red solid points represent bSSFP acquisitions while blue outlined points represent GRE acquisitions. The points were fit to Eq 3, shown as the solid black line, with a physiological noise scale factor, λ, of 0.0197. The lower and upper 95% confidence interval of λ was 0.0173 and 0.0221 and are plotted with their corresponding dotted black lines. The physiological noise covariance factor, α, is 0.817 with 95% confidence interval of 0.383, 1.252. The vertical dashed red line corresponds to the SNR in which the noise ratio is unity and has a value of 50.8 (1/λ). The noise ratio is less than unity when left of the line, and greater than unity when right of the line. The horizontal dashed red line corresponds to the asymptotic value for TSNR which also has a value of 50.8 (1/λ).

thumbnail
Fig 2. Plots of temporal signal-to-noise ratio (TSNR) versus signal-to-noise ratio (SNR).

Separate plots are for A) all LV myocardium, C) septal myocardium and D) lateral myocardium, from mid-short axis slices. Septal and lateral segments are illustrated in B). Each point represents an average over four subjects while the error bars represent the standard deviation within the ROI averaged over four subjects. bSSFP acquisitions are shown in solid red while GRE acquisitions are shown in outlined blue. The solid black line represents the fit to the PNC model along with the lower and upper 95% confidence interval shown in dotted black. The dashed black line indicates the line of unity. Vertical and horizontal dashed red lines represent when the noise ratio is equal to one and the asymptotic limit of TSNR respectively. Both occur at 1/λ. At low SNR, TSNR varies linearly with SNR while at high SNR, TSNR approaches an asymptotic limit. The lateral myocardium also exhibited higher physiological noise (λ = 0.0242, α = 0.5581) and hence lower asymptotic TSNR than the septum (λ = 0.0162,α = 0.7583).

https://doi.org/10.1371/journal.pone.0214566.g002

The data were also analyzed within the lateral and septal walls, which are shown in Fig 2C and Fig 2D respectively. Septal and lateral segments are defined by the AHA 17-segment model [22] and are shown in Fig 2B. λ is statistically different between the lateral and septal wall. It is higher within the lateral walls at a value of 0.0242 with lower and upper 95% confidence interval of 0.0219 and 0.0265 is lower in the septum at a value of 0.0162 with a confidence interval of 0.0141 and0.0183. This leads to an asymptotic value for TSNR of 41.4 and 61.7 for the lateral wall and septum respectively. α within the lateral and septal walls are not significantly different and are 0.5581 with 95% confidence of 0.2023 to 0.9140 and 0.7583 with 95% confidence of 0.3973 to 1.1193 respectively.

Fig 3 contains plots of average SNR, TSNR, and σp/σt within the left ventricular myocardium ROI as a function of the image matrix size and acceleration factor. SNR consistently decreased as the spatial resolution (matrix size) and acceleration factor (R) increased for both GRE and bSSFP imaging. In GRE imaging, TSNR decreased as the spatial resolution and acceleration factor increased. The noise ratio was consistently lower than unity, represented by the dashed red line, for all matrix sizes and acceleration factors. In contrast, TSNR only slightly decreased as the matrix size increased in bSSFP imaging, but remained constant when acceleration factor increased. The noise ratio in bSSFP imaging was consistently above unity and only decreased to approach unity at the largest matrix size and acceleration factor.

thumbnail
Fig 3. Comparison of SNR, TSNR, and the σp/σt for several image matrix sizes and acceleration factors.

A) GRE and B) bSSFP. Matrix sizes of 192×96, 192×128, and 192×192 are shown in blue, green, and yellow respectively. As expected, SNR decreases with increasing matrix size and higher acceleration factors. In GRE, TSNR follows a similar trend and the noise ratio is < 1 for all imaging settings. In bSSFP, TSNR remains relatively constant and the noise ratio is ≥1 at low acceleration factors and approaches 1 only at higher accelerations and larger matrix sizes.

https://doi.org/10.1371/journal.pone.0214566.g003

Discussion

In this work, we investigated how changes in spatial resolution and parallel imaging acceleration factor affect SNR, TSNR, and the noise ratio for quantitative cardiac MR using snapshot GRE and bSSFP imaging. We found that when spatial resolution is made coarser and acceleration factor is reduced, SNR consistently increases while TSNR faces diminishing returns. In particular, we observed that TSNR varies linearly with SNR using GRE imaging (SNR < 20) and approaches an asymptotic limit using bSSFP imaging (SNR > 40). This makes the noise ratio (σp/σt) thermal noise dominant (<1) for GRE and physiological noise dominant (>1) for bSSFP in all but the largest of matrix sizes and acceleration factors (192× 192, R = 2). These results are predicted by the PNC model; physiological noise scales with signal strength and above a certain threshold, further improvements in SNR do not necessarily translate to improvements in TSNR.

The relationship between SNR and TSNR, as descried by the PNC model, can be categorized into three regimes—the linear, transition, and asymptotic regime. The linear regime is defined as the range of SNR where the noise ratio is less than unity while the asymptotic regime is defined as the range of SNR where TSNR is greater than 95% of its asymptotic value. The transition regime lies between the two. One should avoid imaging in the linear regime because TSNR can always be improved by increasing SNR. Similarly, one should avoid imaging in the asymptotic regime because SNR can be sacrificed with little to no loss in TSNR, which will yield improvements in spatial resolution or scan time. Quantitative cardiac MR protocols should be performed in the transition regime, and the experiments described in this manuscript can inform appropriate parameter selection.

In the original KG model, physiological noise can be interpreted as a modulation of signal intensity within a voxel caused by complex physiological processes, such as pulsatile blood flow or the effect of deoxyhemoglobin in the BOLD effect. These effects are local and occur within the voxel. In contrast, with bulk motion, tissue signal intensities are modulated as they move through the spatial gradient of the receiver sensitivity profile over time and can no longer be accounted for by the KG model. In cardiac imaging, the major sources of PN are bulk motion during the cardiac cycle and respiration. The PNC model relaxes the assumption that PN must scale with the signal level and can be used to model the major sources of PN in CMR. While the PNC model was originally developed for long-TR EPI imaging used in BOLD fMRI, we found that it is applicable to snapshot GRE and bSSFP often used for cardiac imaging. Snapshot imaging is performed in a transient state and is designed to capture the longitudinal magnetization prior to imaging.

Previous studies have only investigated the relationship between SNR and TSNR for functional MRI of the brain [13,14]. In those studies, reported values for the physiological noise scaling parameter, λ, (0.009) are significantly lower than values we report for the heart (0.0197). This is likely due to cardiac and respiratory motion found in the heart that are absent in the brain. We also reported regional variations in SNR and TSNR and found that λ is 1.49 times larger in the lateral wall (0.0242) than in the septum (0.0162).

One concern about the results of this study is the precision of the TSNR estimate, which is dependent on the accuracy of the noise standard deviation. The noise standard deviation is determined by the number of images acquired and pixels within the ROI. Using a Monte Carlo simulation for 20 images with 150 pixels within the left ventricular myocardium, 50 pixels for regional analysis, and individual pixel-wise analysis, the relative error of the TSNR measurement was 1.6%, 2.1%, and 12.97% respectively. This is below the inter-subject variation for SNR and TSNR and does not alter the conclusions of the study. The large standard deviation in measured SNR and TSNR is reflective of both spatial variability due to B1 inhomogeneity within the myocardium and inter-subject variation of SNR due to differences in body habitus and chest wall thickness. Sung et. al. found that flip angles within the myocardium can vary as much as 53% within an individual subject [23]. In this work, the standard deviation is dominated by normal inter-subject variability. It is possible to lower the measured standard deviations by carefully choosing subjects with similar body habitus, cardiovascular performance at rest, and controlling for confounds such as coffee intake.

Another concern in our analysis is the existence of physiological noise within an imaging window, which may be more pronounced in longer acquisition windows (e.g. 672 ms for a 192×192 matrix size without acceleration). TSNR can only be measured between images and all noise within an image is assumed to come from thermal noise. Therefore, SNR may be underestimated and the SNR-TSNR curve may be shifted to the right. Rapid acquisition schemes such as EPI or spiral imaging may be evaluated to observe the degree of this bias. We hypothesize that these acquisition schemes, with their shorter imaging window, may even lie on a completely different SNR-TSNR curve with a lower λ and higher asymptotic TSNR.

A natural follow-up to this study would be to investigate whether the relationship between SNR, TSNR, and the noise ratio changes in systolic imaging, during exercise or pharmacologic stress, or with different cardiac pathologies. For instance, arrhythmia and stress may cause more cardiac motion during imaging and decrease TSNR while heart failure and post-acute myocardial infarct scar may reduce cardiac motion and increase TSNR. These studies may be difficult to perform in humans, but could potentially be examined in a large animal model, where respiration can be controlled and many images can be acquired. The drawback is that animals are typically anesthetized and may have a different stress response than humans. Nevertheless, it would be insightful to determine whether image acquisitions parameters can be optimized for cardiac phase, stress, or pathology.

Another limitation of this study is that we only considered equilibrium contrast. Quantitative cardiac MR can involve large variations in contrast due to magnetization preparation and/or passage of contrast agents. A single experiment (i.e. T1 mapping or T2 mapping) can have images with a broad range of SNR and regional contrast. While changes in image acquisition parameters are expected to cause shifts along the SNR vs. TSNR curve, changes in contrast or contrast preparation parameters could cause shifts of the entire curve. The findings of this study are therefore only directly applicable to quantitative cardiac MR techniques that utilize snapshot imaging and have stable contrast between images. This includes T2-prepared BOLD cardiac MR [24] and ASL cardiac MR using PICORE, EPISTAR, or steady-pulsed labeling [25].

In summary, we demonstrated that the SNR and TSNR of equilibrium snapshot cardiac MR images can be parametrized by the PNC model and varies with different image acquisition parameters, including spatial resolution and acceleration factor. However, we have also identified additional imaging parameters that must be considered for cardiac imaging, the precise timing and duration of imaging within the RR interval. Future work needs to be done to characterize these different regimes for systolic imaging, stress imaging, and in different cardiac pathologies.

Acknowledgments

We thank our team of researchers at the Magnetic Resonance Engineering Laboratory including Hung Phi Do, Vanessa Landes, Ahsan Javed, Yongwan Lim, Yannick Bliesener, and Xin Miao.

References

  1. 1. Salerno M, Kramer CM. Advances in Parametric Mapping With CMR Imaging. JACC Cardiovasc Imaging. 2013;6:806–22. pmid:23845576
  2. 2. Iles L, Pfluger H, Phrommintikul A, Cherayath J, Aksit P, Gupta SN, et al. Evaluation of Diffuse Myocardial Fibrosis in Heart Failure With Cardiac Magnetic Resonance Contrast-Enhanced T1 Mapping. J Am Coll Cardiol. 2008;52:1574–80. pmid:19007595
  3. 3. Flett AS, Hayward MP, Ashworth MT, Hansen MS, Taylor AM, Elliott PM, et al. Equilibrium Contrast Cardiovascular Magnetic Resonance for the Measurement of Diffuse Myocardial Fibrosis. Circulation. 2010;122:138–44. pmid:20585010
  4. 4. Jerosch-Herold M, Sheridan DC, Kushner JD, Nauman D, Burgess D, Dutton D, et al. Cardiac magnetic resonance imaging of myocardial contrast uptake and blood flow in patients affected with idiopathic or familial dilated cardiomyopathy. Am J Physiol Heart Circ Physiol. 2008;295:H1234–42. pmid:18660445
  5. 5. Schelbert EB, Testa SM, Meier CG, Ceyrolles WJ, Levenson JE, Blair AJ, et al. Myocardial extravascular extracellular volume fraction measurement by gadolinium cardiovascular magnetic resonance in humans: slow infusion versus bolus. J Cardiovasc Magn Reson. 2011;13:16. pmid:21375743
  6. 6. Kellman P, Wilson JR, Xue H, Bandettini WP, Shanbhag SM, Druey KM, et al. Extracellular volume fraction mapping in the myocardium, part 2: initial clinical experience. J Cardiovasc Magn Reson. 2012;14:64. pmid:22967246
  7. 7. Verhaert D, Thavendiranathan P, Giri S, Mihai G, Rajagopalan S, Simonetti OP, et al. Direct T2 Quantification of Myocardial Edema in Acute Ischemic Injury. JACC Cardiovasc Imaging. 2011;4:269–78. pmid:21414575
  8. 8. Anderson LJ, Holden S, Davis B, Prescott E, Charrier CC, Bunce NH, et al. Cardiovascular T2-star (T2*) magnetic resonance for the early diagnosis of myocardial iron overload. Eur Heart J. 2001;22:2171–9. pmid:11913479
  9. 9. Zia MI, Ghugre NR, Connelly KA, Strauss BH, Sparkes JD, Dick AJ, et al. Characterizing Myocardial Edema and Hemorrhage Using Quantitative T2 and T2* Mapping at Multiple Time Intervals Post ST-Segment Elevation Myocardial InfarctionClinical Perspective. Circ Cardiovasc Imaging. 2012;5:566–72. pmid:22744938
  10. 10. Hsu L-Y, Groves DW, Aletras AH, Kellman P, Arai AE. A Quantitative Pixel-Wise Measurement of Myocardial Blood Flow by Contrast-Enhanced First-Pass CMR Perfusion Imaging: Microsphere Validation in Dogs and Feasibility Study in Humans. JACC Cardiovasc Imaging. 2012;5:154–66.
  11. 11. Zun Z, Varadarajan P, Pai RG, Wong EC, Nayak KS. Arterial Spin Labeled CMR Detects Clinically Relevant Increase in Myocardial Blood Flow With Vasodilation. JACC Cardiovasc Imaging. 2011;4:1253–61. pmid:22172781
  12. 12. Arnold JR, Karamitsos TD, Bhamra-Ariza P, Francis JM, Searle N, Robson MD, et al. Myocardial Oxygenation in Coronary Artery Disease: Insights From Blood Oxygen Level–Dependent Magnetic Resonance Imaging at 3 Tesla. J Am Coll Cardiol. 2012;59:1954–64. pmid:22624835
  13. 13. Triantafyllou C, Polimeni JR, Wald LL. Physiological noise and signal-to-noise ratio in fMRI with multi-channel array coils. NeuroImage. 2011;55:597–606. pmid:21167946
  14. 14. Krüger G, Glover GH. Physiological noise in oxygenation-sensitive magnetic resonance imaging. Magn Reson Med. 2001;46:631–7. pmid:11590638
  15. 15. Northrup BE, McCommis KS, Zhang H, Ray S, Woodard PK, Gropler RJ, et al. Resting myocardial perfusion quantification with CMR arterial spin labeling at 1.5 T and 3.0 T. J Cardiovasc Magn Reson. 2008;10:53. pmid:19014709
  16. 16. Do H, Jao TR, Nayak KS. Myocardial arterial spin labeling perfusion imaging with improved sensitivity. J Cardiovasc Magn Reson. 2014;16:15. pmid:24467918
  17. 17. Triantafyllou C, Polimeni JR, Keil B, Wald LL. Coil-to-coil physiological noise correlations and their impact on functional MRI time-series signal-to-noise ratio. Magn Reson Med. 2016;76:1708–19. pmid:26756964
  18. 18. Griswold MA, Jakob PM, Heidemann RM, Nittka M, Jellus V, Wang J, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47:1202–10. pmid:12111967
  19. 19. Roemer PB, Edelstein WA, Hayes CE, Souza SP, Mueller OM. The NMR phased array. Magn Reson Med. 1990;16:192–225. pmid:2266841
  20. 20. Avants BB, Tustison NJ, Johnson HJ. Advanced Normalization Tools [Internet]. Available from: http://stnava.github.io/ANTs/
  21. 21. Robson PM, Grant AK, Madhuranthakam AJ, Lattanzi R, Sodickson DK, McKenzie CA. Comprehensive quantification of signal-to-noise ratio and g-factor for image-based and k-space-based parallel imaging reconstructions. Magn Reson Med. 2008;60:895–907. pmid:18816810
  22. 22. Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart a statement for healthcare professionals from the cardiac imaging committee of the Council on Clinical Cardiology of the American Heart Association. Circulation. 2002;105:539–542. pmid:11815441
  23. 23. Sung K, Nayak KS. Measurement and characterization of RF nonuniformity over the heart at 3T using body coil transmission. J Magn Reson Imaging. 2008;27:643–8. pmid:18306272
  24. 24. Fieno DS, Shea SM, Li Y, Harris KR, Finn JP, Li D. Myocardial Perfusion Imaging Based on the Blood Oxygen Level–Dependent Effect Using T2-Prepared Steady-State Free-Precession Magnetic Resonance Imaging. Circulation. 2004;110:1284–90. pmid:15326062
  25. 25. Capron T, Troalen T, Robert B, Jacquier A, Bernard M, Kober F. Myocardial perfusion assessment in humans using steady-pulsed arterial spin labeling: Human MBF Using Steady-Pulsed Arterial Spin Labeling. Magn Reson Med. 2014; 74: 990–998. pmid:25263761