Determination of rheology and surface tension of airway surface liquid: a review of clinical relevance and measurement techniques

Abstract

By airway surface liquid, we mean a thin fluid continuum consisting of the airway lining layer and the alveolar lining layer, which not only serves as a protective barrier against foreign particles but also contributes to maintaining normal respiratory mechanics. In recent years, measurements of the rheological properties of airway surface liquid have attracted considerable clinical attention due to new advances in microrheology instruments and methods. This article reviews the clinical relevance of measurements of airway surface liquid viscoelasticity and surface tension from four main aspects: maintaining the stability of the airways and alveoli, preventing ventilator-induced lung injury, optimizing surfactant replacement therapy for respiratory syndrome distress, and characterizing the barrier properties of airway mucus to improve drug and gene delivery. Primary measuring techniques and methods suitable for determining the viscoelasticity and surface tension of airway surface liquid are then introduced with respect to principles, advantages and limitations. Cone and plate viscometers and particle tracking microrheometers are the most commonly used instruments for measuring the bulk viscosity and microviscosity of airway surface liquid, respectively, and pendant drop methods are particularly suitable for the measurement of airway surface liquid surface tension in vitro. Currently, in vivo and in situ measurements of the viscoelasticity and surface tension of the airway surface liquid in humans still presents many challenges.

Background

Human airways, from the trachea through the bronchioles to the alveoli, are lined on the inside with a continuous film of surface liquid, which increases in thickness from ~ 0.1 μm in the alveoli to ~ 10 μm in the trachea [1,2,3,4,5]. In the airways, this film of liquid is a bilayer consisting of a periciliary layer and a mucous or ‘gel’ layer atop. The mucous layer is composed of 97% water and 3% solids (mucins, nonmucin proteins, salts, and cellular debris), which determine the linear and nonlinear viscoelasticity and diffusive properties of the mucus [6, 7]. The periciliary layer is of low viscosity containing water and solutes and in which the cilia reside [8]. The cilia beat 12 to 15 times per second, with cilia tips intermittently gripping the underside of the mucous layer, thus propelling it and entrapped particles towards the mouth, where it is swallowed or expectorated. In this way, the airways are kept clean. In addition to acting as a solid physical barrier to most pathogens, the airway lining fluid also contains lysozymes and a range of defensins that are capable of chemically inactivating inhaled pathogens [1].

In airway generations beyond 15 or 16, the primary secretory cells are Club cells (originally known as Clara cells), found in the respiratory bronchioles, and type II epithelial cells, found in the alveoli. These two types of cells have a relatively weaker mucus-secreting capacity compared with the Goblet cells found in the bronchi and conducting bronchioles [1, 4]. As a result, the airway lining fluid transitions from a two-layered fluid to a single layer of fluid that is primarily saltwater, yet with significant concentrations of surfactant [9]. Pulmonary surfactant (PS) is a complex mixture of some 90% lipids and 10% proteins. Most of the lipids are phospholipids, of which 70–80% are dipalmitoyl phosphatidyl choline (DPPC), the main surface-active material responsible for lowering surface tension, while the surfactant proteins (SP) are SP-A, SP-B, SP-C and SP-D [1, 10]. Among these surfactant-associated proteins, SP-B and SP-C are vital to the stabilization of the surfactant monolayer; SP-A and SP-D are involved in the control of surfactant release and possibly in the immunology of the lung. The main function of the pulmonary surfactant film is to reduce surface tension at the air-liquid interface, thus preventing collapse of the alveoli and small airways during end-exhalation [11].

Pulmonary surfactant is present not only in the alveoli but also in the bronchioles and small airways. We hereafter refer to the thin fluid continuum consisting of the airway lining layer and the alveolar lining layer as airway surface liquid [9]. Changes in macro- and microrheological properties of airway surface liquid have a significant impact on normal respiratory mechanics and normal barrier and clearance functions of the lung [12, 13]. For example, an intermediate viscoelasticity of the mucous gel layer, or in other words a viscosity in the range of 12–15 Pa ∙ s(1 Pa ∙ s = 1000 cP) and an elastic modulus of 1 Pa, are essential for optimal mucociliary clearance [14,15,16]. If the viscoelasticity of airway mucus becomes too low, however, the elasticity is not enough for mucus to counteract gravitational action, which likely makes mucus to slide down into the lung and flood the alveoli [13, 17,18,19]. In contrast, pulmonary disease conditions, such as cystic fibrosis (CF), chronic obstructive pulmonary disease (COPD), and asthma, are usually characterized by an increase in the viscoelasticity of mucus. As a result, ciliary action and cough are incapable of effectively clearing the sticky mucus, leading to the accumulation of mucus and even the complete blockage of the airway observed in the above disorders [8, 13, 20].

Likewise, alterations in the surface tension of the alveolar lining fluid also cause severe respiratory diseases, such as neonatal respiratory distress syndrome (NRDS) and acute lung injury (ALI) or the acute respiratory distress syndrome (ARDS) [21,22,23]. Premature infants who suffer from NRDS due to surfactant deficiency exhibit a high lining-fluid surface tension and hence a propensity for prominent atelectasis, decreased lung compliance, increased work of breathing and impaired gas exchange. The postnatal delivery of exogenous surfactant can significantly lower surface tension forces in the lung and has been established as a standard therapeutic intervention in the management of preterm infants with NRDS [24]. Similarly, in ALI/ARDS patients, the normal function of PS is inhibited by protein-rich oedematous fluid present in the airspaces, leading to a dramatic increase in surface tension and hence a decrease in lung compliance [25]. The development of ventilator-induced lung injury (VILI) in mechanically ventilated ALI/ARDS patients is largely attributed to this change in the surface tension-relevant lung micromechanics.

Therefore, the study of the rheological properties of airway surface liquid has both physiological and clinical significance. Unfortunately, due to a lack of suitable in vivo and in situ measurement techniques, thus far, all rheological measurements of human respiratory mucus came from in vitro studies that may not give a true picture of in vivo conditions. Moreover, reported literature values of the viscoelasticity of human respiratory mucus show large (orders of magnitude) intersubject, intrasubject, and even within the same mucus -sample variations (Table 1). It is imperative to heighten collaborations between clinicians, biomedical engineers, and applied scientists to explain these variations in perspective of both physiology and experimental techniques, to further develop tools to assess the quantitative properties of airway surface liquid, and finally to correlate the biophysical properties of airway surface liquid with healthy versus diseased states. This article reviews the importance of airway surface liquid rheology and surface tension measurements in: (1) maintaining the stability of small airways and alveoli; (2) preventing ventilator-induced lung injury; (3) optimizing surfactant replacement therapy (SRT); and (4) characterizing lung barrier and clearance functions. Subsequently, new methods and techniques for determining the viscosity and surface tension of airway surface liquid are described.

Table 1 Comparison of Published Viscosities of Airway Surface Liquid

Importance of rheological measurements of airway surface liquid

Maintaining stability of airways and alveoli

The role of the surface tension and viscosity of airway surface liquid in maintaining airway stability is primarily two-fold: retarding small airway closure and preventing alveolar collapse. As described in Section 1, the liquid lining usually forms a thin and relatively uniform layer on the inner surface of the airway, but sometimes it is possible for the airway to become occluded by the liquid, leading to airway closure. Closure of the distal airways at low lung volumes can occur through two mechanisms, “liquid bridge formation” or “compliant collapse” (Fig. 1) [30,31,32]. In the former case, liquid in a uniform film lining on the inner wall of an axisymmetric airway redistributes via a classical fluid-elastic instability known as the Plateau-Rayleigh instability [30, 33, 34]. This process leads to complete occlusion of the airway by a liquid plug or “bridge”, provided that the condition of sufficient liquid for closure is satisfied. A theoretical analysis by White and Heil showed that the growth rate of the film thickness increased with surface tension and decreased with an increase in the fluid’s viscosity [35]. Halpern and coworkers revealed that the growth rate for a viscoelastic layer was larger than for a Newtonian fluid with the same viscosity [36]. The overall timescale required for an occlusion to form is small compared with a single breathing cycle, provided that no surfactant is present. Halpern and Grotberg further demonstrated that the closure time for a pulmonary surfactant-rich film can be approximately five times greater than that for a film free of pulmonary surfactant [37].

Fig. 1
figure 1

Mechanisms of airway closure: (A) Liquid bridge formation (B) compliant collapse

Alternatively, if the surface tension of the airway lining fluid is sufficiently large relative to the airway’s bending stiffness, a fluid-elastic “compliant collapse” is more likely to occur [30, 31, 33]. As lung volume falls during expiration, the radius of the airway is decreased, thus resulting in an increase of the curvature of the air-liquid interface. The initially uniform and axisymmetric liquid lining can become unstable, and pressure gradients are induced in the fluid that drive flows redistributing the fluid. As a result, in the region where the liquid lining film is thickest, surface tension creates a large pressure jump over the highly curved air-liquid interface, causing negative pressure in the liquid. At the same time, parenchymal tethering forces on the external surface of the airway fall because of the gradual increase in lung volumes. This combination of reduced lining fluid pressure and parenchymal tethering subjects the airway wall to a significant compressive load and promotes the propensity of the airway to buckle inward, producing a compliant collapse. In diseased conditions such as pulmonary oedema or neonatal RDS, this compliant collapse of the airways may occur due to an increase in the volume of fluid or in the surface tension.

Surface forces also have a critical effect on airspace stability, as illustrated in Fig. 2. Two connected bubbles (alveoli) with a common pressure and a constant surface tension are blown at the end of a Y-tube [1, 38]. According to the Laplace equation, the pressure generated by surface tension in the small bubble is larger than that in one with a greater diameter, resulting in an inherently unstable system: the smaller alveolus will eventually collapse and the larger one will become over-distended. Of course, this is not the case in a healthy lung. The surface tension of the alveolar lining fluid is variable in situ as a function of expansion and compression of the alveolar surface area due to the presence of pulmonary surfactant. The surface tension drops as the alveolar surface decreases, and it rises when the surface expands, allowing for equal pressure between two different sized alveoli; therefore, system stability is maintained.

Fig. 2
figure 2

Connected alveoli illustrating the driving force collapsing the smaller alveolus in the case of constant surface tension

Preventing ventilator-induced lung injury

A number of theoretical and experimental studies have demonstrated that the increase in viscosity and surface tension of airway surface liquid likely results in VILI. Two main physical mechanisms for VILI are lung tissue overdistention caused by surface tension-induced alterations in interalveolar micromechanics and atelectrauma to the epithelial cells during repetitive airway reopening and closure [39,40,41]. The prediction from an adjoining two-alveoli model by Chen et al. [42] shows that the pattern of alveolar expansion can appear heterogeneous or homogeneous, strongly depending on differences in air-liquid interface tension on alveolar segments. More specifically, if surface tension in the liquid-filled alveolus is much greater than that in the air-filled alveolus, then alveolar expansion is heterogeneous. Consider a pair of juxtaposed alveoli: the maximum stress and strain within the septum shared by the two alveoli may occur at a low alveolar pressure; in contrast, as alveoli inflate to near total lung capacity (TLC), the stress and strain of the alveolar walls may decrease instead. On the other hand, if the surface tensions in two adjacent alveoli are identical, then alveolar expansion is homogenous; that is, the stress and strain of all alveolar septa will appear to linearly increase as alveolar volume varies from functional residual capacity (FRC) to near TLC. These calculations are in good agreement with the experimental phenomenon observed by Perlman and coworkers [43]. Using real-time optical section microscopy, these investigators quantified the micromechanics of an air-filled alveolus that shares a septum with a liquid-filled alveolus. Instilling liquid into the alveolus produced a meniscus that changed the septal curvature and consequently the pressure difference across the septum. As a consequence, the air-filled alveolus bulged into its liquid-filled neighbour even at FRC. Given that liquid-filled and air-filled alveoli can be focal, diffuse or patchy in pulmonary oedema, these findings may provide a novel understanding of segmental heterogeneities and alveolar overdistension during mechanical ventilation.

Using thin-walled polyethylene tubes to mimic bronchial walls held in apposition by airway lining fluid, Gaver III et al. [39] investigated the effect of the tube radius (R) and the surface tension and viscosity of airway lining fluid on the airway opening velocity (U) and the applied opening pressure. They found that increasing the surface tension (γ) or viscosity (μ) resulted in an increase in airway opening pressures. Gaver III et al. further defined a nondimensional parameter capillary number (Ca ≡ μU/γ) to represent the relative importance of viscous and surface tension forces in airway opening. When Ca is small, the applied opening pressure must exceed the “yield pressure” ~8γ/R before airway opening can proceed. When Ca is larger than 0.5, the contribution of viscous forces to the overall opening pressures is non-negligible.

In subsequent studies, Bilek and Kay et al. utilized a parallel-plate flow chamber lined with pulmonary epithelial cells as an idealized model airway to investigate the mechanisms of surface tension-induced epithelial cell damage [44]. The narrow channel of the chamber was filled with either phosphate-buffered saline (high surface tension) or Infasurf (ONY, Buffalo, NY), a biologically derived pulmonary surfactant with low surface tension. Airway reopening was generated by the steady progression of a semi-infinite bubble of air along the length of the channel, which displaced the occlusion fluid. Two bubble progression velocities were investigated, and the results showed that for the saline-occluded channels, both slow and fast bubble velocities resulted in significant cellular injury compared with the control and that for the Infasurf-occluded channels, cellular injury was dramatically reduced at both bubble velocities, indicating that surfactant has a protective effect. A comparison of the experimental and theoretical observations demonstrated that among four potentially injurious components of the stress cycle associated with airway reopening (shear stress, pressure, shear stress gradient or pressure gradient), the pressure gradient was the most predominant mechanism underlying the observed cellular damage [45].

Recently, Chen et al. [46] further estimated in situ the magnitudes of mechanical stresses exerted on the alveolar walls during repetitive alveolar reopening by using the tape-peeling model from McEwan and Taylor [47]. Their calculations showed that 1) for a lung with normal fluid viscosity and surface tension, the predicted maximum shear stresses were less than 15 dyn/cm2 at all alveolar opening velocities that were in the physiological range, whereas for a lung with an elevated viscosity of the alveolar lining fluid, shear stresses may increase by several orders of magnitude, enough to induce epithelial cell injury; 2) similarly, in the case of elevated viscosity, pressure drops across a cell may rise to levels greater than 300 dyn/cm2 and consequently result in hydraulic epithelial cracks [48]; 3) the capillary pressure for alveolar opening ranged from 5 to 30 cmH2O, strongly depending on the initial depth of the alveolar lining fluid, which may explain the clinically high opening pressure in sticky atelectasis; and 4) assuming alveolar lining fluid to be a Newtonian flow, the magnitudes of shear stress were proportional to the alveolar opening velocity, and therefore, the reduction of inspiratory flow rate or respiratory frequency would lead to a decrease in shear stress and a concomitant reduction in atelectrauma on alveolar epithelial cells.

As discussed above, rheological measurements of airway surface liquid during the progression of the disease in many pulmonary disorders have an important role in the management of mechanical ventilation. Measured values of surface tension and viscoelasticity provide clinical data for establishing and validating mathematical models of VILI. Knowing the values of viscosity and surface tension of airway surface liquid enables clinicians to quickly determine individualized inflation pressures and PEEPs and to roughly estimate lung stress and strain based on computational models, thereby adjusting the ventilation settings or therapeutic strategies in time to avoid VILI as much as possible.

Optimizing surfactant replacement therapy

Exogenous SRT has been established as a standard therapeutic intervention for preterm and term neonates with clinically confirmed respiratory distress syndrome since the early 1990s [10, 21, 49]. During traditional SRT, natural or synthetic surfactant is administered via an endotracheal tube either as a bolus or by infusion via a thin catheter inserted into the endotracheal tube. Thereafter, the infants are maintained on mechanical ventilation. The INSURE (INtubtion-SURfactant-Extubation) technique, which features early bolus instillation of surfactant with prompt extubation to nasal CPAP, has also been studied in a number of small randomized trials. The results showed that this strategy reduced the need for mechanical ventilation and improved survival rates [21, 50, 51]. A number of alternatives to the administration of surfactant include the use of aerosolized surfactant preparations, laryngeal mask airway-aided delivery of surfactant, instillation of pharyngeal surfactant, and administration of surfactant using laryngoscopy or bronchoscopy [21, 51].

SRT has also been applied to adults whose surfactant systems are compromised by ARDS, but clear indications of a distinct surfactant-mediated decrease in mortality or improvement in ventilator care of ARDS patients are still lacking [49, 52]. Additionally, recent randomized clinical trials have indicated that preventive surfactant administration to infants with suspected NRDS is no longer effective in groups of infants when CPAP is used routinely [21]. Most previous studies on SRT failures have focused on examining the biophysical mechanisms for surfactant inhibition due to plasma proteins or lipids [49]. However, the three-dimensional model of SRT recently proposed by Filoche and colleagues provides new insights into this issue, as it strongly suggests that inadequate delivery of surfactant may be a major cause of SRT failure [53]. Using similar surfactant mixtures and instilled dose volume, these investigators simulated the delivery of surfactant to neonates and adults in 3D structural models of the lung airway tree. The results revealed well-mixed distributions in the neonatal lungs but very inhomogeneous distributions in the adult lungs.

When liquid surfactant mixtures are instilled into the trachea via an endotracheal tube, they form liquid plugs, which are then blown distally into the branching network of the airways by forced inspirations. Filoche and colleagues simplified the above complicated flow process into two separate steps: step A, deposition of the liquid onto the airway walls into a trailing film; and step B, liquid plug splitting at an airway bifurcation. Step A determines the total amount of liquid reaching the acini, i.e., the delivery efficiency. Theoretical work by Helpern et al. [54] has shown that the thickness (h) of a trailing film in the parent tube is related to the local capillary number by the relation

$$ \frac{\mathrm{h}}{{\mathrm{a}}_1}=0.36\left(1-{\mathrm{e}}^{-2{\mathrm{Ca}}_{\mathrm{p}}^{0.523}}\right) $$
(1)

where a1 is the radius of the parent tube, the capillary number Cap = μUp/γ represents the ratio of viscous force (μ) to surface tension force (γ), and Up is the plug speed. Eq. (1) shows that as the viscosity of the liquid plugs or the airflow rate increases, so does Cap, and thus there is more liquid deposited into the trailing film. Step B governs the homogeneity of delivery. When the liquid plug splits at the bifurcation of the airway, a fraction of the plug’s volume goes down one daughter airway, V1, and the rest goes down the other, V2. Zheng and coworkers [55, 56] defined the ratio of the volumes in the daughter airways as the split ratio, Rs = V1/V2, which is affected by a number of factors, including the physical properties of the liquid (viscosity, density and surface tension), the gravitational orientation, the airway geometry, plug propagation speed, interfacial activity, and the presence of plug blockage in nearby airways from previous instillations. A critical parent tube capillary number was found to exist below which Rs = 0, and above which Rs increased and eventually levelled out with Cap. This feature can be explained by the driving pressure at the bifurcation. When the fluid viscosity or plug velocity is too small, the driving pressure is not large enough to overcome gravity; thus, no liquid enters the upper (gravitationally opposed) daughter airway after bifurcation.

In summary, the viscosity and surface tension of surfactant mixtures have a profound effect on the distribution quality of the delivered surfactant. For example, a computational model of SRT by Filoche and colleagues showed that the synthetic surfactant mixture Exosurf, with a low viscosity of ~ 3 cP, yielded a less homogenous distribution compared with the surfactants Survanta, Curosurf and Infasurf, each with a viscosity of ~ 30 cP, under the same neonatal treatment protocol. Furthermore, in the distal regions of the lung, surface tension gradient-induced Marangoni flows drive the surfactant deeper into the lung. This requires that the surface tension of the endogenous surfactant be above that of the instilled exogenous surfactant [57].

Characterizing barrier properties of mucus

The airway mucus gel layer acts as a solid physical barrier to foreign pathogens, toxins and environmental ultrafine particles while allowing rapid passage of selective small molecules, ions, capsid viruses and many proteins. These selective barrier properties of airway mucus are intimately related to its viscoelasticity, which shows order-of-magnitude variations in healthy versus diseased states. The rheological characterization of airway mucus has contributed greatly to both the understanding of mucocilliary clearance and the quantitation of the severity of airway diseases such as CF, COPD and chronic bronchitis [13].

For example, Hill et al. reported that the mean solids concentration (% solids by weight including salts, denoted wt%) of sputum for normal subjects is 1.7%, whereas the sputum wt% of COPD and CF subjects are ~ 2× higher (3.5%) and ~ 4× higher (7.0%), respectively. Below 3.0 wt%, the loss (viscous) modulus of human bronchial epithelial (HBE) cell culture mucus dominates the storage (elastic) modulus across all frequencies ranging from ~ 0.1 Hz (tidal breathing frequency) to ~ 10 Hz (ciliary beating frequency), implying that low wt% mucus is a viscoelastic solution; at 4.0 wt%, the elastic and viscous moduli were almost equal over the above frequency range, suggestive of the beginning of a transition from solution-like to gel-like behaviour; and at 5.0 wt%, the elastic modulus dominated the viscous modulus, suggesting that high solids wt% mucus is a viscoelastic gel at frequencies above 0.1 Hz [58]. These findings provide key data linking increased mucus solids concentration to the observation of Puchelle and Zahm [59] that when cilia beat against a fluid viscosity of higher than 100 cP, the ciliary beating frequency decreases and mucus clearance slows. In addition, Button and coworkers recently found that mucus concentration was also strongly correlated with the mucus-epithelial surface adhesive and mucus cohesive strengths. The increased mucus concentration and viscous energy dissipation in CF and COPD patients therefore make the cough mechanism fail to effectively clear accumulated mucus from the lungs [60].

The gel-on-brush model of the mucus clearance system by Button et al. demonstrated that the mucus and pericilliary brush layers must be in relative osmotic modulus balance for effective mucociliary clearance [20]. At a mucus solids concentration of ~ 5 wt% or above, the corresponding osmotic modulus of the mucus layer begins to exceed that of the PCL, therefore collapsing the PCL and slowing down the mucus clearance observed in diseases. Kesimer et al. tested the relationships predicted by the gel-on-brush model between total mucin concentration and the increase in severity of chronic bronchitis [61]. The mean total mucin concentrations were higher in current or former smokers with severe COPD than in controls who had never smoked. The relationships between total mucin concentration and prospective annualized respiratory exacerbation showed that mucin concentrations were higher in participants who had exacerbations than in those who had none. These results suggest that airway mucin concentrations may serve as a biomarker for the diagnosis of chronic bronchitis.

Microrheology affords a detailed characterization of the barrier properties of airway mucus at a scale relevant to pathogens, toxins, and foreign particles. When the scale approaches the mesh size of the mucus layer, the diffusion rates of particles are expected to be reduced due to steric or adhesive forces, thus leading to a higher apparent viscosity. A variety of conventional nanoparticle-based drug delivery systems for CF and other pulmonary diseases have been discouraged by the mucus barrier since nanoparticles are usually subjected to mucociliary clearance before they reach airway mucosal surfaces due to the extremely slow diffusion rates of these particles in the mucus. As such, to engineer nanoparticles capable of penetrating this highly viscoelastic and adhesive mucus barrier, it is imperative to characterize the local viscoelasticity of mucus at scales relevant to nanoparticle delivery systems. Suk and coworkers investigated the effect of nanoparticle size and surface chemistry on transport rates in fresh, undiluted CF sputum. They found that the transport rates of 200 nm particles that were densely coated with low molecular weight polyethylene glycol (PEG) were 90-fold faster than the same-sized uncoated particles. On the other hand, the movement of both coated and uncoated 500 nm particles was strongly hindered. Therefore, by tracking the displacement of 200 nm PEG-coated particles, they further showed that the local microviscosities of fresh, undiluted cystic fibrosis sputum were only 5-fold higher than the viscosity of water, while the bulk viscosity was 10,000-fold higher at low shear rates. Additionally, these authors estimated the average mesh spacing of physiological human CF sputum to be 140 ± 50 nm [7]. In light of these findings, investigators have further designed various muco-inert nanoparticles that can rapidly penetrate the mucus layer, thus enhancing the efficacy of drug and gene delivery at mucosal surfaces [62, 63].

Techniques for measuring the rheology of airway surface liquid

When selecting an appropriate technique to investigate the viscosity of airway surface liquid, it is important to keep in mind that airway surface liquid has two particular features: 1) a relatively small available sample volume and 2) large variations in the range of viscosity depending on the patient, the sampling site in the lung, and healthy or diseased conditions [7, 13,14,15,16, 26, 27, 29, 59, 64]. Commonly used instruments for the measurement of viscosity include glass capillary viscometers, falling sphere viscometers, rotational viscometers, magnetic microrheometers and particle tracking microrheometers. Among these instruments, the first four have been used to determine the macroscopic bulk viscosity, while the last one has been applied to the study of microrheology.

Glass capillary viscometers

Glass capillary viscometers are also known as tube-type viscometers, which consist of a U-shaped glass tube with a reservoir bulb on one arm of the U and a measuring bulb with a precise narrow bore (the capillary) on the other. There are two calibrated marks along the length of the capillary. During use, liquid is suctioned into the measuring bulb and then allowed to flow downward through the capillary into the reservoir. The time it takes for the liquid to pass between two marks is a measure of the viscosity η [65]. The principle of the glass capillary viscometer is based on the Poiseuille law [66]:

$$ \eta =\frac{\pi {r}^4\varDelta Pt}{8 LV} $$
(2)

where r is the radius of the capillary tube, ΔP is the pressure difference between the two ends of the capillary tube, t is the time it takes for a volume V of fluid to elute, and L represents the length of the capillary tube. Theoretically, the more viscous the liquid, the longer it takes to flow.

Basch et al. first used a capillary viscometer to measure sputum viscosity, but their results were later demonstrated to be unreliable by Forbes and Wise [67, 68]. Despite a variety of modified versions later, using a wide-bore or horizontal tube, for instance, the measurements with sputum were still widely scattered and not reproducible, as the sputum frequently either slipped through the tube as a solid plug or remained stuck somewhere in the tube. In addition, lung fluids such as mucus and sputum behave as a non-Newtonian viscosity that is dependent on shear rate. Thus, capillary viscometers are further limited since they can only measure viscosity for one shear rate at a time.

Falling-sphere viscometers

Stokes’ law is the basis of the falling-sphere viscometer, in which the fluid under examination is stationary in a vertical or inclined glass tube. A small sphere is allowed to move through the test fluid. As the falling velocity of the sphere increases, the frictional force also increases, and eventually, a terminal velocity Vs is reached when the gravitational force is balanced with the buoyant force and this frictional force. The viscosity η of the test fluid can be calculated by Stokes’ law:

$$ \eta ={d}^2\left({\rho}_s-{\rho}_f\right)g/18{V}_s $$
(3)

where d is the sphere diameter, ρs is the sphere density, ρf is the fluid density, and g is the local gravitational acceleration [69].

Falling sphere viscometers have undergone important modifications over the years; some commercially available instruments, for example, use cylindrical needles or pistons with hemispheric ends instead of spheres [65]. Unlike the traditional falling sphere viscometer that only applies for viscosity measurements of Newtonian fluids, the falling needle also possesses the ability to measure non-Newtonian rheological parameters [70]. In terms of sputum viscosity, there are several drawbacks associated with the falling sphere viscometers, including the requirement of a significant sample volume, operation at low shear rates and poor measurement stability and reproducibility. For instance, Elmes and White measured sputum viscosity employing a rolling ball viscometer and found that the ball moved along the line of least resistance and rolled around the aggregation of viscous material suspended in the sputum [71].

Rotational viscometers

Rotational viscometers use the concept that viscosity is defined as the ratio of shear stress to shear rate. They measure the torque required to rotate an immersed element (the spindle) in a fluid at a known speed. The spindle is driven by a motor through a calibrated spring. By utilizing a multiple speed transmission and interchangeable spindles, a wide range of viscosities can be measured, thus enhancing the versatility of the instruments.

There are two basic types of rotational viscometers, one with two coaxial cylinders and the other with a cone and plate [65, 66]. In the cylinder viscometer, the liquid to be tested is placed in a narrow space between the rotating cylinder and the fixed cylinder. The more viscous the fluid is, the greater the torque required to spin the rotating cylinder. The primary disadvantage of the cylinder viscometer is the relatively large sample volumes required. For example, despite the use of specifically designed Small Sample Adapter, commercial Wells-Brookfield cylinder viscometers still require a sample volume of 2 to 16 mL. Baldry and Josse found that the rotating cylinder did not move at all or rotated with a very low speed when sputum viscosity was relatively high [66]. For these reasons, they are not extensively used in clinical laboratories.

In the cone and plate viscometer, a nearly flat cone (with cone angle between 0.8° and 3°) in close proximity to a plate forms a narrow gap where the liquid is contained (Fig. 3) [65]. This cone and plate spindle geometry requires a sample volume of only 0.5 to 2.0 mL and generates a constant shear rate at all locations under the cone at any given rotational speed. The viscosity can easily be determined from shear stress (the torque) and shear rate (the angular velocity) by the following equation:

$$ \eta =\frac{3M\sin \theta }{2\pi {r}^3\omega } $$
(4)

where M is the torque input by the instrument, θ is the cone angle, r is the cone radius, and ω is the angular velocity.

Fig. 3
figure 3

Schematic representation of a cone and plate viscometer

Furthermore, both the elastic and viscous characteristics of the material can be studied by using a strain-controlled cone and plate rheometer [72, 73]. In this type of device, a motor is designed to impose a sinusoidal strain γ(t) to a material, and the resultant stress σ(t) is measured with a transducer. The phase angle (δ) between stress and strain is used to decompose the measured stress into an in-phase component and to determine the elastic modulus G of a material, defined as the in-phase stress divided by the amplitude of the strain, \( {G}^{\prime }=\frac{\hat{\sigma}}{\hat{\gamma}} cos\delta \), where \( \hat{\sigma} \) and \( \hat{\gamma} \) are the amplitudes of the stress and strain, respectively. Similarly, the viscous modulus G′′ of the material is defined as the out-of-phase stress divided by the amplitude of the strain, \( {G}^{\prime \prime }=\frac{\hat{\sigma}}{\hat{\gamma}} sin\delta \).

The cone and plate viscometer has been widely employed in the measurement of the rheological properties of airway surface liquid [26, 64, 66, 74]. As a result of measurements taken with a cone and plate viscometer, Baldry and Josse showed that comparable readings could be obtained with duplicate sputum samples at different shear rates [66]. Lieberman found that sputum viscosity could reach a relatively steady state after a limited amount shearing in a cone and plate viscometer [74]. Similarly, King et al. investigated the bulk shear viscosities of aqueous dispersions of calf lung surfactant in a cone and plate viscometer, which showed that the lung surfactant exhibited a complex non-Newtonian behaviour, with higher viscosities at low shear rates [75].

Magnetic microrheometer

The magnetic micrcorheometer involves a pair of permanent magnets or electromagnets for generating a rotating magnetic field [13, 76,77,78]. The test fluid sample is placed in a small test tube with a concave and clear bottom. A metal microsphere is inserted in the sample. The tube, sealed to prevent evaporation of the sample, is centred between the two magnets. The rotating magnetic field generates a magnetic driving force that rotates the metal sphere. In the case of low frequencies and small sphere diameters, the sphere inertia can be neglected. Therefore, the angular speed of the sphere is determined by the rotational speed and strength of the magnetic field as well as the viscosity of the sample around the sphere. The motion of the sphere is monitored by a high-resolution video microscope set below the sample cell. The torque exerted on the sphere is proportional to the difference between the angular velocity of the magnetic field ΩB and that of the sphere ΩS, and the shear rate of the flow is linearly proportional to ΩS. Thus, the viscosity of the test sample can be calculated depending on the shear rate by measuring ΩS at different values of ΩB [78]. In another type of displacement magnetic microrheometer, a micron-sized magnetic bead is carefully positioned in the sample and oscillated by means of an electromagnet at a variable frequency (ω) and amplitude [16, 79, 80]. Images of the bead are captured by a charge-coupled device (CCD) camera to measure bead displacement. The amplitude of the displacement of the bead and its phase shift with respect to the magnetic force are determined to calculate the elastic modulus G and viscous modulus G′′ of the sample using:

$$ {G}^{\prime }=\frac{f_0}{6\pi a{X}_0\left(\omega \right)}\cos \varphi \left(\omega \right) $$
(5)
$$ {G}^{\prime \prime }=\frac{f_0}{6\pi a{X}_0\left(\omega \right)}\sin \varphi \left(\omega \right) $$
(6)

where f0 is the amplitude of the applied force, a is the radius of the bead, X0(ω) is the frequency-dependent amplitude of the bead displacement, and φ(ω) is the frequency-dependent phase shift between the force and the bead displacement.

The two remarkable features of the magnetic microrheometer are the need for only microlitre quantities of sample volume and freedom from contamination. Consequently, it is well suited to the investigation of the rheological properties of lung fluids partially because only small lung fluid samples can be obtained in normal or disease conditions. King and coworkers pioneered the use of a magnetic rheometer to determine the viscoelastic properties of normal tracheal mucus from canines and discussed the significance of these rheological behaviours in terms of the clearance of secretions from the lung [77].

Particle tracking microrheometer

Particle tracking microrheology can be used to characterize the linear viscoelasticity of complex fluids with the accuracy of bulk rheology measurements but with smaller sample volumes on the order of picolitres to microlitres required [13, 81, 82]. A modern experimental set up to perform particle tracking microrheology experiments primarily consists of a light source, a colloidal probe, optical microscopy, a fast COMS camera, and specialized particle tracking software.

Colloidal spheres are embedded into a soft viscoelastic fluid, and movies are made of the Brownian motion of the colloidal probes in the sample using the fast COMS camera. The positions of the centroids of the colloidal probes are subsequently matched frame by frame using a specialized routine to identify each particle and generate its trajectory. Then, the mean squared displacements (MSD),〈Δr2(t)〉, of individual particles are calculated from the colloidal sphere trajectories.

Mathematical analysis of MSD can provide a measure of the linear viscoelasticity of the test fluid as a function of time or frequency. The simplest method, for example, is to calculate the creep compliance J(t) in the form [81,82,83,84]:

$$ J(t)=\frac{\pi a}{kT}\left\langle \varDelta {r}^2(t)\right\rangle $$
(7)

where kT is the thermal energy and a is the radius of the particle. A purely viscous liquid of shear viscosity η, such as water or glycerol, subjected to a constant stress, exhibits a creep compliance that increases linearly with time, J(t) = t/η; a highly elastic material of modulus G0 under stress exhibits a time-independent compliance J = 1/G0; the time-dependent compliance of a viscoelastic material such as mucus or cytoplasm shows an intermediate behaviour [13, 83].

In fact, people often prefer to work with the complex shear modulus (G(ω) = G(ω) + iG′′(ω)) since its real and imaginary parts more clearly define the contributions of elasticisty (G) and dissipation (the viscosity, η = G′′/ω as ω → 0) to the viscoelasticity response [81]. The G(ω) of the complex fluids can be obtained from measurements of the time-dependent mean square displacement, 〈r2(t)〉, of thermally driven colloidal spheres suspended in the fluid using a generalized Stokes-Einstein (GSE) equation [85]. The frequency-domain representation of the GSE equation takes the following form:

$$ {\mathrm{G}}^{\ast}\left(\upomega \right)=\frac{\mathrm{kT}}{\uppi \mathrm{a}\mathrm{i}\upomega \mathrm{Fu}\left\{\left\langle \Delta {\mathrm{r}}^2\left(\mathrm{t}\right)\right\rangle \right\}}=\frac{\mathrm{kT}}{6\uppi \mathrm{a}{\mathrm{D}}^{\ast}\left(\upomega \right)} $$
(8)

where D is the time-dependent diffusion coefficient and Fu is the Fourier transform. Consider spheres diffusing in a purely viscous fluid or a viscoelastic material; 〈r2(t)〉 can be calculated by Eq. (9) and (10), respectively,

$$ \left\langle \Delta {\mathrm{r}}^2\left(\mathrm{t}\right)\right\rangle =6\mathrm{Dt} $$
(9)
$$ \left\langle \Delta {\mathrm{r}}^2\left(\mathrm{t}\right)\right\rangle ={\mathrm{r}}_0^2\left[1-\exp \left(-6\mathrm{Dt}/{\mathrm{r}}_0^2\right)\right] $$
(10)

where \( {\mathrm{r}}_0^2 \) is the saturation value of the mean square displacement of the spheres as time approaches infinity. Combining Eq. (8) with (9) or (10), the frequency-independent viscosity is recovered \( \upeta =\frac{\mathrm{kT}}{6\uppi \mathrm{aD}} \).

Dawson et al. used multiple particle tracking to determine the effective viscoelastic properties of human cystic fibrotic sputum at the micron scale. They found that CF sputum microviscosity was an order of magnitude lower than its macroviscosity, suggesting that the enhanced viscoelasticity of CF sputum correlates with the increased microheterogeneity in particle transport [28]. A primary problem with a particle tracking micorheology-based characterization of airway mucus is a possible overestimation of the true mucus viscoelasticity due to adhesive interactions between colloidal probes and mucus. In addition, the maximally achievable viscosity and shear rate ranges are limited due to restrictions on particle sizes and the temporal resolution of tracking, respectively.

Methods for measuring surface tension of airway surface liquid

Over the past few decades, a variety of measuring techniques have been developed for determining the surface activity of surfactant materials derived from the lung. Among these are film balances, bubble methods and drop shape analysis methods. In addition, the surface tension of pulmonary surfactant can be inferred from pressure-volume data. In the following section, we will discuss classical methods and recent techniques in terms of their basic principles, advantages and limitations to help research workers select the method(s) best suited to their needs.

The Langmuir-Wilhelmy balance

Clements first introduced the Langmuir-Wilhelmy surface balance to determine the surface tension-area relationship in his pioneering studies of lung extracts [86, 87]. In this method, lung extracts are dropped onto the surface of a subphase substance (usually normal saline) contained in the trough of the surface balance, while the exposed surface area is varied over a wide range by means of a movable barrier. A roughened and clean platinum plate is attached to a balance with a thin metal wire. When the plate is perpendicularly dipped into the liquid, the downward force (F) on it due to wetting is measured by a balance connected to a transducer. The surface tension (γ) can then be calculated using the following equation:

$$ \upgamma =\frac{F}{L\bullet \cos \left(\theta \right)} $$
(11)

where L is the wetted perimeter of the plate and θ is the contact angle between the liquid phase and the plate. In practice, the platinum plate is assumed to be completely wetted, and therefore, the contact angle is 0°. In this way, Clements and Brown showed that the surface tension of a lung-derived surface approached a lower limiting of 10~15 dyn/cm and an upper limiting of 40~50 dyn/cm [86, 88].

Although the Langmuir-Wilhelmy balance is one of the most commonly used tools for measuring surface tension, there are still some drawbacks to this apparatus in terms of investigating the tension-area behaviour of lung extracts. One of the most intractable problems has been the film leakage that occurs on the surfaces of the restraining walls and barrier, causing experimental artefacts [89]. Furthermore, large sample volumes are required on the Langmiur-Wilhelmy surface balance because of its large size. Finally, the apparatus does not seem readily adaptable to rapid oscillations of surface area at rates corresponding to a normal cycle of breathing [57].

Captive bubble method

The captive bubble method developed by Schürch and coworkers is useful for reproducing the in situ behaviour of lung surfactant monolayers because it eliminates the possibility of surface film leakage [90,91,92,93]. In this method, a lung surfactant suspension is placed into a glass flow-through chamber, and a bubble of atmospheric air is introduced and allowed to float against the slightly concave hydrophilic agarose ceiling of the chamber (Fig. 4) [90, 94]. The bubble volume is compressed and expanded by varying the fluid pressure in the chamber. From the beginning of adsorption measurements, the chamber pressure is rapidly reduced to an estimated level at which the bubble just doubles its diameter from 2~3 mm to 4~6 mm. Then, the captive bubble is submitted to a number of compression-expansion cycles. As the pressure is increased, the bubble volume and surface area are decreased, compressing the absorbed surfactant monolayer at the air-water interface, during which the bubble progressively flattens, indicating a lower surface tension.

Fig. 4
figure 4

Captive bubble chamber. A lung surfactant suspension (A) is placed into a glass flow-through chamber, and a captive bubble (B) is formed by a syringe within the aqueous phase and then allowed to float against the agarose gel (G) ceiling. After the stopcock (S) is closed, B can be compressed or expanded by withdrawing fluid through the pressure control port (P) [94]

Images of the bubble are recorded by a computer digital image system. By measuring the height and diameter of the bubble in the video picture, the surface tension and area can be calculated using the approach of Malcolm and Elliott or the formulas of Rotenberg [94]. One important feature of the captive bubble approach is that low and stable surface tensions of 1~2 mN/m can be obtained upon the first quasi-static or dynamic compression following adsorption, and therefore, it is suitable for studies on the role of surfactant proteins in surfactant formation [90, 94].

Microdroplet method

Schürch et al. reported the use of the microdroplet method to measure alveolar surface tension directly in an excised lung [92, 95,96,97]. The method is based on the observation that a droplet of a nonpolar test liquid resting on top of a monolayer at the air-water interface changes its shape from a sphere to a thin lens as the interfacial tension is raised beyond the value characteristic of the test fluid. In a prior calibration experiment, test fluid droplets with known surface tension are placed onto a surfactant film on a fluid substrate in the surface balance. The shape of the droplets is changed by varying the surface tension of the film and is monitored with a microscope. The relative diameter of the lens (defined as the diameter normalized to its spherical shape) can thus be plotted as a function of the film surface tension. Conversely, if the relative diameter is known, the film surface tension can be determined with this calibration curve.

In the subsequent measurements of alveolar surface tension in situ, the alveolus is punctured by a micropipette with a tip diameter of 1~2 μm, and then a test fluid droplet of fluorocarbon liquid or silicone oil is deposited onto the alveolar surface. The diameter of the spherical drop from the pipette before deposition is usually 10~100 μm [95]. After deposition onto the alveolar surface, the drop spreads out into a lens shape whose diameter depends on the alveolar surface tension. By using the microdroplet method, Schürch et al. confirmed that the surface tension in the excised rat lungs at functional residual capacity was between 1 and 10 mN/m [97] and further showed that the maximum surface tension at TLC was approximately 30 mN/m [96]. Most importantly, Schürch et al. found that in excised cat lungs at a given volume between 40 and 70% of TLC, no difference in surface tension could be detected with respect to alveolar size and location [95].

Pendant drop method

The pendant drop method is extensively used for surface tension measurement, even at elevated temperatures and pressures [98,99,100,101]. In the pendant drop setup, a drop is formed at the tip of a stainless-steel needle. The volume and surface area are controlled by moving the plunger of the syringe connected to a stepper motor. The pendant drop is constrained inside a glass cuvette to maintain controllable environmental conditions (i.e., temperature and pressure). The drop images are magnified by a horizontally mounted microscope and then acquired using an automatic camera. Then, these images are digitalized and stored in a computer for further extraction of the pendant drop profile to calculate the surface tension [102, 103].

In the traditional method, the maximum diameter DE is measured from the drop profile and the diameter DS at a horizontal plane at a distance DE from the bottom of the drop (Fig. 5), after which the surface tension is calculated by the equation

$$ \upgamma =\Delta \uprho \mathrm{g}{D}_E^2/H $$
(12)

where ∆ρ is the density difference between the drop and the surrounding medium, g is the gravitational acceleration, and H is a function of S = DS/DE, which can be read from a table [99, 104]. Axisymmetric drop shape analysis (ADSA) has recently been developed as a more accurate and applicable technique for measuring surface tension based on acquired drop images. Briefly, ADSA matches the experimental drop profile through an optimization procedure to a theoretical profile calculated from the Laplace equation of capillarity. The matched theoretical drop profile is used to find the surface tension of the pendant drop [99, 100, 102, 103].

Fig. 5
figure 5

Geometry of the pendant drop method

The pendant method offers several distinct advantages over the conventional film balance, including a high degree of automation, a small sample size, and the ease with which the sample can be isolated from the environment to protect it from contamination [89, 100]. Since the pendant drop method does not suffer a similar restriction in compression rate compared with the Langmuir-Wilhelmy surface balance, it has been used to examine the rate dependence of the surface pressure-surface area isotherm of a DPPC monolayer. The results obtained by the pendant drop method are in good agreement with those from the conventional film balance measurements [103, 105].

Pressure-volume measurements

Some early studies used pressure-volume data from air- and liquid-filled excised lungs to calculate alveolar surface tension [106,107,108,109,110]. For example, assuming a maximum surface tension and a constant relationship between lung surface area and lung volume, Bachofen and coworkers [110] derived an equation that relates the surface tension (γ) to lung volme (V) and the component of recoil pressure (Ps) caused by surface tension in the form

$$ \upgamma =\frac{3}{2\mathrm{k}}{\mathrm{P}}_{\mathrm{s}}{\mathrm{V}}^{1/3} $$
(13)

where k is the shape factor, which can be found from the maximum surface tension. Thus, for each pair of air and saline P-V curves, the corresponding surface tension-surface area can be obtained by the above equation. Smith and Stamenovic [109] provided another approach to deduce values of alveolar surface tension from pressure-volume data. Their approach rests on the assumption that at a given volume, Ps is uniquely determined by γ. First, the pressure-volume curves of control air-filled lungs and test liquid-filled lungs with a limited set of fixed interfacial tensions are measured. As the intersections between the curves with the normal and test liquid interfaces define states of equal surface tension, by pooling data from all intersections, the surface tension-lung volume relationships are obtained. In addition, Wilson used an energy analysis method to calculate the values of surface tension from pressure-volume loops. The reader is referred to reference [108] for a detailed description.

Conclusions

In recent years, measuring the rheological properties of airway surface liquid has attracted considerable clinical attention due to new developments in microrheology instruments and methods. The quantitative characterization of the viscoelasticity and surface tension of airway surface liquid contributes to a better understanding of physiological processes such as airway mucus trapping and clearance and ciliary action to further identify potential markers for ranking the severity of relevant muco-obstructive lung diseases and to develop muco-inert nanoparticle systems for improved drug and gene delivery to mucosal tissues; a good knowledge of lung surfactant dynamics to improve surfactant replacement therapy for respiratory distress in neonates and even adults; and a deep insight into the micromechanical mechanisms of VILI to prescribe personalized mechanical ventilation in ARDS patients. In terms of measurements of viscosity and surface tension of airway surface liquid, the cone and plate system is currently the most commonly used instrument for determining bulk viscosity in clinical practice, while multiple particle tracking is more suitable for probing the microviscosity of lung fluids. In light of the use of the axisymmetric drop shape analysis algorithm and the rapid development of data acquisition and image processing techniques, pendant drop methods have seen a broad prospective application in the surface tension measurement of airway surface liquid.

Availability of data and materials

Not Applicable.

Abbreviations

ADSA:

Axisymmetric drop shape analysis

ALI:

Acute lung injury

ARDS:

Acute respiratory distress syndrome

CCD:

Charge-coupled device

CF:

Cystic fibrosis

COPD:

Chronic obstructive pulmonary disease

DPPC:

Dipalmitoyl phosphatidyl choline

FRC:

Functional residual capacity

GSE:

Generalized Stokes-Einstein equation

HBE:

Human bronchial epithelial

INSURE:

INtubtion-SURfactant-Extubation

MSD:

Mean squared displacement

NRDS:

Neonatal respiratory distress syndrome

PEG:

Polyethylene glycol

PS:

Pulmonary surfactant

SP:

Surfactant protein

SRT:

Surfactant replacement therapy

TLC:

Total lung capacity

VILI:

Ventilator-induced lung injury

References

  1. Lumb A, PF. Chapter 3 Elastic forces and lung volumes. In: Nunn's applied respiratory physiology. 7th Ed ed: Televise Health Sciences. Churchill Livingstone Elsevier; 2010. ISBN 9780702029967.

  2. Widdicombe JH, Widdicombe JG. Regulation of human airway surface liquid. Respir Physiol. 1995;99(1):3–12.

    Article  CAS  PubMed  Google Scholar 

  3. Widdicombe JH, et al. Regulation of depth and composition of airway surface liquid. Eur Respir J. 2010;201(4):313–8.

    Google Scholar 

  4. Widdicombe JH, Wine JJ. Airway gland structure and function. Physiol Rev. 2015;95(4):1241.

    Article  CAS  PubMed  Google Scholar 

  5. Widdicombe J. Airway and alveolar permeability and surface liquid thickness: theory. J Appl Physiol. 1997;82(1):3–12.

    Article  CAS  PubMed  Google Scholar 

  6. Cribb JA, et al. Nonlinear signatures of entangled polymer solutions in active microbead rheology. J Rheol (N Y N Y). 2013;57:1247.

    Article  CAS  Google Scholar 

  7. Suk JS, et al. The penetration of fresh undiluted sputum expectorated by cystic fibrosis patients by non-adhesive polymer nanoparticles. Biomaterials. 2009;30(13):2591–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  8. John VF, Burton DF. Airway mucus function and dysfunction. N Engl J Med. 2010;363(23):2233–47.

    Article  Google Scholar 

  9. Levy R, et al. Pulmonary fluid flow challenges for experimental and mathematical modeling. Integr Comp Biol. 2014;54(6):985.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  10. Nkadi PO, Merritt TA, Pillers D-AM. An overview of pulmonary surfactant in the neonate: genetics, metabolism, and the role of surfactant in health and disease. Mol Genet Metab. 2009;97(2):95–101.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  11. Veldhuizen EJA, Haagsman HP. Role of pulmonary surfactant components in surface film formation and dynamics. BBA. 2000;1467(2):255–70.

    Article  CAS  PubMed  Google Scholar 

  12. Cone RA. Barrier properties of mucus . Adv Drug Deliv Rev. 2009;61(2):75–85.

    Article  CAS  PubMed  Google Scholar 

  13. Lai SK, et al. Micro- and macrorheology of mucus . Adv Drug Deliv Rev. 2009;61(2):86–100.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  14. Rubin BK, et al. Collection and Analysis of respiratory mucus from subjects without lung disease. Am Rev Respir Dis. 1990;141(1):1040–3.

    Article  CAS  PubMed  Google Scholar 

  15. Puchelle E, Zahm JM, Quemada D. Rheological properties controlling mucociliary frequency and respiratory mucus transport. Biorheology. 1987;24(6):557–63.

    Article  CAS  PubMed  Google Scholar 

  16. Jeanneret-Grosjean, A., ., et al., Sampling technique and rheology of human tracheobronchial mucus. Am Rev Respir Dis, 1988. 137(3): p. 707–710.

    Article  CAS  PubMed  Google Scholar 

  17. Woodring, et al. Types and mechanisms of pulmonary atelectasis : Atelectasis, Part 1. J Thorac Imaging. 1996;11(2):92–108.

    Article  CAS  PubMed  Google Scholar 

  18. Li Bassi G, et al. Following tracheal intubation, mucus flow is reversed in the semirecumbent position: possible role in the pathogenesis of ventilator-associated pneumonia. Crit Care Med. 2008;36(2):518–25.

    Article  PubMed  Google Scholar 

  19. Wong JW, et al. Effects of gravity on tracheal mucus transport rates in normal subjects and in patients with cystic fibrosis. Pediatrics. 1977;60(2):146–52.

    CAS  PubMed  Google Scholar 

  20. Button B, et al. A Periciliary brush promotes the lung health by separating the mucus layer from airway epithelia. Science. 2012;337(6097):937.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  21. Polin RA, Carlo WA. Surfactant replacement therapy for preterm and term neonates with respiratory distress. Pediatrics. 2014;133(1):156.

    Article  PubMed  Google Scholar 

  22. Thompson BT, Chambers RC, Liu KD. Acute respiratory distress syndrome. N Engl J Med. 2017;377(6):562–72.

    Article  CAS  PubMed  Google Scholar 

  23. Sweeney RM, Mcauley DF. Acute respiratory distress syndrome. Lancet. 2016;388(10058):2416–30.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Sweet DG, et al. European consensus guidelines on the Management of Respiratory Distress Syndrome - 2019 update. Neonatology. 2019;115(4):432–50.

    Article  PubMed  Google Scholar 

  25. Papazian L, et al. Formal guidelines: management of acute respiratory distress syndrome. Ann Intensive Care. 2019;9(1):69.

    Article  PubMed  PubMed Central  Google Scholar 

  26. Puchelle E, Zahm JM, Aug F. Viscoelasticity, protein content and ciliary transport rate of sputum in patients with recurrent and chronic bronchitis. Biorheology. 1981;18(3–6):659–66.

    Article  CAS  PubMed  Google Scholar 

  27. Baconnais, S., ., et al., Ion composition and rheology of airway liquid from cystic fibrosis fetal tracheal xenografts. Am J Respir Cell Mol Biol, 1999. 20(4): p. 605–611.

    Article  CAS  PubMed  Google Scholar 

  28. Dawson M, Wirtz D, Hanes J. Enhanced viscoelasticity of human cystic fibrotic sputum correlates with increasing microheterogeneity in particle transport. J Biol Chem. 2003;278(50):50393–401.

    Article  CAS  PubMed  Google Scholar 

  29. Feather EA, Russell G. Sputum viscosity in cystic fibrosis of the pancreas and other pulmonary diseases. Br J Dis Chest. 1970;64(4):192–200.

    Article  CAS  PubMed  Google Scholar 

  30. Heil M, Hazel AL, Smith JA. The mechanics of airway closure. Respir Physiol Neurobiol. 2008;163(1):214–21.

    Article  PubMed  Google Scholar 

  31. Grotberg JB, Jensen OE. Biofluid mechanics in flexible tubes. Annu Rev Fluid Mech. 2004;36(1):121–47.

    Article  Google Scholar 

  32. Kamm RD, Schroter RC. Is airway closure caused by a liquid film instability? Respir Physiol. 1989;75(2):141–56.

    Article  CAS  PubMed  Google Scholar 

  33. Otis DR, et al. Role of pulmonary surfactant in airway closure: a computational study. J Appl Physiol. 1993;75(3):1323–33.

    Article  PubMed  Google Scholar 

  34. Halpern D, Grotberg JB. Nonlinear saturation of the Rayleigh instability due to oscillatory flow in a liquid-lined tube. J Fluid Mech. 2003;492(492):251–70.

    Article  CAS  Google Scholar 

  35. White JP, Heil M. Three-dimensional instabilities of liquid-lined elastic tubes: a thin-film fluid-structure interaction model. Phys Fluids. 2005;17(3):529.

    Article  CAS  Google Scholar 

  36. Halpern D, Fujioka H, Grotberg JB. The effect of viscoelasticity on the stability of a pulmonary airway liquid layer. Phys Fluids. 2010;22(1):2333.

    Article  CAS  Google Scholar 

  37. Halpern, D., . and J.B. Grotberg, Surfactant effects on fluid-elastic instabilities of liquid-lined flexible tubes: a model of airway closure. J Biomech Eng, 1993. 115(3): p. 271–277.

    Article  CAS  PubMed  Google Scholar 

  38. Iii, D.P.G., et al., The significance of air-liquid interfacial stresses on low-volume ventilator-induced lung injury. 2006.

    Google Scholar 

  39. Gaver DP, Samsel RW, Solway J. Effects of surface tension and viscosity on airway reopening. J Appl Physiol. 1990;69(1):74–85.

    Article  PubMed  Google Scholar 

  40. Slutsky AS, Ranieri VM. Ventilator-induced lung injury. N Engl J Med. 2013;369(22):2126–36.

    Article  CAS  PubMed  Google Scholar 

  41. Bates JHT, Smith BJ, Allen GB. Computational models of ventilator induced lung injury and surfactant dysfunction. Drug Discov Today Dis Models. 2015;15:17–22.

    Article  PubMed  Google Scholar 

  42. Zheng-long C, Ya-zhu C, Zhao-yan H. A micromechanical model for estimating alveolar wall strain in mechanically ventilated edematous lungs. J Appl Physiol. 2014;117(6):586–92.

    Article  Google Scholar 

  43. Perlman CE, Lederer DJ, Jahar B. Micromechanics of alveolar edema. Am J Respir Cell Mol Biol. 2011;44(1):34.

    Article  CAS  PubMed  Google Scholar 

  44. Bilek AM, Dee KC, Gaver DP. Mechanisms of surface-tension-induced epithelial cell damage in a model of pulmonary airway reopening. J Appl Physiol. 2003;94(2):770–83.

    Article  PubMed  Google Scholar 

  45. Kay SS, et al. Pressure gradient, not exposure duration, determines the extent of epithelial cell damage in a model of pulmonary airway reopening. J Appl Physiol. 2004;97(1):269–76.

    Article  PubMed  Google Scholar 

  46. Zheng-Long C, et al. An estimation of mechanical stress on alveolar walls during repetitive alveolar reopening and closure. J Appl Physiol. 2015;119(3):190.

    Article  Google Scholar 

  47. McEwan AD, Taylor GI. The peeling of a flexible strip attached by a viscous adhesive. J Fluid Mech. 2006;26(1):1–15.

    Article  Google Scholar 

  48. Laura C, et al. Hydraulic fracture during epithelial stretching. Nat Mater. 2015;14(3):343.

    Article  CAS  Google Scholar 

  49. Zuo YY, et al. Current perspectives in pulmonary surfactant — inhibition, enhancement and evaluation. Biochim Biophys Acta Biomembr. 2008;1778(10):1947–77.

    Article  CAS  Google Scholar 

  50. Speer CP, Sweet DG, Halliday HL. Surfactant therapy: past, present and future. Early Hum Dev. 2013;89:S22–4.

    Article  CAS  PubMed  Google Scholar 

  51. Walsh BK, et al. AARC Clinical Practice Guideline. Surfactant Replacement Therapy: 2013. Respir Care. 2013;58(2):367.

    Article  PubMed  Google Scholar 

  52. Gregory TJ, et al. Bovine surfactant therapy for patients with acute respiratory distress syndrome. Am J Respir Crit Care Med. 1997;155(4):1309–15.

    Article  CAS  PubMed  Google Scholar 

  53. Filoche M, Tai C-F, Grotberg JB. Three-dimensional model of surfactant replacement therapy. Proc Natl Acad Sci. 2015;112(30):9287.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  54. Halpern D, Jensen OE, Grotberg JB. A theoretical study of surfactant and liquid delivery into the lung. J Appl Physiol. 1998;85(1):333–52.

    Article  CAS  PubMed  Google Scholar 

  55. Zheng Y, et al. Effects of inertia and gravity on liquid plug splitting at a bifurcation. J Biomech Eng. 2006;128(5):707–16.

    Article  CAS  PubMed  Google Scholar 

  56. Zheng Y, et al. Effect of gravity on liquid plug transport through an airway bifurcation model. J Biomech Eng. 2005;127(5):798–806.

    Article  CAS  PubMed  Google Scholar 

  57. Uhlig S, Taylor AE. Methods in pulmonary research; 1998.

    Book  Google Scholar 

  58. Hill DB, et al. A biophysical basis for mucus solids concentration as a candidate biomarker for airways disease. PLoS One. 2014;9(2):e87681.

    Article  PubMed  PubMed Central  CAS  Google Scholar 

  59. Puchelle E, Zahm JM. Influence of rheological properties of human bronchial secretions on the ciliary beat frequency. Biorheology. 1984;21(1–2):265.

    Article  CAS  PubMed  Google Scholar 

  60. Button B, et al. Roles of mucus adhesion and cohesion in cough clearance. Proc Natl Acad Sci U S A. 2018;115(49):12501–6.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  61. Kesimer M, et al. Airway Mucin concentration as a marker of chronic bronchitis. N Engl J Med. 2017;377(10):911–22.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  62. Huckaby JT, Lai SK. PEGylation for enhancing nanoparticle diffusion in mucus. Adv Drug Deliv Rev. 2017;124:S0169409X17301813.

    Google Scholar 

  63. Samuel KL, et al. Rapid transport of large polymeric nanoparticles in fresh undiluted human mucus. Proc Natl Acad Sci U S A. 2007;104(5):1482–7.

    Article  CAS  Google Scholar 

  64. Feather EA, Russell G. Effect of tolazoline hydrochloride on sputum viscosity in cystic fibrosis. Thorax. 1970;25(6):732.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  65. Rosencranz R, Bogen SA. Clinical laboratory measurement of serum, plasma, and blood viscosity. Am J Clin Pathol. 2006;125 Suppl(Suppl 1):S78–86.

    PubMed  Google Scholar 

  66. Baldry PE, Josse SE. The measurement of sputum viscosity. Am Rev Respir Dis. 1968;98(3):392.

    CAS  PubMed  Google Scholar 

  67. Basch FP, Holinger P, Poncher HG. Physical and chemical properties of sputum: I. factors determining variations in portions from different parts of the tracheobronchial tree. Am J Dis Child. 1941;62(5):981–90.

    Article  Google Scholar 

  68. John Forbes MDL, FRCP, Wise BSL, Manc MB. Expectorants and sputum viscosity. Lancet. 1957;270(6999):767–70.

    Article  Google Scholar 

  69. Sutterby JL. Falling sphere viscometer. J Phys E Sci Instrum. 1973;6(10):1001.

    Article  CAS  Google Scholar 

  70. Davis AMJ, Brenner H. The falling-needle viscometer. Phys Fluids. 2001;13(10):3086–8.

    Article  CAS  Google Scholar 

  71. White JC, Elmes PC. The rheological problem in chronic bronchitis. Rheol Acta. 1958;1(2–3):96–102.

    Article  CAS  Google Scholar 

  72. Ma L, et al. Keratin filament suspensions show unique micromechanical properties. J Biol Chem. 1999;274(27):19145–51.

    Article  CAS  PubMed  Google Scholar 

  73. Tseng Y, et al. Micromechanics and ultrastructure of actin filament networks crosslinked by human fascin: a comparison with alpha-actinin. J Mol Biol. 2001;310(2):351–66.

    Article  CAS  PubMed  Google Scholar 

  74. Lieberman J. Measurement of sputum viscosity in a cone-plate viscometer. I. Characteristics of sputum viscosity. Am Rev Respir Dis. 1968;97(4):654.

    CAS  PubMed  Google Scholar 

  75. King DM, et al. Concentration-dependent, temperature-dependent non-Newtonian viscosity of lung surfactant dispersions. Chem Phys Lipids. 2001;112(1):11–9.

    Article  CAS  PubMed  Google Scholar 

  76. Chang-Young P, Omar AS. Image-based synchronization of force and bead motion in active electromagnetic microrheometry. Meas Sci Technol. 2014;25(12):125010.

    Article  CAS  Google Scholar 

  77. King M, Macklem PT. Rheological properties of microliter quantities of normal mucus. J Appl Physiol. 1977;42(6):797–802.

    Article  CAS  PubMed  Google Scholar 

  78. Keiji S, Taichi H, Maiko H. Electromagnetically spinning sphere viscometer. Appl Phys Express. 2010;3(1):016602.

    Article  CAS  Google Scholar 

  79. Ziemann F, Rädler J, Sackmann E. Local measurements of viscoelastic moduli of entangled actin networks using an oscillating magnetic bead micro-rheometer. Biophys J. 1994;66(6):2210–6.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  80. Tassieri M, et al. Analysis of the linear viscoelasticity of polyelectrolytes by magnetic microrheometry—pulsed creep experiments and the one particle response. J Rheol. 2010;54(1):117–31.

    Article  CAS  Google Scholar 

  81. Thomas Andrew W. Advances in the microrheology of complex fluids. Rep Prog Phys. 2016;79(7):074601.

    Article  CAS  Google Scholar 

  82. Mason TG, et al. Particle tracking microrheology of complex fluids. Phys Rev Lett. 1997;79(17):3282–5.

    Article  CAS  Google Scholar 

  83. Tseng Y, An KM, Wirtz D. Microheterogeneity controls the rate of gelation of actin filament networks. J Biol Chem. 2002;277(20):18143–50.

    Article  CAS  PubMed  Google Scholar 

  84. Apgar J, et al. Multiple-particle tracking measurements of heterogeneities in solutions of actin filaments and actin bundles. Biophys J. 2000;79(2):1095–106.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  85. Mason TG. Estimating the viscoelastic moduli of complex fluids using the generalized Stokes-Einstein equation. Rheol Acta. 2000;39:371–8.

    Article  CAS  Google Scholar 

  86. Clements JA. Surface tension of lung extracts. Proc Soc Exp Biol Med. 1957;95(1):170–2.

    Article  CAS  PubMed  Google Scholar 

  87. Clements JA, et al. Pulmonary surface tension and alveolar stability. J Appl Physiol. 1961;16(3):444–50.

    Article  CAS  PubMed  Google Scholar 

  88. Brown ES, Johnson RP, Clements JA. Pulmonary surface tension. J Appl Physiol. 1959;14(14):717.

    Article  CAS  PubMed  Google Scholar 

  89. Prokop RM, Neumann AW. Measurement of the interfacial properties of lung surfactant. Curr Opin Colloid Interface Sci. 1996;1(5):677–81.

    Article  CAS  Google Scholar 

  90. Schürch S, et al. Surface properties of rat pulmonary surfactant studied with the captive bubble method: adsorption, hysteresis, stability. Biochim Biophys Acta Biomembr. 1992;1103(1):127–36.

    Article  Google Scholar 

  91. Schoel WM, Schürch S, Goerke J. The captive bubble method for the evaluation of pulmonary surfactant: surface tension, area, and volume calculations. Biochim Biophys Acta. 1994;1200(3):281.

    Article  CAS  PubMed  Google Scholar 

  92. Schürch S, Bachofen H, Possmayer F. Surface activity in situ, in vivo, and in the captive bubble surfactometer. Comp Biochem Physiol A Mol Integr Physiol. 2001;129(1):195–207.

    Article  PubMed  Google Scholar 

  93. Putz, G., ., et al., Evaluation of pressure-driven captive bubble surfactometer. J Appl Physiol, 1994. 76(4): p. 1417–1424.

    Article  CAS  PubMed  Google Scholar 

  94. Schurch S, et al. A captive bubble method reproduces the in situ behavior of lung surfactant monolayers. J Appl Physiol. 1989;67(6):2389–96.

    Article  CAS  PubMed  Google Scholar 

  95. Schürch S. Surface tension at low lung volumes: dependence on time and alveolar size. Respir Physiol. 1982;48(3):339–55.

    Article  PubMed  Google Scholar 

  96. Schürch S, Goerke J, Clements JA. Direct determination of volume- and time-dependence of alveolar surface tension in excised lungs. Proc Natl Acad Sci U S A. 1978;75(7):3417–21.

    Article  PubMed  PubMed Central  Google Scholar 

  97. Schürch S, Goerke J, Clements JA. Direct determination of surface tension in the lung. Proc Natl Acad Sci U S A. 1976;73(12):4698–702.

    Article  PubMed  PubMed Central  Google Scholar 

  98. Berry JD, et al. Measurement of surface and interfacial tension using pendant drop tensiometry. J Colloid Interface Sci. 2015;454:226–37.

    Article  CAS  PubMed  Google Scholar 

  99. Hansen FK, Rødsrud G. Surface tension by pendant drop : I. a fast standard instrument using computer image analysis. J Colloid Interface Sci. 1991;141(1):1–9.

    Article  CAS  Google Scholar 

  100. Saad SMI, Policova Z, Neumann AW. Design and accuracy of pendant drop methods for surface tension measurement. Colloids Surf A Physicochem Eng Aspects. 2011;384(1):442–52.

    Article  CAS  Google Scholar 

  101. Franses EI, Basaran OA, Chang CH. Techniques to measure dynamic surface tension. Curr Opin Colloid Interface Sci. 1996;1(1):296–303.

    Article  CAS  Google Scholar 

  102. Chen P, et al. Axisymmetric drop shape Analysis (ADSA) and its applications. Adv Colloid Interface Sci. 2016;238(98):62.

    Google Scholar 

  103. Jyoti A, Prokop RM, Neumann AW. Manifestation of the liquid-expanded/liquid-condensed phase transition of a dipalmitoylphosphatidylcholine monolayer at the air-water interface. Colloids Surf B Biointerfaces. 1997;8(3):115–24.

    Article  CAS  Google Scholar 

  104. Fordham S. On the calculation of surface tension from measurements of pendant dram. Proc R Soc A. 1948;194(1036):1–16.

    Article  CAS  Google Scholar 

  105. Yu K, Yang J, Zuo YY. Automated Droplet Manipulation Using Closed-Loop Axisymmetric Drop Shape Analysis. Langmuir. 2016;32(19):4820–6.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  106. Bachofen H, et al. Relations among alveolar surface tension, surface area, volume, and recoil pressure. J Appl Physiol. 1987;62(5):1878–87.

    Article  CAS  PubMed  Google Scholar 

  107. Fisher MJ, Wilson MF, Weber KC. Determination of alveolar surface area and tension from in situ pressure-volume data. Respir Physiol. 1970;10(2):159–71.

    Article  CAS  PubMed  Google Scholar 

  108. Wilson TA. Surface tension-surface area curves calculated from pressure-volume loops. J Appl Physiol Respir Environ Exerc Physiol. 1982;53(6):1512.

    CAS  PubMed  Google Scholar 

  109. Smith JC, Stamenovic D. Surface forces in lungs. I. Alveolar surface tension-lung volume relationships. J Appl Physiol. 1986;60(4):1341–50.

    Article  CAS  PubMed  Google Scholar 

  110. Bachofen, H., ., J. Hildebrandt, ., and M. Bachofen, . Pressure-volume curves of air- and liquid-filled excised lungs-surface tension in situ. J Appl Physiol, 1970. 29(4): p. 422–431.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

The authors would like to thank Mr. Chunyuan Zhang for preparing the figures in the manuscript.

Funding

This study was supported by the National Natural Science Foundation of China.

(81770075).

Author information

Authors and Affiliations

Authors

Contributions

All authors substantially contributed to the drafting and review of this article. All authors read and approved the final manuscript.

Corresponding authors

Correspondence to Zhaoyan Hu or Yuanlin Song.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Chen, Z., Zhong, M., Luo, Y. et al. Determination of rheology and surface tension of airway surface liquid: a review of clinical relevance and measurement techniques. Respir Res 20, 274 (2019). https://doi.org/10.1186/s12931-019-1229-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12931-019-1229-1

Keywords