Low-Impedance, High Surface Area Pt-Ir Electrodeposited on Cochlear Implant Electrodes

, , , and

Published 22 May 2018 © The Author(s) 2018. Published by ECS.
, , Citation Curtis D. Lee et al 2018 J. Electrochem. Soc. 165 G3015 DOI 10.1149/2.0031812jes

1945-7111/165/12/G3015

Abstract

A high surface area, Pt-Ir coating was electrodeposited onto the contacts of Advanced Bionics HiFocus 1j electrode arrays. The coating was imaged using optical and scanning electron microscopy (SEM) and the impedance was directly measured using clinically relevant voltage transients of biphasic pulses as short as 17 μs. The coating, which SEM showed to completely cover the underlying substrate, decreased the polarization impedance of 17 μs and 32 μs pulses by 91% and 93%, respectively.

Export citation and abstract BibTeX RIS

This is an open access article distributed under the terms of the Creative Commons Attribution 4.0 License (CC BY, http://creativecommons.org/licenses/by/4.0/), which permits unrestricted reuse of the work in any medium, provided the original work is properly cited.

Cochlear implants have successfully restored functional hearing to hundreds of thousands of individuals; nevertheless, there is still room from improvement. Battery life is limited, tissue encapsulation can insulate the electrode from the target spiral ganglion cells,1 and the spectral resolution of cochlear implants does not match that of natural hearing.2 Lowering the impedance of cochlear implant electrodes with a coating, such as electrodeposited Pt-Ir, could help mitigate these issues.

For example, low impedance electrodes can increase the amount of current injected without exceeding polarization potentials that can lead to corrosion, the production of potentially toxic byproducts, or tissue damage.3 This is useful because the cochlear implantation procedure and the electrode insertion process initiate a foreign body response, resulting in tissue encapsulation around part or all of the electrode array. This tissue capsule can partially insulate electrodes, leading to an increase in impedance and a weaker electric field at target neurons,4 which then require higher currents to activate.

Reducing electrode impedance could also enable the use of small, high-density microelectrode arrays, including those made with thin-film fabrication technology, which simulations have shown could improve spectral resolution and sound quality.5 Microelectrodes have not commonly been used for stimulation because their small geometric surface areas are associated with high impedance and—when injecting currents required for neural activation–high polarization potentials, which can cause delamination and corrosion.6 Lowering the impedance of electrodes would reduce polarization, which would enable the use of high-density microelectrode arrays and open up the possibility of using thin-film fabricated electrodes7 with their concomitant advantages of batch fabrication and high dimensional precision.

Electrodeposited Pt-Ir is a coating that has been shown to lower impedance and increase charge storage capacity.8 It combines many of the favorable qualities of other low impedance coatings, while avoiding their shortcomings (Table I). Pt-Ir also exhibits the enhanced mechanical properties of alloys over softer single component metals or brittle ceramics. In this communication, Pt-Ir was deposited onto commercial cochlear implant electrodes for the first time. The impedance was measured and compared to non-coated electrodes directly from waveforms of clinically relevant pulses, which avoids the overestimation of charge storage capacity measurements that may occur when using cyclic voltammetry9 and is more relevant than using electrochemical impedance spectroscopy (EIS) where low-amplitude, continuous sine-wave stimuli are utlized, and often only a single frequency, such as 1 kHz, is used to compare impedances of different electrode materials.

Table I. Comparison of Pt-Ir to other low impedance coatings.

  Deposited Possesses a    
  at ambient fractal high   Alloy
  atmospheric surface area Contains mechanical
  pressures morphology IrOx properties
TiN12        
Pt gray13    
AIROF14    
SIROF15    
PEDOT16      
Pt-Ir8

Experimental

Electrodes

The Advanced Bionics HiFocus 1j Electrode is a commercially available cochlear implant electrode array consisting of 16 rectangular platinum stimulating contacts. Each contact measures approximately 0.4 mm wide by 0.5 mm long (0.2 mm²). On each array, Pt-Ir was electrodeposited onto 12 electrode contacts using a method described previously8 and 4 were left uncoated as a control.

Optical microscopy and SEM

Optical images were taken using a digital microscope (Keyence VHX-1000). Scanning electron microscopy (SEM) and EDS scans were performed using a JEOL JSM-6480 (accelerating voltage: 10 kV, working distance: 9 to 12 mm). Electrodes were grounded to the SEM stage, thus no additional sample preparation was required.

Electrochemical evaluation

A custom-built Advanced Bionics current pulse generator based on the HiRes 90K implant electronics was used to control current pulses. Voltage transients were captured using a Tektronix DPO 3014 oscilloscope, which measured the voltage across the working electrode (HiFocus 1j contacts) and a large graphite rod counter electrode (Princeton Applied Research, G0091) while trains of biphasic, cathodic-first, charge-balanced pulses were injected through each electrode contact. The electrolyte used was prepared using Type 2 purified water (EMD Millipore Elix Advantage, resistivity >5 MΩ-cm, TOC < 30 ppb) and consisted of a physiologically relevant 0.01 M phosphate-buffered saline solution (Sigma-Aldrich, P5368) composed of 2.7 mM KCl, 138 mM NaCl, 1.5 mM KH2PO4, and 8.5 mM Na2HPO4. All experiments were performed at room temperature (∼22°C) in open air (dissolved O2 was present in solution). (Cyclic voltammetry and electrochemical impedance spectroscopy was also performed. Details of which can be found in the supplemental materials.) Four pulse designs within a clinically relevant parameter space10 were tested (Table II). Waveforms were recorded once the waveform shape reached a steady state (generally after 10–20 s).

Table II. Decrease in impedance for biphasic pulses with increasing charge densities.

Pulse Parameters   Mean Impedance Decrease P-value
Pulse width (μs)/Amplitude (μA) /Charge density (μC·cm−2)   Uncoated (n = 3) Coated (n = 6) (%) (2-sample t-test, α = 0.05)
17/498/4.2 Ra (Ω) 910 844 7 0.081
  |Zp| (Ω) 550 56 90 0.003
17/937/8.0 Ra (Ω) 925 860 7 0.109
  |Zp| (Ω) 507 46 91 0.003
32/498/8.0 Ra (Ω) 904 843 7 0.022
  |Zp| (Ω) 921 65 93 0.006
32/937/15.0 Ra (Ω) 918 854 7 0.039
  |Zp| (Ω) 752 67 91 0.004

The access resistance (Ra) (1) and polarization impedance (|Zp|) (2) were calculated based on the access voltage (Va) and polarization potential (Ep) taken from voltage transients measured in response to biphasic current pulses11 (also see supplemental material), with ic being the amplitude of the cathodic current pulse.

Equation ([1])

Equation ([2])

Results and Discussion

Morphology and composition

SEM (Figures 1a and 1c) and optical images (Figures 1b and 1d) of bare Pt (Figures 1a and 1b) and Pt-Ir coated Pt (Figures 1c and 1d) show the Pt-Ir coating has an anisotropic, rough appearance with microscopic nodules characteristic of electrodeposited Pt-Ir.8 The Pt-Ir coating masked the striations visible on the surface of the bare electrode and did not extend beyond the edge of the base Pt electrode, indicating a well-controlled plating process.

Figure 1.

Figure 1. SEM (a and c) and optical (b and d) images of uncoated (a and b) and coated (c and d) electrodes. The coating darkens the appearance of the electrode (d) due to its rough, nodular morphology (visible on both SEM (c) and optical (d) images). The roughness of the coating masks horizontal striations and reflective morphological features visible on the uncoated SEM (a) and optical (b) images. (Scale bars are 100 μm.)

Electrochemical evaluation

Electrode impedance was directly compared by measuring voltage transients during clinically-relevant biphasic pulses. The Ra 1 of the coated and uncoated electrodes were not statistically different (Table II). The Ra is proportional to the Va which is the near instantaneous drop in voltage during the leading edge of the pulse train and are nearly identical when comparing coated and uncoated electrodes (Figure 2). This is to be expected because access voltage/resistance is generally determined by the geometric area of the electrode11 and the electrolyte resistivity, which the coating did not change either.

Figure 2.

Figure 2. Mean values of voltage transients measured across coated (solid, black line, n = 3) and uncoated (dashed, red line, n = 6) electrodes during biphasic pulses of pulse widths and amplitudes. The standard deviation of the voltage transients, which is represented with shading above and below solid and dotted lines, is small and therefore not visible on some curves.

In contrast, |Zp|—generally determined by the electrochemical surface area (ESA) and the electrode material11—was decreased by 91% for the long, low amplitude pulses and 93% for short, high amplitude pulses (Table II). The polarization impedance is proportional to the Ep, defined as the drop in potential between the onset (after the access potential drop) and the end of the cathodic phase of the pulse. For uncoated arrays, the drop appears as a negatively sloped line, whereas the polarization is small enough for the coated arrays that the potential appears to remain constant during the pulse (Figure 2). Ep is associated with the electrochemical processes occurring at the electrode interface such as double-layer charging and redox reactions. Since these processes are dependent on electrode material and the ESA of the electrode, Ep for the Pt-Ir coated samples is greatly reduced due to the presence of Ir and the coating's 3-dimensional structure. Many other electrode materials/coatings listed in Table I show low polarization under slow rates of charge transfer (e.g. during cyclic voltammetry or under biphasic stimulation with long pulse widths), but their impedance advantage does not hold up under biphasic pulsing at short pulse widths like those used in the present study. This may be due to the porous nature of other electrode materials, which does not allow sufficient ion flux in and out of pores at fast timescales.

Summary

Pt-Ir was coated onto commercial cochlear implant electrodes and significantly improved their electrochemical performance by decreasing |Zp| during clinically relevant pulses. Decreased |Zp| may improve sound quality and spectral resolution for future cochlear implants by enabling the use of smaller, more closely packed microelectrodes. In addition, the increase in impedance caused by tissue encapsulation will be less impactful since overall impedance will be lower. Finally, lower impedance electrodes can improve the battery life of the implantable pulse generator (IPG) (or allow for a smaller battery) by using less power to deliver the same amount of current.

Acknowledgments

Weiland, Whalen, Petrossians, and Lee acknowledge a financial interest in Platinum Group Coatings, LLC. Hudak is an employee and has a financial interest in Advanced Bionics. The authors acknowledge the support of the Coulter Program at USC.

ORCID

Curtis D. Lee 0000-0001-8295-7786

Eric M. Hudak 0000-0003-2570-2060

John J. Whalen III 0000-0002-5202-4747

Artin Petrossians 0000-0003-2967-8205

James D. Weiland 0000-0003-3453-9074

Please wait… references are loading.
10.1149/2.0031812jes