23 February 2009 Calibration of redox scanning for tissue samples
He N. Xu, Baohua Wu, Shoko Nioka, Britton Chance, Lin Z. Li
Author Affiliations +
Abstract
The fluorescence properties of reduced nicotinamide adenine dinucleotide (NADH) and oxidized flavoproteins (Fp) such as flavin adenine dinucleotide (FAD) in the respiratory chain are sensitive indicators of intracellular redox states and have been applied to the studies of mitochondrial function with energy-linked processes. The redox scanner, a threedimensional (3D) redox cryo-imager previously developed by Chance et al., can quantitatively determine the metabolic properties of tissue samples by acquiring the fluorescence images of NADH and Fp. The redox ratios, i.e., Fp/(Fp+NADH) and NADH/(Fp+NADH), obtained on the basis of relative signal intensity ratios, provide a sensitive index of steady-state of the mitochondrial metabolism that has been determined for a variety of biological tissues. This paper presents the further development of the instrument by establishing a calibration method to quantify the concentration of the fluorophores and facilitate the comparison of redox images obtained at different time or with different instrument functions. Calibration curves of both NADH and Fp have been obtained using snap-frozen standard references with NADH concentration ranging from 150-1400 μM and Fp from 80-720 μM. Snap-freeze tissue samples such as human breast tumors xenografted in mice, normal mouse pancreases and spleens were imaged. The NADH and Fp concentrations as well as the redox ratios in the tissue samples were quantified based on the adjacent solution standards of NADH and Fp. The obtained multi-slice redox images revealed high heterogeneity of the tissue samples which can be quantitatively interpreted.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
He N. Xu, Baohua Wu, Shoko Nioka, Britton Chance, and Lin Z. Li "Calibration of redox scanning for tissue samples", Proc. SPIE 7174, Optical Tomography and Spectroscopy of Tissue VIII, 71742F (23 February 2009); https://doi.org/10.1117/12.809778
Lens.org Logo
CITATIONS
Cited by 18 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Tissues

Calibration

Tumors

Luminescence

Liquids

Optical filters

Pancreas

Back to Top