Paper
19 March 2013 Soft-tissue imaging in low-dose, C-arm cone-beam CT using statistical image reconstruction
Author Affiliations +
Proceedings Volume 8668, Medical Imaging 2013: Physics of Medical Imaging; 86681F (2013) https://doi.org/10.1117/12.2008421
Event: SPIE Medical Imaging, 2013, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
C-arm cone-beam CT is an emerging tool for intraoperative imaging, but currently exhibits modest soft-tissue imaging capability. This work adapts a spectrum of statistical iterative reconstruction approaches to C-arm CBCT and investigates performance in imaging of low-contrast tasks pertinent to soft-tissue surgical guidance. Experiments involved a mobile C-arm and phantoms and cadavers presenting soft-tissue structures imaged using 3D FBP and penalized likelihood reconstruction. Statistical reconstruction - especially non-quadratic PL - boosted soft-tissue image quality through reduction of noise and artifacts, therefore presenting promise for interventional imaging. Further investigation of task-specific performance may overcome conventional tradeoffs in noise, resolution, and dose.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Adam S. Wang, Sebastian Schafer, J. Webster Stayman, Yoshi Otake, Marc S. Sussman, A. Jay Khanna, Gary L. Gallia, and Jeffrey H. Siewerdsen "Soft-tissue imaging in low-dose, C-arm cone-beam CT using statistical image reconstruction", Proc. SPIE 8668, Medical Imaging 2013: Physics of Medical Imaging, 86681F (19 March 2013); https://doi.org/10.1117/12.2008421
Lens.org Logo
CITATIONS
Cited by 8 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Spatial resolution

Image quality

Head

Tissues

Bone

Sensors

Computed tomography

Back to Top