Abstract
In vivo quantification of vessel wall cyclic strain has important applications in physiology and disease research and the design of intravascular devices. We describe a method to calculate vessel wall strain from cine PC-MRI velocity data. Forward–backward time integration is used to calculate displacement fields from the velocities, and cyclic Green–Lagrange strain is computed in segments defined by the displacements. The method was validated using a combination of in vitro cine PC-MRI and marker tracking studies. Phantom experiments demonstrated that wall displacements and strain could be calculated accurately from PC-MRI velocity data, with a mean displacement difference of 0.20±0.16 mm (pixel size 0.39 mm) and a mean strain difference of 0.01 (strain extent 0.20). A propagation of error analysis defined the relationship between the standard deviations in displacements and strain based on original segment length and strain magnitude. Based on the measured displacement standard deviation, strain standard deviations were calculated to be 0.015 (validation segment length) and 0.045 (typical segment length). To verify the feasibility of using this method in vivo, cyclic strain was calculated in the thoracic aorta of a normal human subject. Results demonstrated nonuniform deformation and circumferential variation in cyclic strain, with a peak average strain of 0.08±0.11. © 2002 Biomedical Engineering Society.
PAC2002: 8761-c, 8719Uv, 8719Rr, 8757Gg
Similar content being viewed by others
REFERENCES
Bayliss, W. M. On the local reactions of the arterial wall to changes of internal pressure. Proc. Physiol. Soc. 26:220-231, 1901.
Draney, M. T., K. L. Wedding, C. A. Taylor, and N. J. Pelc. An in vivo method for measuring vessel wall motion and cyclic strain using magnetic resonance imaging. Proceedings of the 1st Joint BMES/EMBS Conference, Atlanta, GA, 1999.
Draney, M. T. Biomechanics of the human aorta and aortic aneurysms. Proceedings of the Frontiers in Vascular Disease Conference, Pebble Beach, CA, 2000.
Draney, M. T., F. R. Arko, M. T. Alley, M. Markl, R. J. Herfkens, N. J. Pelc, and C. K. Zarins. In vivo quantification of porcine aortic wall motion using cine PC-MRI. Proceedings of the 10th Annual International Society for Magnetic Resonance in Medicine Conference, HI, 2002.
Dzau, V. J., and G. H. Gibbons. Vascular remodeling: Mechanisms and implications. J. Cardiovasc. Pharmacol. 21 Suppl. 1, S1-S5, 1993.
Friedman, M. H., G. M. Hutchins, and C. B. Bargeron. Correlation between intimal thickness and fluid shear in human arteries. Atherosclerosis 39:425, 1981.
Fronek, K., G. Schmid-Schoenbein, and Y. C. Fung. A noncontact method for three-dimensional analysis of vascular elasticity in vivo and in vitro. J. Appl. Physiol. 40:634-637, 1976.
Fung, Y. C., and S. Q. Liu. Change of residual strains in arteries due to hypertrophy caused by aortic constriction. Circ. Res. 65:1340-1349, 1989.
Glagov, S., C. K. Zarins, D. P. Giddens, and D. N. Ku. Hemodynamics and atherosclerosis. Insights and perspectives gained from studies of human arteries. Arch. Pathol. Lab. Med. 112:1018-1031, 1988.
Hansen, B., A. H. Menkis, and I. Vesely. Longitudinal and radial distensibility of the porcine aortic root (see comments). Ann. Thorac. Surg. 60:S384-S390, 1995.
Hardt, S. E., A. Just, R. Bekeredjian, W. Kubler, H. R. Kirchheim, and H. F. Kuecherer. Aortic pressure-diameter relationship assessed by intravascular ultrasound: Experimental validation in dogs. Am. J. Physiol. 276:H1078-H1085, 1999.
Hokanson, D. E., D. J. Mozersky, D. S. Summer, and D. E. Strandness, Jr. A phase-locked echo tracking system for recording arterial diameter changes in vivo. J. Appl. Physiol. 32:728-733, 1972.
Imura, T., K. Yamamoto, K. Kanamori, T. Mikami, and H. Yasuda. Noninvasive ultrasonic measurement of the elastic properties of the human abdominal aorta. Cardiovasc. Res. 20:208-214, 1986.
Kamiya, A., and T. Togawa. Adaptive regulation of wall shear stress to flow change in the canine carotid artery. Am. J. Physiol. 239:H14-H21, 1980.
Lyon, R. T., A. Runyon-Hass, H. R. Davis, S. Glagov, and C. K. Zarins. Protection from atherosclerotic lesion formation by reduction of artery wall motion. J. Vasc. Surg. 5:59-67, 1987.
Meyer, S. L. Propagation of error and least squares. In: Data Analysis for Scientists and Engineers. New York: Wiley, 1975 pp. 39-48.
Moore, C. C., E. R. McVeigh, and E. A. Zerhouni. Quantitative tagged magnetic resonance imaging of the normal human left ventricle. Top. Magn. Reson. Imaging. 11:359-371, 2000.
Moreno, M. R., J. E. Moore, Jr., and R. Meuli. Crosssectional deformation of the aorta as measured with magnetic resonance imaging. J. Biomech. Eng. 120:18-21, 1998.
Pelc, N. J., R. J. Herfkens, A. Shimakawa, and D. R. Enzmann. Phase contrast cine magnetic resonance imaging. Magn. Reson. Q. 7:229-254, 1991.
Pelc, N. J., M. Drangova, L. R. Pelc, Y. Zhu, D. C. Noll, B. S. Bowman, and R. J. Herfkens. Tracking of cyclical motion using phase contrast cine MRI velocity data. J. Magn. Reson. Imaging. 5:339-345, 1995.
Thubrikar, M. J., and F. Robicsek. Pressure-induced arterial wall stress and atherosclerosis. Ann. Thorac. Surg. 59:1594-1603, 1995.
Tropea, B. I., S. P. Schwarzacher, A. Chang, C. Asvar, P. Huie, R. K. Sibley, and C. K. Zarins. Reduction of aortic wall motion inhibits hypertension-mediated experimental atherosclerosis. Arterioscler. Thromb. Vasc. Biol. 20:2127-2133, 2000.
Wedding, K. L., M. T. Draney, R. J. Herfkens, C. K. Zarins, C. A. Taylor, and N. J. Pelc. Measurement of vessel wall strain using cine phase contrast MRI. J. Magn. Reson. Imaging. 15:418-428, 2002.
Wedeen, V. J. Magnetic resonance imaging of myocardial kinematics. Technique to detect, localize, and quantify the strain rates of the active human myocardium. Magn. Reson. Med. 27:52-67, 1992.
Xu, C., S. Glagov, M. A. Zatina, and C. K. Zarins. Hypertension sustains plaque progression despite reduction of hypercholesterolemia. Hypertension 18:123-129, 1991.
Zarins, C. K., D. P. Giddens, B. K. Bharadvaj, V. S. Sottiurai, R. F. Mabon, and S. Glagov. Carotid bifurcation atherosclerosis: Quantitative correlation of plaque localization with flow velocity profiles and wall shear stress. Circ. Res. 53:502-514, 1983.
Zerhouni, E. A., D. M. Parish, W. J. Rogers, A. Yang, and E. P. Shapiro. Human heart: Tagging with MR imaging: A method for noninvasive assessment of myocardial motion. Radiology 169:59-63, 1988.
Zhu, Y., M. Drangova, and N. J. Pelc. Fourier tracking of myocardial motion using cine-PC data. Magn. Reson. Med. 35:471-480, 1996.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Draney, M.T., Herfkens, R.J., Hughes, T.J.R. et al. Quantification of Vessel Wall Cyclic Strain Using Cine Phase Contrast Magnetic Resonance Imaging. Annals of Biomedical Engineering 30, 1033–1045 (2002). https://doi.org/10.1114/1.1513566
Issue Date:
DOI: https://doi.org/10.1114/1.1513566