Controlled environment for reciprocity calibration of laboratory standard microphones and measurement of sensitivity pressure correction

and

Published under licence by IOP Publishing Ltd
, , Citation G S K Wong and L Wu 1999 Metrologia 36 275 DOI 10.1088/0026-1394/36/4/5

0026-1394/36/4/275

Abstract

The operation of an environmental chamber for high-precision calibration of laboratory standard microphones and acoustical measurements is described. The nominal ranges for the environmental control system are: temperature from -10 °C to 50 °C; pressure from 80 kPa to 110 kPa; and relative humidity (RH) from 10 % to 90 %. The estimated uncertainties (expanded, with 95 % confidence level) of the pressure, temperature and humidity measurements were 20 Pa, 0.05 °C and 5 % RH, respectively. The sensitivity pressure correction for Bruel & Kjaer type 4160 laboratory standard microphones varies with frequency. Over the barometric pressure range 94 kPa to 106 kPa, the sensitivity of B&K 4160 laboratory standard microphones changes by approximately 0.2 dB at 1 kHz and 0.35 dB at 10 kHz. An empirical equation for this microphone sensitivity pressure correction for the B&K 4160 microphone has been developed.