Introduction

One of the most popular physical sensors is a mass sensor1,2,3,4,5 relying on resonance frequency shift of the structure. With the aid of the development of micro- and nanofabrication techniques, there are numerous efforts and ongoing progress towards performance enhancement via miniaturization of sensing devices. Nanoelectromechanical systems (NEMS) resonators have achieved zeptogram-scale mass resolution almost a decade ago6 and bottom-up synthesized nanomaterials based resonators have enabled single atom7 and proton8 resolution. Encouragingly, nanomechanical resonators have recently enabled single-molecule mass spectrometry in real time9. However, exceptional resolutions achieved with NEMS and nanomaterials based resonators are in general limited to ideal conditions such as high vacuum levels and cryogenic temperatures. Such experimental environments are far from being compatible to portable field applications and minuscule matters suspended in liquid, not to mention that the maintenance of such operating conditions requires intensive time, cost and labor.

Hollow resonators10,11,12,13,14 have become one of the most promising platforms for inertial sensing of liquids and particle suspensions with the hollow core intrinsically serving as a sample delivery and transport channel. The embedded channel localizes liquid samples and makes suspended particles guided towards the mass sensing region on demand. This is a unique advantage of hollow resonators over solid ones that are fully exposed to liquid sample and passively observe stochastic landing of suspended particles. Most hollow resonators reported to date have been made via sacrificial processes10,13 and fusion bonding11,12. Hollow resonators have shown numerous applications with unprecedented performance such as mass11,15,16,17,18,19, density16,17, volume17, surface charge20, and deformability21,22 measurements of single micro-/nanoparticles or cells, however, slow and complicated fabrication hampers widespread adaptation of hollow resonators. The fabrication time and complexity are further increased if an electrical readout for the resonance frequency is harnessed into hollow resonators23.

A novel approach by atomic migration of silicon recently reported has significantly simplified and expedited the overall fabrication of hollow resonators24. Nevertheless, researchers who have no or limited access to standard microfabrication facilities housed in a cleanroom hardly take advantage of the simple process. Interestingly, a commercially available microcapillary suspended over a pre-defined trench seems very promising since it is potentially available to anyone. Suspended microcapillary resonators with optical detection were employed for liquid density measurements25 and differentiation of X- and Y- chromosome bearing sperm cells26. Despite the absence of commercially available nanoscale capillary tubes, suspended nanocapillary resonators27 have been simply fabricated by using polymer nanofibers deposited over lithographically pre-defined trenches as sacrificial templates and used with an optical interferometry. Optical detection, however, makes the overall system bulky and requires a precise alignment. In addition, small inner and outer diameters of micro-/nanocapillaries are not directly compatible with existing standard fluidic tubings and connectors.

Here, we introduce an innovative solution for aforementioned issues—a hollow microchannel mass sensing system which exhibits user friendly fluidic interconnection and alignment-free electrical readout by using pulled glass microcapillaries and quartz tuning fork (QTF) without relying on any conventional microfabrication process. After the proposed system was thoroughly tested and validated, mass densities of various liquid samples and sizes of oil microdroplets suspended in water were measured with the responsivity comparable to microfabricated hollow resonators.

Results

Materials and fabrication

To fabricate microcapillary tube resonators without relying on conventional microfabrication in a cleanroom, we employed a capillary pulling process with a borosilicate glass capillary of which inner and outer diameters are 0.58 and 1.00 mm, respectively, (Fig. 1a(i)) (1B100F-3, World Precision Instruments) and a commercial pipette puller (P-2000, Sutter Instrument). The laser-heated pipette puller defines the microcapillary tube through sequential heating and pulling processes which locally stretch the middle section of the glass capillary while maintaining the cross-sectional area of both ends (Fig. 1a(ii)). This is somewhat different from the typical use of a pipette puller which is used to produce two symmetric pipettes with a nanoscale fluidic access hole. Once the glass capillary was partially pulled, it was flexible enough to be bent 180 degree and make a U-shaped loop (Fig. 1a(iii)). Furthermore, it could be wound up and make a 360-degree loop as shown in Fig. S1. Inner and outer diameters of the pulled section of the glass microcapillary can be varied by adjusting parameters such as laser power and pulling force. The smallest and largest inner diameters of the pulled section were 14 and 148 μm, respectively, while the ratio of inner to outer diameter of the pulled section was 0.623 ± 0.025 which is similar to the initial ratio of 0.58 (Fig. 1a(iv)). Hereafter, pulled microcapillaries of which inner and outer diameters are 30 and 50 μm and overall elongation length is 58 mm were used.

Figure 1: Off-the-shelf glass capillary and quartz tuning fork (QTF) components simply modified and assembled toward a microfabrication-free microscopic sensing platform.
figure 1

(a) A glass capillary with inner and outer diameters of 0.58 and 1.0 mm, respectively (top). Schematic showing the capillary pulling to decrease the inner and outer diameters around the middle section of the capillary (bottom left). A pulled capillary (bottom middle). Scanning electron micrographs of cross-section areas of pulled capillaries with various inner and outer diameters. Scale bars are 100, 10, 10, and 10 μm, respectively (bottom right). (b) A QTF before (top) and after (bottom) the package is opened and removed. The scale bar is 1 mm. (c) A pulled micro capillary tube resonator (PμTR) of which one nodal point is defined by a physical contact to one of two prongs of a QTF. (d) Assembly process of PμTR-QTF system (not to scale). (e) Photographs of an assembled PμTR-QTF sensing platform ready for testing.

For simple and robust sensing of the resonance frequency of pulled microcapillary tubes, a QTF (AB26T-32.768 kHz, Abracon) was employed after its hermetic metal can packaging was removed. Figure 1b(i,ii) show the QTF before and after the removal of the hermetic packaging, respectively. For actuation of pulled microcapillary tubes, a piezo actuator (TA0505D024W, Thorlabs) was employed.

Figure 1c illustrates the overall schematic of the pulled microcapillary tube resonator (PμTR). A microcapillary tube with axial tension applied makes contact with a solid block on top of a piezo actuator and one prong of a QTF. The contact point above the piezo actuator becomes the actuation node while the contact point with the QTF becomes the sensing node. The node-to-node distance (i.e. distance between two contact points) defines the length of the (PμTR). Unpulled sections at both ends facilitate interfacing with standard fluidic components. To realize the proposed PμTR, parts were assembled by following the procedures in Fig. 1d. First, 0.3 N of axial tension was applied to a pulled microcapillary using a tension gauge (110 g, OHBA SIKI) to remove slacks and also increase the resonance frequency of PμTR. Effect of the applied tension on the amplitude signal and maximum axial tension allowed can be found in Supplementary information section 1 (Figs S2 and S3). With the applied tension maintained, the pulled microcapillary was brought over and attached to a custom machined aluminum jig with an epoxy. Then, the jig holding the attached microcapillary was fixed to a base block having a nodal block on the piezo actuator with two standard M4 bolts. After this step, the pulled microcapillary made a solid contact with the actuation node. Finally, using a motorized stage (MTS25-Z8, Thorlabs), the QTF mounted on a custom printed circuit board (PCB) approached precisely with a 10-μm step towards the pulled microcapillary and made another solid contact which defined the sensing node. Finally, the PCB was bonded to the jig assembly by using the epoxy. Figure 1e shows several pictures of the final PμTR-QTF coupled system. The typical setting for the node-to-node distance was 6.5 mm throughout this work unless otherwise stated.

Theory and operation

The resonance frequency, f, of the PμTR under axial tension, T, is given by28

where d is the PμTR’s length, E is the Young’s modulus, I is the area moment of inertia, and md is the mass per unit length of the PμTR, respectively. When the operation of the actuating node induces out-of-plane vibration of the PμTR, the vertical reflection force, Fv, which is the vertical component of the applied axial tension, is felt by the QTF and is given by

where θ is the tangent angle of the PμTR at the QTF sensing node, A is the vibration amplitude, fE is the excitation frequency, t is time, and ϕ is the phase delay (Fig. 2a). This vertical reflection Fv, of which magnitude is proportional to the vibration amplitude of the PμTR is transmitted to the QTF, then, induces a forced vibration at one of two prongs with the frequency of fE . Although the proposed actuation scheme is asymetric, the response from the PμTR is symmetric (Fig. S4d). Dynamic amplitude of the equivalent spring-mass-damper system for the QTF with an external input of Fv is transduced to dynamic current generation at the RLC components in a Van-Dyke equivalent circuit due to the intrinsic piezoelectric effect of the QTF29 (Fig. 2b). A transimpedance amplifier was configured using an operational amplifier (OPA129, Texas Instruments) with a 200 kΩ gain resistor and connected to one electrode of the QTF for current-to-voltage conversion with a gain while the other electrode of the QTF was grounded.

Figure 2: Operation scheme of the PμTR-QTF coupled sensing system.
figure 2

(a) Free body diagram for the PμTR-QTF. (b) Mechanical and electrical equivalent models for the QTF which generates electrical currents via piezo-electric transduction of the vertical reflection force, Fv . (c) Open and closed loop drive circuits for the PμTR-QTF system. (d) Fluidic setup diagram to deliver a liquid sample into the PμTR.

The PμTR-QTF system can be operated in either open or closed loop modes (Fig. 2c). In the open loop mode, a function generator drives the piezocrystal actuator at the actuation node near the resonance frequency of the PμTR while the piezoelectrically transduced vibration of the PμTR by the QTF sensing node is measured with a lock-in amplifier (Model 7265, Signal Recovery) through the aforementioned transimpedance amplifier. In the closed loop mode, the output from the transimpedance amplifier is low pass filtered, amplified, phase-shifted and fed back to the piezo actuator at the actuation node. The filter used is a second-order active low-pass filter (LPF) with the cut-off frequency of 30 kHz (Voltage pre-amplifier SR560, Stanford Research Systems). For sustaining oscillation of the PμTR, the insertion of the LPF is crucial for the piezoelectric readout where two distinct resonance peaks with comparable amplitudes are present (see Open loop operation Section for details). For real-time tracking of the resonant frequency of the PμTR, the feedback signal is mixed-down to approximately 1 kHz by an analog multiplier (AD734, Analog Devices) and measured by using a 100 MHz frequency counter equipped in the data acquisition board (USB-6361, National Instruments). For both open and closed loop operations, the voltage signal fed into the piezo actuator at the actuation node is restricted within ±15 V in order to prevent excessive oscillation amplitude of the PμTR which may affect the contact status between the actuation node and PμTR and induce problems associated with nonlinearities such as bi-stabilities. For our entire measurements, the vibration amplitudes are far below the onset of nonlinearity (see Supplementary information section 2) and the contact status was stable. Even at the maximum drive condition in our current experimental setup, we were not able to get into the nonlinear dynamics regime. We have not seen any noticeable changes in phase spectra as the drive amplitude was varied (Fig. S5). This is also because the PμTR was operated in the linear regime.

To introduce liquid samples into the PμTR, a pneumatic system was installed with electronic pressure regulators (ITV-0030, SMC), pressurizable vials, standard tubings (1/16″ FEP, Upchurch Scientific) and standard connectors as shown in Fig. 2d. Such interfacing with standard tubings and connectors is a unique advantage of the PμTR over fluidic resonators based on microcapillary tubes or microfabricated hollow resonators.

Open loop operation

For proof-of-concept experiments, a couple of basic measurements were conducted in the open loop operation mode. After PμTRs with various lengths were prepared by adjusting the node-to-node distance and filled with water, their resonance frequencies were measured sequentially using the QTF sensing node via piezoelectric transduction and plotted as a function of the node-to-node distance (Fig. 3a). Their amplitude spectra and quality factors can be found in Fig. S6. Measurements show good agreement with theoretical calculation, the solid line in Fig. 3a, by Equation (1). To validate the piezoelectric QTF sensing mechanism, a complementary optical-lever measurement was performed for a 6.5-mm long PμTR with a partially deposited metal (gold) reflector (see Supplementary information section 3 and Fig. S7a). While the piezoelectric readout captures two resonance peaks, one near 24.5 kHz for PμTR and the other near 33.0 kHz for QTF, the optical-lever readout measures a single resonance peak at 24.5 kHz for PμTR (Fig. 3b). This difference in two readout schemes is explained in detail by the block digram (see Supplementary information section 3 and Fig. S7b,c). The resonance peak for the PμTR was universally detected by two different readout schemes. Therefore, the expensive and bulky optical setup which accompanied an additional metal coating for reflective laser signal and precise alignment is circumvented once the simple QTF sensing node based on piezoelectric transduction. Moreover, the piezoelectric readout can be superior to the optical-lever readout in terms of the frequency stability unless significant amount of effort is made towards precise optical alignment or high-end optical components are used (Fig. S8).

Figure 3: Open loop operation of the PμTR-QTF system intended for density calibration and measurements.
figure 3

(a) Resonance frequency as a function of node-to-node distance (or length) of the PμTR filled with water. (b) Amplitude (top) and phase (bottom) spectra of a water filled 6.5-mm long PμTR measured by optical-lever (red dashed line) and piezoelectric (solid black line) readouts. While the resonance peak for the PμTR around 24.5 kHz is universally detected, the QTF resonance around 33 kHz is exclusively measured by the piezoelectric readout. (c) Resonant spectra of a 6.5-mm long PμTR filled with water, ethanol, and glycerol-water mixture (glycerol:water = 1:1 in weight ratio). (d) Resonance frequency shift of the 6.5-mm long PμTR measured while water, ethanol-water binary mixtures, and glycerol-water binary mixtures are sequentially introduced. The density responsivity of −3,088 Hz g−1 cm3 is obtained from the linear regression.

Next, another 6.5-mm long PμTR was prepared and its resonance frequency was measured by using the QTF sensing node while water, ethanol, and glycerol-water mixture (glycerol:water = 1:1 in weight ratio) were sequentially introduced into the PμTR (Fig. 3c). Then, for more precise calibration, various ethanol-water and glycerol-water binary mixtures were prepared and injected in the PμTR while the resonance frequency of PμTR was measured (Fig. 3d). From the linear regression of the frequency shift vs. density plot, the density responsivity of −3,088 Hz-g−1 cm3 was extracted at room temperature.

Closed loop operation

With the closed loop operation mode shown in Fig. 2c, the event which occurs at a much faster time scale can be examined with the PμTR-QTF system. First, the fluidic configuration shown in Fig. 4a was implemented. While deionized (DI) water was filled in one pressurizable vial, mineral oil (Mineral oil light, Sigma Aldrich) was loaded in another pressurizable vial. Due to the immiscible nature of two liquid samples, the water-oil interface was maintained. By alternating the pressure difference of 90 kPa between two vials, a transition from DI water to mineral oil or vice versa could be monitored as shown in Fig. 4b. The difference in the resonance frequency of PμTR due to the water-mineral oil exchange was approximately 450 Hz which corresponds to 0.851 g/cm3 based on the calibration shown in Fig. 3d. This is in good agreement with the specification, 0.838~0.854 g/cm3 (25 °C) given by the supplier. As expected, the transition time from water to mineral oil was found to be a function of the pressure difference, τTR ΔP−1. Second, a microscale droplet generator was configured at one entrance of the PμTR as shown in Fig. 4d. The interior of the droplet generator was hydrophilic so mineral oil was used as a disperse phase and water was used as a continuous phase (i.e. mineral oil droplets in water). In the water, a surfactant (Tween20, Sigma Aldrich) was mixed at a volume fraction of 0.5 percents. Pressure settings for the droplet generator were set to produce oil droplets of which mean radii are approximately 28 μm. When a oil droplet passed through the PμTR, it was squeezed to form a plug since its diameter is larger than the inner diameter of the PμTR as shown in Fig. 4e (top). Once they came out of the pulled section of the PμTR, their shapes restored back to spheres and collected in the outlet vial (Fig. 4e (bottom)). Since the length of squeezed plugs are in general much shorter than the node-to-node distance, the assumption of a point mass traveling through the long string resonator is valid. Figure 4f shows resonance frequency shifts of the PμTR recorded when oil plugs with three different lengths passed through the PμTR. To quantify equivalent radii of oil plugs, buoyant mass of each plug was first calculated from fitting measured resonance shift data with the theoretical estimations given by30

Figure 4: Closed loop operation of the PμTR-QTF system intended for real-time applications.
figure 4

(a) Fluidic configuration modified for switching between two immiscible liquid samples. (b) Resonance frequency shift measured at 1-kHz bandwidth and the pressure difference of 90 kPa while water and mineral oil alternately fills the pulled micro capillary tube. (c) Transition time, τTR, from water to mineral oil as a function of the pressure difference, ΔP, showing the expected inverse relationship. Inset shows a typical transition at the pressure difference of 90 kPa. (d) A micro droplet generator configured at the inlet of the PμTR for real-time monitoring of droplet size. (e) Optical micrographs of a squeezed oil plug passing through the PμTR (top) and collected spherical droplets after measurements (bottom). Scale bars in both images are 100 μm. (f) Resonance frequency shifts fitting upon a transit of mineral oil plugs of which equivalent radii are 25.4, 28.5 and 31.5 μm, respectively. (g) Histogram for the equivalent radius of oil droplets (cumulative counts are 310).

where Δf is the resonance frequency shift due to an oil plug transiting the PμTR, fw is the resonance frequency of the PμTR entirely filled with water, mop is the buoyant mass of the oil plug, mr is the mass of the PμTR, and x is the lateral position of the traveling oil plug. The simple harmonic mode shape is valid for PμTR longer than 5 mm (Fig. S10). Then, an equivalent spherical radius, req, can be given by

where ρw and ρo are densities of water and mineral oil, respectively. When either the buoyant mass or the equivalent radius of an oil plug are extracted, its velocity is simultaneously obtained (see Methods and Supplementary information section 5). Measurements were continued for several hundreds of oil plugs and a histogram of the equivalent radius was acquired with the average of 28.47 μm and the standard deviation of 4.03 μm (Fig. 4g). Simultaneously obtained velocity histogram and velocity vs. radius scatter plot are shown in Fig. S12. Interestingly, spherical oil droplets collected after measurements show a uniform distribution (Fig. 4e (bottom)). Unless a higher magnification is employed, small difference in the droplet size can not be easily observed. Even with a higher magnification, conventional far-field optical measurements are diffraction limited. The PμTR-QTF system truly offers more precise and more quantitative measurements for droplet sizing than general optical microscopy. In addition, the PμTR-QTF system exclusively provides density and mass of droplets.

Performance metrics of PμTR systems

Figure 5a shows the Allan deviation σA of the normalized resonance frequency of the PμTR filled with water as a function of the gate time (Fig. S8 to compare Allan deviations from the piezoelectric and optical-lever readouts). While the water-filled PμTR was operated in closed loop, its resonance frequency was sampled and recorded approximately every 1 ms. As the gate time increases, the Allan deviation measured decreases, reaches a minimum around 0.01 s, and increases. Therefore, the best frequency stability 1.66 ppm was obtained at the gate time of 0.01 s. This frequency stability is only one order of magnitude lower than that of the microfabricated hollow channel resonator with integrated piezoresistive readout23.

Figure 5: Frequency noise and density responsivity of the PμTR-QTF system.
figure 5

(a) Allan deviation of the 6.5-mm long PμTR with water filled as a function of gate time. The piezoelectric readout is used. The best frequency stability is obtained at the gate time of 0.01 s. (b) Normalized density responsivity of the 6.5-mm long PμTR in comparison with those of other micro fabricated density sensors found from literatures plotted (band plot) as a function of area ratio (density of wall, ρwall, is set to be 2.25 ± 0.25 g/cm3). The dashed line along with pink shade shows the theoretically estimated responsivity. (c) Detection resolution of the difference in mineral oil droplet radius as a function of the equivalent oil droplet radius (calculated from frequency stabilities, 1 and 3 σA). Figure 4g is included as inset figure to show the radius distribution of measured oil plugs. The limit of detection (LOD) for the smallest droplet radius is 4.25 μm.

Normalized density responsivity of our PμTR-QTF system was plotted as a function of the ratio of resonator wall cross-sectional area to channel cross-sectional area (area ratio) and compared with those of other density sensors based on microfabricated hollow resonators10,11,12,14,24,25,31,32,33,34,35,36 (Fig. 5b). The solid line in Fig. 5b is theoretical estimation by

where fw and f are resonance frequencies of a PμTR filled with water and actual sample, respectively, ρfluid and ρwall are densities of the fluid sample and resonator wall material, respectively, and Awall and Afluid are cross-sectional areas of the resonator wall and channel, respectively. Normalized density responsivity of our PμTR-QTF system is better than several microfabricated hollow channel density sensors (Fig. S9 for data from three PμTR-QTF pairs). With different to particle mass (i.e. point mass) responsivity inversely proportional to the resonator’s effective mass, normalized density responsivity depends on the area ratio, Awall/Afluid only (Equation (5)). Therefore, miniaturization does not necessarily improve the normalized density responsivity. Once a thinning process of the resonator wall such as the selective exterior etching is fully developed, the normalized density responsivity of the PμTR-QTF system can be further improved. For example, the outer wall of the pulled section can be exposed to and etched in glass etching solutions such as hydrofluoric acid and buffered oxide etch solution thus the area ratio can be decreased (see Supplementary information section 6 and Fig. S13), with the aid of the flexibility which offers 180 degree bending or 360 degree wound up loop shown in Fig. 1a and Fig. S1, respectively. However, after the selective exterior etching, PμTR became very fragile thus easily broken with the piezoelectric readout which requires hard contact between the microcapillary and the QTF for the etched PμTR of which wall thickness is less than 4 μm. We suggest that the proposed actuation scheme would be paired with the capacitive detection with a remote sensing electrode or with the optical detection which was already used for proof-of-concept experiments (see Supplementary information section 3).

Figure 5c shows the detection resolution of the difference in mineral oil droplet radius, Δreq, plotted as a function of the equivalent oil droplet radius. This resolution was calculated from frequency stabilities, 1 and 3 σA (see Equation (8) in the Methods section). For the average droplet radius of 28.47 μm shown in Fig. 4g, we can differentiate oil droplets if their radius differ by approximately 31.66 and 94.99 nm based on the 1 and 3 σA criteria, respectively. This resolution is far superior to the diffraction limit of general optical microscopy based techniques. As the droplet gets smaller, the resolution of Δreq becomes negatively affected. The colored area above each solid line represents the “Measurable regime” based on each criterion. The limit of detection in the equivalent oil droplet radius (i.e. the smallest oil droplet to be measured) is 4.25 μm.

Discussion

Here, we proposed a microfabrication-free approach to realize microscale fluidic resonators for gravimetric sensing applications by configuring three off-the-shelf parts; glass capillary, piezocrystal actuator, and quartz tuning fork. Commercially available glass capillary with a millimeter scale diameter were pulled to reduce their diameters to several tens of micrometers and increase overall lengths by a few centimeters via laser heating. As a result, microcapillary tubes were locally defined in the middle of glass capillary while two ends maintained original millimeter scale dimensions. Fabricated microcapillary tubes were mounted on a custom machined jig and made contact with two nodes to complete the fabrication process. One contact point was the actuation node, a machined aluminum block on a piezocrystal actuator which applied out-of-plane vibration. The other contact point was the sensing node, a prong of a commercially available QTF. This PμTR-QTF hybrid system offers the following benefits; simple fabrication of microfluidic resonators, electrical readout of the resonance frequency and facile connection to standard tubings and fluidic components.

After manufacturing the PμTR-QTF hybrid system, two gravimetric applications were demonstrated. In the first application, the PμTR-QTF hybrid system was operated in open loop to sense the mass density of water, ethanol, glycerol and their binary mixtures introduced into the PμTR. In the second application, the system was operated in closed loop to weigh and size mineral oil droplets passing through the PμTR. With the Allan deviation separately acquired, we found out that the smallest radius mineral oil droplet measurable is 4.25 μm.

Encouragingly, the normalized density responsivity of our PμTR-QTF hybrid system is higher than a group of microfabricated hollow channel density sensors. Diameters and wall thickness of the PμTR can be further downsized by optimization of the pulling process or selective etching from either inner or outer surfaces. The structural material of the PμTR is transparent to visible light so that many optical and spectroscopic techniques can be easily combined with the PμTR. In addition, other capillary experimental setup based separation and detection schemes such as capillary electrophoresis37 and whispering gallery mode optical cavity38 can be integrated to realize micro total analysis systems or multimodal sensing platforms.

Methods

Data processing for equivalent radius and velocity of oil plugs

Frequency change data due to traveling oil plugs recorded in time were processed to extract equivalent radius and velocity of oil plugs as illustrated in Fig. S11. First, recorded raw data were Savitzky-Golay filtered (window of 50 and polynomial order of 3). Then, first time derivatives of the frequency change were compared with a pre-configured threshold to locate each peak resulting from an oil plug. For each selected peak, a small window was masked to get a baseline to be compensated. With the node-to-node distance (or resonator length) and the constant velocity assumption of traveling oil plugs, filtered and baseline compensated frequency change data were fitted by Equation (3) to obtain the buoyant mass and velocity of oil plugs. The buoyant mass was converted to the equivalent radius by Equation (4) where densities of water and oil separately measured with the PμTR were used. The constant velocity assumption for oil plugs traveling through the PμTR is valid because oil plugs do not exhibit motion relative to the surrounding water and the vibration amplitude of the PμTR is negligibly small.

Calculation for resolution of the difference in oil plug equivalent radius

The detection resolution of the difference in equivalent radii of oil plugs is obtained by dividing the Allan deviation (σA) of the PμTR from the radius responsivity . The chain rule is applied to the radius responsivity to include the buoyant mass of oil plugs (mop) as follows.

The derivative of derived by spherical mass-volume relationship is given by

Finally, by employing Equations (6) and (7), the relationship between the mass and radius resolutions is obtained as

The mass responsivity is acquired by fitting data of transiting oil plugs shown in Fig. 4f,g. and used here. Equation (8) indicates that the radius resolution is inversely proportional to the square of its own radius.

Additional Information

How to cite this article: Lee, D. et al. Pulled microcapillary tube resonators with electrical readout for mass sensing applications. Sci. Rep. 6, 33799; doi: 10.1038/srep33799 (2016).