Abstract
There is increasing concern regarding the biological consequences of metal release from implants. However, the mechanisms underpinning implant surface degradation, especially in the absence of wear, are often poorly understood. Here the synergistic effect of albumin and H2O2 on corrosion of Ti6Al4V in physiological saline is studied with electrochemical methods. It is found that albumin induces a time-dependent dissolution of Ti6Al4V in the presence of H2O2 in physiology saline. Potentiostatic polarisation measurements show that albumin supresses dissolution in the presence of H2O2 at short times (<24 h) but over longer time periods (120 h) it significantly accelerates corrosion, which is attributed to albumin-catalysed dissolution of the corrosion product layer resulting in formation of a thinner oxide film. Dissolution of Ti6Al4V in the presence of albumin and H2O2 in physiological saline is also found to be dependent on potential: the titanium ion release rate is found to be higher (0.57 µg/cm2) at a lower potential (90 mV), where the oxide capacitance and resistance inferred from Electrochemical Impedance Spectroscopy also suggests a less resistant oxide film. The study highlights the importance of using more realistic solutions, and considering behaviour over longer time periods when testing corrosion resistance of metallic biomaterials.
Similar content being viewed by others
Introduction
Implantable medical devices are commonly fabricated from titanium (Ti) alloys. Ti alloys have suitable mechanical properties for many biomedical applications and possess excellent corrosion resistance which is dependent on the presence of a stable passive oxide film on its surface1,2,3. However, the peri-implant environment contains a variety of biological species which can act to modify corrosion behaviour, including proteins such as albumin and reactive oxygen species (ROS) such as hydrogen peroxide (H2O2)4,5,6. The effect of albumin alone, or H2O2 alone, on the corrosion of Ti alloys has been investigated7,8,9,10,11,12,13,14,15,16. However, the effect of combined exposure to both H2O2 and albumin on corrosion of Ti6Al4V has only recently been considered17. Yu et al. demonstrated that a combination of H2O2 and albumin can lead to significantly higher levels of corrosion of Ti6Al4V than either species alone17. In this paper, the mechanism proposed by Yu et al. to explain the synergistic action of albumin and H2O2 on Ti corrosion is explored, taking into account how the corrosion behaviour may change as a function of time.
Inflammation is ubiquitous with healing following implant insertion, but can also be present chronically in the peri-implant tissues associated with failing devices. Inflammation is associated with the production (by immune cells that have infiltrated the peri-implant tissues) of ROS species such as superoxide anions and H2O218,19. Quantification of the levels of H2O2 in different environments in the body is complex and therefore there is a great variation in the way H2O2 exposures are simulated in in-vitro corrosion tests. ROS species have short half-lives and cellular production of ROS can take place over a wide range of time-scales, making quantification challenging20. However, it is accepted that peak concentrations of ROS can be reached when ROS release is confined in environmental niches, such as sub-cellular phagosomes or extracellularly between the cell and a material surface. H2O2 is known as a strong oxidant and can act to modify the corrosion resistance of Ti surfaces. Studies have been carried out to investigate the effect of H2O2 on corrosion of Ti alloys using concentrations ranging from 33 mM to 300 mM7,8,9,10,17,21. A number of studies have shown that H2O2 increases corrosion of Ti and its alloys by forming Ti(IV)-H2O2 complexes6,7,8,9,17,21,22, leading to formation of rougher, thicker and more porous surface oxide films.
Proteins are known to interact with metal oxides and modify the corrosion resistance of implants. Albumin is the most abundant protein (4.2–5.3%) in blood plasma and extracellular tissue fluid4. It is generally agreed that albumin inhibits cathodic reactions on Ti by adsorbing onto the metal surface and covering the reaction sites and/or blocking mass transport of dissolved O211,13,15,16,17. However, it has also been claimed that albumin increases11,12,17,23, decreases15,16,23, or has no effect14,24,25 on anodic dissolution of Ti. Similar contradictory results have been found for other metallic biomaterials26,27,28. Metal-protein interactions are generally attributed to two effects: adsorption and chelation. In the case of adsorption, metal ion release is inhibited by a layer of adsorbed protein15,16. In the case of chelation, soluble metal-protein complexes are formed, enhancing the rate of dissolution11,29,30. Whilst albumin is the most prevalent protein in the tissue fluid, a wide range of proteins will interact with metal implant surfaces in-vivo. Different proteins can variably influence the corrosion susceptibility of the same metal surface29,31. For example, it has recently been shown that sequential exposure of albumin and fibrinogen to stainless steel modifies the corrosion rate32, and was attributed to displacement and exchange of the surface adsorbed albumin by the larger protein fibrinogen32.
Exposing Ti surfaces to H2O2 has been shown to modify subsequent interactions with adsorbed proteins, including albumin. Nagassa et al. showed that exposing Ti surfaces to H2O2 enhanced surface roughness and resulted in greater serum albumin adsorption33. Sousa et al. demonstrated differential binding affinities of albumin to a sputtered TiO2 surface when compared with a TiO2 surface that had been modified by H2O2 immersion34. Yu et al. found a synergistic effect of albumin and H2O2 on Ti6Al4V corrosion by measuring the rate of metal ion release after a 2 week immersion in physiological saline containing both species17. It was shown that albumin accelerated dissolution of Ti6Al4V in presence of H2O2 during the long term exposure. However, short term electrochemical tests showed contradictory findings where albumin appeared to supress the anodic reaction of Ti6Al4V in presence of H2O2. To explain this contradictory behaviour, the author suggested a mixed potential theory where an “active” dissolution region was proposed at low potentials and adsorption of albumin would supress the corrosion potential by taking it into this more “active” region of Ti17. However, the hypothesis needs to be validated carefully and more direct measurements reflecting temporal effects on dissolution rates need to be carried out.
The aim of this study is to understand the mechanism underpinning the synergistic effect of albumin and H2O2 on corrosion of Ti6Al4V as a function of time. Despite the various simulated physiological media for corrosion testing, 0.9% NaCl was chosen as a basic testing electrolyte (directed by ASTM F1875–98 and F1801-97) to control surface interactions that may further impact on interpretation of the interaction between albumin, H2O2 and the Ti6Al4V surface35. Both short term (<24 h) and long term exposures (up to 120 h) are investigated by electrochemical techniques in order to reflect possible changes in degradation of the metal surface, which is often projected as a slow and continuous process following the implantation in the body. Additionally, to test the “active” dissolution theory proposed by Yu et al.17, ion release was measured as a function of potential and parallel EIS measurements were carried at the static potentials to characterise the metal/oxide/solution interface.
Results
Corrosion under open circuit conditions
Figure 1 shows the OCP (open circuit potential or corrosion potential) as a function of time in different solutions. In all cases, the corrosion potential gradually increased, consistent with a decrease in the anodic current density as a consequence of thickening of the oxide film and/or corrosion product layer on the metal surface. The OCP was consistently higher in a solution containing 0.1% H2O2 in 0.9% NaCl compared with that in 0.9% NaCl alone; this is likely to be due to an increased cathodic current since H2O2 is a strong oxidant8,17. In a solution containing 4% albumin, in addition to 0.1% H2O2 in 0.9% NaCl, the OCP was consistently lower than that in the presence of NaCl and H2O2 alone, but higher than in NaCl alone. This is consistent with previous work by Yu et al.17, showing that the presence of albumin can decrease the rate of the cathodic reduction of H2O2, probably by adsorption on the surface.
Potentiodynamic polarisation curves can be used to confirm the interpretation of the OCP behaviour. Figure 2 shows anodic polarisation curves in the same set of solutions as those in Fig. 1. The order of the open circuit potentials is consistent with that in Fig. 1. The lowest OCP was observed in NaCl alone, with addition of H2O2 leading to a higher potential associated with a higher cathodic reaction rate (consistent with cathodic polarisation curves shown in reference17). Addition of H2O2 to NaCl also led to an increase in the anodic reaction rate above 400 mV (sufficiently above the OCP for the cathodic reaction to be negligible), with a small plateau from ~+400 to +600 mV followed by a steady increase. In the presence of NaCl, H2O2 and albumin, the anodic reaction above +400 mV was the same as that for NaCl and H2O2 alone, suggesting that albumin caused little or no change in the anodic reaction, but the cathodic reaction appears to be inhibited, explaining the relative values of the OCP shown in Fig. 1.
Figure 3 shows Bode and Nyquist plots of Ti6Al4V after immersion at OCP for 24 h and 120 h. Two time constants corresponding to the oxide film and likely the (double) layer were observed in the Bode plot after 24 h immersion, although this was less obvious for that in 0.9% NaCl than in the presence of 0.1% H2O2 and combination of 0.1% H2O2 and 4% albumin. Nevertheless, after 120 h immersion, more pronounced two time constants were presented in all solutions, which could also be associated with a two layer structured passive film. It can also be found that the real part of the impedance (diameter of semicircle in Nyquist plot, indicating the combined resistances of the passive film and the electrochemical reaction rates at OCP), increased with immersion time which is consistent with their OCP behaviour as observed in Fig. 1. It can also be found that resistance of Ti6Al4V in the presence of both albumin and H2O2 in 0.9% NaCl was lower than that the rest of the solutions at both measured time periods.
The data shown in Fig. 3 were fitted using the equivalent circuit shown in supplementary Figure S1. The polarisation resistance (Rp) (Fig. 4) is the sum of two resistive components Rhf (oxide) and Rlf (inversely proportional to electrochemical reaction rate); it should be noted that Rp is dominated by Rhf (oxide resistance). Figure 4 shows that in all solutions, Rp gradually increased with immersion time and subsequently approached steady state at around 70 h. It can be seen that the presence of H2O2 reduced Rp by more than an order of magnitude after the initial measurement. At 24 h, the value of Rp in the presence of both albumin and H2O2 was similar to that in the presence of H2O2. However, at longer times, Rp in the presence of albumin and H2O2 was significantly lower than that in the presence of H2O2 alone.
Figure 5(a) shows the as-polished surface of Ti6Al4V with a distribution of β phase (identified with EDX, supplementary Table S1) in an α matrix. Figure 5(b) shows a secondary electron image of Ti6Al4V after immersion in 0.9% NaCl + 10% H2O2 for 3 days at 37 °C. The mud-crack morphology of corrosion products suggests the formation of a thick corrosion product layer. Figure 5(c) and (d) shows the morphology surfaces that were pre-treated in the same way as Figure (b) and then subsequently immersed 0.9% NaCl + 0.1% H2O2 without (c) or with (d) 4% albumin for 7 days at 37 °C. Figure 5(c) shows a similar mud-crack morphology to that in Fig. 5(b), but it can be seen that in the presence of albumin (Fig. 5(d)) the thick corrosion product layer was dissolved, and microstructural features such as β phase can be observed (detected with EDX, supplementary Table S1).
Time-dependent corrosion behaviour at controlled potential
Figure 6 shows the current as a function of time for Ti6Al4V initially immersed in 0.9% NaCl at a constant potential of 400 mV with additions of 0.9% NaCl containing H2O2 or albumin. The potential was selected so that the effect on the anodic current density could be evaluated, but it should be noted that this value is a little higher than the interfacial potential under in vivo conditions. It can be seen that at an early stage (~1 h, see inset Figure), the current density increased after addition of solution to give a concentration of 0.1% H2O2 in 0.9% NaCl. For times up to 20 h, the solution with the additional 4% albumin had a lower current density compared with the control that contained only 0.1% H2O2 in 0.9% NaCl.
However, differences were revealed over longer exposure times. In the presence of H2O2 alone, the initial increase in current density was found to reach a maximum value of >1 µA/cm2 at ~1.7 h followed by a steady decay of almost two orders of magnitude, reaching ~0.03 µA/cm2 at 70 h. This is likely to be a result of the accumulation of corrosion products similar to those shown in Fig. 5(c). However, in the presence of both albumin and H2O2, the current density was found to be relatively constant after the initial increase without much of degradation. This is consistent with the observation of a much thinner corrosion product layer in Fig. 5(d), suggesting that albumin can promote dissolution of the H2O2-induced corrosion product, so the current reaches a steady state relatively quickly. A cross-over point in the current densities in these two solutions was observed at ~22 h and at 70 h, the current density in the presence of albumin and H2O2 was almost an order of magnitude greater than that in H2O2 alone.
Metal ion release and EIS measurements at constant potential
Yu et al.17 proposed that the higher metal release in the presence of both H2O2 and albumin when compared with H2O2 alone, might be a consequence of albumin adsorption onto the metal surface, inhibiting the cathodic reaction and decreasing the potential to a proposed ‘active’ region for dissolution Ti6Al4V. To test this hypothesis, Ti6Al4V was exposed to 0.9% NaCl containing 0.1% H2O2 with and without 4% albumin for 24 h under potentiostatic control. Two potentials were chosen based on the OCP values observed in the two solutions after 24 h in Fig. 1(b). The OCP value in the presence of 0.1% H2O2 alone was close to 300 mV and in the presence of 0.1% H2O2 and 4% albumin a value closer to 90 mV was observed. Potential values of 90 mV and 200 mV were chosen so that they could be in the potential range of active/passive transition as proposed by Yu et al.17.
Table 1 shows that at both measured potentials, Ti ion release was significantly higher in the presence of albumin by almost a factor of 2, indicating that the effect of albumin in dissolving corrosion products dominates the synergistic effect of H2O2 and albumin in accelerating Ti ion release. However, in the presence of albumin and H2O2, Ti release was significantly higher at 90 mV than that at 200 mV (p < 0.001), supporting the hypothesis of Yu et al.17. In presence of H2O2 alone, slightly higher metal Ti ion release was observed at the lower potential, but the difference was much smaller (p < 0.01).
The metal/oxide/solution interface was also characterised by EIS measurements following potentiostatic polarisation for 20 h to develop stable passive film. Figure 7(a) shows the polarisation resistance as a function of potential, obtained by fitting the EIS spectra (supplementary Figure S2) with a two time constant equivalent circuit (supplementary Figure S1). The polarisation resistance increased steadily with potential from ~360 kΩ-cm2 at 40 mV to ~930 kΩ-cm2 at 200 mV. Figure 7(b) shows the oxide resistance (Rhf) and capacitance (calculated from Yhf using equation (2)) as a function of potential. The oxide resistance showed little change from 40 to 90 mV then increased significantly with potential up to 200 mV. The oxide capacitance showed some variation with potential, but appeared to give a maximum value at 90 mV.
Discussion
It can be seen from potentiodynamic polarisation (Fig. 2) that the addition of H2O2 increased the passive current density of Ti6Al4V in 0.9% NaCl. Similarly, the potentiostatic current density (Fig. 6) increased after addition of H2O2 in 0.9% NaCl. H2O2 is known to increase the corrosion rate of Ti22, and its alloys7, via a complexation reaction with Ti ions6,8,9,21.
At short time exposures (t < ~22 h), addition of 4% albumin suppressed the current density in the presence of H2O2 (Fig. 6). Rapid adsorption of albumin onto the Ti6Al4V oxide surfaces has previously been observed with Quartz Crystal Microbalance (QCM) measurements and X-ray photoelectron spectroscopy (XPS)34,36. In addition, using 125I radio labelling, it has been shown that albumin adsorbs on oxidised Ti surfaces from the early stages of immersion with most of the adsorption taking place within the first few minutes37. To date, most studies have reported that albumin inhibits corrosion of commercially pure (CP) Ti and Ti alloys15,16,23, and also protects it from dissolution induced by e.g. fluoride ions38. Adsorption of albumin films have been proposed to block dissolution. However, all of these experiments have taken place over short time periods, typically over no more than a few hours.
After longer term exposures (up to 72 h), it can be seen from Fig. 6 that the potentiostatic current density in the presence of H2O2 alone decreased significantly after the initial increase, most likely as a result of blocking of the surface by the peroxide corrosion product layer shown in Fig. 5(b,c). It has been proposed that the Ti oxide film forms a two-layer structure in phosphate buffered saline, which consists of a thin barrier type inner layer and thick porous type outer layer, and the presence of H2O2 reduces corrosion resistance and thickens the outer porous layer8,9,21.
In contrast, prolonged exposure in the presence of both H2O2 and albumin resulted in a higher current density than that in the presence H2O2 alone, with a cross-over point at ~22 h. This is consistent with the observation of a much thinner corrosion product layer in Fig. 5(d), suggesting that albumin promotes dissolution of the peroxide corrosion product layer at longer time periods.
The interaction between albumin and peroxide corrosion layer has been studied by Sousa et al. via a kinetic perspective of albumin adsorption, adhesion and exchangeability (desorption)34. It was found that the more hydrophilic surface of TiO2 after H2O2 immersion adsorbed less albumin but with higher work of adhesion i.e. it was more strongly attached on the surface, compared with that of the more hydrophobic surface of sputtered TiO234. After the adsorption step, it was suggested that the exchangeability of albumin was changing with time: after 24 h the adsorbed albumin molecules on the H2O2-treated TiO2 surface seem to be less exchangeable than those adsorbed on the sputtered TiO2 surface; however after 72 h nearly all the adsorbed albumin molecules effectively exchanged with other albumin molecules, which suggested that longer time is needed for exchange of albumin molecules adsorbed on surface of H2O2-TiO2 complexes34. Therefore, the time dependent dissolution behaviour observed in the presence work is consistent with the desorption of albumin-metal complexes as it is a slow, rate determining process34, and can result in thinner oxides.
The time-dependent data were confirmed by the EIS data shown in Fig. 4. The polarisation resistance (Rp) of Ti6Al4V in in the presence of H2O2, Rp was significantly lower than that in NaCl alone after 24 h and both increased gradually with time, consistent with the formation of a thicker corrosion product layer. At 120 h, Rp in presence of H2O2 was over an order of magnitude lower than that in NaCl alone. Comparable values were measured for CP Ti in PBS where Rp decreased from 2 × 107 ohm-cm2 to 8 × 105 ohm-cm2 in the presence of 50 mM H2O2 (compared with 33 mM in the present work) after 125 h39. The decreased corrosion rate of Ti6Al4V in the presence of H2O2 over time has been attributed to growth of a corrosion product layer.
In the presence of albumin and H2O2, Rp showed little change over time compared with the behaviour in H2O2 alone, and it can be seen from Fig. 4 that after 120 h, Rp in presence of H2O2 and albumin was significantly lower than that of H2O2 alone, consistent with the presence of a thinner corrosion product layer and increased dissolution. This observation correlates well with previous ICP-MS measurements carried out by Yu et al.17, where metal ion release was measured in identical solutions after 2-week incubation and the combination of H2O2 and albumin was found to give much higher Ti release than that of H2O2 alone. It suggests that the explanation for this observation is likely to be the increased dissolution of the peroxide corrosion product in the presence of albumin.
In the work of Yu et al., short term electrochemistry tests (t ~ 1 h) and long term (2 weeks) ion release measurements showed contradictory results regarding the effect of albumin on dissolution of Ti6Al4V in the presence of H2O2 in physiological saline. The investigation explained the observation using a proposed mixed potential theory, suggesting that albumin suppresses the cathodic reaction, decreasing the corrosion potential into a more “active” region for dissolution of Ti at lower potential17. In the present work, this hypothesis was tested by measuring metal ion release from Ti6Al4V at constant applied potentials. Table 1 shows that in the presence of albumin and H2O2, the rate of Ti ion release was higher at lower potentials, consistent with the proposed “active dissolution” effect17. However, it is clear that the increased metal ion release at both potentials is substantially greater in the presence of albumin, indicating that the synergistic effect of albumin and peroxide observed at longer exposure times is dominated by the role of albumin in dissolving the peroxide corrosion product.
Metal implants are exposed to an environment that is more complex than the physiological saline (0.9% NaCl) that is used in standard tests such as ASTM F1801-97 and ASTM F1875-9840,41. In the present work on Ti6Al4V, the major protein in the body, albumin, is shown to interact with the corrosion products of Ti in the presence of H2O2, which is produced in inflammatory processes. The level of H2O2 used in this work (0.1% = 33 mM) is at the lower end of those used in previous in vitro corrosion studies7,8,9,10,17,21. Concentrations of H2O2 measured in extracellular environments are frequently reported to be in the µM range18,19,20,42,43. However, both immune cells and microbial biofilms (such as those found on oral implants) create local micro-environments where H2O2 can be considerably concentrated reaching mM concentrations. For example, measurements of H2O2 concentration in surface biofilms of the oral microbe Streptococcus gordinni have been shown to increase with proximity to the biofilm with quasi-steady-state concentrations of 1.4 mM measured at 100 µm from the exposed microbial surface43. It was proposed that the concentration of H2O2 may reach much higher levels within the biofilm itself and at the biofilm substrate interface.
More recently, the effect inflammatory cell induced corrosion has been recognised as an important mechanism for in vivo metal degradation42,44,45. It has been found that during inflammation, the cellular physiology produces a localized acidic45 and concentrated ROS environment42, which can lead to accelerated corrosion of implants and generation of metallic particulates and ions. These implant derived products have themselves been shown to elicit innate and adaptive cellular responses including release of pro-inflammatory cytokines46, leading in some cases to persistent peri-implant inflammation, bone resorption and aseptic loosening of the device46,47,48.
This work also highlights the need to examine the time-dependence of corrosion behaviour of metals used for biomedical implants, since in this case there is a clear switch in the observed behaviour at ~1 day. Similar time-dependent corrosion behaviour in the presence of albumin was also found in biodegradable Mg alloys in simulated body fluid49. It was found that albumin blocks dissolution in the first few hours of exposure but increased corrosion was observed over longer exposures49. As biomolecules such as albumin and H2O2 are present in inflamed peri-implant tissue, these observations raise the question of whether standard testing should be complemented by testing in more realistic environments, over more realistic time periods, and using enhanced corrosion characterisation methods rather than standard polarisation curves50.
Methods
Materials and surface preparation
Ti6Al4V (ASTM Grade 5) was commercially sourced in disc shape specimens of 14 mm diameter and 1.2 mm thickness (Titanium Products Ltd, UK). The composition (max. wt%) was 0.05N, 0.08C, 0.015H, 0.3 Fe, 0.2O, 3.5–4.5V, 5.5–6.8 Al, with the balance Ti. Samples for electrochemical tests were cold-mounted in a non-conducting setting resin. To ensure a reproducible surface finish, all samples were wet polished from 800 to 4000 grit using SiC abrasive paper and finished with an OP-S colloidal silica suspension (0.04 μm) on MD-Chem cloth (Struers, Ballerup, Denmark). Samples were cleaned thoroughly with deionised water (Millipore, >15 MΩcm), rinsed with 100% methanol and finally dried in cool air stream. To ensure consistency in the surface oxide layer between different samples, each sample was exposed in ambient lab air for a fixed time of 15 min following surface preparation prior to electrochemical experiments.
Test solutions and electrochemical cells
Physiological saline (0.9% NaCl, 0.15 M) solutions were prepared by dissolving NaCl (Sigma Aldrich, UK) in deionised water. H2O2 (30 wt.% in H2O, Sigma Aldrich, UK) was added to 0.9% NaCl to give a concentration of 0.1% (33 mM). Bovine serum albumin (≥98%, lyophilized powder, Sigma Aldrich) was dissolved in 0.9% NaCl containing 0.1% H2O2 at concentration of 4%. Both reagent chemicals i.e. 30 wt% H2O2 and bovine serum albumin powder (\({\rm{\ge }}\)98%) were stored at 4 ± 2 °C and were used to prepare fresh solutions for each electrochemical experiment. A double-wall three-electrode cell was used for both DC and AC electrochemical tests. The cold mounted Ti6Al4V samples were used as working electrode, Pt mesh as counter electrode and saturated calomel electrodes (SCE) as a reference electrode. All potentials are referred to this scale in this paper. The cell was maintained at a temperature of 37 ± 1 °C by a circulating heated water jacket.
DC electrochemical tests
A Gill AC potentiostat (ACM instruments) was used for DC electrochemical tests. Samples were immersed in a test solution at the OCP for 1 h, followed by potentiodynamic polarisation from −50 mV (vs. OCP) at a scan rate of 1 mV/s. Each test was repeated three times to confirm consistency. In addition to potentiodynamic tests, potentiostatic polarisation at 400 mV was conducted after 1 h immersion at OCP in 0.9% NaCl, followed by sequential additions of 0.1% H2O2 with and without 4% albumin. Solutions were stirred during addition of species to ensure complete mixing. Solution of 0.9% NaCl containing H2O2 was added at ~0.6 h to give final concentration of 0.1% H2O2 and solution of 0.9% NaCl containing albumin was added at ~0.74 h to give final concentration of 4% albumin. The control solution without any addition of species was stirred at the same times.
Electrochemical Impedance Spectroscopy (EIS)
A Gamry PCI4 potentiostat (Gamry instruments) was used to measure EIS. EIS measurements were performed at open circuit potential perturbing with 10 mV RMC AC signal with a wide frequency range from 10−3 to 105 Hz. The acquisition rate was 10 points per decade. Measurements were taken after immersion in the solution for 24 h intervals up to 120 h. Additional samples were held at constant potentials for 20 h after which EIS was measured at the constant potential. A two time constant electrical circuit (supplementary Figure S1) was used to fit EIS data, which consists of two parallel Constant Phase Element (CPE) and resistor pairs in series with solution resistance (Rs)51,52. Fitting of EIS data was performed by Gamry Echem Analyst software with simplex algorithm method and the quality of fitting was judged by chi squared value (χ2) < 1 × 10−3. Quantitative parameters were obtained. Polarisation resistance was calculated from the sum Rhf and Rlf, corresponding to resistances of inner compact and outer porous passive layer respectively. The impedance of a CPE is given by
The capacitance was converted from the CPE using the Mansfield equation53:
where \({{\omega }^{^{\prime\prime} }}_{m}\) is the angular frequency at which the imaginary part of impedance has a maximum53.
Ion release measurements
Samples were potentiostatically polarised at 90 or 200 mV for 24 h after 1 h immersion at OCP in 0.9% NaCl with presence of 0.1% H2O2 or combination of 0.1% H2O2 and 4% albumin. Solutions were then extracted and released ions of interests (i.e. Ti, Al, V) were analysed by Butterworth Laboratories (UK) using inductive coupled plasma mass spectroscopy (ICP-MS) (Agilent 8800 ICPMS triple quad). The working quantitation limit for Ti is 10 µg/L, Al is 100 µg/L and V is 10 µg/L. 5 mL sample aliquots were measured and diluted up to 1000 mL with water and diluent containing 1.5 mL of Triton-x-100, 0.15 mL of 20% ammonia and 15 mL of 1.4 mM EDTA. The ion release results for each solution were compared across the potentials through one-way analysis of variance (ANOVA) at a significance level of α = 0.05. A total number of four repeats were performed for each condition.
Incubation test and SEM surface analysis
Ti6Al4V discs were mirror polished on both sides and ultrasonically cleaned with deionised water and methanol for 10 min respectively. Surface prepared discs were incubated in saline with 10% H2O2 solution for 3 days in thermostatic chamber at 37 ± 1 °C. The H2O2 treated samples were then transferred to 0.1% H2O2 saline solutions with and without addition of 4% albumin for 7 days. Samples before and after incubation were examined by SEM JEOL 6060 at an acceleration voltage of 20 kV in both secondary electron (SE) and back scattered electron (BSE) mode.
Data availability
The datasets generated and analysed during the current study are available from the corresponding author on reasonable request.
Conclusions
Albumin was found to have a time dependent effect on dissolution of Ti6Al4V in the presence of H2O2 in 0.15 M NaCl.
At short times (t < ~22 h), albumin supressed dissolution rates, but at longer times, the rate of dissolution increased to values greater than those found in absence of albumin.
The long term dissolution behaviour in the presence of albumin and H2O2 can be attributed to the enhanced dissolution of peroxide corrosion products in the presence of albumin, resulting in formation of a thinner oxide layer on the metal surface.
The rate of metal ion release in the presence of albumin and H2O2 was found to depend on potential. At lower potential, a higher metal ion release rate was observed via ICP-MS measurements, and the oxide resistance and capacitance determined from EIS was consistent with a thinner oxide layer than that found at higher potentials.
The corrosion rate of Ti6Al4V in the presence of albumin and H2O2 in 0.15 M NaCl was observed to be significantly higher than that in 0.15 M NaCl alone, emphasising the need to test alloys for biomedical implants in more realistic physiological solutions. It is also essential to make measurements over longer time periods than those typically performed in standard tests involving conventional polarisation curves.
References
Ask, M., Lausmaa, J. & Kasemo, B. Preparation and surface spectroscopic characterization of oxide films on Ti6Al4V. Appl. Surf. Sci. 35, 283–301 (1989).
Milošev, I., Metikoš-Huković, M. & Strehblow, H. H. Passive film on orthopaedic TiAlV alloy formed in physiological solution investigated by X-ray photoelectron spectroscopy. Biomaterials 21, 2103–2113 (2000).
Pound, B. G. Passive films on metallic biomaterials under simulated physiological conditions. J. Biomed. Mater. Res., Part A 102, 1595–1604 (2014).
Peters, T. Jr In Advances in Protein Chemistry (ed. Anfinsen, C. B., Edsall, J. T. & Richards, F. M.) 161–245 (Academic Press, 1985).
Addison, O. et al. Do ‘passive’ medical titanium surfaces deteriorate in service in the absence of wear? J. R. Soc., Interface 9, 3161–3164 (2012).
Tengvall, P. & Lundström, I. Physico-chemical considerations of titanium as a biomaterial. Clin. Mater. 9, 115–134 (1992).
Takemoto, S., Hattori, M., Yoshinari, M., Kawada, E. & Oda, Y. Discoloration of titanium alloy in acidic saline solutions with peroxide. Dent. Mater. J. 32, 19–24 (2013).
Pan, J., Thierry, D. & Leygraf, C. Electrochemical and XPS studies of titanium for biomaterial applications with respect to the effect of hydrogen peroxide. J. Biomed. Mater. Res. 28, 113–122 (1994).
Pan, J., Thierry, D. & Leygraf, C. Hydrogen peroxide toward enhanced oxide growth on titanium in PBS solution: Blue coloration and clinical relevance. J. Biomed. Mater. Res. 30, 393–402 (1996).
Tengvall, P., Elwing, H., Sjöqvist, L., Lundström, I. & Bjursten, L. M. Interaction between hydrogen peroxide and titanium: a possible role in the biocompatibility of titanium. Biomaterials 10, 118–120 (1989).
Cheng, X. L. & Roscoe, S. G. Corrosion behavior of titanium in the presence of calcium phosphate and serum proteins. Biomaterials 26, 7350–7356 (2005).
Williams, R. L., Brown, S. A. & Merritt, K. Electrochemical studies on the influence of proteins on the corrosion of implant alloys. Biomaterials 9, 181–186 (1988).
Karimi, S. & Alfantazi, A. M. Electrochemical corrosion behavior of orthopedic biomaterials in presence of human serum albumin. J. Electrochem. Soc. 160, C206–C214 (2013).
Karimi, S., Nickchi, T. & Alfantazi, A. M. Long-term corrosion investigation of AISI 316L, Co–28Cr–6Mo, and Ti–6Al–4V alloys in simulated body solutions. Appl. Surf. Sci. 258, 6087–6096 (2012).
Ouerd, A., Alemany-Dumont, C., Berthomé, G., Normand, B. & Szunerits, S. Reactivity of titanium in physiological medium: I. electrochemical characterization of the metal/protein interface. J. Electrochem. Soc. 154, C593–C601 (2007).
Contu, F., Elsener, B. & Böhni, H. Characterization of implant materials in fetal bovine serum and sodium sulfate by electrochemical impedance spectroscopy. I. Mechanically polished samples. J. Biomed. Mater. Res. 62, 412–421 (2002).
Yu, F., Addison, O. & Davenport, A. J. A synergistic effect of albumin and H2O2 accelerates corrosion of Ti6Al4V. Acta Biomater. 26, 355–365 (2015).
Fantone, J. C. & Ward, P. A. Role of oxygen-derived free radicals and metabolites in leukocyte-dependent inflammatory reactions. Am. J. Pathol. 107, 395–418 (1982).
Nakagawara, A., Nathan, C. F. & Cohn, Z. A. Hydrogen peroxide metabolism in human monocytes during differentiation in vitro. J. Clin. Invest. 68, 1243–1252 (1981).
Dupré-Crochet, S., Erard, M. & Nüβe, O. ROS production in phagocytes: why, when, and where? J. Leukocyte Biol. 94, 657–670 (2013).
Pan, J., Thierry, D. & Leygraf, C. Electrochemical impedance spectroscopy study of the passive oxide film on titanium for implant application. Electrochim. Acta 41, 1143–1153 (1996).
Mabilleau, G. et al. Influence of fluoride, hydrogen peroxide and lactic acid on the corrosion resistance of commercially pure titanium. Acta Biomater. 2, 121–129 (2006).
Karimi, S., Nickchi, T. & Alfantazi, A. Effects of bovine serum albumin on the corrosion behaviour of AISI 316L, Co-28Cr-6Mo, and Ti-6Al-4V alloys in phosphate buffered saline solutions. Corros. Sci. 53, 3262–3272 (2011).
Okazaki, Y., Tateishi, T. & Ito, Y. Corrosion resistance of implant alloys in pseudo physiological solution and role of alloying elements in passive films. Mater. Trans., JIM 38, 78–84 (1997).
Clark, G. C. F. & Williams, D. F. The effects of proteins on metallic corrosion. J. Biomed. Mater. Res. 16, 125–134 (1982).
Kocijan, A., Milosev, I. & Pihlar, B. The influence of complexing agent and proteins on the corrosion of stainless steels and their metal components. J. Mater. Sci.: Mater. Med. 14, 69–77 (2003).
Sousa, S. R. & Barbosa, M. A. Electrochemistry of AISI 316L stainless steel in calcium phosphate and protein solutions. J. Mater. Sci.: Mater. Med. 2, 19–26 (1991).
Wagener, V., Faltz, A. S., Killian, M. S., Schmuki, P. & Virtanen, S. Protein interactions with corroding metal surfaces: comparison of Mg and Fe. Faraday Discuss. 180, 347–360 (2015).
Hedberg, Y. et al. Surface-protein interactions on different stainless steel grades: effects of protein adsorption, surface changes and metal release. J. Mater. Sci.: Mater. Med. 24, 1015–1033 (2013).
Grebner, T. & Hansen, D. C. The Effect of Galvanic Coupling Between Modern Fracture Fixation Constructs in Physiologically Relevant Solutions: The Role of Serum Proteins on the Corrosion of Ti6Al4V and 316L Alloys. ECS Trans. 11, 13–27 (2008).
Zhang, C., Xiao, Y. M., Liu, C. F. & Liu, C. H. The effect of mucin, fibrinogen and IgG on the corrosion behaviour of Ni-Ti alloy and stainless steel. Biometals 30, 367–377 (2017).
Hedberg, Y. et al. Interaction of Albumin and Fibrinogen with Stainless Steel: Influence of Sequential Exposure and Protein Aggregation on Metal Release and Corrosion Resistance. Corrosion 73, 14 (2017).
Nagassa, M. E. et al. Optimisation of the hydrogen peroxide pre-treatment of titanium: surface characterisation and protein adsorption. Clin. Oral Implants Res. 19, 1317–1326 (2008).
Sousa, S. R., Moradas-Ferreira, P., Saramago, B., Viseu Melo, L. & Barbosa, M. A. Human serum albumin adsorption on TiO2 from single protein solutions and from plasma. Langmuir 20, 9745–9754 (2004).
Tamilselvi, S., Raman, V. & Rajendran, N. Corrosion behaviour of Ti–6Al–7Nb and Ti–6Al–4V ELI alloys in the simulated body fluid solution by electrochemical impedance spectroscopy. Electrochim. Acta 52, 839–846 (2006).
Kusakawa, Y., Yoshida, E. & Hayakawa, T. Protein adsorption to titanium and zirconia using a quartz crystal microbalance method. BioMed Res. Int., https://doi.org/10.1155/2017/1521593 (2017).
Lima, J., Sousa, S. R., Ferreira, A. & Barbosa, M. A. Interactions between calcium, phosphate, and albumin on the surface of titanium. J. Biomed. Mater. Res. 55, 45–53 (2001).
Takemoto, S., Hattori, M., Yoshinari, M., Kawada, E. & Oda, Y. Corrosion behavior and surface characterization of titanium in solution containing fluoride and albumin. Biomaterials 26, 829–837 (2005).
Fonseca, C. & Barbosa, M. A. Corrosion behaviour of titanium in biofluids containing H2O2 studied by electrochemical impedance spectroscopy. Corros. Sci. 43, 547–559 (2001).
F1875-98, A. (ASTM International, West Conshohocken, United States, 2014).
F1801-97, A. (ASTM International, West Conshohocken, United States, 2014).
Gilbert, J. L. & Kubacki, G. W. In Oxidative Stress and Biomaterials (ed. Butterfield, D. A.) 59–88 (Academic Press, 2016).
Liu, X. et al. Real-time mapping of a hydrogen peroxide concentration profile across a polymicrobial bacterial biofilm using scanning electrochemical microscopy. Proc. Natl. Acad. Sci. USA 108, 2668–2673 (2011).
Fonseca-Garcia, A. et al. The effect of simulated inflammatory conditions on the surface properties of titanium and stainless steel and their importance as biomaterials. Mater. Sci. Eng., C 66, 119–129 (2016).
Brooks, E. K., Brooks, R. P. & Ehrensberger, M. T. Effects of simulated inflammation on the corrosion of 316L stainless steel. Mater. Sci. Eng., C 71, 200–205 (2017).
Gibon, E. et al. The biological response to orthopaedic implants for joint replacement: Part I: Metals. J. Biomed. Mater. Res., Part B 105, 2162–2173 (2017).
Di Laura, A. et al. Clinical relevance of corrosion patterns attributed to inflammatory cell-induced corrosion: A retrieval study. J. Biomed. Mater. Res., Part B 105, 155–164 (2017).
Rodrigues, D. C. et al. Titanium Corrosion Mechanisms in the Oral Environment: A Retrieval Study. Materials 6, 5258–5274 (2013).
Rettig, R. & Virtanen, S. Time-dependent electrochemical characterization of the corrosion of a magnesium rare-earth alloy in simulated body fluids. J. Biomed. Mater. Res., Part A 85A, 167–175 (2008).
Wang, J. L. et al. A closer look at the in vitro electrochemical characterisation of titanium alloys for biomedical applications using in-situ methods. Acta Biomater. 54, 469–478 (2017).
Yu, S. Y., Brodrick, C. W., Ryan, M. P. & Scully, J. R. Effects of Nb and Zr alloying additions on the activation behavior of Ti in hydrochloric acid. J. Electrochem. Soc. 146, 4429–4438 (1999).
Rondelli, G., Torricelli, P., Fini, M. & Giardino, R. In vitro corrosion study by EIS orthopaedic of a nickel-free stainless steel for applications. Biomaterials 26, 739–744 (2005).
Hsu, C. H. & Mansfeld, F. Technical Note: Concerning the conversion of the constant phase element parameter Y0 into a capacitance. Corrosion 57, 747–748 (2001).
Acknowledgements
Y.Z. received a postgraduate research scholarship by School of Metallurgy and Materials in University of Birmingham. O.A was funded by the National Institute for Health Research [award No. NIHR/CS/010/001]. This article presents independent research in part funded by the National Institute for Health Research (NIHR). The views expressed are those of the author(s) and not necessarily those of the NIHR. EIS measurements were carried out at the University of Virginia.
Author information
Authors and Affiliations
Contributions
A.J.D., O.A., J.R.S., Y.Z. conceived the experiments, Y.Z. conducted the experiments and Y.Z., B.C.R.T., F.Y., A.J.D. analysed the results. All authors reviewed the manuscript.
Corresponding author
Ethics declarations
Competing Interests
The authors declare no competing interests.
Additional information
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons license, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons license and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this license, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Zhang, Y., Addison, O., Yu, F. et al. Time-dependent Enhanced Corrosion of Ti6Al4V in the Presence of H2O2 and Albumin. Sci Rep 8, 3185 (2018). https://doi.org/10.1038/s41598-018-21332-x
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-018-21332-x
This article is cited by
-
Electrochemical corrosion behavior of β-Ti alloy in a physiological saline solution and the impact of H2O2 and albumin
Journal of Solid State Electrochemistry (2023)
-
A review of the bio-tribology of medical devices
Friction (2022)
-
Particle release from implantoplasty of dental implants and impact on cells
International Journal of Implant Dentistry (2020)
-
Alternating current oxidation of Ti–6Al–4V alloy in oxalic acid for corrosion resistant surface finishing
SN Applied Sciences (2020)
-
General review of titanium toxicity
International Journal of Implant Dentistry (2019)
Comments
By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.