Abstract
A new version of enzyme immunoassay with the use of an amperometric immunosensor was proposed for the determination of an aminoglycoside antibiotic gentamicin. The biosensing part of the enzyme immunosensor simultaneously incorporates immobilized enzyme cholinesterase and antibodies against gentamicin. The detection limit for gentamicin in this method is 1 × 10–9 mg/mL. The time of determination is no longer than 20 min. The stability constants of the immunocomplex and the kinetic parameters of the enzymatic reaction in its presence were determined. Gentamicin was determined in samples of pharmaceutical preparations and foodstuffs (milk).
Similar content being viewed by others
REFERENCES
Astakhova, A.B. and Bogomolova, N.S.,Antibiot. Khimioter., 1991, vol. 36, no. 2, p. 53.
Nouws, J., van Edmond, H., Smulder, I., Loeffen, G., Schouten, J., and Stageman, H., Int. Dairy J., 1999, vol. 9, no. 2, p. 85.
Stead, D.A., J. Chromatogr., B: Biomed. Appl., 2000, vol. 747, no. 1, p. 69.
Posyniak, A., Zmudzki, J., and Niedzielska, J., J. Chromatogr., 2001, vol. 914, no. 1, p. 59.
Kijak, P.J., Jackson, J., and Shaikin, B., J. Chromatogr., B: Biomed. Appl., 1997, vol. 691, no. 2, p. 377.
Adams, E., Roelants, W., and De Paepe, R., J. Pharm. Biomed. Anal., 1998, vol. 18, no. 2, p. 689.
Prey, M., Guyot, D., and Petz, M., J. Chromatogr., 1998, vol. 818, no. 1, p. 95.
Kaale, E., Leonard, S., van Schepdael, A., Roets, E., and Hoogmartens, J., J. Chromatogr., 2000, vol. 895, no. 1, p. 67.
Van Es, R.M., Blankwater, Y.I., and Mejjer, D., Anal. Chim. Acta, 2001, vol. 429, nos. 1–3, p. 37.
Kolosova, A.Yu., Blintsov, A.N., Samsonova, J.V., and Egorov, A.M., Fresenius' J. Anal. Chem., 1998, vol. 361, p. 329.
Baoxin, L., Zhang, Z., and Manli, W.U., Anal. Lett., 2000, vol. 33, no. 8, p. 1577.
Medyantseva, E.P., Babkina, S.S., Vertlib, M.G., and Budnikov, G.K., Zh. Anal. Khim., 1993, vol. 48, no. 10, p. 1632.
Medyantseva, E.P., Budnikov, G.K., and Babkina, S.S., Zh. Anal. Khim., 1990, vol. 45, no. 7, p. 1386.
Kuznetsov, B.A. and Shumakovich, G.P., Metody sovremennoi biokhimii (Methods of Modern Biochemistry), Moscow: Nauka, 1975, p. 102.
Heyrovsky, J. and Kuta, J., Zaklady polarografie (Fundamentals of Polarography), Praha: CNTL, 1962. Translated under the title Osnovy polyarografii, Moscow: Mir, 1965.
Plambeck, J., Electroanalytical Chemistry: Basic Principles and Applications, New York: Wiley, 1982. Translated under the title Elektrokhimicheskie metody analiza: Osnovy teorii i primenenie, Moscow: Mir, 1985, p. 387.
Medyantseva, E.P., Li, F., Fedoseeva, O.V., et al., Zh.Prikl. Khim., 1993, vol. 29, no. 4, p. 619.
Budnikov, G.K., Medyantseva, E.P., Volkov, A.V., and Aronzon, S.S., Zh. Anal. Khim., 1983, vol. 38, no. 7, p. 1283.
Krupyanko, V.I., Vektornyi metod predstavleniya fermentativnykh reaktsii (A Vector Method for Presentation of Enzymatic Reactions), Moscow: Nauka, 1990.
Krupyanko, V.I., Collect. Czech. Chem. Commun., 1988, vol. 53, no. 1, p. 161.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Khaldeeva, E.V., Medyantseva, E.P., Imanaeva, N.A. et al. Determination of Gentamicin with an Amperometric Enzyme Immunosensor. Journal of Analytical Chemistry 57, 1097–1102 (2002). https://doi.org/10.1023/A:1021480500400
Issue Date:
DOI: https://doi.org/10.1023/A:1021480500400