A comparison of phantom scatter from flattened and flattening filter free high-energy photon beams
Introduction
Megavoltage flattening filter free (FFF) treatment beams are typified by the forward-peaked nature of their fluence distributions and have recently become treatment options offered on modern linear accelerators. The requirement to have a flattened beam profile for treatment delivery is no longer paramount when intensity-modulated radiotherapy or intensity-modulated arc therapy is used. The patient dose distribution can instead be sculpted by the multileaf collimator (MLC) movement and dose rate control to create the desired clinical effect. The advantages of a FFF beam are reduced head scatter and head leakage and a more uniform photon energy spectrum away from the central axis. Further advantages of FFF beams are the intrinsically high dose rate at which treatment can be delivered, which can lead to reduced treatment times. This is especially important where large doses per fraction are prescribed, e.g., stereotactic ablative body radiotherapy or where intrafraction organ/patient motion might affect the efficacy of the delivery or both.
The FFF beam produced by an Elekta Agility linac is generated by replacing the conical flattening filter with a flat attenuating disk made of 2-mm thick stainless steel in the beam line below the level of the x-ray target. Varian units use a 1-mm thick steel plate. This filter serves a number of functions. Firstly, it prevents direct exposure of the patient to the electron beam if there should be a catastrophic target failure. Secondly, it generates a flux of secondary electrons at the level of the monitor ionization chambers, making the beam feedback mechanism sufficiently responsive to maintain energy and symmetry. Finally, it removes low-energy photons that only increase the superficial dose and not contribute to the target dose. The energy of a particular megavoltage photon beam is a matter of departmental choice and is usually influenced by previous treatment machines. If the same linac operating parameters, e.g., bending magnet current and microwave power, were used for a flattened and FFF beam, then the quality index of the 2 would be quite different because of the beam-hardening effect of the flattening filter. The 6-MV FFF beams on Elekta linacs are initially set up to best match the conventional flattened 6-MV beam percentage depth dose at 100-cm source-to-surface distance (SSD), with collimator setting of 10 × 10 cm2. To obtain a FFF treatment beam with similar depth dose characteristics requires a higher initial beam energy provided by the bending magnet and magnetron power characteristics closer to that of an 8-MV beam. The philosophy of Varian is different in that the same beam generating parameters are used for flattened and FFF beams. The FFF beam is therefore much softer than its flattened counterpart. FFF beams in general are characterized by a forward-peaked intensity profile and a small variation in energy spectrum off axis.
Dosimetric data requirements for treatment planning systems and independent monitor unit check systems usually include variation of descriptors with field size. These may include the total scatter (Scp) output factors that are defined as the ratio of dose for the field of interest to that of a reference field for the same delivered monitor units measured under full scatter conditions in a large water tank at the reference depth, e.g., 10-cm deep isocentrically. Collimator scatter (Sc) factors, also known as head scatter factors, account for the variation in beam output with field size from changes in direct and indirect radiation from the head of the linac.1 It is recommended that they are determined in a miniphantom that has a smaller size than the smallest radiation field to be measured2 but provides sufficient charged particle equilibrium from the side walls.
It has been noted by Sauer3 that the conventional notion of equivalent square field size for high-energy FFF photon beams is different from that of conventional flattened beams. Data for phantom scatter (Sp) in conventional flattened beams have been published in the literature4 as a function of both field equivalent square and beam quality index. There is a need for similar dosimetric reference data for FFF beams. If these reference data are to be generally applicable, then they should be measured for a standard dosimetry system such as that advocated by European SocieTy for Radiotherapy and Oncology (ESTRO)5 and more recently the American Association of Medical Physicists,6 e.g., isocentrically at 90-cm SSD and 10-cm deep.
Section snippets
Measurements
An Elekta linear accelerator with an Agility MLC (Elekta AB, Sweden) and a Varian TrueBeam STx (Varian Inc, Palo Alto) operating at nominal 6 and 10 MV in flattened and FFF modes were used for all measurements in this study. The quality indices of the beams, as described by the TPR20/10 for a reference field of 10 × 10 cm2, were determined by direct measurement in water and are listed in Table 1. The percentage depth dose values for a field of 10 × 10 cm2 at 10-cm depth, 100-cm SSD, are also given
Sc data and fit
The measured Sc data for each beam investigated are listed in Table 2. It can be seen that these change with square field size, with a weaker variation with collimator setting for the FFF beams than that noted in their flattened counterparts. For example, the variation shows a range of 0.966 to 1.026 for the Elekta 6-MV flattened beam and 0.986 to 1.010 for the equivalent unflattened beam for square field sizes of 4 to 40 cm. Similar ranges are seen for the measurements with the Varian linac,
Discussion
The agreement between published and calculated Sp factor data for the 3 flattened beams was within 1.0% for all beams and field sizes, except for the field of 40 × 40 cm2 of the Elekta 6-MV beam, which was + 1.2% different. The mean difference (± 1 standard deviation) in our experimentally determined Sp and published data was + 0.08 ± 0.43%, + 0.06 ± 0.20%, and − 0.04 ± 0.38% for the Varian 6-MV, Varian 10-MV, and Elekta 6-MV beams, respectively. It can therefore be concluded from this
Conclusions
The dosimetric quantities Sc, Scp, and hence Sp required for planning system measurement, or a monitor unit check methodology, were easily and accurately parameterized for our flattened and FFF beams using a simple mathematical expression. The data reproduced may be used as expectation values for comparison when commissioning similar beams, as there are scant published data on FFF beams from these accelerator types. As has been highlighted here, the use of published standard flattened field Sp
References (7)
- et al.
In-air output ratio, Sc, for megavoltage photon beams
Med. Phys.
(2009) - Aspradakis M., Byrne J, Palmans H; et al. Small field MV photon dosimetry, IPEM report no. 103. IPEM (York);...
Determination of the quality index (Q) for photon beams at arbitrary field sizes
Med. Phys.
(2009)
Cited by (6)
Experimental and Monte Carlo-based determination of the beam quality specifier for TomoTherapyHD treatment units
2018, Zeitschrift fur Medizinische PhysikCitation Excerpt :The non-uniformity of the dose profile in a beam generated without a flattening filter (FFF beam) results to differences in phantom scatter that would influence the dose variation with increasing depth. These can be of the order of 4% in 6MV field size of 40 × 40 cm2 [14,15]. The addendum to the 1990 United Kingdom dosimetry code of practice recommends the application of Eq. (3) for TomoTherapy machines [16] for a field of 10 × 5 cm2 and ignores the influence of the non-flattened beam.
Zero field PDD and TMR data for unflattened beams in conventional linacs: A tool for independent dose calculations
2016, Physica MedicaCitation Excerpt :The question is whether BJR formulas are directly applicable to FFF beams. Indeed, the applicability to fields larger than 15–20 cm2 fields has been questioned by other authors [10,11]. Our paper aims at investigating the applicability of the above mentioned standard analytical formulas/methods to non-conventional FFF beams, using experimental and MC simulation data to provide independent validations.
Empirical determination of collimator scatter data for use in Radcalc commercial monitor unit calculation software: Implication for prostate volumetric modulated-arc therapy calculations
2016, Medical DosimetryCitation Excerpt :Collimator scatter data for the independent MU check program were measured as part of linac commissioning and input into the Radcalc software as a function of square field size. These data have previously been published in the literature.5 The VMAT treatment plans for the 101 patients with prostate cancer discussed in the previous section were verified by point dose measurements on the linac delivered through Integrity Version 3.2 software (Elekta AB, Stockholm).
Comparison of absolute dose calibration for linear accelerator in FFF mode based on IAEA 277 report versus IAEA 398 report
2020, Chinese Journal of Medical PhysicsA full quantitative analysis of 18 MV photon beam from 2100 C/D-Varian clinical linear accelerator with and without flattening filter
2019, International Journal of Radiation ResearchA comparison between direct TMR measurements and TMRs calculated from PDDs using BJR Supplement 25 data for flattened and unflattened photon beams
2015, Australasian Physical and Engineering Sciences in Medicine