Technical note
Effect of serum content and diluent selection on assay sensitivity and signal intensity in multiplex bead-based immunoassays

https://doi.org/10.1016/j.jim.2007.09.011Get rights and content

Abstract

Introduction

In the Luminex multiplexed immunoassay system, complex samples such as human serum are diluted to minimize disturbing matrix effects with a specific diluent. This diluent has to imitate the sample matrix to allow interpolation and has to provide optimal cytokine–antibody binding for all cytokines. Because diluents influence multiplex immunoassay results, this paper explores several methods to determine the quality of a chosen diluent.

Materials and methods

Two commercially available diluents, DY997 and RD6 from R&D Systems, were compared in a 19-plex immunoassay setup from Luminex.

Results

Using diluent DY997, multiplex signal intensity was reduced by 55% when spiked samples (chemokines and cytokines at 100 pg/mL) contained 50% v/v human serum, compared to samples containing 25% v/v. When using diluent RD6, signal intensity was reduced by 20% when samples contained 50% v/v human serum, compared to 25% v/v human serum. Diluent DY997 showed decreasing multiplex assay sensitivity with increasing protein concentrations, but not as low as in the presence of 50% v/v human serum.

Conclusions

In a 19-plex setup, this paper describes signal intensity, assay sensitivity and background signal levels in relation to the total volume-fraction of serum and protein concentration. For the determination of cytokines in serum samples with the multiplexed system Luminex the diluent RD6 seems more appropriate than the diluent DY997.

Introduction

There is increasing demand for multiplexed immunoassay systems that allow multiple cytokine determination by using a small amount of samples (Kurkjian et al., 2006, Martins et al., 2006, Nolan and Mandy, 2006, Skogstrand et al., 2005).

The Luminex-100 bead-based system combines the principle of ELISA and fluorescent bead-based flow cytometry and is theoretically capable of measuring up to 100 analytes simultaneously.

An important factor in multiplex assay optimization is the sample diluent; a protein mixture that minimizes disturbing influences such as heterophilic antibodies and unspecific binding. In addition, the diluent optimizes cytokine detection by sample dilution and, at the same time, mimics the sample matrix in the calibration curve (i.e. human serum) to allow accurate interpolation of unknown sample response in an external calibration curve. Diluents, of course, lack potentially disturbing factors such as endogenous cytokines and heterophilic antibodies (de Jager et al., 2005, Muller et al., 2002).

In a 19-plex setup, this paper describes signal intensity, assay sensitivity and background signal levels in relation to the total volume-fraction of serum and protein concentration. We present several methods to determine the quality of a diluent. As an example we used two widely used diluents from R&D Systems; DY997 and RD6. We demonstrate the importance of using a diluent that adequately imitates the sample matrix.

Section snippets

Luminex multiplex assay

Proteins and antibodies used were purchased from R&D Systems (Wiesbaden, Germany) with the exception of recombinant proteins IL-1beta, IL-2, IL-5, IL-6, IL-8, IL-10, IL-13, IFN-gamma which were from Strathmann (Hamburg, Germany), capture antibody IL-4 was from Bioscience (Cambridge, United Kingdom), IL-6, IL-10, IL-13 were from Pelikine (Amsterdam, The Netherlands) and detection antibodies IL-4 and IL-10 were from Bioscience and Pelikine, respectively. All cytokines and proteins were

Influence of different diluents on sensitivity of calibration curves

We compared calibration curves prepared in 100% v/v DY997 and in 50% v/v DY997 + 50% v/v human serum, representative of sample volume fraction used in Luminex assays (Herder et al., 2006). Calibration curves for IL-8, as an example for the effects observed in all investigated cytokines, were calculated using a 5-parameter logistic (PL) function (Fig. 1A). Interestingly, the IL-8 background mean fluorescence intensity (MFI) signal of 50% v/v DY997 + 50% v/v human serum was actually lower than DY997

Conclusions

As shown in this paper for human serum, a diluent needs to mimic the sample matrix when establishing an external calibration curve to allow accurate interpolation of unknown samples. With respect to our specific assay requirements, diluent RD6 from R&D Systems appears to be a more appropriate candidate for cytokine determination using the Luminex system. Our recommendation would be that one should compare several diluents in order to optimize a multiplex setup. More and more companies are

Acknowledgements

We are grateful to Mrs. G. Gornitzka and Mr. Y. Sagik for expert help with the Luminex analyses. We thank Dr. B. Rose, Institute for Clinical Diabetes Research, German Diabetes Center, Duesseldorf, for fruitful discussions.

References (8)

There are more references available in the full text version of this article.

Cited by (0)

View full text