Evaluating MISER chromatography as a tool for characterizing HILIC column equilibration

https://doi.org/10.1016/j.chroma.2020.460931Get rights and content

Highlights

  • Evaluation of MISER for use in HILIC column characterization.

  • Characterization of one HILIC column using three different mobile phases.

  • Percentage water in the mobile phase drastically effects column equilibration.

Abstract

Hydrophilic Interaction Liquid Chromatography (HILIC) is a technique for retaining polar analytes that uses polar stationary phases and acetonitrile-rich mobile phases. While this technique has several advantages over reversed-phase liquid chromatography (RPLC), one main drawback is the reported need for longer column equilibration. The reason for this is not fully understood and is a topic of current investigation. In order to better understand and reduce the equilibration needs, accurate characterization of column equilibration under varying conditions is required. The current method of characterizing HILIC column equilibration produces limited data points per test, or low time resolution, and is highly dependent on the column and probe compounds being used. There is a need for an improved method for characterizing HILIC column equilibration, especially if trends across stationary phases are to be observed. In this work, MISER, or Multiple Injections in a Single Experimental Run, is evaluated as a possible tool for characterizing HILIC column equilibration. MISER improves time resolution by allowing for replicate injections without interruption of data collection, enabling a more thorough evaluation of column equilibration compared to traditional techniques. Experimental results gathered using MISER show that equilibration of a BEH Amide column is notably shorter when equilibrating from acetonitrile to mobile phases containing higher percentages of water. Column equilibration to a 10% aqueous mobile phase was found to be approximately 5-fold faster than equilibration to a 3% aqueous mobile phase.

Introduction

Hydrophilic Interaction Chromatography (HILIC) is an analytical technique used to separate polar analytes. Since its naming by Alpert in 1990 [1], HILIC has grown in popularity in many fields of study including metabolomics, biological research, food and environmental testing, and pharmaceutical analyses, with the number of published papers almost doubling in the last ten years [2]. Articles covering method development for challenging polar compounds as well as investigations of the retention mechanisms in HILIC have been published [3], [4]. The advantages of HILIC compared to reversed phase chromatography include the ability to retain highly polar analytes, different selectivity, greater sensitivity for mass spectrometry detection and lower operating pressures. One of the disadvantages of this technique is the need for greater equilibration volumes. While reversed-phase columns typically equilibrate in less than 10 column volumes (VC) [5], HILIC columns can require 20 VC or more to achieve full equilibration [6], [7]. While the notion of equilibration typically applies to gradient runs, wherein the analyst is trying to bring the column back to starting conditions as fast as possible, the same mechanisms are involved for isocratic analyses. The difference for isocratic analyses is that equilibration only takes place when transitioning from storage conditions to analysis conditions. Similar to gradient runs, reduced equilibration times would lead to an overall increase in throughput.

The equilibration of HIILC columns has been studied by several scientists including McCalley [6], Heaton [7], Rappold [8], and Shollenberger [9]. Rappold found that for targeted metabolomics using gradient elution, retention and peak shape of analytes was dependent on the amount of re-equilibration between gradient injections [8]. Peak shapes would deteriorate at low equilibration volumes, leading to poor integration and quantitation. Additionally, retention times would shift if different equilibration volumes were used. The work also indicated that when controlling equilibration between gradients, stable retention times are possible, even if full equilibration was not achieved. McCalley investigated factors that impact HILIC equilibration, including flow rate, the storage conditions of the column, and the presence of buffer in the storage solvent. The water content of the storage solvent was shown to have a strong influence on the speed of equilibration, while the presence or absence of buffer in the storage solvent was of little consequence for isocratic separations [6]. Heaton found that for their assay the selected column required 40 VC or more to achieve full equilibration, but the volume could be reduced by partially equilibrating the column [7]. This made re-equilibrating the column between gradient runs considerably faster without sacrificing retention reproducibility. Lastly, Shollenberger found that the presence of ion-exchange mechanisms slows equilibration for both isocratic and gradient runs [9].

HILIC equilibration is linked to the formation of an adsorbed aqueous layer on the surface of the stationary phase [10], [11]. This requires the bulk mobile phase to penetrate the pores of the particle and stabilize in the surface layer. Understanding the factors that affect this mechanism is critical to being able to reduce equilibration times and ultimately speed up HILIC analyses. Factors like the water content of the storage solvent and mobile phase, flow rate, and stationary phase all impact the speed at which the column equilibrates [6]. Accurate characterization of the equilibration of a column under differing conditions is crucial to understand how to reduce the equilibration needs of a column. Unfortunately, current methods for characterizing HILIC equilibration are limited. Conventional methods include flushing a column with a storage solvent and then injecting a test sample in replicate to track retention time as the column equilibrates to assay conditions. When the retention times match a target value, or stabilize, the column is considered equilibrated. While a good starting point, this approach has several limitations and an improved method of characterizing HILIC equilibration is needed.

MISER, or multiple injections in a single experimental run, is a technique for high-throughput analyses [12]. It has been utilized for applications including reaction monitoring [13], sample screening [14] and scouting ideal reaction conditions [15]. The advantage of MISER is that many injections may be made while the separation is occurring, without waiting for the initial separation to be completed. When applied to studies of column equilibration, this allows for considerably more data points per test than conventional techniques. The work presented here shows the limitations of the conventional method for characterizing equilibration, the application of MISER to study column equilibration, and comparisons to the conventional method. We also used the MISER method to characterize the equilibration under different mobile phase conditions for an ACQUITY UPLC BEH Amide column, a widely-used HILIC column that has been shown to have moderate equilibration needs.6

Section snippets

Chemicals

LC-MS grade Acetonitrile and MS grade formic acid were purchased from Fisher Scientific (Hampton, NH). Ammonium formate (≥99.995% trace metals basis), nortriptyline, maleic acid, uridine, and bretylium tosylate were purchased from Millipore-Sigma (Burlington, MA). Deionized water was produced using a Millipore Milli-Q system.

Instrumentation and columns

Tests were performed on an ACQUITY UPLC H-Class instrument from Waters Corporation (Milford, MA) consisting of a quaternary solvent manager (QSM), a sample manager with a

Choice of test probes and test descriptors

Evaluating HILIC column equilibration requires a test mixture containing a range of probes, ideally including neutral, acidic and basic compounds. This provides information on the behavior of each type of probe. As demonstrated by Shollenberger [9], the presence of ionic interactions can slow column equilibration. In the first experiment, three probes were chosen: uridine (a neutral nucleoside), bretylium (quaternary amine), and tosylate (sulfonic acid). Both bretylium and tosylate are charged

Conclusions

MISER was shown to be a useful tool for evaluating HILIC column equilibration, mitigating the limitations of the conventional technique, most notably the poor time resolution. A five-fold increase in time resolution was demonstrated and further increases may be possible. MISER was demonstrated to give reproducible equilibration volumes, with relative standard deviations ≤6%. The technique was used to characterize the equilibration of an ACQUITY BEH Amide column from acetonitrile to three

CRediT authorship contribution statement

Kenneth D. Berthelette: Conceptualization, Methodology, Validation, Formal analysis, Investigation, Resources, Writing - original draft, Writing - review & editing, Visualization. Thomas H. Walter: Conceptualization, Methodology, Formal analysis, Resources, Writing - review & editing, Visualization, Supervision. Martin Gilar: Conceptualization, Methodology, Formal analysis, Writing - review & editing, Visualization. Fabrice Gritti: Conceptualization, Methodology, Formal analysis, Writing -

Declaration of Competing Interests

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Cited by (8)

  • An approach to high throughput measurement of accurate retention data in liquid chromatography

    2022, Journal of Chromatography A
    Citation Excerpt :

    Other groups have demonstrated improvements in throughput of LC separations by focusing on “overhead” associated with the instrument itself. For example, the MISER approach introduced by Welch et al. eliminates between-analysis software processing time by making multiple injections during the course of data acquisition stored in a single datafile [16–18], however to the best of our knowledge this approach has not been used for building retention databases. Our view is that a primary reason that extensive, open retention databases do not currently exist is that acquiring these data is very resource intensive (both in terms of instrument and person time).

  • Method for evaluation of ionic interactions in liquid chromatography

    2020, Journal of Chromatography A
    Citation Excerpt :

    For example, the anionic interaction of Charged Surface Hybrid RP sorbents with analytes is expected to be minor [7,8], while Primesep or Acclaim mixed-mode family columns exhibit relatively strong retention for ionic analytes [12–14,16,17,19]. Selected HILIC stationary phases based on silica or immobilized charged ligands provide for ionic interactions that participate in the overall analytes retention [28–31]. Contribution of silanols to retention of basic compound in RP LC is well known [2,32–34]; it has been also recognized as one of the retention modes in HILIC not only with silica columns, but also with derivatized silica surfaces [28,30,35–40].

  • Essentials in Modern HPLC Separations

    2022, Essentials in Modern HPLC Separations
View all citing articles on Scopus
View full text