Original contributionAccurate and sensitive measurements of magnetic susceptibility using echo planar imaging
Introduction
Susceptibility differences in an object can cause many problems in magnetic resonance imaging (MRI) and magnetic resonance spectroscopy (MRS). Near interfaces, the magnetic field is disturbed; therefore, the spatial encoding in MRI is affected. The signal displacements can appear as high-intensity spots and spots with signal void in the image due to accumulation or subtraction of signal. In MRS, a magnetic field change can lead to a shift in resonance frequency and peak broadening.
If it is possible to determine the susceptibility of phantom and implant materials, then it is also possible to minimize the artifacts in advance by choosing materials with small susceptibility differences. This can become even more important as the trend for clinical MR goes toward higher field strength, 3 T and beyond, and higher field strengths lead to more pronounced susceptibility effects.
By means of MRS, a susceptibility-induced shift in resonance frequency can be used to determine the volume susceptibility of signal-giving liquids [1]. The method is accurate and easy to use. However, phantoms are often made of plastics or other solid materials, which is why the use of the MRS method is not always applicable. Beuf et al. [2] devised an MRI method that can be used for any sort of material because it is not the object itself but the effect it has on a reference liquid that is imaged. The susceptibility difference between an outer compartment and an inner compartment in a coaxial circular cylindrical phantom was determined by the size of the artifact in an MR image. For ordinary imaging sequences, the sensitivity was limited and the susceptibility differences caused by materials usually used in phantoms (e.g., plastics) were not detectable.
The aims of this study were (1) to increase the sensitivity of the MRI measurements of susceptibility differences to make it possible to also measure plastics and tissue-like materials, (2) to improve the sensitivity and accuracy in data evaluation by designing an automatic evaluation program based on model fitting and (3) to apply the improved method to measure the volume susceptibilities of polymethyl methacrlyate (PMMA) and polyethylene.
Section snippets
Theory
Not only the artifacts but also the susceptibility parameter itself can be very confusing. Volume (χ), mass (χm) and molar (χM) susceptibilities, given in either SI or centimeter-gram-second (cgs) units, are all present in the literature. Volume susceptibility is defined as the dimensionless proportionality coefficient between magnetization, M, and magnetizing field strength, H; it describes the contribution to the magnetic flux density present, B, made by a substance when subjected to a
Results
The results from the MRS measurements showed good agreement with the volume susceptibilities of the CuSO4 solutions as calculated using Eq. (1) (Fig. 2).
MR images using EPI exhibited larger artifacts as compared with those using SE (Fig. 3); that is, they gave a higher sensitivity in susceptibility determination. The smallest equivalent susceptibility difference that gave an observable artifact in EPI images was Δχe=0.02 ppm. The lowest equivalent susceptibility difference among the tested
Discussion
All results were in accordance with the theory. The MRI method using EPI and model fitting for evaluation proved to be comparable with the MRS method with regard to accuracy and sensitivity but with the strong advantage that it could be used for any type of sample.
Model fitting was found to be superior to the manual image analysis because it utilized the entire artifact instead of only the three intensity spots and therefore the evaluation was less sensitive to other image artifacts. It also
Conclusion
On a clinical MR scanner, the use of EPI instead of standard implementation of SE sequence increases sensitivity significantly. Increased sensitivity gives an opportunity not only to detect smaller susceptibility differences but also to decrease the sample radius and volume. Evaluation by model fitting increased the sensitivity as well as the accuracy further as compared with manual image analysis. The results were verified by the MRS measurements.
Acknowledgments
This work was supported by grants from the Swedish Research Council (14039) and the Lundberg Foundation (Göteborg, Sweden).
References (8)
- et al.
Magnetic resonance imaging for the determination of magnetic susceptibility of materials
J Magn Reson
(1996) - et al.
Bulk magnetic susceptibility shifts in NMR studies of compartmentalized samples: use of paramagnetic reagents
Magn Reson Med
(1990) - Gupta. Landolt-Bornstein Zahlenwerte und Funktionen. Volume New Series II/16. p...
The role of magnetic susceptibility in magnetic resonance imaging: MRI magnetic compatibility of the first and second kinds
Med Phys
(1996)
Cited by (5)
Magnetic properties of materials for MR engineering, micro-MR and beyond
2014, Journal of Magnetic ResonanceCitation Excerpt :A well-known method to measure the susceptibility of materials that do not produce an MR signal is to place the measurement samples in a reference material (for example water) with a known susceptibility and measure the field distortion in the reference material outside the sample. Typically, the sample is a long cylinder, for which the distortion is a dipole [12,13,11,14,15]. [2] uses a slightly special method, as there, the authors place two rings of the sample material in a Helmholtz configuration, such that the field distortion in the centre between the rings is homogeneous.
Differentiation between recurrent gliomas and radiation necrosis using arterial spin labeling perfusion imaging
2016, Experimental and Therapeutic MedicineIsomagnetophoresis to discriminate subtle difference in magnetic susceptibility
2008, Journal of the American Chemical Society