1 Introduction

Parkinson’s disease (PD) is a neurodegenerative disease affecting about 6.1 millions of people worldwide [1]. Over the years, as a result of an increasing number of older people, the global burden of PD has become more critical [1, 2]. As worldwide life expectancy has increased it will likely continue to grow. The projected number of people with PD in the five most populous nations of Western Europe and in the ten most populous nations of the world for 2030 is 8.67 million, considering individuals over 50 years old [3].

PD is a very complex condition with several motor and non-motor symptoms that evolve over time for specialists to treat optimally. Its diagnosis is purely clinical and relies on medical history and neurological evaluation [4]. Newly, the guidelines for clinical diagnosis of PD have been revised [5]. Nonetheless, the motor symptoms still reflect the traces of the disorder and are one of the most important features for PD diagnosis.

The Unified Parkinson’s Disease Rating Scale (UPDRS) is a structured assessment tool that helps with disease progression and treatment response documentation [6]. The scale has four parts and measures complications of daily living non-motor experiences, daily living motor experiences, motor impairment, and complications of treatment. Such tools are still subjective despite the existence of scales to evaluate PD, and there is a strong dependence on the experience of the specialist.

These limitations in the diagnosis and follow-up of PD have motivated the development of new methods for PD assessment [7]. Wearable motion sensors, such as accelerometers, gyroscopes, magnetometers, and many other microelectronic devices have been used to obtain motor- and non-motor-related physiological information (i.e. anxiety, impaired decision-making, sleep disturbance) aspects of the disease [7]. These techniques are attempting to provide an objective way to measure the disease’s motor symptoms with the intention of improving diagnosis, assessing disease progression, and evaluating the effectiveness of care and therapy.

Although several studies using these sensors show relevant results and a strong correlation with severity scales [7], the use of contact sensors presents challenges [8]. These sensors can cause skin irritation, require careful preparation of the skin and may not be used in situations with high risk of contact contamination [8, 9]. On the other hand, the non-contact capacitive (NCC) sensor is an option for motion and biopotential analysis. This technology offers many advantages such as replicability, stability in long-term task monitoring, patient comfort and the possibility of application for daily monitoring [10].

There are some studies that introduced the use of NCC sensors in the context of motor quantification of PD. For example, the authors in [11] evaluated the motor tasks performed by a healthy person and one with PD. That paper describes the fundamental technology and provides some preliminary results. In a follow-up study [12], while healthy subjects were performing wrist movements, we compared the NCC sensor with gyroscope. The guiding hypothesis was that it is possible to quantify the main motor signs of Parkinson’s disease, similar to one of the most commonly used inertial sensors, i.e., the gyroscope, by using the NCC electrical potential integrated circuit (EPIC) sensor.

In this study, it is presented the development of hardware and a wider application of this technology, based on the technology of electrical potential sensing (EPS) [13], for the evaluation of hand motor tasks performed by subjects with and without PD.

This research aims to address two issues: (1) modeling the relationship between NCC responses and displacement in a controlled scenario; (2) experimental evaluation of the implemented system by comparing results obtained from a gyroscope, whereas healthy individuals and people with PD performed different hand motor tasks.

The paper is structured as follows. Section 2 presents the materials and methods with the Sects. 2.1, 2.2, 2.3, 2.4, 2.5, 2.6 and 2.7. The results are presented in Sect. 3. In Sect. 4, the discussion about the 4.1 NCC system, 4.2 Waveforms and 4.3 Frequency analysis is presented. Lastly, conclusion and direction of future research are presented in Sect. 5.

2 Materials and methods

2.1 Non-contact capacitive sensor

The human body has dielectric properties and when it moves the electrical field is disturbed and NCC sensors are designed to detect this type of event [10]. By using this tool, different human tasks can be differentiated. Several studies have documented using this sensing technique for multiple purposes, including non-invasive respiration and cardiac monitoring [14], wearable devices to detect hair touch and restless leg movement [15] and passive tracking of movements [16]. However, we have not found a report on the use of capacitive sensors for the monitoring of hand movements in the context of PD.

The PS25454 EPIC capacitive sensor (see Fig. 1), used in our system, incorporates several features to enhance the measured signals, allowing the detection of field disturbance due to the movement of a nearby object. The main features of the employed sensor are ultra-high input resistance, typically 20 GΩ; input capacitance as low as 15 pF; lower − 3 dB point typically 0.2 Hz and upper − 3 dB point typically 20 kHz. It works with bipolar power supply from ± 2.4 V to ± 5.5 V.

Fig. 1
figure 1

PS25454 EPIC sensor (Plessey semiconductors, UK), a front and b rear views

This sensor (see Fig. 1) was designed with the specifications that made it suitable for the detection of movement in an open environment. When a person moves, the electrical field (E-field) is disturbed so that a relative signal can be used to estimate various position and motion-related information. One of the main advantages of this sensor technology is its ability to operate fully passively, without the need to emit waves in the environment, such as in the case of radar [17].

2.2 System design

Four PS25454 sensors were used in pairs in our system (National Institute of Intellectual Property—Brazil—BR 10 2018 069864 8). The directional axis (x or y) is defined by each pair. A custom-made structure was designed using four sensors positioned at the four centered edges of a squared area of size 0.21 m × 0.21 m. Figure 2 shows this kind of arrangement.

Fig. 2
figure 2

2-D array of PS25454 sensors. Each pair of sensors provides information on movements in the x-axis and the y-axis. In the background, the marker board was specifically designed to assist the subject during the execution of experimental tasks. The orange arrow points to the four EPIC sensors

The sensors were configured in differential mode to measure the local electrical field. In order to avoid common mode noise, INA333 (Texas Instruments, Dallas, TX, USA) instrumentation amplifier was used, which has high common mode rejection, low power consumption and low price.

The sensors require a power supply voltage of ± 5.5 V, which is also appropriate for the instrumentation amplifier. In order to reduce environmental electrical noise, a battery was used as a power source.

The signals were fed to the instrumentation amplifier from both EPIC sensors. The output of the amplifier is then buffered by the voltage divider to adjust the signal level for analog-to-digital conversion (ADC). Figure 3 displays a simplified circuit diagram. The system was grounded to minimize interference with environmental noise.

Fig. 3
figure 3

Simplified circuit diagram to use PS25454 EPIC sensors in a differential mode for movement quantification

2.3 System validation

A micro-machine unit (Minitech CNC Mini-Mill/GX, USA) with a resolution of 1 μm was used as a guide for device validation. Attached to the reference system, the plaster hand model was fixed in order to simulate the use scenario. Figure 4 displays the experimental setup for the calibration of the device.

Fig. 4
figure 4

Experimental configuration for testing of the NCC framework. The orientation of the axes is indicated by the red arrows and the orange arrows point to the four EPIC sensors. Each sensor is encapsulated in a customized framework

Two scenarios were considered in order to establish the relationship between the displacement and the voltage amplitude range. For the first, the hand was fixed 8 cm apart from the sensor indicated by a red asterisk (see Fig. 4) and for the second, the hand was fixed 4 cm apart from the same sensor. In both scenarios, the movements were along the y-axis and were executed three times each in the following sequence: (1) move 10 mm; (2) move 5 mm; (3) move 2.5 mm; (4) move 1.5 mm and (5) move 0.5 mm; always going back and forth.

The aim of this stage of the analysis was to establish the relationship between hand displacement and the response of the EPIC sensor. To do this, a linear mixed model [18] was used to describe the relationship between the variables measured,

$$y \sim x + \left( {1 |setup} \right) +\upvarepsilon$$
(1)

where x is our fixed effect term related to the hand displacement, 1|setup is the random effect to model the two possible scenarios (i.e., the hand fixed at 4 cm and 8 cm apart from the reference sensor), and Ɛ is the general error term.

2.4 Participants

A total of 57 subjects were recruited, being 30 with PD (SPD) and 27 controls without PD (SC). All the subjects signed a written informed consent, and the study procedure was approved by the National Research Ethics Committee (65165416.4.0000.5152). Each group of subjects was paired in terms of age (equal medians with p = 0.56) and sex. Subjects with PD were evaluated using MDS-UPDRS by three qualified evaluators and the disease progression to different Hoehn and Yahr (HY) stages was also evaluated by three qualified evaluators. Table 1 shows characteristics of the subjects with PD. The information summary of controls group (SC) are: age = 60.7 ± 8.5 and sex = 17 M, 10 F. The results of UPDRS and HY scores provided in the table are the overall mean between evaluators.

Table 1 Clinical data of subjects with PD (SPD)

MDS-UPDRS Part III concerns the motor test with a total score ranging from 0 to 132 [6]. The overall mean score of Part III provided in Table 1 reflects the PD severity of our sample from the SPD group. As a figure, this is 25% ± 9%. In clinical terms, this score is correlated with a moderate to mild PD motor test severity. “Moderate” refers to signs with sufficiently low frequency or intensity to cause no impact on function and “mild” refers to signs of frequency or intensity sufficient to cause a modest impact on function [6].

2.5 Experimental setup

The experimentsFootnote 1 were performed in an open unshielded environment (Fig. 5). A three-axial gyroscope (L3GD20H, STMicroelectronics, Switzerland) was placed on the dorsal region of the hand along with a laser pointer to assist the subject in performing a set of motor tasks (Fig. 5). The TREMSEN (Precise Tremor Sensing Technology) system (National Institute of Intellectual Property—Brazil—BR 10 2014 023282 6) was used for data acquisition and real-time visualization of the signals [19].

Fig. 5
figure 5

The experimental setup. (left) The 3-axial gyroscope is positioned along with the laser point on the dorsal part of the hand. The orange arrow is pointing to the gyroscope sensor. The y-axes and z-axes are marked. The forearm is on the support. During the experiment, the squared board is used to guide the participant. (right) Set of tasks performed. The number of repetitions is shown for each task

Subjects were instructed to start a motor task with a visual laser point in the central region of the board (highlighted by a crossed circle). The white circles on the board were set as targets for each type of experimental task (Fig. 5). For task T1, subjects were instructed to hold a hand in a position against gravity with a visual laser point in the central region of the board for 10 s. For tasks T2 to T5, subjects were asked to perform hand movements 5 times in 10 s, while for tasks T6 to T9, subjects were asked to perform the same type of tasks 15 times in 10 s. This sequence was performed three times with each execution having a rest.

While the selected motor tasks are not in the UPDRS, they are closely related to those performed by patients during the UPDRS evaluation [6]. Gyroscopes are also suitable for evaluating the selected movements as they detect angular velocity, which may vary depending on the task.

Signals collected from the gyroscope were sampled at 50 Hz (with sensitivity of ± 500 º/s) [20] and from the NCC system at 3 kHz. This sample rate was chosen because of the sensor behavior, in a way that, after carrier removal, 50 points are left for every second. The signals were synchronized by an external trigger. The x and y axes of the NCC system correspond to the z and y coordinates of the gyroscope, respectively.

2.6 Signal pre-processing

The NCC sensor signal comes with an amplitude modulation carrier at 60 Hz (power line frequency). Thus, the carrier was removed using an envelope technique to extract the relevant information from the output sensor signal. The mains frequency (50 Hz/60 Hz) can be measured by each sensor. When a dielectric object, such as a hand, is positioned among two sensors, the level of the main signal detected by the sensor is altered.

The signals collected from the gyroscope were pre-processed by using a moving average filter and a nonlinear detrending technique described in [21]. For the signals detected from the NCC system, the sequence of steps was employed, first the signal envelope of the peaks at 60 Hz component was estimated to remove the carrier signal, followed by the same sequence of steps applied to the gyroscope.

2.7 Frequency analysis

The NCC sensor detects disturbances in the local E-field and translates this into a signal consisting of the sum of all disturbances detected in the E-field. The following steps were applied to the signals from both sensors (i.e. NCC and gyroscope): (1) estimate the instantaneous mean frequency (IMNF) based on Hilbert spectrum (HS) [22]; (2) calculate some of the key values of the descriptive statistics (i.e., min—minimum value excluding outliers, q1—first quartile, median, q3—third quartile and max—maximum value excluding outliers). Points are treated as outliers if they are greater than q3 + w × (q3 − q1) or less than q1 − w × (q3 − q1), where w is the maximum whisker length, and q1 and q3 are the 25th and 75th percentiles of the sample data, respectively; (3) apply one-sample Kolmogorov–Smirnov test to check the normality of the distribution; (4) apply a statistical test to check the statistical significance of the differences between the SPD and SC groups, considering the estimated descriptive statistical features for each axis.

3 Results

3.1 System validation

As a result of the experiment described in Sect. 2.3, Fig. 6 shows the model response of NCC EPIC sensor.

Fig. 6
figure 6

Relationship between displacement and amplitude range of NCC EPIC sensor. The dashed lines represent upper and lower 95% confidence intervals. In red on the upper left corner is the line equation of the linear mixed model

As shown in Fig. 6, the model presented strong linearity. R-squared and adjusted R-squared were nearly the same (R2 = 0.7692 and \(R_{adj}^{2}\) = 0.7631).

The following linear equation

$$y = 0.003074x + 0.001767$$
(2)

was estimated by means of the fitted model, where x is the displacement in millimeters and y the sensor output value in volts. From (2) it is possible to estimate the sensor behavior.

3.2 Typical waveforms

Figure 7 illustrates characteristic waveforms obtained for a subject from SC and SPD groups while executing the tasks shown in Fig. 5.

Fig. 7
figure 7

Typical signals collected. a subject without PD, b subject with PD. Each rectangular window highlighted in red represents the signal from a task

In Fig. 7, the signals from both sensors can be visually compared and a high similarity between their waveform can be seen. Each type of movement occurs most intensely on a specific axis. For instance, T4 and T5, i.e., flexion and extension, respectively, occur mostly on the y-axis. Nevertheless, it is also possible to observe smaller scale contribution on the z-axis to the gyroscope sensor, and on the x-axis to the NCC sensor, latter if the signal is zoomed in.

3.3 Frequency analysis

The estimation of IMNF for frequency analysis was made by means of the HS, for all subjects considering three trials and the nine tasks. Figure 8 shows an example of the estimation of IMNF, for signals collected from the task 2—abduction.

Fig. 8
figure 8

Typical IMNFs (in green), estimated through the HS, for task 2. a Signals and IMNFs for a subject without PD, for the x-axis of both sensors. b Signals and IMNFs for a subject with PD, for the x-axis of both sensors. Selected subjects are matched in age and sex

The differences between signal waveforms for SC and SPD subjects can be seen in Fig. 8. In the first 2 s of Fig. 8b, it is possible to see oscillations in the signals of gyroscope and NCC sensor, latter to a lesser extent. These oscillations come from hand shaking, and they are also present throughout the signal, similar to transients.

Table 2 presents a comparison between SC and SPD groups considering the estimated IMNF values for both sensors (i.e., NCC and gyroscope) and for each task. The p-values highlighted in bold help in visualization of which sensor axes and features not reached a significance level of 95% (p < 0.05).

Table 2 Comparison of descriptive statistics based on IMNF values between SC and SPD groups by Mann–Whitney U-test

In most cases, both sensors were able to discriminate features from signals of each group (SC and SPD). For NCC sensor, tasks 1 and 3 were not discriminated completely by considering both axes. The gyroscope sensor, in turn, did not show statistical significance for tasks 1, 3, 4, 5, 6, 8, and 9.

4 Discussion

This research proposed the use of non-contact capacitive sensors for the assessment of people with Parkinson’s disease. The results showed that the NCC sensor acts as well as a gyroscope sensor for discrimination against movements performed by people with and without PD.

The fact that PD has yet to be cured challenges science, and this is not the only challenge. Diagnosis and follow-up of PD is another, and over the years several studies have focused on the detection of motor and non-motor PD symptoms. PD was originally considered a pure motion disorder characterized by its main motor symptoms, also known as cardinal signs, i.e. bradykinesia, stiffness, tremor and postural instability [23] and these motor signs worsen over time.

Therefore, many sensors and techniques are employed to quantify the cardinal signs of PD. Most studies [7, 24, 25] address this challenge by using contact sensors (e.g., inertial sensors), and this method has limitations. As a result, the alternative use of NCC sensors offers new possibilities, in addition to overcoming certain drawbacks related to the contact approach.

The analysis of the tasks depicted in Fig. 5 can support the development of a score to quantify the severity of some of cardinal signs. For example, Amanda et al. [20] studied bradykinesia by quantifying the flexion and extension of the wrist. Pose versus Gravity was evaluated in [26] for the estimation of hand tremor parameters.

In addition, NCC systems can be helpful in many ways: (1) for people with sensitive skin, such as the elderly, the majority of whom suffer from PD [8]; (2) for invisible application, once they work without contact and even through the wall [27]; (3) as a tool to allow examination in environments with high risk of contamination by contact [9]; (4) less interference due to cable movements; (5) privacy concerns and (6) less amount of data and processing time while compared with image-based systems.

4.1 NCC system

To date, there have been only a few studies using EPIC sensors, especially non-contact versions. The data sheet of these sensors does not contain enough information, and the lack of application notes makes its use difficult.

González-Sánchez et al. [14] used EPIC sensors to develop an electrocardiographic (ECG) monitoring system that does not require skin preparation or direct contact. From a sensor instrumentation circuit point of view, their study presented very well the circuit design and some faced issues. The PS25251 EPIC sensor was used in their research and, according to its datasheet, this variant of the sensor was designed to be used as an ECG dry contact sensor. However, their first test setup of two differential sensors was used in a contactless approach. A laboratory mouse was placed between the sensors in this setup. In the end, a plastic tube was used with ground aluminum foil to reduce the electrical noise detected by the sensors.

Two approaches have been tested in our study. In the first, the NCC system was designed to operate in a single-ended mode. As a result, each sensor individually captured the E-field disturbance in such a way that the amplitude of the signal corresponding to the sensor unit increases when the subject’s hand approaches it. After that, in order to determine each axis, the signals, after digitization, were subtracted from each pair of sensors. This approach was presented as not stable and the signals were often swamped by environmental noise.

A differential mode between a pair of sensors was adopted in the second approach. Most of the time, using this method, the signal of interest was not swamped by environmental noise. After digitization, the 60 Hz signal appeared as a carrier, and the signal of interest varied the amplitude of the carrier, as did the amplitude modulation (AM). However, even using differential mode design, the signal of interest (i.e., voluntary and involuntary movements of the hand) sometimes seemed weak and was heavily swamped by environmental noise. In order to overcome this limitation, the grounding technique also referred to in [14] was used, i.e. the support of the arm was covered by aluminum foil connected to the ground (GND). Using this grounding technique, the signals were stable during all data collections.

As far as device validation is concerned, the calculated model shown in Fig. 6 provided an agreement with the measurements made by Aydin et al. [28]. The differences in amplitude range between our results and those presented in [28] are due to measured phenomena. In our experiment, the electrical field disturbance due to hand displacement was measured and, in their experiment, the goal was to calibrate the EPIC sensor as a field meter, thus establishing a known electrical field in the research laboratory.

Although the variations between the movements performed in the device validation (i.e. hand translation) and the experimental process (i.e. hand rotation), the basic principle of the NCC EPIC sensor was modeled and illustrated in Fig. 7. This phase was important to confirm the previous result [28] and to move forward. This kind of device testing could be expanded, considering, for example, a number of hand movements.

4.2 Waveforms

The results obtained suggest that both sensors, the NCC and the gyroscope, can measure all assessed movements (i.e., T1 to T9). The signal shapes of the NCC and the gyroscope sensor are very similar to those shown in Fig. 7. For example, a correlation study was carried out in [12], and the results showed a high degree of agreement between the two sensors. This similar behavior between them is important once the gyroscope sensor is commonly used and well known in the field of motion analysis, including the PD motor symptom assessment [7].

On the other hand, the NCC system is highly dependent on the design and calibration of the sensors. This problem can be seen in Fig. 7b, where the signal amplitude of the NCC sensor for the x-axis is higher than for the y-axis. There are at least two possible reasons for this phenomenon: (1) the subject’s hand is closer to the sensors of x-axis, once the human hand is wider in transverse plane than in the sagittal plane; (2) in Fig. 7b, the subject has PD and he suffers from bradykinesia (score 3 for item 3.4 of MDS-UPDRS) and rigidity (score 2 for item 3.3 of MDS-UPDRS), which usually leads to slow movement and decreased motion range. These two symptoms directly affect the movements of the subject’s hand that are reflected in the NCC sensor signal captured. The latter reason can also be noted, but to a lesser degree, in the gyroscope sensor signal.

4.3 Frequency analysis

Frequency domain information is important and commonly used for evaluating PD motor symptoms [24, 25, 29].

In Fig. 8, for example, IMNFs have been estimated over time. Figure 8b shows the task signal and its IMNF for the PD subject. Tremor symptom appears while analyzing signals from both sensors, and the frequency estimation method used was able to track signal frequency oscillation. This phenomenon can be visually observed in the first 2 s of Fig. 8b for both sensors where the subject had a severe hand tremor.

A comparison of descriptive statistics based on IMNF values between SC and SPD groups is shown in Table 2. Both sensors distinguish very well the motor patterns of both groups. Overall, the NCC sensor was performing better. In many cases where the NCC or gyroscope did not have statistical significance for a particular axis, the possible reason is that the movement analyzed occurred mainly on the other axis. For example, task 3 occurs primarily on the x-axis. The result of the NCC sensor in task 1 on the y-axis should be highlighted. In task 1, the subject was asked to maintain a posture against gravity, and in this posture, even if the subject’s hand shakes, it would be closer to the x-axis sensors and proportionally away from the y-axis sensors, given the anatomical dimensions of the hand.

5 Conclusion and future work

This paper described a new application of the non-contact sensing technology, which is completely passive and is working to detect variations in the electrical field of the ambient. We have shown that, with the custom-made arrangement of electrical potential sensors, hand movements can be quantified. The system was calibrated and applied for the assessment of wrist motor tasks performed by subjects with and without Parkinson’s disease. This technology and the application presented in this paper open up new possibilities for the field of movement analysis.

A comparison was made using the estimated frequency features of the NCC and gyroscope signals, together with a statistical significance check. In view of this scenario, the NCC system has been shown to be as successful as the gyroscope-based system.

In addition to the practicality of using the proposed system on a daily basis, it is also suitable for people with skin problems, such as sensitive skin in the case of the elderly. It could be used in environments with a high risk of contact contamination. While compared to image-based systems, there are no privacy concerns once the proposed system does not require the capture and/or storage of data to be processed.

In addition, other experimental configurations could be evaluated in future studies, as the literature shows that distance and number of sensors impact directly on the acquired signal [14, 28]. Studies to turn this system into wearable could also be relevant to the field.