Abstract
In areas like adaptive therapy, multi-phase radiotherapy, and single fraction palliative treatment or in the treatment of patients with metal implants where megavoltage(MV) CT could be considered as a treatment planning modality, the reduced contrast in the MV CT images could lead to limited accuracy in localization of the structures. This would affect the precision of the treatment. In this study, as an extension our previous work on bespoke MV cone beam CT (MV CBCT), we propose to register the MV CBCT with kilovoltage (kV) CT for treatment planning. The MV CBCT images registered with kV CT would be effective for treatment planning as it would account for the inadequate soft tissue information in the MV CBCT and would allow comparison of changes in patient dimensions and assist in localization of the structures. The intensity based registration algorithm of the BrainSCAN therapy planning software was used for image registration of the MV CBCT and kV CT images. The accuracy of the registration was validated using qualitative and quantitative measures. The effect of image quality on the level of agreement between the contouring done on both the MV CBCT and kV CT was assessed by comparing the volumes of six structures delineated. To assess the level of agreement between the plans after the registration, two independent plans were generated on the MV CBCT and the planning CT using the posterior fossa of the skull as the target. The dose volume histograms and conformity indices of the plans were compared. The results of this study show that treatment planning with MV CBCT images would be effective, using additional anatomical structure information derived from registering the MV CBCT image with a standard kVCT.
Similar content being viewed by others
Abbreviations
- CT:
-
Computed tomography
- MV CBCT:
-
Megavoltage cone beam CT
- MV CT:
-
Megavoltage computed Tomography
- kV CT:
-
Kilovoltage computed tomography
- kV CBCT:
-
Kilovoltage cone beam computed tomography
- EPID:
-
Electronic portal imaging device
- HU:
-
Hounsfield units
- TPS:
-
Treatment planning system
- DVH:
-
Dose volume histogram
- CI:
-
Conformity index
References
Jaffray DA (2007) Kilovoltage volumetric imaging in the treatment room. Front Radiat Ther Oncol 40:116–131
Pouliot J (2007) Megavoltage imaging, megavoltage cone beam CT and dose-guided radiation therapy. Front Radiat Ther Oncol 40:132–142
Jaffray DA et al (2002) Flat-panel cone-beam computed tomography for image-guided radiation therapy. Int J Radiat Oncol Biol Phys 53(5):1337–1349
Yan D et al (2000) An off-line strategy for constructing a patient-specific planning target volume in adaptive treatment process for prostate cancer. Int J Radiat Oncol Biol Phys 48(1):289–302
Kaus MR et al (2007) Assessment of a model-based deformable image registration approach for radiation therapy planning. Int J Radiat Oncol Biol Phys 68(2):572–580
Yan D et al (2005) Computed tomography guided management of interfractional patient variation. Semin Radiat Oncol 15(3):168–179
Broderick M et al (2007) A comparison of kilovoltage and megavoltage cone beam CT in radiotherapy. J Radiother Prac 6:173–178
Ruchala KJ et al (1999) Megavoltage CT on a tomotherapy system. Phys Med Biol 44(10):2597–2621
Aubin M et al (2006) The use of megavoltage cone-beam CT to complement CT for target definition in pelvic radiotherapy in the presence of hip replacement. Br J Radiol 79(947):918–921
Kachnic L, Berk L (2005) Palliative single-fraction radiation therapy: how much more evidence is needed? J Natl Cancer Inst 97(11):786–788
van den Hout WB et al (2003) Single- versus multiple-fraction radiotherapy in patients with painful bone metastases: cost-utility analysis based on a randomized trial. J Natl Cancer Inst 95(3):222–229
Hansen EK et al (2006) Image-guided radiotherapy using megavoltage cone-beam computed tomography for treatment of paraspinous tumors in the presence of orthopedic hardware. Int J Radiat Oncol Biol Phys 66(2):323–326
Lewis DG et al (1992) A megavoltage CT scanner for radiotherapy verification. Phys Med Biol 37(10):1985–1999
Mosleh-Shirazi MA et al (1998) A cone-beam megavoltage CT scanner for treatment verification in conformal radiotherapy. Radiother Oncol 48(3):319–328
Newhauser WD et al (2008) Can megavoltage computed tomography reduce proton range uncertainties in treatment plans for patients with large metal implants? Phys Med Biol 53(9):2327–2344
Pouliot J et al (2005) Low-dose megavoltage cone-beam CT for radiation therapy. Int J Radiat Oncol Biol Phys 61(2):552–560
Thomas TH et al (2009) The adaptation of megavoltage cone beam CT for use in standard radiotherapy treatment planning. Phys Med Biol 54(7):2067–2077
Song WY et al (2006) Prostate contouring uncertainty in megavoltage computed tomography images acquired with a helical tomotherapy unit during image-guided radiation therapy. Int J Radiat Oncol Biol Phys 65(2):595–607
McQuade P et al (2005) Investigation into 64Cu-labeled Bis(selenosemicarbazone) and Bis(thiosemicarbazone) complexes as hypoxia imaging agents. Nucl Med Biol 32(2):147–156
Kessler ML (2006) Image registration and data fusion in radiation therapy. Br J Radiol 79:S99–S108
Kak AC, Slaney M (1988) Principles of computerized tomographic imaging. IEEE Press, New York
Lawson JD et al (2007) Quantitative evaluation of a cone-beam computed tomography-planning computed tomography deformable image registration method for adaptive radiation therapy. J Appl Clin Med Phys 8(4):2432
Feuvret L et al (2006) Conformity index: a review. Int J Radiat Oncol Biol Phys 64(2):333–342
Wu QR et al (2003) Quality of coverage: conformity measures for stereotactic radiosurgery. J Appl Clin Med Phys 4(4):374–381
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Thomas, T.H.M., Devakumar, D., Balukrishna, S. et al. Validation of image registration and fusion of MV CBCT and planning CT for radiotherapy treatment planning. Australas Phys Eng Sci Med 34, 441–447 (2011). https://doi.org/10.1007/s13246-011-0092-2
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s13246-011-0092-2