Visualization of the needle is important for safe and effective ultrasound-guided peripheral nerve block (PNB).1 The most common error of novice and experienced operators while performing ultrasound-guided PNBs is failure to visualize the needle prior to advancement.2 Previous studies have shown that needle visibility becomes poor as the angle of needle insertion increases in relation to the ultrasound probe due to reflective signal losses.3,4 Several techniques are now available to improve needle visibility, including echogenic needle design and electronic beam steering.5-8 Echogenic needles have reflective surfaces like Cornerstone reflectors that increase the number of ultrasound waves returning to the transducer; however, they are more expensive than the corresponding non-echogenic needles. Electronic beam steering is a technique that steers the incident ultrasound beam perpendicular to the angle of needle insertion to increase the amount of reflected ultrasound waves and can be an alternative to the use of echogenic needles. According to the law of reflection, the angle of incidence equals the angle of reflection; therefore, by directing the ultrasound beam perpendicular to the needle, the ultrasound waves can be reflected back to ultrasound probe, leading to enhanced needle visibility. Electronic beam steering, a software upgrade (known as EMB for the SonoSite ultrasound machine) for a multibeam transducer, entails a one-time installation cost. Beam steering has yet to be evaluated for its applicability in regional anesthesia as compared with the echogenic needle design. The aim of our study was to compare the effect of beam steering on the visibility of the echogenic and non-echogenic needles.

Methods

The University of Western Ontario Health Sciences Research Ethics Board (HSREB) deemed that ethical approval was unnecessary for this laboratory-based study (April 2013). Previous studies have shown that the PAJUNK® (NanoLine PAJUNK, Medizintechnologie GmbH, Geisingen, Germany) echogenic needles scored the best amongst all current commercially available echogenic needles.9,10 We therefore chose to compare the PAJUNK UniPlex NanoLine (non-echogenic, 22G, 80 mm) and PAJUNK SonoPlex Stim (echogenic, 22G, 80 mm) needles for the study. While the PAJUNK SonoPlex Stim echogenic needles were obtained from the manufacturer, the manufacturer had no influence on any aspect of the study design or data analysis. Electronic beam steering was already installed in the ultrasound machine available in our institution and was not specifically purchased for the purpose of the study. Both the echogenic and non-echogenic needles were inserted into uncooked pork meat (pork roast) at 20°, 40°, 60°, and 70° angles to a length of 60 mm. The angle of needle insertion was defined as the angle between the needle and the surface of pork meat. A high-frequency linear probe (7-13 MHz) was prepared using transparent Tegaderm™ dressing, taking care that no air was trapped between the ultrasound probe and the dressing. Using ultrasound gel as a coupling medium, the ultrasound probe was positioned on the surface of the pork meat in such a manner that the needle shaft and the tip were visible in-plane to the ultrasound beam. An experienced clinician (S.G.) performed all needle insertions and aligned the ultrasound probe to obtain the best possible needle images (snapshots) for storage to reduce the confounding factors of image acquisition. This ensured the best possible needle visibility results during the study. The angle of needle insertion was confirmed by using the built-in protractor of the ultrasound machine (SonoSite M-Turbo®, Bothel, WA, USA) as shown in Fig. 1. The ultrasound image of the non-echogenic needle was recorded to serve as a control. Beam steering was then applied at shallow, medium, and steep angles to obtain the best possible needle images. The same process was repeated for the echogenic needle.

Fig. 1
figure 1

Figure showing confirmation of the angle of needle insertion between the needle and the surface of the pork roast by using the small parts calculations “hip angle” built in the ultrasound machine (SonoSite M-Turbo, Bothel, WA, USA)

At each angle of needle insertion, eight ultrasound images of the needle were obtained:

  1. 1.

    Non-echogenic needle without beam steering (one image);

  2. 2.

    Non-echogenic needle with beam steering: shallow, medium, and steep (three images);

  3. 3.

    Echogenic needle without beam steering (one image);

  4. 4.

    Echogenic needle with beam steering: shallow, medium, and steep (three images).

Four needle insertion angles (20, 40, 60, and 70°) were studied, producing a total of 32 images. The images were transferred to a laptop computer and evaluated by eleven anesthesiologists (five consultants, five fellows, and one resident) who were blinded to the type of needle and use of beam steering. All anesthesiologists who regularly worked in the block room during the time of the study were approached for participation in the study. The sample size was one of convenience.

To enable the blinding process, multiple ultrasound images were presented to the anesthesiologists for scoring needle visibility. The experience of the evaluators in performing ultrasound-guided nerve blocks was as follows: consultants > 500; anesthesia fellows < 200, but routinely used ultrasound for central venous catheter insertions; and the anesthesia resident < ten. This allowed scoring by anesthesiologists of varied levels of experience with use of ultrasound guidance for needle visualization. The anesthesiologists were asked to score needle visibility in the recorded images (0 = no needle visible; 10 = excellent needle tip and shaft visibility). They were shown an example of an image that would score 10 and an example that would score 0. They were asked to give a higher score if they considered the needle tip to be clearly visible. Mean scores were used to classify needles as poor visibility (mean score 0-3.3), intermediate visibility (mean score 3.4-6.6), and good visibility (mean score 6.7-10).

In clinical practice, when the angle of needle insertion is approximately 20°, use of a shallow angle of beam steering is likely to be optimal to achieve best return of the ultrasound beam; whereas, for steeper angles of needle insertion, such as 60° or 70°, beam steering at a steep angle may optimize the image. Therefore, for analysis, shallow beam steering was used for a 20° angle of needle insertion, medium beam steering was used for a 40° angle of needle insertion, and steep beam steering was used for 60° and 70° angles of needle insertion. Thus, at each angle of needle insertion four images were analyzed: non-echogenic needle without beam steering, non-echogenic needle with beam steering, echogenic needle without beam steering, and echogenic needle with beam steering.

For each angle of needle insertion, needle visibility scores were analyzed using 2 × 2 repeated-measures analysis of variance to evaluate the use of needle type (echogenic or non-echogenic) and beam steering (yes or no). Pairwise comparisons were evaluated using Tukey’s test, for which corrected confidence intervals and P values are reported. All reported P values are two-sided. Statistical analyses were performed using SAS® version 9.3 (SAS Institute Inc., Cary, NC, USA).

Results

Eleven anesthesiologists with varied levels of experience in performing ultrasound-guided PNBs and central venous catheter insertions scored the images obtained. Visibility scores were similar for experienced (consultants) and less experienced (trainees) anesthesiologists (results not shown).

Figure 2 shows the individual needle trajectory visibility scores over the different angles for each factor combination, i.e., beam steering (BS) and echogenic needle (E). Solid lines represent visibility scores for experienced (consultants) and dashed lines represent visibility scores for less experienced (trainees) anesthesiologists.

Fig. 2
figure 2

The individual trajectories of outcome over different angles for each factor combination, BS (beam steering) and E (echogenic needle). Solid lines represent visibility scores for experienced (consultants), and dotted lines represent visibility scores for less experienced (trainees) anesthesiologists. 1 = Non-echogenic needle, no beam steering 2 = Non-echogenic needle, beam steering used. 3 = Echogenic needle, no beam steering. 4 = Echogenic needle, beam steering used

The visibility scores for 20°, 40°, 60°, and 70° angles are shown in Fig. 3 and are summarized in Table 1. Pairwise comparisons for the four angles of needle insertion are reported in Table 2.

Fig. 3
figure 3

Dot plots for needle visibility score at 20°, 40°, 60°, and 70° angles of needle insertion. Connecting lines indicate mean value. NE/NBS = non-echogenic needle, no beam steering; NE/BS = non-echogenic needle, beam steering used; E/NBS = echogenic needle, no beam steering; E/BS = echogenic needle, beam steering used

Table 1 Visibility scores of non-echogenic and echogenic needles with or without electronic beam steering
Table 2 Pairwise comparison of visibility scores of non-echogenic with echogenic needles with or without electronic beam steering, as scored by 11 anesthesiologists

Needle visibility at a 20° angle of needle insertion

For the 20° angle of needle insertion, the control group (non-echogenic needle with no beam steering) had an intermediate visibility score, while the mean needle visibility score for all other groups was good (Table 1). There was a statistically significant interaction between beam steering and needle type (P = 0.028), and pairwise comparisons showed that the control group was significantly different from the other three groups, while differences among the other three groups were not statistically significant (Table 2).

Needle visibility at a 40° angle of needle insertion

For the 40° angle of needle insertion, the non-echogenic needle with beam steering had a good visibility score (Table 1). The echogenic needle with or without beam steering had intermediate visibility scores. The control group scored poorly at this angle. There was a statistically significant interaction between beam steering and needle type (P < 0.001). Pairwise comparison showed that the non-echogenic needle with beam steering group was significantly different from the other three groups, while differences among the other three groups were not statistically significant (Table 2).

Needle visibility at a 60° angle of needle insertion

The echogenic needle with or without beam steering had an intermediate visibility score. The mean visibility scores of the non-echogenic needle were poor with or without beam steering. There was a statistically significant interaction between beam steering and needle type (P = 0.045). Pairwise comparison showed that the visibility scores of the echogenic needle were significantly higher than those of the non-echogenic groups (Table 2). Application of beam steering to the echogenic needle did not change visibility scores of this needle (P = 0.068). For the non-echogenic needle, application of beam steering made a statistically significant change in the visibility scores (0.013).

Needle visibility at a 70° angle of needle insertion

The mean needle visibility scores were poor for the non-echogenic needle with or without beam steering, while the echogenic needle showed intermediate visibility scores with or without beam steering. There was no interaction between beam steering and needle type (P = 0.821). Pairwise comparison showed that the visibility scores of the echogenic needle with or without beam steering were higher than those of the non-echogenic needle with or without beam steering (Table 2). Beam steering did not influence the visibility scores of either the echogenic or the non-echogenic needles.

Discussion

This study shows the effect of different angles of beam steering on the visibility of echogenic and non-echogenic needles at various angles of needle insertion. The visibility of the PAJUNK echogenic needle, with or without beam steering, was better than that of the non-echogenic needle at 60° and 70° angles of insertion. At a 40° angle of insertion, the visibility of the non-echogenic needle with beam steering was better than that of the echogenic needle with or without beam steering.

It is well known that needle visibility decreases as the angle of needle insertion increases.3,4 Various methods have been used to improve needle visualization, including use of a needle with a larger diameter, use of an echogenic needle, mechanical and optical needle guides, Doppler signals, oscillating air columns, electronic beam steering, and spatial compound imaging.11-15 When direct needle visualization is difficult, surrogate markers of needle tip location have to be used, such as tissue movement, hydrolocation, colour Doppler, and microbubble injection.16-18 Nevertheless, surrogate markers have their own disadvantages. For example, the use of microbubbles can cause deterioration of the image quality of target structures. Therefore, it is reasonable to use currently available technology to optimize needle visualization to facilitate regional anesthesia procedures.

These findings may have important implications regarding the selection of block needles and use of beam steering technology while performing ultrasound-guided needling procedures (PNBs and other clinical procedures, including needle-localized breast biopsy). At a 20° angle of needle insertion, both echogenic and non-echogenic needles had moderate to good visibility with or without electronic beam steering due to specular reflection of the incident rays. Consequently, use of a non-echogenic needle is appropriate either with or without beam steering. This concept will apply to superficial blocks such as axillary and interscalene brachial plexus blocks. At a 40° angle of needle insertion, electronic beam steering improved the visibility of the non-echogenic needle from poor to good visibility. In other words, electronic beam steering was an effective modality when the angle of needle insertion was 40° in relation to the ultrasound probe. The type of needle used did not affect needle visibility at this angle of needle insertion. Interestingly, the beam-steered non-echogenic needle was more visible than the echogenic needle with or without beam steering. The lower visibility scores for the echogenic needle at this angle of needle insertion may be due to scattering of the ultrasound beam by the Cornerstone reflector, whereas the smooth surface of the non-echogenic needle may lead to more uniform reflection of ultrasound waves back to the transducer when the angle of incidence of ultrasound waves is perpendicular to the needle. It is also possible that the higher visibility of the beam-steered non-echogenic needle compared with the echogenic needle (at this particular angle) may be a “random high”, and this may require confirmation through replication of the experiment. The electronic beam steering did not improve the visibility of the echogenic needle. This is an important finding, as the non-echogenic needle with the aid of beam steering may be adequate with little role from an echogenic needle if the estimated needle trajectory is expected to be about 40° as assessed during the pre-procedural scan. Nevertheless, if electronic beam steering is not available, the echogenic needle can provide intermediate needle visibility on its own.

At 60° and 70° angles of needle insertion, the visibility of the non-echogenic needle was poor and remained poor even with electronic beam steering. At a 60° angle of insertion, the echogenic needle showed intermediate visibility without beam steering, which improved to good visibility with application of beam steering. At a 70° angle of needle insertion, the echogenic needle showed intermediate needle visibility which showed little improvement with the application of electronic beam steering. Therefore, we can recommend that echogenic needles should be used when the angle of needle insertion is expected to be 60° or 70° (e.g., while performing ultrasound-guided infraclavicular brachial plexus block), particularly in obese patients.19 Electronic beam steering might have little role in such circumstances. The interaction between the angle of needle insertion, beam steering, and needle type is yet to be assessed with the use of a low-frequency curved array probe often used for sciatic blocks.

Cheung and Rohling have described the algorithm and principles of physics used to enhance needle visibility by electronic beam steering.20 The angle of needle insertion in clinical practice is dependent on the needle entry point and the depth of the target structure. The current commercially available ultrasound machines steer the ultrasound beams at three fixed angles (shallow, medium, and steep). This is an important limitation, as the steered ultrasound beam may not be exactly perpendicular to the needle shaft surface, which is required for best needle visibility. In the future, the technology may allow us to steer the ultrasound beam directly perpendicular to the needle, enabling better needle visualization. Until such technology is available that allows the ultrasound beam to be steered exactly perpendicular to the angle of needle insertion, the clinical utility of beam steering may be limited except at particular angles of needle insertion. Nevertheless, if the electronic beam steering is pre-installed in the ultrasound machine, it may be prudent to use it if the needle visibility is poor at certain angles.

Our study has several limitations. The study was not designed to replicate a real patient scenario, as needle visibility is a dynamic process. We used pork meat for the study because previous studies recommend that meat-based models are superior to synthetic models (e.g., gel phantoms or water baths) when comparing needle visibility.21,22 The pork meat also had adequate thickness to allow for needle insertion at a steep angle to the desired length. Secondly, we did not look at needle tip and shaft visibility separately; instead, we requested the scorers to score higher if the tip was seen more clearly. This methodology has been used previously by Guo et al. and allows comparison of the overall visibility of the needles.9 Needle tip visibility is a subject of interest with no established single best method of identification. The overall improvement in needle visibility with the use of a new technology may translate to better needle tip visibility with control of other factors of needle visibility. Thirdly, we did not compare different types of echogenic needles and the effect of beam steering on them. Echogenic needles from different manufactures may show differences with the application of beam steering.10 Finally, we did not evaluate whether or not electronic beam steering caused deterioration of the image quality of target structures.

In conclusion, the PAJUNK echogenic needle was more visible compared with the non-echogenic needle at 60° and 70° angles of insertion with or without beam steering; whereas, at a 40° angle of needle insertion, the non-echogenic needle with beam steering was more visible compared with the echogenic needle.