Abstract
Purpose
The cardiac gating signal (CGS) in positron emission tomography (PET) studies is usually obtained from an electrocardiography (ECG) monitor. In this work, we propose a method to obtain the CGS in small-animal PET using the acquired list-mode data without using any hardware or end-user input.
Procedures
The CGS was obtained from the number of coincidences over time acquired in the lines-of-response connected with the cardiac region. This region is identified in the image as its value changes with frequencies in the range of 3 to 12 Hz. The procedure was tested in a study with 29 Wistar rats and 6 mice injected with 2-deoxy-2-[18F]fluoro-d-glucose, which underwent a 45-min single-bed list-mode PET scan of the heart syncronized with an ECG. The estimated signals and the reconstructed images using eight-gated frames were compared with the ones obtained using the ECG signal from the monitor.
Results
The differences of the PET-based CGS with respect to the ECG relative to the duration of the heartbeats were 5.6 % in rats and 11.0 % in mice. The reconstructed gated images obtained from the proposed method do not differ qualitatively with respect to the ones obtained with the ECG. The quantitative analysis of both set of images were performed measuring the volume of the left ventricle (LV) of the rats in the end-of-systole and end-of-diastole phase. The differences found in these parameters between both methods were below 12.1 % in the ESV and 9.3 % in the EDV with a 95 % confidence interval, which are comparable to the accuracy (7 %) of the method used for segmenting the LV.
Conclusion
The proposed method is able to provide a valid and accurate CGS in small-animal PET list-mode data.
Similar content being viewed by others
References
Di Carli MF, Lipton MJ (2007) Cardiac PET and PET/CT Imaging. Springer Science and Business Media
Wu JC, Inubushi M, Sundaresan G et al (2002) Positron emission tomography imaging of cardiac reporter gene expression in living rats. Circulation 106:180–183
Yang Y, Rending S, Siegel S et al (2005) Cardiac PET imaging in mice with simultaneous cardiac and respiratory gating. Phys Med Biol 50:2979–2989
Croteau E, Bénard F, Cadorette J et al (2003) Quantitative gated PET for the assessment of the left ventricular function in small animals. J Nucl Med 44:1655–1661
Cussó L, Hernandez-Porras I, Montesinos P et al (2013) Phenotyping K-RasV141 knock-in mouse by cardiac-MRI: a pilot study. World Mol Imag Cong (WMIC): Program Book. pp. 79
Cussó L, Santa Marta C, Benito M et al (2009) Multimodal assessment of myocardial infarction in rats: comparison of late gadolinium enhanced MRI and PET. Abstract Eur Soc Mol Imag (ESMI). pp. 104
Böning G, Todica A, Vai A et al (2013) Erroneous cardiac ECG-gated PET list-mode trigger events can be retrospectively identified and replaced by an offline reprocessing approach: first results in rodents. Phys Med Biol 58:7937–7959
Dineley J (2010) Fully automated gating: a new era for PET?, http://medicalphysicsweb.org/cws/article/research/44273
He J, O’Keefe GJ, Gong SJ et al (2008) A novel method for respiratory motion gated with geometric sensitivity of the scanner in 3D PET. IEEE Trans Nucl Sci 55:2557–2565
Büther F, Dawood M, Wübbeling F et al (2009) List mode-driven cardiac and respiratory gating in PET. J Nucl Med 50:674–681
Kesner AL, Bundschuh RA, Detorieet NC et al (2009) Respiratory gated PET derived in a fully automated manner from Raw PET data. IEEE Trans Nucl Sci 56:677–687
Kesner AL, Kuntner C (2010) A New fast and fully automated software based algorithm for extracting respiratory signal from raw PET data and its comparison to other methods. Med Phys 37:5550–5559
Wang Y, Seidel J, Tsui BMW et al (2006) Performance evaluation of the GE healthcare eXplore VISTA dual-ring small-animal PET scanner. J Nucl Med 47:1891–1900
Herraiz JL, España S, Vaquero JJ et al (2006) FIRST: fast iterative reconstruction software for (PET) tomography. Phys Med Biol 51:4547–4565
Kadrmas DJ (2004) LOR-OSEM: statistical PET reconstruction from raw line-of-response histograms. Phys Med Biol 49:4731–4744
Siddon RL (1985) Fast calculation of the exact radiological path for a three-dimensional CT array. Med Phys 12:252–255
Mancosu P, Sghedoni R, Bettinardi V et al (2009) 4D-PET data sorting into different number of phases: a NEMA IQ phantom study. J Appl Clin Med Phys 10:220–231
Heiberg E, Sjogren J, Ugander M et al (2010) Design and validation of segment—freely available software for cardiovascular image analysis. BMC Med Imaging 10:1 (http://segment.heiberg.se)
Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476):307–310
Herraiz JL, España S, Cal-González J et al (2011) Fully 3D GPU PET reconstruction. Nucl Inst Methods Phys Res A 648:S169–S171
Acknowledgments
This work was supported in part by Consejería de Educación, Juventud y Deporte de la Comunidad de Madrid (Spain) through the Madrid-MIT M+Visión Consortium, Comunidad de Madrid (S2013/MIT-3024 TOPUS-CM), UCM (Grupos UCM, 910059), CPAN (Consolider-Ingenio 2010, CSPD-2007-00042), RIC-RETIC network, Spanish MINECO (RD12/0042/0057), Ministerio de Ciencia e Innovación, Spanish Government (ENTEPRASE grant, PSE-300000-2009-5 and TEC2007-64731/TCM), and European Regional funds.
Conflict of Interest
All authors declare that they have no competing interest.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Herraiz, J.L., Herranz, E., Cal-González, J. et al. Automatic Cardiac Self-Gating of Small-Animal PET Data. Mol Imaging Biol 18, 109–116 (2016). https://doi.org/10.1007/s11307-015-0868-y
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11307-015-0868-y