Abstract

Purpose

Accessory pathway (AP) mapping is currently based on point-by-point mapping and identifying if a local electrogram’s origin is atrial, pathway, or ventricular, which is time-consuming and prone to insufficient mapping. We sought to determine the feasibility of automated and high-density mapping to define AP location using open-window mapping (OWM), which does not rely on defining the electrogram’s origin but simply detects the sharpest local signal at each point.

Methods

We enrolled 23 consecutive patients undergoing catheter ablation for atrioventricular reentrant tachycardia. High-density mapping was performed using OWM and ablation was performed. The successful site of ablation was determined by the loss of pathway function.

Results

OWM was 100% effective at identifying the successful site of ablation (average mapping time 7.3 ± 4.3 min.) Permanent AP elimination was achieved using a mean radiofrequency energy time of 18.5 ± 24.5 s/patient. Transiently successful ablations were 4.0 ± 1.8 mm from permanently successful sites and had lower contact force (5.1 ± 2.5 g vs. 11.7 ± 9.0 g; P = 0.041). Unsuccessful sites had similar contact force to permanently successful sites (12.2 ± 9.2 g vs. 11.7 ± 9.0 g; P = 0.856) but were 6.4 ± 2.0 mm away from successful sites.

Conclusion

A novel technique of high-density, automated, and open-window mapping (OWM) effectively localizes APs without the need to differentiate the signal’s site of origin. These findings suggest that OWM can be used to rapidly and successfully map and ablate APs. Both distances from the pathway and contact force were shown to be important for pathway ablation.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

Availability of data and material

All raw data are available upon request.

References

  1. Blomstrom-Lundqvist C, Scheinman MM, Aliot EM, Alpert JS, Calkins H, Camm AJ, et al. ACC/AHA/ESC guidelines for the management of patients with supraventricular arrhythmias--executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for the Management of Patients With Supraventricular Arrhythmias). Circulation. 2003;108(15):1871–909. https://doi.org/10.1161/01.CIR.0000091380.04100.84.

    Article  PubMed  Google Scholar 

  2. Scheinman M, Calkins H, Gillette P, Klein R, Lerman BB, Morady F, et al. NASPE policy statement on catheter ablation: personnel, policy, procedures, and therapeutic recommendations. Pacing Clin Electrophysiol. 2003;26(3):789–99. https://doi.org/10.1046/j.1460-9592.2003.00139.x.

    Article  PubMed  Google Scholar 

  3. Morady F. Catheter ablation of supraventricular arrhythmias: state of the art. J Cardiovasc Electrophysiol. 2004;15(1):124–39. https://doi.org/10.1046/j.1540-8167.2004.03516.x.

    Article  PubMed  Google Scholar 

  4. Shen MJ, Knight BP. Value of high-density mapping in the electrophysiology laboratory. Curr Opin Cardiol. 2019;34(1):6–15. https://doi.org/10.1097/HCO.0000000000000576.

    Article  PubMed  Google Scholar 

  5. Schaeffer B, Akbulak RO, Jularic M, Moser J, Eickholt C, Schwarzl JM, et al. High-density mapping and ablation of primary nonfocal left atrial tachycardia: characterizing a distinct arrhythmogenic substrate. JACC Clin Electrophysiol. 2019;5(4):417–26. https://doi.org/10.1016/j.jacep.2019.02.002.

    Article  PubMed  Google Scholar 

  6. Maury P, Rollin A, Cardin C, Mondoly P, Guerrero F. High density mapping of pre-excitation 3D-illustration of anatomical features. J Atr Fibrillation. 2018;10(6):1824. https://doi.org/10.4022/jafib.1824.

    Article  PubMed  PubMed Central  Google Scholar 

  7. Cappato R, Schluter M, Mont L, Kuck KH. Anatomic, electrical, and mechanical factors affecting bipolar endocardial electrograms. Impact on catheter ablation of manifest left free-wall accessory pathways. Circulation. 1994;90(2):884–94.

    Article  CAS  Google Scholar 

  8. Xie B, Heald SC, Camm AJ, Rowland E, Ward DE. Successful radiofrequency ablation of accessory pathways with the first energy delivery: the anatomic and electrical characteristics. Eur Heart J. 1996;17(7):1072–9. https://doi.org/10.1093/oxfordjournals.eurheartj.a015004.

    Article  CAS  PubMed  Google Scholar 

  9. Otomo K, Gonzalez MD, Beckman KJ, Nakagawa H, Becker AE, Shah N, et al. Reversing the direction of paced ventricular and atrial wavefronts reveals an oblique course in accessory AV pathways and improves localization for catheter ablation. Circulation. 2001;104(5):550–6. https://doi.org/10.1161/hc3001.093499.

    Article  CAS  PubMed  Google Scholar 

  10. Takenaka S, Yeh SJ, Wen MS, Yeh KH, Wang CC, Lin FC, et al. Characteristics and radiofrequency ablation in posteroseptal and left free-wall subepicardial accessory pathways. J Electrocardiol. 2005;38(1):69–76. https://doi.org/10.1016/j.jelectrocard.2004.09.003.

    Article  PubMed  Google Scholar 

  11. Weiss C, Cappato R, Willems S, Meinertz T, Kuck KH. Prospective evaluation of the coronary sinus anatomy in patients undergoing electrophysiologic study. Clin Cardiol. 1999;22(8):537–43. https://doi.org/10.1002/clc.4960220810.

    Article  CAS  PubMed  Google Scholar 

  12. Gatzoulis KA, Apostolopoulos T, Costeas X, Zervopoulos G, Papafanis F, Sotiropoulos H, et al. Radiofrequency catheter ablation of posteroseptal accessory pathways--results of a step-by-step ablation approach. J Interv Card Electrophysiol. 2001;5(2):193–201.

    Article  CAS  Google Scholar 

  13. Gaeta S, Bahnson TD, Henriquez C. Mechanism and magnitude of bipolar electrogram directional sensitivity: characterizing underlying determinants of bipolar amplitude. Heart Rhythm. 2019;17:777–85. https://doi.org/10.1016/j.hrthm.2019.12.010.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgments

We would like to thank Dr. R. Hardwin Mead and Dr. Girish Narayan for contributing patients to this study. We also thank John Crowell and Brady Stevens for their assistance with data collection.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Christopher E. Woods.

Ethics declarations

Conflicts of interest

Mr. Suchomel and Mr. Moskovitz are employees of Abbott Laboratories. Dr. Engel serves as a consultant and advisor to Medtronic. Dr. Woods serves as a non-paid consultant for Abbott Laboratories and Edwards Lifesciences. Dr. Winkle has an equity interest in Farapulse and is an investigator for Abbott. All other authors have no conflicts of interest to disclose.

Ethics approval

Not applicable.

Consent to participate

As permitted by our Institutional Review Boards, the requirement to obtain consent was waived.

Consent for publication

All authors have given explicit consent to submit this manuscript for publication.

Code availability

Not applicable.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Electronic supplementary material

ESM 1

(MP4 667 kb)

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Schricker, A.A., Winkle, R., Moskovitz, R. et al. Open-window mapping of accessory pathways utilizing high-density mapping. J Interv Card Electrophysiol 61, 525–533 (2021). https://doi.org/10.1007/s10840-020-00850-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10840-020-00850-7