Abstract
European national metrology institutes use calibration systems of various types for calibrating thermometers in air. These were compared to each other for the first time in a project organized by the European Association of National Metrology Institutes (EURAMET). This EURAMET P1061 comparison project had two main objectives: (1) to study the equivalence of calibrations performed by different laboratories and (2) to investigate correlations between calibration methods and achievable uncertainties. The comparison was realized using a pair of 100 \(\Omega \) platinum resistance thermometer probes connected to a digital thermometer bridge as the transfer standard. The probes had different dimensions and surface properties. The measurements covered the temperature range between \(-40\,^{\circ }\mathrm{{C}}\) and \(+150\,^{\circ }\mathrm{{C}}\), but each laboratory chose a subrange most relevant to its scope and performed measurements at five nominal temperature points covering the subrange. To enable comparison between the laboratories, comparison reference functions were determined using weighted least-squares fitting. Various effects related to variations in heat transfer conditions were demonstrated but clear correlations to specific characteristics of calibration system were not identified. Calibrations in air and liquid agreed typically within \(\pm 0.05\,^{\circ }\mathrm{{C}}\) at \(+10\,^{\circ }\mathrm{{C}}\) and \(+80\,^{\circ }\mathrm{{C}}\). Expanded uncertainties determined by the participants ranged from \(0.02\,^{\circ }\mathrm{{C}}\) to \(0.4\,^{\circ }\mathrm{{C}}\) and they were shown to be realistic in most cases.
Similar content being viewed by others
References
S. Bell, R. Benyon, N. Böse, M. Heinonen, Int. J. Thermophys. 29, 1537 (2008)
V. Fernicola, S. Bell, R. Benyon, N. Böse, E. Georgin, M. Heinonen, D. Hudoklin, M. Sargent, A European Roadmap for Humidity and Moisture (European Association of National Metrology Institutes - EURAMET, 2012), http://www.euramet.org/fileadmin/docs/Publications/roadmaps/TC_T_Humidity_Roadmap_2012_text.pdf. Accessed 13 June 2013
S. Friederici, E. Tegeler, in Proceedings of TEMPMEKO 2004, 9th International Symposium on Temperature and Thermal Measurements in Industry and Science, ed. by D. Zvizdić, L.G. Bermanec, T. Veliki, T. Stašić (FSB/LPM, Zagreb, 2004), pp. 795–800
E. Renaot, Y. Hermier, M.H. Valin, G. Bonnier, F. Adunka, A. Van Der Linden, M. Dobre, M. Šmíd, A. Bronnum, T. Weckstrom, E. Tegeler, U. Noatsch, M. Anagnostou, E. Turzo-Andras, S. Nemeth, M. White, P. Marcarino, P. Steur, A. Pauzha, V. Augevicius, M. de Groot, J. Nielsen, F. Helgesen, W. Roszkowski, E. Grudniewicz, R. Kuna, E. Filipe, I. Lobo, S. Gaita, V. Chimenti, J. Ivarsson, S. Duris, J. Ranostaj, J. Bojkovski, A. Steiner, S. Ugur, J. Gray, S. Bruce, R. Rusby, D. Head, Int. J. Thermophys. 29, 991 (2008)
M.J. Ballico, M.K. Nguyen, U. Norranim, Metrologia 44(Tech. Suppl.), 03005 (2007)
IEC 60751 (1995–07), Industrial Platinum Resistance Thermometer Sensors (International Electrotechnical Commission (IEC), Geneva, 1995)
Acknowledgments
The authors want to thank Hannu Räsänen for his extensive efforts in carrying out most of the measurements at MIKES.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Heinonen, M., Anagnostou, M., Bartolo, J. et al. Comparison of Air Temperature Calibrations. Int J Thermophys 35, 1251–1272 (2014). https://doi.org/10.1007/s10765-014-1654-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10765-014-1654-9