A magnetic beads-based portable flow cytometry immunosensor for in-situ detection of marine biotoxin

Biomedical Microdevices Aims and scope Submit manuscript

Abstract

Okadaic acid (OA), a representative diarrhetic shellfish poisoning toxin, mainly produced by toxigenic dinoflagellates, has significant hazard to public health. Traditional methods for detection of OA can not give the consideration to the need of rapid, high sensitive, quantitative and in-situ detection at the same time. Herein, a new effective detection method of OA was developed based on fluorescence immunosensor and flow cytometry (FCM). In this assay, Streptavidin-coated magnetic beads (MBs) were used as the supporter to immobilize the biotinylated OA. Modified MBs competed with the free OA in the sample solution to bind with the anti-OA monoclonal antibody (OA-MAb). The R-phycoerythrin (R-PE) dye labeled IgG was served as a secondary antibody to perform fluorescence detection. A portable flow cytometry was applied for the in-situ fluorescence quantification. The results showed that the OA concentration was inversely proportional to the R-PE fluorescence intensity. The detection method took within 50 min with a limit of detection (LOD) was 0.05 μg/L and range from 0.2 to 20 μg/L for OA detection. Moreover, the matrix effect and the recovery rate were assessed during real sample measurement, showing a high recovery. Performance features such as high sensitivity, low LOD, speediness and simplicity of the analysis protocol, shows this biosensing-systems as a promising tool for routine use.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  • R. Andrew, Clinical measurement of steroid metabolism. Best Pract. Res. Clin. Endocrinol. Metab. 15, 1–16 (2001)

    Article  Google Scholar 

  • L.B. Bangs, New developments in particle-based immunoassays: Introduction. Pure Appl. Chem. 68, 1873–1879 (1996)

    Article  Google Scholar 

  • U. Bilitewski, Peer reviewed: Can affinity sensors be used to detect food contaminants? Anal. Chem. 72, 692 A–701 A (2000)

    Article  Google Scholar 

  • M. Campàs, J.-L. Marty, Enzyme sensor for the electrochemical detection of the marine toxin okadaic acid. Anal. Chim. Acta 605, 87–93 (2007)

    Article  Google Scholar 

  • M. Campàs, D. Szydlowska, M. Trojanowicz, J.-L. Marty, Towards the protein phosphatase-based biosensor for microcystin detection. Biosens. Bioelectron. 20, 1520–1530 (2005)

    Article  Google Scholar 

  • M. Campas, B. Prieto-Simón, J.-L. Marty, Biosensors to detect marine toxins: Assessing seafood safety. Talanta 72, 884–895 (2007)

    Article  Google Scholar 

  • S. Centi, S. Laschi, M. Mascini, Improvement of analytical performances of a disposable electrochemical immunosensor by using magnetic beads. Talanta 73, 394–399 (2007)

    Article  Google Scholar 

  • EFSA, Opinion of the scientific panel on contaminants in the food chain on a request from the European Commission on marine biotoxins in shellfish—Okadaic acid and analogues. EFSA J. 589, 1–62 (2008)

    Google Scholar 

  • J.V. Forment, S.P. Jackson, A flow cytometry-based method to simplify the analysis and quantification of protein association to chromatin in mammalian cells. Nat. Protoc. 10, 1297–1307 (2015)

    Article  Google Scholar 

  • F. Gessler, K. Hampe, M. Schmidt, H. Böhnel, Immunomagnetic beads assay for the detection of botulinum neurotoxin types C and D. Diagn. Microbiol. Infect. Dis. 56, 225–232 (2006)

    Article  Google Scholar 

  • H. Goto, T. Igarashi, M. Yamamoto, M. Yasuda, R. Sekiguchi, M. Watai, et al., Quantitative determination of marine toxins associated with diarrhetic shellfish poisoning by liquid chromatography coupled with mass spectrometry. J. Chromatogr. A 907, 181–189 (2001)

    Article  Google Scholar 

  • J.L. Guesdon, T. Ternynck, S. Avrameas, J.L. Guesdon, J. Ternyck, S. Avrameas, The use of avidin-biotin interaction in immunoenzymatic techniques. J. Histochem. Cytochem. 27, 1131–1139 (1979)

    Article  Google Scholar 

  • B.I. Haukanes, C. Kvam, Application of magnetic beads in bioassays. Bio/Technology 11, 60–63 (1993)

    Google Scholar 

  • A. Hayat, L. Barthelmebs, J.-L. Marty, Enzyme-linked immunosensor based on super paramagnetic nanobeads for easy and rapid detection of okadaic acid. Anal. Chim. Acta 690, 248–252 (2011)

    Article  Google Scholar 

  • A. Hayat, L. Barthelmebs, A. Sassolas, J.-L. Marty, Development of a novel label-free amperometric immunosensor for the detection of okadaic acid. Anal. Chim. Acta 724, 92–97 (2012)

    Article  Google Scholar 

  • T. Jung, U. Schauer, C. Heusser, C. Neumann, C. Rieger, Detection of intracellular cytokines by flow cytometry. J. Immunol. Methods 159, 197–207 (1993)

    Article  Google Scholar 

  • M.P. Kreuzer, C.K. O'Sulliva, G.G. Guilbault, Development of an ultrasensitive immunoassay for rapid measurement of okadaic acid and its isomers. Anal. Chem. 71, 4198–4202 (1999)

    Article  Google Scholar 

  • N.M. Llamas, L. Stewart, T. Fodey, H.C. Higgins, M.L.R. Velasco, L.M. Botana, et al., Development of a novel immunobiosensor method for the rapid detection of okadaic acid contamination in shellfish extracts. Anal. Bioanal. Chem. 389, 581–587 (2007)

    Article  Google Scholar 

  • A.P. Louppis, A.V. Badeka, P. Katikou, et al., Determination of okadaic acid, dinophysistoxin-1 and related esters in Greek mussels using HPLC with fluorometric detection, LC-MS/MS and mouse bioassay. Toxicon 55(4), 724–733 (2010)

    Article  Google Scholar 

  • B. Prieto-Simón, H. Miyachi, I. Karube, H. Saiki, High-sensitive flow-based kinetic exclusion assay for okadaic acid assessment in shellfish samples. Biosens. Bioelectron. 25, 1395–1401 (2010)

    Article  Google Scholar 

  • C. Riccardi, I. Nicoletti, Analysis of apoptosis by propidium iodide staining and flow cytometry. Nat. Protoc. 1, 1458–1461 (2006)

    Article  Google Scholar 

  • S. Siena, M. Bregni, B. Brando, N. Belli, F. Ravagnani, L. Gandola, et al., Flow cytometry for clinical estimation of circulating hematopoietic progenitors for autologous transplantation in cancer patients. Blood 77, 400–409 (1991)

    Google Scholar 

  • K. Su, X. Qiu, J. Fang, Q. Zou, P. Wang, An improved efficient biochemical detection method to marine toxins with a smartphone-based portable system − bionic e-eye. Sensors Actuators B Chem. 238, 1165–1172 (2016)

    Article  Google Scholar 

  • A.X. Tang, M. Pravda, G.G. Guilbault, S. Piletsky, A.P. Turner, Immunosensor for okadaic acid using quartz crystal microbalance. Anal. Chim. Acta 471, 33–40 (2002)

    Article  Google Scholar 

  • T. Yasumoto, M. Murata, Marine toxins. Chem. Rev. 93, 1897–1909 (1993)

    Article  Google Scholar 

  • L. Zou, Q. Wang, M. Tong, H. Li, J. Wang, N. Hu, et al., Detection of diarrhetic shellfish poisoning toxins using high-sensitivity human cancer cell-based impedance biosensor. Sensors Actuators B Chem. 222(, 205–212 (2016)

    Article  Google Scholar 

Download references

Acknowledgements

This work was supported by a key project of the Natural Science Foundation of China (No. 31627801), International Cooperation Project of Natural Science Foundation of China (No. 61320106002,31661143030,), National 973 Project of China (No.2015CB352101), and Natural Science Foundation of China (No. 31571004).

Author information

Authors and Affiliations

Authors

Corresponding authors

Correspondence to Ning Hu or Ping Wang.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Pan, Y., Wei, X., Liang, T. et al. A magnetic beads-based portable flow cytometry immunosensor for in-situ detection of marine biotoxin. Biomed Microdevices 20, 60 (2018). https://doi.org/10.1007/s10544-018-0304-6

Download citation

  • Published:

  • DOI: https://doi.org/10.1007/s10544-018-0304-6

Keywords

Navigation