Abstract
White blood cell (WBC) and neutrophil counts are important laboratory tests used by clinicians to assess a variety of conditions. However, current methods to measure WBC and neutrophil counts are difficult to perform at the point of care, being either cost or labor prohibitive. To meet this need, we developed the LeukoScope: a portable, imaging-based system to measure WBC and neutrophil counts from a drop of blood. Here, we present the performance of the LeukoScope in 136 pediatric and 164 neonatal subjects at a central hospital in Malawi. For pediatric patients, 95.4, 66.7, and 80.0% of samples with normal, low, and high WBC counts, respectively, were correctly identified, and 88.6, 100.0, and 89.3% of samples with normal, low, and high neutrophil counts, respectively, were correctly identified. Accuracy was lower overall for neonatal samples; 92.1, 64.3, and 26.7% of samples with normal, low, and high WBC counts, respectively, were correctly identified, and 73.2 and 78.6% of samples with normal and high neutrophil counts, respectively, were correctly identified. Results of this study show that the LeukoScope can help meet need for point-of-care measurement of WBC counts in pediatric patients and highlight the challenges of point-of-care assessment of WBC counts in neonatal patients.
Similar content being viewed by others
Change history
22 July 2021
A series of 0's was removed from the Table 4 reference.
References
Bland, J. M., and D. G. Altman. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 327:307–310, 1986.
Bond, M. M., and R. R. Richards-Kortum. Drop-to-drop variation in the cellular components of fingerprick blood implications for point-of-care diagnostic development. Am. J. Clin. Pathol. 144:885–894, 2015.
Bui, H. N., J. P. A. M. Bogers, D. Cohen, T. Njo, and M. H. Herruer. Evaluation of the performance of a point-of-care method for total and differential white blood cell count in clozapine users. Int. J. Lab. Hematol. 38:703–709, 2016.
Houwen, B. The differential cell count. Lab. Hematol. 7:89–100, 2001.
Karawajczyk, M., S. Haile, M. Grabski, and A. Larsson. The HemoCue WBC DIFF system could be used for leucocyte and neutrophil counts but not for full differential counts. Acta Paediatr. 106:974–978, 2017.
Laboratories, M. C. CBC with Differential, Blood. (2019). Available at: https://www.mayocliniclabs.com/test-catalog/Clinical+and+Interpretive/9109. (Accessed: 5th February 2019)
Laudicina, R. J., and Y. Simonian. The Leukocyte. In: Clinical Laboratory Hematology, edited by S. B. McKenzie, and J. L. Williams. London: Pearson, 2010.
Majors, C. E., M. E. Pawlowski, T. Tkaczyk, and R. R. Richards-Kortum. Low-cost disposable cartridge for performing a white blood cell count and partial differential at the point-of-care. Health Innovations and Point-of-Care Technologies Conference 10–13, 2014.
Majors, C. E., et al. Clinical training and validation of the LeukoScope: a low-cost, point-of-care device to perform white blood cell and neutrophil counts. RSC Adv. 9:27324–27333, 2019.
Russcher, H., N. VanDeursen, T. Ermens, and R. DeJonge. Evaluation of the HemoCue WBC DIFF system for Point-of-Care counting of total and differential white cells in pediatric samples. Ned. Tijdschr. voor Klin. Chemie en Lab. 38:140–141, 2013.
Spaeth, B. A., M. D. S. Shephard, B. McCormack, and G. Sinclair. Evaluation of HemoCue white blood cell differential counter at a remote health centre in Australia’s Northern Territory. Pathology 47:91–95, 2015.
Zheng, S., J. Lin, H. Kasdan, and Y. Tai. Fluorescent labeling, sensing, and differentiation of leukocytes from undiluted whole blood samples. Sensors Actuators B Chem. 132:558–567, 2008.
Acknowledgments
This work was supported by Grant #2013138 from the Doris Duke Charitable Foundation. This work was funded by Grant #OPP1028843 from the Bill and Melinda Gates Foundation through the Grand Challenges in Global Health Initiative. This material is based upon work supported by the National Science Foundation Graduate Research Fellowship under Grant No. 1450681. The authors would like to acknowledge the members of the Rice360 Institute for Global Health team in Blantyre, Malawi for their support in administration and logistics in running this clinical trial, notably the assistance of Prince Mtenthaonga for his assistance with patient enrollment and sample collection, and Cynthia Wong for her assistance with optical system performance analysis.
Author information
Authors and Affiliations
Corresponding author
Additional information
Associate Editor Stefan M. Duma oversaw the review of this article.
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Majors, C.E., Pawlowski, M.E., Tkaczyk, T.S. et al. Evaluation of the LeukoScope for Point-of-Care Measurement of White Blood Cell and Neutrophil Counts in Malawi. Ann Biomed Eng 49, 2566–2578 (2021). https://doi.org/10.1007/s10439-021-02827-7
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10439-021-02827-7