1 Introduction

In general, a cell is the minimum unit with self-determined responses based on genetic and epigenetic information, e.g., their interaction histories, and based on fluctuations in its environmental conditions. If we are to understand such things as the variations between cells with identical genetic information, the inheritance of non-genetic information from the previous cell generation, the adaptation of cells to environmental change, and the community effect of cells, we need to analyze, using identified single cells, their epigenetic information, starting from twin complementary viewpoints of cell regulation: as an ‘algebraic’ system with emphasis on the temporal aspects, such as adaptation from one generation to the next, and as a ‘geometric’ system with emphasis on spatial aspects, such as the spatial pattern-dependent community effect.

We are conducting a series of studies to analyze epigenetic information in sequential cell generations and in the spatial structures of cell networks to expand our understanding of how the fates of living systems are determined. Our objective is the development of a system for disposable contamination-free recultivation that can be operated simply and safely in both research and clinical settings. To facilitate our efforts, we have developed an on-chip microfabrication method that is based on a strategy we call “on-chip cellomics.” As shown in Fig. 1, it comprises three steps. First, cells from tissue are purified one by one in a non-destructive manner (Yasuda 2000). The cells are then cultivated and observed under fully controlled conditions (i.e., the cell population, the network pattern, and the nutrient conditions are controlled) using an on-chip single-cell cultivation chip (Inoue et al. 2001a, b, 2004; Wakamoto et al. 2001, 2003; Umehara et al. 2003; Takahashi et al. 2003; Hattori et al. 2003; Matsumura et al. 2003a, b) or an on-chip agarose microchamber system (Moriguchi et al. 2002, 2004; Hattori et al. 2004; Sugio et al. 2004; Suzuki et al. 2004a,b; Kojima et al. 2003, 2005). Finally, single-cell-based genome/proteome analysis is performed using photothermal denaturation and single-molecule-level analysis (Yasuda et al. 2000).

Fig. 1
figure 1

On-chip cellomics: on-chip single cell-based analysis of cellular networks temporally and spatially

The purification of particular target cells from a crude sample mixture of cells is the first and most important step for on-chip experiments. Efficient and rapid purification of cells has been accomplished with techniques such as fluorescence-activated cell sorting (FACS) (Ogawa et al. 1994; Hulett et al. 1973; Bonner et al. 1972; Herzenberg and Sweet 1976; Assenmacher et al. 1995; Johnson et al. 2007), magnetic-activated cell separation (MACS) (Imamura et al. 1995; Owen and Sykes 1984; Zborowski and Chalmers 2005, 2011), automated single-cell sorting using dual-beam optical trapping (Grover et al. 2001), differential adhesion cell sorting (Hsu et al. 2008), and disposable microfabricated fluorescence-activated cell sorting (μFACS) (Liu et al. 2011; Yoon et al. 2011; Azuma et al. 2007). In addition, recent advancements in optofluidic flow cytometry, in which optics and microfluidics are used together to create novel functionalities on a small chip, hold great promise for lab-on-a-chip flow cytometry (Yasuda et al. 2006; Hatayama et al. 1994). The development of a low-cost, compact, handheld flow cytometer and μFACS system would have a significant impact on point-of-care diagnostics and thereby improve health care.

In measurements using a conventional flow cytometer, the target cells for FACS are usually labeled with fluorescent dyes using specific antibodies or other molecules, and the optical signals (angular light scattering or emitted fluorescence) are detected and analyzed to identify individual cells or biological samples. These tools greatly facilitate the study of both the physical properties (e.g., size, shape) and biochemical properties (e.g., cell cycle distribution, DNA contents) of biological samples such as cells. Information about the cells of interest is obtained optically in a non-destructive and quantitative manner (Yamashita et al. 2007; Takahashi et al. 2006).

However, FACS can damage cells during droplet generation, and detection based on non-direct scattering in principle carries the risk of poor cell recognition. If the cells are to be used for cultivation after purification, cell damage due to sorting must be minimized. MACS, e.g., enables cells to be easily separated by incubating them with magnetic particles coated with antibodies against a particular surface antigen. This incubation promotes the attachment of cells expressing a specific antigen for the magnetic particles. The cell solution is then transferred into a column placed in a strong magnetic field. The cells attached to the particles (expressing the antigen) remain in the column, while the other cells (not expressing the antigen) flow out. While this method is simple and economical, the attached antibodies are difficult to remove from the cell surface and can change the properties of the target cells. Other conventional techniques also have disadvantages with respect to cost, efficiency, response speed, separation resolution, and/or adaptability.

In contrast, μFACS offers several advantages, including reduced sample volume, shorter handling time, smaller space requirement, and less operational cost. In particular, the smaller space required and potential lower cost of a μFACS system would make it more portable and affordable, which are particularly important to individual laboratories and point-of-care clinics. There are three core components of a μFACS system that need to be miniaturized while maintaining performance: (1) the fluidic system for introduction and placement of biological samples, (2) the optical system for illumination of samples and collection of emitted light (both scattered and fluorescent), and (3) the sorting system for collection of samples of interest, which requires both real-time control and a rapid-response actuation system. The key challenge to developing a high-performance μFACS system is to figure out how to reduce the volume and cost of these three components while maintaining adequate performance (e.g., detection sensitivity and sorting throughput).

However, even if such a μFACS system were developed, it would be hard to apply it to non-invasive collection of target cells due to its use of fluorescent antibodies for labeling and identifying target cells. To achieve the above advantages and to promote deeper penetration of current flow cytometry technology into the clinical area, a detection method must be developed that does not rely on fluorescence-based target detection so that cells can be detected without seriously damaging them.

One approach to overcoming this labeling problem is to use optical microscope images instead of fluorescent intensities to identify target cells from the morphological viewpoint. Various challenges have been met and overcome, and target cells have been detected using optical images without any labeling, both in vitro (Schonbrun et al. 2012) and in vivo (Golan et al. 2012). These studies were limited in terms of imaging flow cytometry, i.e., a cell-sorting function was not included, so the next step is to develop an imaging cell-sorting system in which target cells are recognized and sorted using morphological characteristics like cell size for the collection of label-free intact cells.

As a step towards achieving such a system, we have developed an on-chip high-speed imaging cell-sorting system for label-free collection of cells (Takahashi et al. 2004; Hattori and Yasuda 2010; Sawaki et al. 2007). The developed system has five core features: (1) fully automated sorting and collection of samples using a 3-D micropipetting device, (2) arrangement of the entire fluidic system in a disposable plastic chip, enabling biological samples to be lined up in a laminar flow using hydrodynamic focusing, (3) direct observation-based cell identification using specific image indexes and an optical system comprising phase-contrast/fluorescence microscopy and real-time image processing, (4) non-destructive wider dynamic range sorting using mild electrostatic force in a laminar flow (agarose gel electrodes are used to prevent electrode loss and electrolysis bubble formation), and (5) contamination-free recultivation of collected target cells in a microreservoir. The most important characteristics of this system are real-time image processing and the sorting of target cells on the basis of morphological indexes (e.g., cellular area or perimeter). The system also has a fluorescence detection unit, so cells can be sorted using a combination of image processing and conventional fluorescence detection. The cells are separated by applying a mild electrostatic force to only those cells to be removed for disposal; i.e., the targeted cells remain in the laminar flow and do not receive any stimulation from the electrostatic force.

In this paper, we describe the concept and present a complete picture of our automated imaging flow cell-sorting system. We do this by summarizing previously reported descriptions and data and presenting new descriptions and data that together give an overall picture of the integrated systems. We also discuss potential improvements that overcome the limitations of existing technologies.

2 Fully automated on-chip imaging flow cell-sorting system

Cell sorting is a target-cell purification technology consisting of (1) sample cell preparation, (2) target cell identification, (3) target cell collection, and finally (4a) recultivation of target cells or (4b) analysis of cells such as by DNA/mRNA analysis, protein analysis, or functional measurement (Fig. 2). Target cell identification and target cell acquisition are two major areas in cell-sorting technology development. The post-cell-sorting procedures (recultivation or analysis) should also be included in an integrated cell-sorting system.

Fig. 2
figure 2

Cell sorting procedure in fully automated on-chip imaging flow cytometry system: 1 cells in a crude sample mixture are dispersed and roughly purified using conventional preparation method, 2 target cells are identified on the basis of their configuration information, 3 identified cells are collected, and 4a identified target cells are recultivated for further experimentation or 4b used for analysis

For the purpose of non-invasive target cell identification and target cell acquisition, we have developed a fully automated system for on-chip imaging flow cell sorting (Hayashi et al. 2011a). As shown in Fig. 3, it comprises (1) a cell-sorting chip attached to a cell-sorting chip holder, (2) a holder for sample tubes and collection tubes, (3) an automatic 3-D motion pipetting device (60 × 20 cm motion in xy plane, and 20 cm motion in z axis), (4) a holder for pipetting chips, (5) an ejection port for used pipetting chips, (6) an air pressure unit connected to a syringe pump, (7) an automatic alignment set of phase-contrast/fluorescent optical microscopy modules, (8) an image-processing unit consisting of a high-speed camera and an image-processing and system control computer, and (9) a switching DC power supply.

Fig. 3
figure 3

Fully automated on-chip imaging flow cytometry system with disposable contamination-free plastic recultivation chip: a schematic diagram, b photograph (reproduced from Hayashi et al. 2011a)

3 Cell-sorting chips

3.1 Design

We tested two previously reported types of cell-sorting chips made of PMMA-based plastic (Hayashi et al. 2011a; Hattori et al. 2011; Takahashi et al. 2004): a two-way outlet chip (Fig. 4a–c) and a three-way-outlet chip (Fig. 4d–f). Both types have inlet pathways for the sample and buffer flows, hydrodynamic-focusing pathways, a cell-sorting area where the sample and buffer flows join together while maintaining their respective boundaries, gel channels (a pair of agarose microelectrode storage spaces) attached to the cell-sorting area, and outlet pathways for the target samples and waste samples. The cell-sorting area was 50 μm wide by 25 μm high. The two agarose microelectrode storage spaces were filled with 1 % sol-state agarose containing 0.5 M NaCl as electrolyte. The use of this electroconductive solvent to form the electrodes prevented electrolysis bubble formation. The layer of PMMA film on the bottom of each chip was 100 μm thick (which is within the depth of focus of a × 100 objective lens) and was attached to the chip by heat bonding.

Fig. 4
figure 4

Two- (ac) and three- (df) way-outlet cell-sorting chips. Photographs (a and d), schematic images (b and e), and magnified images (c and f) of chips. g Overall view and h side view of a recultivation chip (reproduced from Hayashi et al. 2011a; Hattori et al. 2011)

The agarose was injected into the gel inlets immediately before each trial. Small acrylic tubes were attached to both the sample inlet and outlet for chip collection, and large acrylic tubes were attached to the inlet area upstream and to the outlet area downstream to serve as a buffer reservoir (Fig. 4g–h). In the two-way-outlet chip, a raw sample was introduced from the sample reservoir into one of the two microfluidic pathways fabricated in the chip, and a voltage was applied to the gel microelectrode storage spaces to shift the cells to the other buffer pathway. In the three-way-outlet chip, a raw sample was introduced from the sample reservoir into the center pathway, and an appropriate voltage was applied to the gel electrodes to shift the cells to the left or right laminar buffer flow.

3.2 Fabrication

We selected poly(methyl methacrylate) (PMMA) as the base material for both types of cell-sorting chips due to its visible light transparency, low self-fluorescence, easy disposability after use, compatibility with glue-free heat bonding, ability to be flexibly formed by heat, hydrophilicity in a microfluidic pathway, and cost efficiency. As shown in Fig. 5a–c, PMMA has a high transmittance (~90 %) in the range of visible light from 400 to 600 nm, which is approximately the same as that of a glass slide and a polycarbonate culture dish.

Fig. 5
figure 5

Transmittance of a a glass slide, b a polystyrene culture dish, and c a PMMA sample plate 1 mm thick. Magnified images of PMMA-based cell-sorter chip fabricated by d hot embossing molding and e, f injection molding. Chip shown in (f) had added narrow channels between flow channels and gel channels (adapted from Hayashi et al. 2011a)

We first investigated the use of hot embossing to mold the chips. As shown in Fig. 5d, the fabricated microchannels had rounded edges and did not reflect the detailed edge structures of the microfluidic pathways in microscopic resolution. This was because the PMMA was not soft enough to permeate into the edges of the square-shaped microchannels at the temperature used in the hot embossing process. Then, as an alternative, we investigated the use of injection molding for the same microchannel design used for the hot embossing. The edge structures of the microchannels were improved—they had a sharp square shape. However, the melted PMMA resin did not permeate into the thin wall structure at the junction of the flow channels and gel channels (Fig. 5e). We thus redesigned the injection mold to thicken the wall structure by adding narrow channels (15 μm wide; 45 μm long) between the flow channels and gel channels (Fig. 5f). This enabled the PMMA resin to permeate deep into the walls around all the microchannels, resulting in junctions with a fine structure. This new design was used for the cell-sorter chips. The flow channel in the cell-sorting area was 50 μm wide and 25 μm high. The molded PMMA plate was sealed with a thin PMMA sheet (100 μm thick) by hot lamination without glue.

3.3 Spontaneous filling of agarose solution

To enable filling of the agarose into the microchannels, the chip was preheated on a hot plate at 42 °C for 5 min. A melted solution of 1 % agarose and 0.5 M NaCl was then loaded into the gel inlets. Unlike with the poly(dimethylsiloxane) (PDMS)-based microfluidics chip previously reported (Hattori and Yasuda 2010; Sawaki et al. 2007), the melted gel spontaneously permeated into the gel channels due to capillary action—additional pressure was not required. The injected solution flowed around the junctions of the flow channels and gel channels, permeated the narrow channel structure, and stopped flowing at the tip of the narrow channel. The agarose solution became a gel as it cooled at room temperature, and the gel remained filled to the edges of the flow channels (Fig. 6). This spontaneous permeation is attributed to the PMMA having higher wettability than PDMS.

Fig. 6
figure 6

Filling of agarose electrode gel in microelectrode structure. Agarose was labeled with fluorescent dye (Rhodamine B) for clear visualization, and a fluorescent image of the cell-sorting area was captured (reproduced from Hayashi et al. 2011a)

Channel length was an important factor in the complete filling of the agarose solution to the flow channel edges. Channels that were 70 μm or shorter were completely filled (Fig. 7a, b), while those that were longer than 70 μm had portions that were not filled (Fig. 7c).

Fig. 7
figure 7

Effect of flow channel length on precise control of gel electrode edge position: a 30 μm, b 70 μm, and c 90 μm. Yellow arrows indicate tips of agarose gel electrodes in gel channel. Scale bar 100 μm (reproduced from Hattori and Yasuda 2010)

The technique described above for forming gel electrodes in the microfluidic flow channels can be easily expanded for use in forming an array of electrodes to increase the effective area of electrostatic force irradiation. As shown in Fig. 8, 100 pairs of 15-μm wide gel electrodes were arranged with a 35-μm interval in a three-way-outlet sorting chip. The pathway of flowing living cells was continuously shifted by sequential application of electrostatic force to the electrode array (Fig. 8c).

Fig. 8
figure 8

100-step gel electrode array in microfluidic flow channel on a chip. a Array overview. In three-way-outlet sorting chip (see Fig. 4d–f), a 100-step gel electrode array was arranged on both side walls of a microfluidic pathway. b, c Continuous shift of cell flow pathway using the array. Time-course movements of human promyelocytic leukemia cells (HL60) were taken and overlapped. Cell flow pathways are shown as white lines. c When 10 V DC was applied to gel electrodes, cells shifted to anode gel electrode array wall because cell surface had a negative charge in general. Agarose gel containing 0.5 M NaCl formed electrolyte. Width of electrode was 15 μm, and gap between electrodes was 35 μm. Scale bar 200 μm

3.4 Cell-sorting process

The cell-sorting process begins with cell separation. An electric field created by DC voltage is used to generate electrostatic force in the cell-sorting area, causing cells to move from one laminar flow to another. As cells usually are charged negatively in water, the plus charge gel electrode is set at the cell collection streamline side. The use of agarose gel electrodes enables a high voltage to be applied between a pair of electrodes without bubble formation due to electrolysis. However, heat generation in an agarose gel electrode is problematic because its melting point is about 80 °C. Application of a high voltage to the gel electrodes could cause them to melt. This means that selection of the electrolyte component is important.

To select a suitable electrolyte, we estimated the resistance (Ω/mm) of an agarose gel electrode containing various types of 0.5 M electrolyte (Fig. 9). The resistance of a gel electrode filling a 1-mm-diameter capillary tube was measured with an impedance analyzer (Agilent 4294A). An agarose gel containing CoCl2 had the lowest resistance, and ones containing CaCl2 and FeCl3 had relatively low resistances. Although one containing NaCl did have slightly higher resistance, we selected NaCl as the electrolyte in consideration of the possible effects of current carrier ions on cells during cell sorting. For example, Na+ and Cl do not stimulate cells as do other types of ions such as Cu2+, K+, Ca2+, and Mg2+. Furthermore, agarose containing high-ionic-strength electrolytes, such as MgSO4, CaCO3, or ZnSO4, is not suitable as a gel electrode because it tends not to harden at room temperature.

Fig. 9
figure 9

Resistance of 1 % agarose gel electrode containing various electrolytes with 0.5 M concentration (reproduced from Hattori and Yasuda 2010)

3.5 Focusing technology

3.5.1 Hydrodynamic focusing of buffer streams in microchannel of PMMA-based cell-sorting chip for lining up samples

To examine the efficacy of laminar flow generation in the microchannel of a PMMA-based cell-sorting chip for lining up biological samples, we loaded a solution containing red ink into the sample inlet and clear buffer into the buffer inlets (Fig. 10). The stream of sample solution was narrowed from 50 to 5 μm using a hydrodynamic-focusing mechanism created by hydrodynamic-focusing inlets on both sides of the sheath stream, as described in previous reports (Takahashi et al. 2004; Sawaki et al. 2007; Hattori and Yasuda 2010; Hattori et al. 2011). The laminar flows of the sample and sheath solutions were observed without turbulence or intermixture at the flow velocity in the cell-sorting area (0.1–200 mm/s; Reynolds number 0.005–10.0) (data not shown). The velocity of the microfluidic flow was measured by tracing 2-μm polystyrene spheres in the laminar flow using a sequence of camera images taken at 1/200–1/10,000 s. The flow velocity in the cell-sorting area was about ten times that in the sample inlet channel, and the linear density of the sample particles in the cell-sorting area was ten times lower than in the sample inlet channel because the ratio of the widths of the laminar flow between the sample channel and the sheath channels was 1:10 (Sawaki et al. 2007), i.e., the width of the central streamline was less than the size of a cell. The PMMA-based cell-sorting chip, thus, had microchannels shaped well enough for on-chip cell sorting that was exact as that with a PDMS-based chip.

Fig. 10
figure 10

Laminar flow generated by three-forked sheath structure. Sheath buffer containing red ink was used as sample solution. Flow velocity was 1 mm/s (reproduced from Hayashi et al. 2011a)

We also examined the effectiveness of hydrodynamic focusing combined with hydrodynamic-focusing side sheath inlets quantitatively. First, we examined the hydrodynamic focusing with three simple equal-flow stream lines, as shown in Fig. 11a–c, in which a 25-μm sample flow stream line was confined by two 25-μm flanking sheath flow lines, resulting in a combined 25-μm flow line. A width of 25 μm is exactly that needed to fit the width of the microfluidic pathways. Comparison of the cell distributions in the flow lines upstream (Fig. 11b) and downstream (Fig. 11c) showed that a distribution of red blood cells (diameter 5 μm) was successfully focused within a 10 μm width, which is twice the sample size.

Fig. 11
figure 11

Laminar flow generated by three-forked sheath structure. Sheath buffer containing red ink was used as sample solution. Flow velocity was 1 mm/s. Evaluation of shapes of two side sheath flow channels. a Micrograph of trial microfluidic chip used to evaluate effect of side sheath flows. Widths of sample channel and sheath flow channels were 20 μm. b Distribution of horse red blood cells in sample channel at A in a. c Distribution of red blood cells at B in a. d Micrograph of another trial microfluidic chip in which width of two side sheath flow channels changed smoothly from 200 to 20 μm before reaching junction. e Distribution of red blood cells at C in d. f Distribution of red blood cells at D in d. g Micrograph of proposed cell-sorting chip in which width of two side sheath flow channels changed smoothly from 200 to 20 μm before reaching junction. h Distribution of 2-μm polystyrene beads at E in g. i Distribution of 2-μm polystyrene beads at F in g. j Phase-contrast image of 2-μm polystyrene beads in image recognition area of g. k Distribution of 2-μm beads for 20-μm range in h. l Distribution of 2-μm beads for 20-μm range in i. Scale bar 100 μm (reproduced from Hattori et al. 2011)

Next, we examined the contribution of accelerated hydrodynamic focusing using a wider flow line with less flow capillary resistance (Chip design B) (Fig. 11d–f). The design of the hydrodynamic-focusing inlets was changed to increase the widths of the side sheath flow channels from 25 to 200 μm and focused again to 20 μm immediately before focusing. We found that the distribution of cells at the center streamline was further narrowed to 5 μm in the merged lower streamline (Fig. 11e–f). As this width matches the size of the target cells (5 μm), the differential flow velocity of the hydrodynamic focusing is sufficiently effective to prevent particles from lining up perpendicular to the flow direction. It should be noted that we simply needed to use a single common pressure at the inlets of the center streamline and the two side streamlines to generate differential flow velocities at the hydrodynamic-focusing point because we can generate relative differences in flow velocity among streamlines using the hydrodynamic-focusing inlets. The flow velocity depends on the capillary tube’s width and height. Hence, we adjusted the shape of our three-way-outlet cell-sorting chip, as shown in Fig. 4.

Finally, we applied Chip design B to our original 50-μm wide two-way-outlet cell-sorting chip (Fig. 11g–l). We used 50-μm wide sample flow lines for practical application (Fig. 4). Although the width of the sample flow lines was twice that of the original chip, the width of the sample distribution after hydrodynamic focusing reached 5 μm even when another sheath flow was attached to the sample flow line in the cell-sorting area (see Fig. 11j). Hence, combining two chips with design B did not affect each set of streamlines, meaning that multiple chips with design B can potentially be combined.

3.5.2 Focusing of microparticles by dielectrophoretic force using V-shaped electrode array

In addition to particle focusing using a sheath flow, another method based on dielectrophoretic (DEP) force (Hayashi and Yasuda 2010; Hayashi et al. 2011b) was also examined. The DEP force, F DEP, acting on a particle in a non-uniform electric field is given by the following equations (Hughes 2002):

$$ F_{\text{DEP}} = (\pi D^{ 3} \varepsilon_{\text{m}} {\text{Re[}}f_{\text{CM}} ]\nabla E^{ 2} )/ 4, $$
(1)
$$ f_{\text{CM}} = \left( {\varepsilon_{\text{p}}^{*} - \varepsilon_{\text{m}}^{*} } \right)/\left( { \varepsilon_{\text{p}}^{*} + 2\varepsilon_{\text{m}}^{*} } \right), $$
(2)

where D is the particle diameter, ε m is the permittivity of the medium, f CM is the Clausius–Mossotti (CM) factor, E is the applied electric field, and \( \varepsilon_{\text{m}}^{*} \) and \( \varepsilon_{\text{p}}^{*} \) are the complex permittivities of the medium and the particle, respectively. The complex permittivity is defined as

$$ \varepsilon^{*} = \varepsilon - \left( {i\sigma } \right)/\omega , $$
(3)

where ε is the electric permittivity, σ is the electric conductivity, and ω is the angular frequency of the applied electric field. When the real part of the CM factor is positive, a particle moves toward stronger electric fields, and when the real part is negative, it moves toward weaker ones.

To focus microparticles including cells toward the center of a microfluidic channel using the DEP force, a V-shaped electrode array was placed at the center of the microchannel, as shown in Fig. 12 (Hayashi and Yasuda 2010; Hayashi et al. 2011b). An electrode array was placed on the bottom of the microfluidic channel, and alternate electrodes were connected to the opposite terminals of an alternating current (AC) power supply. The principle of particle focusing is as follows. Application of a voltage causes a non-uniform electric field to form. The strongest region is at the edge of the electrodes at the bottom of the channel, and the weakest region is at the top, immediately above the electrodes. If the DEP force (gray arrows in Fig. 12a) on a particle is positive, the particle is pulled toward one edge of the electrodes. In total of forces between DEP and fluid friction, the particle moves in the tip end direction of the electrode in horizontal and bottom in vertical. If the DEP force on a particle is negative (Fig. 12b), the particle moves in the tip end direction in horizontal and top in vertical in the same manner as with a positive charge. In short, all particles are aligned in the tip-end direction of the V-shaped electrode in the horizontal direction (Fig. 12e) and also focused toward the bottom (positive DEP) or the top (negative DEP) of the microchannel in the vertical direction (Fig. 12c). The sign of the DEP force depends on the real part of the CM factor, as shown in Eq. (1), and it generally depends on the particle type and AC frequency. These properties have been well studied, enabling the DEP force to be used for concentrating and separating particles, including cells (Urdaneta and Smela 2007; Huang et al. 2011; Unni et al. 2012; Wu et al. 2012).

Fig. 12
figure 12

Use of DEP force to focus microparticles. ae Schematic images of DEP focusing. Section views a, b, c and top views d, e of microchannel equipped with V-shaped electrode arrays (gray rectangles). Solid lines in a, b, and c represent electric flux lines. Direction of flow is from left to right. Circles represent microparticles with positive (black) or negative (white) DEP force applied. Arrows indicate resultant force (white arrows), fluid friction (black arrows), and DEP force (gray arrows). fh Detail of geometry for fabricated V-shaped electrode array. Scale bar 500 μm. i, j Focusing of Bacillus spores using V-shaped electrode arrays. Images when AC voltage was i not applied and j applied are shown. Scale bar 100 μm. kp Experimental results for PS bead focusing using DEP force generated by a V-shaped electrode array. k, l AC voltage was not applied to electrodes; n, o AC voltage was applied. m Distance of particle from bottom of microchannel was estimated by pattern matching of defocused microscopic images using standard images. Scale bar 100 μm for k and n and 20 μm for l, m, and o. p Z-positions of PS beads before (filled circles) and after (open circles) flowing through V-shaped electrode array (reproduced from Hayashi and Yasuda 2010; Hayashi et al. 2011b)

The focusing ability of a method based on the DEP force was tested using Bacillus spores (about 1 μm in diameter) as a cell model with the application of 20 Vpp and 1 MHz of AC voltage (Hayashi and Yasuda 2010; Hayashi et al. 2011b). The spores flowed in the microchannel at a velocity of 1 mm/s. They moved across the stream along the lines of the V-shaped electrode array from the periphery to the center of the microchannel, as shown in Fig. 12j. For the vertical direction, the focusing position of the spores was at the bottom of the microchannel because the spores were subjected to positive DEP forces under the experimental conditions. The vertical distribution of particles was measured using polystyrene (PS) beads (1 μm in diameter) under the same electric field conditions as that for the Bacillus spores (Hayashi and Yasuda 2010). The distribution was measured by pattern matching of their defocused microscopic images (Fig. 12k–p). The CM factor of the PS beads was negative (i.e., negative DEP); Fig. 12p shows that randomly distributed PS beads (filled circles) were concentrated on the ceiling of the microchannel (open circles).

As described above, we tested two focusing methods: sheath flow focusing and DEP force focusing. Both methods have advantages and disadvantages, as summarized in Table 1. Sheath flow focusing is suitable for high-flow velocity because a stable laminar flow forms even when the flow velocity is quite fast. DEP focusing is suitable for small amount of samples without dilution of sample buffer concentration. The flow velocity is easier to control with DEP focusing, because DEP focusing do not request any balance control of flow velocities of additional side sheath buffers. However, the suitability of DEP focusing for high-flow velocity is, in principle, limited because of the electrolysis of the solvents used for application of AC voltage to the electrodes placed only on the bottom of the microfluidic channel. Taking these points into account, we are using sheath flow focusing for our high-speed cell-sorting system until the problems with V-shaped DEP focusing are solved.

Table 1 Comparison of two focusing methods investigated

3.6 Particle sorting and separation using two- and three-way-outlet cell-sorting chips

We fabricated both types of PMMA-based cell-sorting chips described above. The sample particles were aligned on the centerline of an upstream microchannel using the sheath flow mechanism, and they passed through the cell-sorting area between the gel electrodes. The outlet channel was switched using electrophoresis by switching the DC voltage applied to the electrodes, while particles were passing between the tips of gel electrodes.

With the two-way-outlet chip, the inlet channels had a symmetrical structure (Figs. 4a–c, 13a–b). The sample inlet flow was sheathed by two sheath channels. The sheathed channels and their mirror-image-like channels joined at the gel electrode connection point and branched into two symmetrical outlet channels. When the DC voltage between the gel electrodes was turned off, the stream of sample particles exited the same side outlet (Fig. 13a); when the DC voltage was turned on, it exited the other outlet (Fig. 13b). The separation efficiency with a two-way-outlet cell-sorting chip when a DC voltage was continuously applied depended on the voltage and flow velocity (Fig. 13l–m). At a constant flow velocity, the separation frequency increased with the DC voltage and reached a plateau: the efficiency was 90 % at 45 V for a flow velocity of 200 mm/s (Fig. 13l). On the other hand, at a constant voltage, the separation frequency decreased as the flow velocity was increased, and eventually no particles were separated; the efficiency was less than 50 % at 45 V for a flow velocity of 500 mm/s (Fig. 13m). Moreover, the two-way-outlet chip did not separate the small molecules and charged particles; rather it moved the charged particles away from the small molecules, i.e., the polystyrene particles and fluorescein in the sample mixture were completely separated (Figs. 13j–k).

Fig. 13
figure 13

Particle separation using two- and three-way outlet cell-sorter chips. Switching mechanism using electrophoresis of gel electrodes: a, b two-way type and dg three-way type chip. c Color codes for charge state of sample particles and buffers. a, d Without DC voltage application, sample particles flowed along path marked by red line. b, e, f Application of DC voltage switched the outlet channel depending on particle’s charge and DC voltage polarity. g Mixed particles with negative, positive, or no charge were sorted correctly. Tracks of 20-μm polystyrene particle on two-way outlet chip h without and i with DC voltage applied between gel electrodes. Separation of 2-μm polystyrene particles (green track) and fluorescein (red track) j without and k with DC voltage applied between gel electrodes. Dependence of sorting efficiency on l DC voltage between gel electrodes and m flow velocity. Tracks of a 2-μm polystyrene particle on three-way outlet chip n without and o, p with DC voltage applied in opposite directions between gel electrodes. Direction of electric field is indicated by plus and minus icons. Flow direction in all figures was from top to bottom (reproduced from Hayashi et al. 2011a)

With the three-way-outlet chip, there were two sets of sheathed triplet channels symmetrically arranged as inlet and outlet channels on the chip (Figs. 4d–f, 13d–g). When the DC voltage between the gel electrodes was turned off (Fig. 13d, n), the stream of sample particles exited the center outlet; when the DC voltage was turned on, each particle exited one of the other outlets depending on the particle’s charge and the direction of the applied DC electrical field (Fig. 13e–f, o–p). The particles in a mixed suspension containing negatively charged, positively charged, and non-charged particles were separated into the three outlet channels depending on the charge state of each particle (Fig. 13g). This means that the three-way chip can be used for purification of target cells from a crude sample mixture containing various kinds of undesired particles such as non-target cells, debris, and inanimate particles.

4 Image analysis

4.1 Real-time recognition of image of flowing particles

The image-processing unit recognized the image of each flowing particle at a rate of up to 10,000 frames/s and 1/200- to 1/10,000-s analysis intervals, as shown in Fig. 14. A background image of the observation area was acquired at the end of the initialization process. To determine the boundary of a particle image during the sorting process, the background image was subtracted from the original image in each frame, and the resulting image was binarized using a preset threshold. The parameters of a particle’s image, the mean intensity, area, lengths of major and minor axes, degree of circularity, etc., were calculated about the pixel value within the boundary. When one of, or a set of, the parameters satisfied that of a target or non-target particle, the image-processing unit switched the DC voltage between the gel electrodes of the cell-sorter chip by switching the DC power supply on or off in real time.

Fig. 14
figure 14

Analysis of microscopic image in real time using image-processing unit for 1/10,000-s interval

4.2 Real-time size-discrimination of two particles close in size using microscopic images and improvement of depth of field using zoom lens optics for deeper microfluidic pathways

To evaluate the image-recognition ability of the cytometry system, we used a mixed suspension containing polystyrene particles close in size (2.5 or 3.0 μm in diameter) (Fig. 15). Their images were clear, but the sizes were hard to discriminate with the human eye without a scale or a size-comparison object (Fig. 15a–b). However, the image recognition unit of the cytometry system discriminated between the two sizes with an error rate of less than 2 % at 200 frames/s in real time at a flow velocity of 10 mm/s (Fig. 15c–e). For a mixture of 5- and 10-μm polystyrene particles, the error rate was less than 1 % at a flow velocity of 200 mm/s. For a mixture containing particles with various specific diameters, the error rate increases with the flow velocity (Hattori and Yasuda 2012; Hattori et al. 2011). These results suggest that imaging cytometry has many advantages for precise and high-throughput separation of target cells from undesired cells with a difference in size that is difficult to discriminate with the human eye.

Fig. 15
figure 15

Real-time size-discrimination of polystyrene particles close in size (2.5 or 3.0 μm in diameter) using microscopic images bright-field microscopic images of particles with diameter of a 2.5 μm or b 3.0 μm. Histograms of particle areas measured with imaging cytometry system at flow velocity of 10 mm/s: sample suspension containing c only 2.5-μm particles, d only 3.0-μm particles, and e mixed suspension with 2.5- and 3.0-μm particles (reproduced from Hayashi et al. 2011a)

We also examined the resolution of the depth of field when the height of a polystyrene bead was changed in an inverted microscope (Olympus IX70) with three objective lenses [Olympus UPlanFl 10× (NA = 0.3); LCPlanFl 20× (0.4); LCPlanFl 40× (0.6)]. Figure 16a–c show micrographs of a 15-μm bead acquired with the zoom optics (10× objective lens plus zoom optics) and with the inverted microscope for every 5-μm depth at total magnifications of 10×, 20×, and 40× (upper micrographs) and with 10×, 20×, and 40× objective lenses (lower micrographs). The total magnifications of the optics were adjusted by changing the zoom ratio in the zoom optics and exchanging the objective lenses in the inverted microscope. There was almost no variation in the hues of the bead at bead depths of 0–25 μm in the zoom optics. In contrast, defocusing increased with the bead depth in the microscope images. The amount of focus change was particularly large at magnifications of 20× and 40×. From the results shown in Fig. 16, the depth of field was estimated to be 25 μm or greater at any magnification in the zoom optics. It was estimated to be 15 μm at 10× and 5 μm at 20× and 40× in the microscope images. These results indicate that the depth of field is improved at any magnification from 10× to 40× and that an extended depth of field optics for image-based microfluidic pathways (i.e., wider depth resolution in deep microfluidic pathways) can be constructed using a low NA objective lens and a zoom lens.

Fig. 16
figure 16

Extended depth of field optics for precise image analysis in microfluidic imaging flow cytometry. Micrographs of 15-μm polystyrene bead acquired by zoom optics (upper) and inverted microscope (lower) for every 5-μm depth at total magnifications of a 10×, b 20×, and c 40× (reproduced from Hattori and Yasuda 2012)

4.3 Indexes acquired from imaging flow cell-sorting system

As described above, the parameters of a particle’s image, including the mean intensity, the area, the lengths of the major and minor axes, and the degree of circularity, were calculated about the pixel value (Fig. 14). Figure 17a shows example data analyzed using the calculated parameters. The analyzed roundness and size of each cell image was plotted, and the original images were stored on a hard disk drive and cross-referenced to the plots. This enabled detailed images from the plotting to be called up and used for checking detailed shapes of samples. Figure 17b shows examples of acquired micrographs in which multinucleated cells and M-phase cells were identified, illustrating one of the advantages of imaging flow cell sorting compared to conventional diffraction-based cell identification analysis.

Fig. 17
figure 17

Two-dimensional mapping of acquired cell images. a Analyzed roundness and size (cross-sectional area) of each cell image is plotted on map at center. Original images were stored and cross referenced to plots. b Magnified micrographs of acquired cell images. Even at a rate of 10,000 images/s, the resolution was good enough to identify M-phase cells, multinucleated cells, and rough surface cells

4.4 Purification of cardiomyocytes using index of cell surface roughness

In imaging flow cell sorting, the particular index of each cell image is essential for separating the images. Previous examination showed that most of the round smooth cells in a mixture containing cardiomyocytes are cardiomyocytes (Kaneko et al. 2011). We thus defined index R of cell surface roughness for use in comparing the actual and estimated cell surface perimeters by approximation of a spherical model, enabling the surface roughness of cardiomyocyte cells and other cells to be distinguished quantitatively.

$$ R = \frac{l}{{\sqrt {4\pi S} }}, $$
(4)

where l is the cell surface perimeter, and S is the cell cross-sectional area (both l and S are acquired from pixel-based digitized-image cell analysis). For example, a cell with reduced radius r has R (=2πr/(4π × πr 2)1/2) equal to 1, and a cell with higher roughness and thus a larger perimeter has a proportionally larger R. Index R is used to quantify the smoothness or roughness of cell surfaces. Smooth cells typically have an R less than 1.1 (Fig. 18a), and rough cells typically have an R larger than 1.1 (Fig. 18b).

Fig. 18
figure 18

Separation of cardiomyocyte cells using on-chip imaging flow cytometry system. Phase-contrast images and field programmable gate array (FPGA) calculated indexes of a target smooth cells (cardiomyocytes) and b non-cardiomyocyte cells (reproduced from Nomura et al. 2011)

There are two major requirements for achieving non-labeled separation of cardiomyocyte cells from a mixture of enzyme-treated heart extracts. One is the development of a reliable index for quantifying the surface roughness of cells for use in distinguishing the target cells, and the other is the development of an enzymatic method for visualizing the differences in surface shape for use in distinguishing individual cells.

Experience has shown that the collagenase-digestion process is effective in cardiomyocyte purification for distinguishing between smooth and rough cell surfaces and thus distinguishing between cell types. This process must be followed by incubation for at least 20 min before cell sorting for the differences between smooth and rough surfaces to be distinguishable. With the conventional trypsinization cell preparation procedure, the differences between smooth and rough surfaces are indistinguishable.

In an experiment, a sample mixture of cells was acquired from collagenase-digested heart tissue isolated from 13-day-old mouse embryos, and the cell-sorting system was used to distinguish the smooth surface cells (Fig. 18a) from the rough surface cells (Fig. 18b). Index R for the smooth cells was 1.08 or less, and that for the rough cells was greater than 1.1. These results show that an R value of 1.1 is a suitable threshold for distinguishing smooth and rough surface cells in our on-chip imaging flow cell-sorting system.

Using this index and our label-free on-chip imaging flow cell-sorting system, we succeeded in separating cardiomyocyte cells. As shown in Table 2, of the 242 smooth cells, 178 (73.6 %) were beating cells, 19 (7.9 %) were non-beating cells, 12 (5.0 %) were shrunk cells, and 33 (13.6 %) were unattached cells (Table 2). There were 195 cardiac myosin positively stained cells (99.99 % of healthy cultivated cells; 80.6 % of all collected smooth cells), and there were two negatively stained cells (0.01 % of healthy cultivated cells; 0.8 % of all cells). This means that more than 80 % of the smooth cells were beating cardiomyocyte cells and that more than 99 % of the healthy recultivated cells were cardiomyosine positive cells. These results suggest that non-labeling simple phase-contrast image-based cell sorting using the index of cell surface roughness can potentially be used to separate cardiomyocyte cells from a mixture of heart cells with an error rate of less than 1 % (Nomura et al. 2011).

Table 2 Results of separating cardiomyocyte cells

5 Application of imaging cell sorter

5.1 Selection of membrane filter for preventing contamination

The transfer of sorted cells from the cell-sorting chip into a culture vessel involves the risks of cell contamination, damage, and loss. To overcome these risks, we developed an on-chip culture system using a PMMA-based cell-sorting chip. Although the sorted cells in the cell-sorting chip can be cultured by transferring the whole chip into an incubator, there is still the risk of contamination because the medium in the outlet tubes is exposed to outside air. Moreover, since the two or three outlet tubes on the cell-sorting chip are linked by the culture medium, waste cells and/or particles can contaminate the collection outlet due to convection or self-propulsion during cultivation. To prevent such contamination, we need to isolate the outlet tubes from the downstream reservoir using a membrane filter. The membrane filter needs to be permeable to nutrients and gasses and impermeable to cells and bacteria. A tight seal between the edge of each outlet tube and the filter is created by thermal bonding. Since the softening point of PMMA is about 100 °C, the filter should have resistance at the temperature used for thermal bonding. The membrane surface must be ruck-free to achieve tight sealing. For the filter to be permeable to nutrients and gasses, the membrane must have good hydrophilia. The filter must be transparent to visible light to enable the cells in the outlet tubes to be observed without breaking the seal. We compared the seven commercial membrane filters listed in Table 3 on the basis of these requirements and selected the first four for testing.

Table 3 Seven candidate commercial membrane filters for use as isolation filter

We used E. coli to test the permeability of the four selected filters. A sample filter was bonded to the edges of dummy outlet tubes. The Omnipore filter for all pore sizes and the Isopore filter with pore sizes of 0.1 and 0.2 μm were not permeable to E. coli, while the Isopore filter with pore sizes greater than 0.4 μm and the other filters were permeable (Fig. 19).

Fig. 19
figure 19

Permeability of membrane filters to E. coli: a experimental setup, b number of colonies on lysogeny broth (LB) agar plate after removal from medium following injection of E. coli suspension into dummy outlet tube (reproduced from Hayashi et al. 2011a)

Given the impermeability of the Omnipore filters to E. coli, we tested their use on our cell-sorter chip (Fig. 20). Cardiomyocyte cells adhered to the bottom of the membrane-sealed outlets and remained beating over 2 days (Fig. 20a–c). Contamination by E. coli from the downstream reservoir into the outlet tubes during cultivation was observed for filters with pore sizes of 0.2 and 0.45 μm but not for that with 0.1-μm pores (Fig. 20d–i). Therefore, the use of a 0.1-μm-pore Omnipore filter with our PMMA-based cell-sorting chip enables on-chip cultivation without contamination.

Fig. 20
figure 20

Estimation of cell viability and contamination from downstream reservoir. ac Cardiomyocyte cells at bottom of outlet tube after cultivation for 2 days. Most cells were beating. LB agar plate exposed to medium in df collection reservoir and gi outer reservoir after cultivation for 2 days. Cultivation chip outlet tubes were sealed with Omnipore filter with pore size of 0.1 μm (a, d, g), 0.2 μm (b, e, h), or 0.45 μm (c, f, i) (reproduced from Hayashi et al. 2011a)

5.2 Integration of cell-sorting system with sample preparation unit and ultra-high-speed droplet PCR module

As illustrated in Fig. 2, one of the final destinations of cell sorting is analysis of the collected target cells. To accomplish this, we combined a stand-alone on-chip imaging flow cell-sorting module with a sample preparation module and an ultra-high-speed droplet polymerase chain reaction (PCR) module in a rack, which serves as an integral load-bearing structure (Fig. 21). Each module carries out a specific task related to rapid nucleotide detection: the sample preparation module concentrates the sample cells, (2) the cell-sorting module sorts the cells on the basis of image analysis, (3) and the PCR module analyzes the cells. A modular approach was taken so as to facilitate upgrading of the individual modules as more advanced ones become available. As shown in Fig. 21b, the modules are arranged in the rack so that they can easily be slid out horizontally for easy access and replacement.

Fig. 21
figure 21

Integrated system: a schematic drawing and b photograph. Ultra-high-speed droplet PCR module is located at bottom; immediately above it is on-chip imaging flow cell-sorting module. Sample preparation module (c) is at top; in it, the particulate solution is forced through a dyeing/concentration nozzle (d). Particulates are concentrated by membrane filter attached to bottom of filter housing

In the sample preparation module (Fig. 21c–d), 1,000 μL of the sample solution is forced through a silicone tube under pressure into a glass vessel. A single silicone tube is reused and is washed between trials. Once the vessel is filled with sample solution, it is pressurized to force the solution through a filter cartridge. This is accomplished by sealing the vessel with a lid equipped with two tubes; pressure is applied through one tube, and the solution is forced out through the second, which is long enough to reach residual fluid at the bottom of the vessel. When the solution is forced through the filter cartridge, particulates are trapped on the filter surface. A dispenser is inserted into the cartridge, and negative pressure is applied to recover these particulates. A filter with a smooth surface punctuated with discrete pores was used to facilitate their removal. It would be more difficult to recover the trapped particulates with a fiber filter. The internal shape of the cartridge was designed so as to minimize the volume between the filter and dispenser. It is important that the amount of fluid used for particulate recovery is kept at a minimum, because the concentration ratio is higher when the amount of recovered fluid is greater. The recovered particulates are transferred to another container that is accessible to the inter-module sample transfer dispenser. This module contains 12 sets of glass vessels, filter cartridges, and dispensers. The vessels are mounted on a rotating turret as are the filter cartridges with their own rotating turret. A dye can be injected into the vessels if the pathogen needs to be labeled. After sample preparation, the cells are transported and sorted in the second module, and the collected target cells are then analyzed in the third module.

DNA/mRNA analysis of the collected cells is done using a newly developed simple ultra-high-speed real-time μl sample droplet PCR module featuring rapid water-based heating of sample droplets by quick switching between two hot water circuits (one for denaturation and one for annealing and extension). The module consists of two hot water circuits, a reaction chamber, and a light-fiber fluorescent detector. As shown in Fig. 22, each hot water circuit has a tank with an internal heater, two diaphragm valves, a fine-adjustment heater, a check valve, and a geared pump and is connected to a PCR reaction chamber made of PDMS and containing a thin-film aluminum chip, on which a 3 × 8 array of micro indentations are embossed to hold μl-sized water droplets. The water in each circuit is circulated continuously by the pump during reaction to maintain the temperature of the hot water in each tank. It is transferred to the reaction chamber only when the temperature of the reaction chamber needs to be increased to that of the hot water. Fine temperature control of the circulated hot water is achieved by adjusting the appropriate fine-adjustment heater. One circuit maintains the water temperature at 95 °C; this water is used for denaturation of DNA fragments. The other maintains the water temperature at 50–70 °C; this water is used for extension of DNA fragments. Diaphragm valves are used simultaneously to control the circulation of the water and to transfer it when needed to the reaction chamber, where it is circulated under the aluminum chip, resulting in direct transfer of the heat from the water to the chip and then to the droplets. When the water is no longer needed, it is returned to the appropriate hot water tank and again circulated.

Fig. 22
figure 22

Configuration of ultra-high-speed real-time μl sample droplet PCR module: a schematic image and b photograph

Rapid PCR amplification of a set of droplets containing Bacillus anthracis lined up on the aluminum chip was evidenced by the increase in fluorescent intensity, as shown in Fig. 23a. It was accomplished using the sample preparation module and on-chip imaging flow cell-sorting module within 3.5 min (40 cycles of 1-s denaturation and 3-s extension). As shown in Fig. 23b, this is an order of magnitude faster than with a conventional fast PCR system. This integrated system, thus, enables rapid detection of DNA fragments and is potentially applicable to the measurement of multiple samples simultaneously in a miniaturized microfluidic chip.

Fig. 23
figure 23

a Results of ultra-high-speed real-time PCR amplification, b comparison between ultra-high-speed PCR and conventional PCR (air-thermo PCR instruments) (reproduced from Terazono et al. 2010)

6 Discussion

We can estimate the performance of the imaging flow cell-sorting system as follows. Suppose the maximum sampling rate of image acquisition and recognition is 200 events/s and the sorting efficiency is 90 % at a flow rate less than 200 mm/s in the cell-sorting area. The cell-sorting area is 50 μm wide, 25 μm high, and 100 μm long. The flow rate in the cell-sorting area is ten times that in the sample inlet because of the sheath structure. The discrimination error is less than 2 % for a mixture of 2.5- and 3.0-μm polystyrene particles. All operations after sample setting are automated and take less than 5 min except for the cell-sorting process.

Since the maximum sampling rate is 200 events/s, the maximum sample-processing rate is 25 nL/s (=50 μm(W) × 25 μm(H) × 100 μm(L) × 200 events/s). Under these conditions, the flow rate in the cell-sorting area is 20 mm/s (=100 μm × 200 events/s). For example, 50 μL of the sample will be processed within 33 min (=50 μL/25 nL/s). Therefore, the total operation time is less than 40 min.

Because the unit volume of sample processing is very small (125 pL/event, the volume of the cell-sorting area per event), sample dilution is not necessary for cytometry; the maximum density of sample particles is 8 × 106 particles/mL (=1/125 pL). For example, if the sample density is 1 × 106 particles/mL, which is 1/8th the maximum density, the sample particles are processed one by one with an entanglement error in cell sorting of less than 0.8 %, as shown by the Poisson distribution.

Therefore, analysis of a 50-μL sample at 1 × 106 particles/s requires that the error for discriminating target cells in the collection tube be less than 2 % and that the database of microscopic images be fully accessible within 40 min. This estimated performance demonstrates that our cell-sorting system is useful for biological research and clinical diagnosis.

The sorting time could be reduced by increasing the sorting force using improved technology such as an array of agarose gel electrodes (Fig. 8) to increase the effective area of electrostatic force irradiation. The recognition rate could be increased by improving the efficiency of cell sorting.

Figure 24 compares four typical cell-sorting technologies. Magnetic bead-based acquisition of target cells is the most simple and economical (Anzai et al. 2007; Terazono et al. 2010). However, the magnetic-bead interactions with target cells are probabilistic, so the population count is not precise. The on-chip flow cytometer uses the apoptosis phenomenon for cell purification; i.e., infrared laser irradiated cells are removed by apoptosis during cell recultivation. Although this technology is simple and fast, recultivation is essential for purification. The on-chip cultivation technology described in this review combines on-chip multielectrode array (MEA) measurement and a digestible polyarginine layer; it performs non-invasive electrophysiological functional identification of target cells and single-cell-level collection and recultivation (Terazono et al. 2012). The on-chip imaging flow cell-sorter system is the non-invasive shape-based detection technology of target cells as we described in this review.

Fig. 24
figure 24

Comparison of four typical cell-sorting technologies

One of the advantages of the on-chip imaging flow cell-sorting technology is quick and complete sample measurement based on image information for the samples. In particular, as shown in Figs. 17 and 18, it can obtain detailed information on cell cycles and conditions, which is difficult with conventional diffraction-based cell-sorting technologies. Complete sample measurement is important for estimating the probability of events and populations. Imaging flow cell-sorting technology is thus well suited for detailed analysis of cell groups.

We introduced two potential applications of our on-chip imaging flow cell-sorting: contamination-free recultivation and integrated PCR measurement system. In contrast to conventional flow cytometry, the most important feature of cell-sorting technology is acquisition of target cells from samples. Hence, further cell treatment procedures such as cell recultivation and cell analysis should be considered for practical cell-sorting technology development. We have demonstrated that a membrane filter lid with a disposable cell-sorting chip works well for contamination-free recultivation and that integration of an on-chip imaging flow cell-sorting module with a droplet PCR module enables identification of rod-shaped B. anthracis from a crude sample mixture. This means that one of the next steps in cell-sorting technology development should be the development of an integrated system for further cell treatment and analysis.

7 Summary

We have developed an on-chip high-speed imaging flow cell-sorting system with five core features: (1) optical image-based real-time cell sorting; (2) inclusion of all the cell-sorting procedures in an economical and disposable (burnable after use) plastic chip; (3) cell separation electrodes consisting of stable and bubble formation-resistant gel electrodes; (4) contamination-free recultivation; and (5) fully automated operation. This system can be modularized and combined with additional cell analysis modules such as a PCR analysis module. Testing demonstrated that cells and fluorescent dyes can be separated using this system. The time required for cell sorting for a 20-μL sample with 1 × 107 particles was within 30 min. This simple, safe, fast, and label-free flow cell-sorting system for cell purification is thus practical and potentially useful for biological research and clinical diagnosis.