SCAR R&D Symposium 2003: Comparing the Efficacy of 5-MP CRT Versus 3-MP LCD in the Evaluation of Interstitial Lung Disease

  • Published:
Journal of Digital Imaging Aims and scope Submit manuscript

Abstract

The efficacy of two medical-grade, self-calibrating, gray scale displays were compared with regard to impact on sensitivity and specificity for the detection of interstitial lung disease (ILD) on computed radiographs (CR). The displays were a 5-megapixel (MP) cathode ray tube (CRT) device and a 3-MP liquid crystal display (LCD). A sample consisting of 230 anteroposterior (AP), posteroanterior (PA), and lateral views of the chest with CT-proven findings characteristic for ILD as well as 80 normal images were compared. This double-blinded trial produced a sample sufficient to detect if the sensitivity of the LCD was 10% or more reduced (one-sided) from the “gold standard” CRT display. Both displays were calibrated to the DICOM gray scale standard and the coefficient of variation of the luminance function varied less than 2% during the study. Five board-certified radiologists specializing in thoracic radiology interpreted the sample on both displays and the intraobserver Az (area under the ROC curve) showed no significant correlation to the display used. In addition, an interobserver kappa analysis showed that the relative disagreement between any observer pair remained relatively constant between displays, and thus was display invariant. This study demonstrated there is no significant change in observer performance sensitivity on 5-MP CRT versus 3-MP LCD displays for CR examinations demonstrating ILD of the chest.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Figure 1
Figure 2
Figure 3
Figure 4
Figure 5

Similar content being viewed by others

References

  1. RJ Landis GG Koch (1977) ArticleTitleThe measurement of observer agreement for categorical data Biometrics 33 159–174 Occurrence Handle1:STN:280:CSiC3srhvFI%3D Occurrence Handle843571

    CAS  PubMed  Google Scholar 

  2. JR Mathieson JR Mayo CA Staples et al. (1989) ArticleTitleChronic diffuse infiltrative lung disease: comparison of diagnostic accuracy of CT and chest radiography Radiology 171 111–116 Occurrence Handle1:STN:280:BiaC1crks1I%3D Occurrence Handle2928513

    CAS  PubMed  Google Scholar 

  3. SP Padley DM Hansell CD Flower et al. (1991) ArticleTitleComparative accuracy of high resolution computed tomography and chest radiography in the diagnosis of chronic diffuse infiltrative lung disease Clin Radiol 44 IssueID4 222–226 Occurrence Handle1:STN:280:By2D2s%2FgsVE%3D Occurrence Handle1959296

    CAS  PubMed  Google Scholar 

  4. SC Ackerman JN Gitlin RW Gayler et al. (1993) ArticleTitleROC analysis of fracture and pneumonia detection: comparison of laser digitized workstation images and conventional analog radiographs Radiology 186 263–268 Occurrence Handle1:STN:280:ByyC3c7otVM%3D Occurrence Handle8416576

    CAS  PubMed  Google Scholar 

  5. BS Slasky WF Gur WF Good et al. (1990) ArticleTitleROC analysis of chest image interpretation with conventional, laser printed and high-resolution workstation images Radiology 174 IssueID3 Pt 1 775–780 Occurrence Handle1:STN:280:By%2BC287htFU%3D Occurrence Handle2305061

    CAS  PubMed  Google Scholar 

  6. WW Scott DA Bluemke WK Mysko et al. (1995) ArticleTitleInterpretation of ED radiographs by radiologists and emergency medicine physicians: teleradiology workstation vs. radiograph Readings Radiology 195 223–229 Occurrence Handle7892474

    PubMed  Google Scholar 

  7. GG Cox LT Cook JH McMillan et al. (1990) ArticleTitleChest radiography: comparisons of high-resolution digital displays with conventional and digital film Radiology 176 IssueID3 771–776 Occurrence Handle1:STN:280:By%2BA2MnnvVc%3D Occurrence Handle2389035

    CAS  PubMed  Google Scholar 

  8. Siegel E, Reiner, B, Moffet, BR, et al.: Differences in Image Quality of CR Images Displayed on LCD and CRT Monitors. Presentation at ARRS, 2002, Atlanta, GA

  9. Siegel E, Reiner B, Moffet, BR, et al.: Differences in Image Quality of CR Images Displayed on LCD and CRT Monitors. Presentation at SCAR 2002, Cleveland, OH

  10. ER DeLong DM DeLong DL Clarke-Pearson (1988) ArticleTitleComparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach Biometrics 44 837–845 Occurrence Handle1:STN:280:BiaD2snpslc%3D Occurrence Handle3203132

    CAS  PubMed  Google Scholar 

  11. J Cohen (1960) ArticleTitleA coefficient of agreement of nominal scales Educ Psychol Measure 20 37–46

    Google Scholar 

  12. Digital Imaging and Communications in Medicine, Part 14: Grayscale Standard Display Function, Rosslyn, VA: National Electrical Manufacturers Association, 2001

  13. American Association of Physicists in Medicine Task Group 18, Assessment of display performance for medical imaging systems (draft), available at http://deckard.mc.duke.edu/~samei/tg18; accessed July 8, 2003

  14. Z Wang AC Bovic (2002) ArticleTitleA universal image quality index IEEE Signal Process Lett 3 81–84 Occurrence Handle10.1109/97.995823

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Steve Langer PhD.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Langer, S., Bartholmai, B., Fetterly, K. et al. SCAR R&D Symposium 2003: Comparing the Efficacy of 5-MP CRT Versus 3-MP LCD in the Evaluation of Interstitial Lung Disease . J Digit Imaging 17, 149–157 (2004). https://doi.org/10.1007/s10278-004-1013-7

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10278-004-1013-7

Keywords

Navigation