Skip to main content
Log in

Direct puncture of a prematurely detached balloon in the supraclinoid ICA via the optic canal during embolisation of a traumatic carotid-cavernous fistula

  • Case Report
  • Published:
Acta Neurochirurgica Aims and scope Submit manuscript

Summary

We report a patient with traumatic carotid-cavernous fistula associated with an isolated internal carotid artery in whom, after trans-arterial balloon embolisation, premature balloon detachment occurred with balloon migration to the supraclinoid carotid artery, leading to total occlusion of the blood flow. The carotid flow was eventually restored by direct puncture of the detached balloon via the optic canal and by deploying a coronary stent to fix the balloon in the vascular wall. The fistula was eventually occluded by using detachable coils. He was discharged with a mild hemiparesis and decreased acuity in the left eye.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Bendorf G, Bender A, Campi A, Memeking H, Lanksch WR (2001) Treatment of a cavernous sinus dural arterio-venous fistula by deep orbital puncture of the superior orbital fissure. Neuroradiology 43:499–502

    Article  Google Scholar 

  2. Hallbach VV, Higashida RT, Barnwell SL, Dowd CF, Hieshima GB (1991) Trans-arterial platinum coil embolisation of carotid-cavernous fistulae. AJNR Am J Neuroradiol 12:429–433

    Google Scholar 

  3. Luo CB, Teng MMH, Chang FC, Chang CY (2006) Trans-arterial balloon-assisted N-butyl-2-cyanoacrylate embolisation of direct carotid cavernous fistulae. AJNR Am J Neuroradiol 27:1535–1540

    PubMed  Google Scholar 

  4. Luo CB, Chang FC, Teng MMH, Chen SS, Lirng JF (2001) Enodvascular emoblisation of carotid-cavernous fistulae using the posterior communicating artery approach: a case report. Kaohsiung J Med Sci 17:112–115

    CAS  PubMed  Google Scholar 

  5. Teng MMH, Lirng JF, Chang T et al (1995) Embolisation of carotid-cavernous fistulae by means of direct puncture of superior orbital fissure. Radiology 194:705–711

    CAS  PubMed  Google Scholar 

  6. White JB, Layton KF, Evans AJ et al (2007) Trans-orbital puncture for the treatment of cavernous sinus dural arterio-venous fistulae. AJNR Am J Neuroradiol 28:1415–1417

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to C.-B. Luo.

Additional information

This work was supported in part by a grant from the Taipei Veterans General Hospital (V96C1-073, V97C1-139) and NSC 97-2314-B-075-062-my2.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Luo, CB., Teng, MM.H., Lin, CJ. et al. Direct puncture of a prematurely detached balloon in the supraclinoid ICA via the optic canal during embolisation of a traumatic carotid-cavernous fistula. Acta Neurochir 152, 321–324 (2010). https://doi.org/10.1007/s00701-009-0319-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00701-009-0319-0

Keywords

Navigation