Abstract
Purpose
The 2010 American Heart Association or European Resuscitation Council guidelines for cardiopulmonary resuscitation emphasize that rescuers should minimize interruption of chest compressions, even for endotracheal intubation. Cervical stabilization should also be maintained during traumatic cardiac arrest. The utility of the Pentax-AWS Airwayscope (AWS) video laryngoscope and Airtraq (ATQ) optic laryngoscope for airway management has been reported under cervical stabilization. We first evaluated ATQ utility during chest compression with or without cervical stabilization and then compared the AWS, ATQ, and Macintosh laryngoscope (McL) during chest compressions under cervical stabilization in a manikin.
Methods
In the first trial, 19 novice doctors performed tracheal intubation with ATQ during chest compression with or without cervical stabilization. In the second trial, 21 novice doctors performed tracheal intubation on a manikin with cervical stabilization using AWS, ATQ, and McL with or without chest compression in a manikin. The rate of successful intubation, time to intubation, and subjective difficulty of use (visual analog scale) were recorded.
Results
In the first trial, intubation time during chest compression was significantly shortened under cervical stabilization compared to without cervical stabilization (P < 0.05). In the second trial, using McL, 3 participants failed to perform tracheal intubation without chest compression and 11 failed during chest compression (P < 0.05). Using ATQ, all intubations were successful without chest compression, but 5 failed during chest compression (P < 0.05). Intubation time was significantly prolonged by chest compression using McL or ATQ (P < 0.05). All participants successfully secured the airway with AWS regardless of chest compression, and chest compression did not prolong intubation time. Chest compression worsened the score on the visual analog scale of laryngoscopy in the McL trial (P < 0.05), but not in ATQ or AWS trials. Difficulty of tube passage through the glottis increased with chest compression with the McL and ATQ (P < 0.05) but not with AWS.
Conclusion
The AWS was superior to McL and ATQ for endotracheal intubation during simulated cervical stabilization and chest compression.
Similar content being viewed by others
References
Nolan JP, Soar J, Zideman DA, Biarent D, Bossaert LL, Deakin C, Koster RW, Wyllie J, Böttiger B. European Resuscitation Council Guidelines for Resuscitation 2010: Section 1. Executive summary. Resuscitation. 2010;81:1219–76.
Hazinski MF, Nolan JP, Billi JE, Böttiger BW, Bossaert L, de Caen AR, Deakin CD, Drajer S, Eigel B, Hickey RW, Jacobs I, Kleinman ME, Kloeck W, Koster RW, Lim SH, Mancini ME, Montgomery WH, Morley PT, Morrison LJ, Nadkarni VM, O’Connor RE, Okada K, Perlman JM, Sayre MR, Shuster M, Soar J, Sunde K, Travers AH, Wyllie J, Zideman D. Part 1: executive summary: 2010 international consensus on cardiopulmonary resuscitation and emergency cardiovascular care science with treatment recommendations. Circulation. 2010;122:S250–75.
Langeron O, Birenbaum A, Amour J. Airway management in trauma. Minerva Anesthesiol. 2009;75:307–11.
Dupanovic M, Fox H, Kovac A. Management of the airway in multitrauma. Curr Opin Anaesthesiol. 2010;23:276–82.
Davis DP, Peay J, Sise MJ, Kennedy F, Simon F, Tominaga G, Steele J, Coimbra R. Prehospital airway and ventilation management: a trauma score and injury severity score-based analysis. J Trauma. 2010;69:294–301.
Maharaj CH, Buckley E, Harte BH, Laffey JG. Endotracheal intubation in patients with cervical spine immobilization: a comparison of Macintosh and Airtraq™ laryngoscopes. Anesthesiology. 2007;107:53–9.
Enomoto Y, Asai T, Arai T, Kamishima K, Okuda Y. Pentax-AWS, a new videolaryngoscope, is more effective than the Macintosh laryngoscope for tracheal intubation in patients with restricted neck movements: a randomized comparative study. Br J Anaesth. 2008;100:544–8.
Sadamori T, Kusunoki S, Ishida M, Otani T, Tanigawa K. Video laryngoscopy for emergency tracheal intubation during chest compression. Resuscitation. 2008;77:155–6.
Corso RM, Piraccini E, Agnoletti V, Gambale G. The Airtraq laryngoscope for emergency tracheal intubation without interruption of chest compression. Am J Emerg Med. 2010;28:971–2.
Komasawa N, Ueki R, Nomura H, Itani M, Kaminoh Y. Comparison of tracheal intubation by Macintosh laryngoscope and Pentax-AWS (Airway Scope) during chest compression: a manikin study. J Anesth. 2010;24:306–8.
Maruyama K, Tsukamoto S, Ohno S, Kobayashi K, Nakagawa H, Kitamura A, Hayashida M. Effect of cardiopulmonary resuscitation on intubation using a Macintosh laryngoscope, the AirWay Scope, and the gum elastic bougie: a manikin study. Resuscitation. 2010;81:1014–8.
Koyama J, Iwashita T, Okamoto K. Comparison of three types of laryngoscope for tracheal intubation during rhythmic chest compressions: a manikin study. Resuscitation. 2010;81:1172–4.
Komasawa N, Ueki R, Itani M, Nishi S, Kaminoh Y. Validation of Pentax-AWS Airwayscope utility for intubation device during cardiopulmonary resuscitation on the ground. J Anesth. 2010;24:582–6.
Neumar RW, Otto CW, Link MS, Kronick SL, Shuster M, Callaway CW, Kudenchuk PJ, Ornato JP, McNally B, Silvers SM, Passman RS, White RD, Hess EP, Tang W, Davis D, Sinz E, Morrison LJ. Part 8: Adult advanced cardiovascular life support: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2010;122:S729–67.
Katz SH, Falk JL. Misplaced endotracheal tubes by paramedics in an urban emergency medical services system. Ann Emerg Med. 2001;37:32–7.
Helm M, Hossfeld B, Schafer S, Hoitz J, Lampl L. Factors influencing emergency intubation in the pre-hospital setting: a multicentre study in the German Helicopter Emergency Medical Service. Br J Anaesth. 2006;96:67–71.
Cobas MA, De la Peña MA, Manning R, Candiotti K, Varon AJ. Prehospital intubations and mortality: a level 1 trauma center perspective. Anesth Analg. 2009;109:489–93.
Sawin PD, Todd MM, Traynelis VC, Farrell SB, Nader A, Sato Y, Clausen JD, Goel VK. Cervical spine motion with direct laryngoscopy and orotracheal intubation: an in vivo cinefluoroscopic study of subjects without cervical abnormality. Anesthesiology. 1996;85:26–36.
Maruyama K, Yamada T, Hara K, Nakagawa H, Kitamura A. Tracheal intubation using an AirWay Scope in a patient with Halo-Vest Fixation for upper cervical spine injury. Br J Anaesth. 2009;102:565–6.
Maharaj CH, Higgins B, Harte BH, Laffey JG. Evaluation of ease of intubation with the Airtraq or Macintosh laryngoscope by anaesthetists in easy and simulated difficult laryngoscopy: a manikin study. Anaesthesia. 2006;61:469–77.
Suzuki A, Toyama Y, Katsumi N, Kunisawa T, Sasaki R, Hirota K, Henderson JJ, Iwasaki H. The Pentax-AWS rigid indirect video laryngoscope: clinical assessment of performance in 320 cases. Anaesthesia. 2008;63:641–7.
Koyama J, Aoyama T, Kusano Y, Seguchi T, Kawagishi K, Iwashita T, Okamoto K, Okudera H, Takasuna H, Hongo K. Description and first clinical application of Airway Scope for tracheal intubation. J Neurosurg Anesthesiol. 2006;18:247–50.
Asai T, Liu EH, Matsumoto S, Hirabayashi Y, Seo N, Suzuki A, Toi T, Yasumoto K, Okuda Y. Use of the Pentax-AWS in 293 patients with difficult airways. Anesthesiology. 2009;110:898–904.
Liu L, Tanigawa K, Kusunoki S, Tamura T, Ota K, Yamaga S, Kida Y, Otani T, Sadamori T, Takeda T, Iwasaki Y, Hirohashi N. Tracheal intubation of a difficult airway using Airway Scope, Airtraq, and Macintosh laryngoscope: a comparative manikin study of inexperienced personnel. Anesth Analg. 2010;110:1049–55.
Hirabayashi Y, Seo N. Tracheal intubation by non-anesthesia residents using the Pentax-AWS airway scope and Macintosh laryngoscope. J Clin Anesth. 2009;21:268–71.
Hirabayashi Y. Airway Scope versus Macintosh laryngoscope: a manikin study. Emerg Med J. 2007;24:357–8.
Jordan GM, Silsby J, Bayley G, Cook TM, Difficult Airway S. Evaluation of four manikins as simulators for teaching airway management procedures specified in the Difficult Airway Society guidelines, and other advanced airway skills. Anaesthesia. 2007;62:708–12.
Conflict of interest
The authors have no affiliation with any manufacturer of any device described in the manuscript and declare no financial interest in relation to the material described in the manuscript. Financial support for the study was provided by our institution and department.
Author information
Authors and Affiliations
Corresponding author
About this article
Cite this article
Komasawa, N., Ueki, R., Kohama, H. et al. Comparison of Pentax-AWS Airwayscope video laryngoscope, Airtraq optic laryngoscope, and Macintosh laryngoscope during cardiopulmonary resuscitation under cervical stabilization: a manikin study. J Anesth 25, 898–903 (2011). https://doi.org/10.1007/s00540-011-1218-0
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00540-011-1218-0