Background: Malfunction of peritoneal catheters due to mechanical outflow problems is an annoying complication in patients undergoing chronic peritoneal dialysis (PD). Correction often involves catheter replacement or revision via laparotomy.

Methods: Twenty-five patients undergoing PD who developed mechanical catheter flow restriction underwent 28 laparoscopic procedures. Preoperative diagnoses were made by contrast catheter radiography and were: catheter sequestration (36%), omental wrap (64%). Pneumoperitoneum was induced after general anesthesia and laparoscopy was performed using a Storz laparoscope. The catheter was then identified and manipulation was attempted using instruments placed percutaneously.

Results: In 26 cases (93%), the catheter was freed and function restored. In two cases (7%), adhesions were so numerous and dense that the distal catheter could not be visualized. Four episodes of peritonitis occurred in the perioperative period. Four patients developed subcutaneous leakage of peritoneal fluid which responded to cessation of PD for 2 weeks. Four patients had recurrent occlusions; three of these were managed laparoscopically. Two patients developed late hernias at the site of insertion of the laparoscope. Catheter patency averaged 9.2 months postoperatively.

Conclusions: Laparoscopic revision is a successful technique for salvage of occluded peritoneal catheters.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Received: 27 March 1996/Accepted: 30 May 1996

Rights and permissions

Reprints and permissions

About this article

Cite this article

Amerling, R., Maele, D., Spivak, H. et al. Laparoscopic salvage of malfunctioning peritoneal catheters. Surg Endosc 11, 249–252 (1997). https://doi.org/10.1007/s004649900336

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s004649900336

Navigation