Abstract
Introduction
The widespread adoption of laparoscopic surgery has put new physical demands on the surgeon leading to increased musculoskeletal disorders and injuries. Shoulder, back, and neck pains are among the most common complaints experienced by laparoscopic surgeons. Here, we evaluate the feasibility and efficacy of a non-intrusive progressive arm support exosuit worn by surgeons under the sterile gown to reduce pain and fatigue during surgery.
Methods and procedures
This is a prospective randomized crossover study approved by the Internal Review Board (IRB). The study involves three phases of testing. In each phase, general surgery residents or attendings were randomized to wearing the surgical exosuit at the beginning or at the crossover point. The first phase tests for surgeon manual dexterity wearing the device using the Minnesota Dexterity test, the Purdue Pegboard test, and the Fundamentals of Laparoscopic Surgery (FLS) modules. The second phase tests the effect of the device on shoulder pain and fatigue while operating the laparoscopic camera. The third phase rates surgeon experience in the operating room between case-matched operating days.
Results
Twenty subjects were recruited for this study. Surgeons had the similar dexterity scores and FLS times whether or not they wore the exosuit (p value ranges 0.15–0.84). All exosuit surgeons completed 15 min of holding laparoscopic camera compared to three non-exosuit surgeons (p < 0.02). Exosuit surgeons experienced significantly less fatigue at all time periods and arm pain (3.11 vs 5.88, p = 0.019) at 10 min. Surgeons wearing the exosuit during an operation experienced significant decrease in shoulder pain and 85% of surgeons reported some form of pain reduction at the end of the operative day.
Conclusion
The progressive arm support exosuit can be a minimally intrusive device that laparoscopic surgeons wear to reduce pain and fatigue of surgery without significantly interfering with operative skills or manual dexterity.
Similar content being viewed by others
References
Nilsson G, Larsson S, Johnsson F (2000) Randomized clinical trial of laparoscopic versus open fundoplication: blind evaluation of recovery and discharge period. Br J Surg 87(7):873–878. doi:10.1046/j.1365-2168.2000.01471.x
Mori KM, Neubauer NL (2013) Minimally invasive surgery in gynecologic oncology. ISRN Obstet Gynecol. doi:10.1155/2013/312982
Esposito C, A El Ghoneimi, Yamataka A et al (2013) Work-related upper limb musculoskeletal disorders in paediatric laparoscopic surgery. A multicenter survey. J Pediatr Surg. 48(8):1750–1756. doi:10.1016/j.jpedsurg.2013.01.054
Park A, Lee G, Seagull FJ, Meenaghan N, Dexter D (2010) Patients benefit while surgeons suffer: an impending epidemic. J Am Coll Surg 210(3):306–313. doi:10.1016/j.jamcollsurg.2009.10.017
Szeto GPY, Ho P, Ting ACW, Poon JTC, Cheng SWK, Tsang RCC (2009) Work-related musculoskeletal symptoms in surgeons. J Occup Rehabil 19(2):175–184. doi:10.1007/s10926-009-9176-1
Soueid A, Oudit D, Thiagarajah S, Laitung G (2010) The pain of surgery: pain experienced by surgeons while operating. Int J Surg 8(2):118–120. doi:10.1016/j.ijsu.2009.11.008
Berguer R, Forkey DL, Smith WD (1999) Ergonomic problems associated with laparoscopic surgery. Surg Endosc 13(5):466–468. doi:10.1007/PL00009635
Wauben LSGL, Van Veelen MA, Gossot D, Goossens RHM (2006) Application of ergonomic guidelines during minimally invasive surgery: a questionnaire survey of 284 surgeons. Surg Endosc Other Interv Tech 20(8):1268–1274. doi:10.1007/s00464-005-0647-y
Berguer R, Smith WD, Chung YH (2001) Performing laparoscopic surgery is significantly more stressful for the surgeon than open surgery. Surg Endosc 15(10):1204–1207. doi:10.1007/s004640080030
Reyes DAG, Tang B, Cuschieri A (2006) Minimal access surgery (MAS)-related surgeon morbidity syndromes. Surg Endosc Other Interv Tech 20(1):1–13. doi:10.1007/s00464-005-0315-2
Grandjean E, Hünting W, Pidermann M (1983) VDT workstation design: preferred settings and their effects. Hum Factors 25(2):161–175
Nguyen NTP, Ho HSP, Smith WDP et al (2001) An ergonomic evaluation of surgeons’ axial skeletal and upper extremity movements during laparoscopic and open surgery. Am J Surg 182(6):720–724
Cuschieri A (1995) Whither minimal access surgery: tribulations and expectations. Am J Surg 169(1):9–19. doi:10.1016/S0002-9610(99)80104-4
Uhrich ML, Underwood RA, Standeven JW, Soper NJ, Engsberg JR (2002) Assesment of fatigue, monitor placement, and surgical experience during simulated laparoscopic surgery. Surg Endosc Other Interv Tech 16(4):635–639. doi:10.1007/s00464-001-8151-5
Bergmann G, Graichen F, Bender A et al (2011) In vivo gleno-humeral joint loads during forward flexion and abduction. J Biomech 44(8):1543–1552. doi:10.1016/j.jbiomech.2011.02.142
Gupta S, Van Der Helm FCT (2004) Load transfer across the scapula during humeral abduction. J Biomech 37(7):1001–1009. doi:10.1016/j.jbiomech.2003.11.025
Dubey VN, Agrawal SK (2011) Study of an upper arm exoskeleton for gravity balancing and minimization of transmitted forces. Proc Inst Mech Eng H 225(11):1025–1035. doi:10.1177/0954411911420664
Aitchison LP, Cui CK, Arnold A, Nesbitt-Hawes E, Abbott J (2016) The ergonomics of laparoscopic surgery: a quantitative study of the time and motion of laparoscopic surgeons in live surgical environments. Surg Endosc Other Interv Tech 30(11):5068–5076. doi:10.1007/s00464-016-4855-4
Steenbrink F, Meskers CGM, van Vliet B, Slaman J, Veeger HEJ, De Groot JH (2009) Arm load magnitude affects selective shoulder muscle activation. Med Biol Eng Comput 47(5 Spec. Iss.):565–572. doi:10.1007/s11517-009-0482-8
Zihni AM, Cavallo JA, Ray S, Ohu I, Cho S, Awad MM (2016) Ergonomic analysis of primary and assistant surgical roles. J Surg Res 203(2):301–305. doi:10.1016/j.jss.2016.03.058
Steinhilber B, Seibt R, Reiff F, Rieger MA, Kraemer B, Rothmund R (2016) Effect of a laparoscopic instrument with rotatable handle piece on biomechanical stress during laparoscopic procedures. Surg Endosc Other Interv Tech 30(1):78–88. doi:10.1007/s00464-015-4164-3
Erdelyil A, Sihvonen T, Helin P, Hänninen O (1988) Shoulder strain in keyboard workers and its alleviation by arm supports. Int Arch Occup Environ Health 60(2):119–124. doi:10.1007/BF00381492
Quick NE, Gillette JC, Shapiro R, Adrales GL, Gerlach D, Park AE (2003) The effect of using laparoscopic instruments on muscle activation patterns during minimally invasive surgical training procedures. Surg Endosc 17:462–465. doi:10.1007/s00464-002-8530-6
Julius A, Lees R, Dilley A, Lynn B (2004) Shoulder posture and median nerve sliding. BMC Musculoskelet Disord 5:23. doi:10.1186/1471-2474-5-23
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Disclosures
Shanglei Liu, Daniel Hemming, Ran B. Luo, Jonathan C. Delong, Bryan J. Sandler, Garth R. Jacobsen, and Santiago Horgan have no conflicts of interest or financial ties to disclose.
Rights and permissions
About this article
Cite this article
Liu, S., Hemming, D., Luo, R.B. et al. Solving the surgeon ergonomic crisis with surgical exosuit. Surg Endosc 32, 236–244 (2018). https://doi.org/10.1007/s00464-017-5667-x
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00464-017-5667-x