Abstract
Objective
The purpose of this retrospective study was to evaluate the outcome of medial malleolar fractures treated with magnesium (MgYREZr) bioabsorbable compression screw fixation.
Materials and methods
Eleven patients with a medial malleolar fracture (either isolated or accompanied by bimalleolar or trimalleolar ankle fractures) who were treated with magnesium bioabsorbable compression screws between 2015 and 2016 in our hospital were retrospectively evaluated. Patients were monitored with a mean follow-up of 17.3 ± 4.1 months (range 12–24 months). The mechanism of injury was ground level falls in all patients. All fractures were classified as closed fractures. American Orthopedic Foot and Ankle Society’s (AOFAS) scale and the visual analog scale (VAS) were used to evaluate the clinical results during the final follow-up. Bone union and a possible loss of reduction were assessed with serial radiographs. Potential complications including revision surgery and infection were recorded and reported.
Results
There were 11 patients (4 female, 7 male) with a mean age of 41 ± 21.9 years (range 20–78 years). Six patients had Herscovici type C and five patients had type B fractures. At the final follow-up the mean AOFAS score was 94.9 ± 5.7 points (range 85–100 points) and the mean VAS score was 0.4 ± 1.2 points (range 0–4 points). Radiographic solid union was achieved in all cases. No complications were seen during the follow-up. No patients required implant removal or revision surgery.
Conclusions
This is the first study that investigates the use of bioabsorbable magnesium compression screws in medial malleolar fractures. The results of this study revealed that fixation of medial malleolar fractures with bioabsorbable magnesium compression screws provides adequate fixation with good functional results.
Level of evidence
Level IV, therapeutic, retrospective case series.
Similar content being viewed by others
References
Court-Brown CM, Caesar B (2006) Epidemiology of adult fractures: a review. Injury 37(8):691–697
Herscovici D Jr, Scaduto JM, Infante A (2007) Conservative treatment of isolated fractures of the medial malleolus. J Bone Jt Surg 89B:89–93
Lareau CR, Bariteau JT, Paller DJ, Koruprolu SC, DiGiovanni CW (2015) Contribution of the medial malleolus to tibiotalar joint contact characteristics. Foot Ankle Spec 8:23–28
Hanhisuanto S, Kortekangas T, Pakarinen H, Flinkkilä T, Leskelä HV (2017) The functional outcome and quality of life after treatment of isolated medial malleolar fractures. Foot Ankle Surg 23(4):225–229
Maniar H, Kempegowda H, Tawari AA, Rutter MR, Borade A, Cush G, Horwitz DS (2017) Medial malleoli fractures: clinical comparison between newly designed sled device and conventional screws. Foot Ankle Spec 10(4):296–301
Collinge C, Dombroski D, Heier K (2016) Ankle fractures and dislocations. In: Stannard JP, Schmidt AH (eds) Surgical treatment of orthopaedic trauma. Thieme Publishers, New York, MD, pp 1018–1023
Brown OL, Dirschl DR, Obremskey WT (2001) Incidence of hardware-related pain and its effect on functional outcomes after open reduction and internal fixation of ankle fractures. J Orthop Trauma 15(4):271–274
Naumann MG, Sigurdsen U, Utvåg SE, Stavem K (2016) Incidence and risk factors for removal of an internal fixation following surgery for ankle fracture: a retrospective cohort study of 997 patients. Injury 47(8):1783–1788
Böstman O, Pihlajamäki H (1996) Routine implant removal after fracture surgery: a potentially reducible consumer of hospital resources in trauma units. J Trauma 41(5):846–849
Raikin SM, Ching AC (2005) Bioabsorbable fixation in foot and ankle. Foot Ankle Clin 10(4):667–684
Seitz JM, Lucas A, Kirschner M (2016) Magnesium-based compression screws: a novelty in the clinical use of implants. JOM 68:1177–1182
Plaass C, von Falck C, Ettinger S et al (2017) Bioabsorbable magnesium versus standard titanium compression screws for fixation of distal metatarsal osteotomies—3 year results of a randomized clinical trial. J Orthop Sci. https://doi.org/10.1016/j.jos.2017.11.005
Plaass C, Ettinger S, Sonnow L et al (2016) Early results using biodegradable magnesium screw for modified chevron osteotomies. J Orthop Res 34(12):2207–2214. https://doi.org/10.1002/jor.23241
Windhagen H, Radtke K, Weizbauer A et al. (2013) Biodegradable magnesium-based screw clinically equivalent to titanium screw in hallux valgus surgery: short term results of the first prospective, randomized, controlled clinical pilot study. Biomed Eng Online. 12:62. https://doi.org/10.1186/1475-925X-12-62
Biber R, Pauser J, Brem M, Bail HJ (2017) Bioabsorbable metal screws in traumatology: a promising innovation. Trauma Case Rep (8):11–15
Kellgren JH, Lawrence JS (1957) Radiological assessment of osteoarthrosis. Ann Rheum Dis 16(4):494–502.15
Biber R, Pauser J, Geßlein M, Bail HJ (2016) Magnesium-Based Absorbable Metal Screws for Intra-Articular Fracture Fixation. Case Rep Orthop 2016:9673174
Leonhardt H, Franke A, McLeod NMH, Lauer G, Nowak A (2017) Fixation of fractures of the condylar head of the mandible with a new magnesium-alloy biodegradable cannulated headless bone screw. Br J Oral Maxillofac Surg 55(6):623–625
Meier R, Panzica M (2017) First results with a resorbable MgYREZr compression screw in unstable scaphoid fractures show extensive bone cysts. Handchir Mikrochir Plast Chir 49(1):37–41
Wichelhaus A, Emmerich J, Mittlmeier T (2016) A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws. Case Rep Orthop 2016:7049130
Waizy H, Seitz JM, Reifenrath J et al (2013) Biodegradable magnesium implants for orthopedic applications. J Mater Sci 48:39–5020
Böstman OM (1996) Metallic or absorbable fracture fixation devices. A cost minimization analysis. Clin Orthop Relat Res 329:233–239
Weber BG, Colton C (1991) Malleolar fractures. In: Mueller ME, Allgower M, Schneider R, Willenegger H (eds) Manual of internal fixation: techniques recommended by the AO-ASIF Group, 3rd edn. Springer, New York, pp 595–612
Barnes H, Cannada LK, Watson JT (2014) A clinical evaluation of alternative fixation techniques for medial malleolus fractures. Injury 45(9):1365–1367
Tekin A, Çabuk H, Dedeoğlu SS et al (2016) Anterograde headless cannulated screw fixation in the treatment of medial malleolar fractures: evaluation of a new technique and its outcomes. Med Princ Pract 25(5):429–434
Funding
No funds have been received for this study.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
Authors have no conflict of interest to declare.
Rights and permissions
About this article
Cite this article
Kose, O., Turan, A., Unal, M. et al. Fixation of medial malleolar fractures with magnesium bioabsorbable headless compression screws: short-term clinical and radiological outcomes in eleven patients. Arch Orthop Trauma Surg 138, 1069–1075 (2018). https://doi.org/10.1007/s00402-018-2941-x
Received:
Published:
Issue Date: