Abstract
Background
Minimally invasive, battery-powered drilling systems have become the preferred tool for obtaining representative samples from bone lesions. However, the heat generated during battery-powered bone drilling for bone biopsies has not yet been sufficiently investigated. Thermal necrosis can occur if the bone temperature exceeds a critical threshold for a certain period of time.
Purpose
To investigate heat production as a function of femur temperature during and after battery-powered percutaneous bone drilling in a porcine in vivo model.
Methods
We performed 16 femur drillings in 13 domestic pigs with an average age of 22 weeks and an average body temperature of 39.7 °C, using a battery-powered drilling system and an intraosseous temperature monitoring device. The standardized duration of the drilling procedure was 20 s. The bone core specimens obtained were embedded in 4% formalin, stained with haematoxylin and eosin (H&E) and sent for pathological analysis of tissue quality and signs of thermal damage.
Results
No significant changes in the pigs’ local temperature were observed after bone drilling with a battery-powered drill device. Across all measurements, the median change in temperature between the initial measurement and the temperature measured after drilling (at 20 s) was 0.1 °C. Histological examination of the bone core specimens revealed no signs of mechanical or thermal damage.
Conclusion
Overall, this preliminary study shows that battery-powered, drill-assisted harvesting of bone core specimens does not appear to cause mechanical or thermal damage.
Similar content being viewed by others
Introduction
Bone biopsies are obtained to determine the aetiology of bone lesions by subsequent histological examination. Minimally invasive, percutaneous image-guided biopsy techniques have become the preferred method for obtaining representative samples from bone lesions [1]. In recent years, a device for battery-powered drill-assisted bone tissue harvesting was introduced as a valuable alternative to conventional manual drilling [2,3,4]. Advantages of this battery-powered drill device include higher rotation speed with shorter intervention time, easier application and the possibility to collect larger core samples with less patient discomfort [3].
However, the heat generation during battery-powered single-hole bone drilling for bone sampling has not yet been sufficiently examined. Thermal necrosis has been reported in operative fracture treatment and reconstructive surgery due to an excessive increase in local temperature during extended drilling procedures [4]. If the temperature exceeds a critical threshold for a certain period of time, a denaturing process begins to develop in the tissue, which has a negative effect on the mechanical properties of the bone and may delay or even prevent subsequent bone healing [5]. No final consensus has yet been reached regarding the critical temperature threshold and duration of the procedure for thermal necrosis to occur [6].
With regard to battery-powered drill-assisted bone biopsies, it is currently unclear whether faster drilling with higher rotation speed will lead to greater heating of the needle and the adjacent bone and whether the resulting bone biopsies will show signs of thermal damage. Even during the heating of the battery-powered drilling needle, the tissue sample may be damaged because there is no heat sink effect within the needle.
Therefore, our aim was to investigate heat production as a function of bone temperature during and after battery-powered bone drilling. We used an in vivo porcine model to examine whether battery-powered drill-assisted bone tissue harvesting from the femur causes temperatures in the target tissue that could cause bone tissue damage.
Materials and methods
Animals, housing and care
At our facility, bone drillings (n = 16) were performed on 13 domestic pigs with an average age of 22 weeks. The animals were kept under controlled conditions in the stables of our facility and provided with food and drinking water. All procedures were performed in accordance with the local guidelines and rules for the implementation of the Animal Welfare Act. Before the drilling procedure, the pigs were narcotized and anaesthetized by an intramuscular and subsequently intravenous administration of ketamine, azaperone, xylazine and atropine as previously described by Vahldiek et al. [7]. The pigs received fentanyl as pain medication and a suprapubic catheter was placed in the bladder. Vital parameters (i.e. heart rate, oxygen saturation) were continuously monitored throughout the experiments by a veterinarian. The study and all interventions performed were carried out in accordance with the guidelines and rules of the Federation of Laboratory Animal Science Associations (FELASA) and approved by the National Office for Health and Social Affairs (LaGeSo, Berlin, Germany, G0281/12) and designed according to the ARRIVE guidelines.
Temperature measurement during bone drilling
To ensure temperature monitoring, a hole was drilled percutaneously into the pig’s femur, using a powered bone biopsy device (Arrow® OnControl® Bone Lesion Biopsy System, Teleflex, Shavano Park, TX, USA). The FDA-cleared and CE-certified ARROW® OnControl® battery-powered bone access device is a sealed, hand-held, lithium battery-powered medical device with a driver and biopsy needle set. The needle set consists of two parts—an outer cannula with a length of 102 mm and a bevel-tipped inner stylet, which is used to penetrate the cortex. After removing the bone access needle, we advanced a calibrated 11-gauge temperature monitoring device (EN60751-normed PT-100 with an active 20-mm measuring zone at the needle tip) 3 cm into the bone (Fig. 1). A second drill was placed directly beneath the measuring device (Fig. 1). The median distance between the two probes was 0.96 mm (1.14 mm – 0.78 mm). During drilling, we continuously monitored and recorded the temperature of the tissue at 1-s intervals for a total of 60 s.
Bone drilling procedure
The bone drilling was performed using a battery-powered percutaneous drilling system (Arrow® OnControl® Bone Lesion Biopsy System, Teleflex, Shavano Park, TX, USA). This device operates at a single speed of approximately 420 rotations per minute. The 11-gauge bone access needle (Fig. 2) was inserted into the bone superficially next to and parallel to the previously inserted temperature monitoring probe. For bone tissue sampling, we used a 13-gauge bone lesion biopsy needle to harvest a 3-cm bone cylinder, following 20 s of active drilling in the pig’s femur. This is the drilling time recommended by the manufacturer for single procedures in the instructions for use. The correct parallel needle position and the correct drilling vector were visualized by CT fluoroscopy on an 80-slice multi-detector CT scanner (Aquilion PRIME, Canon Medical Systems, Otawara, Japan) (Fig. 1).
Sample quality check and histological analysis
The bone core specimens obtained were embedded in 4% formalin and sent for pathological analysis of tissue quality and signs of thermal damage. For histological evaluation, the probes were decalcified with 10% ethylenediaminetetraacetic acid (EDTA) [8], conventionally dehydrated and embedded in paraffin wax. Finally, histological sections of 3–5 µm were prepared from the bone samples and stained with haematoxylin and eosin (H&E).
Statistical analysis
The temperatures of each probe were recorded in the PT-100 device and digitally transferred as comma-separated values to a standard PC system. All statistical analysis was performed using “R” version 3.6.3 and the “tidyverse” library [9, 10]. Temperatures were not normally distributed and therefore expressed as medians and interquartile ranges (IQR). Ranges of temperature change were displayed as boxplots, while relative temperature changes over time compared to the initial measurement were displayed as line graphs applying a smoothing factor of 0.25. Temperature measurements were compared using the Wilcoxon rank-sum test. A p-value of < 0.05 was considered statistically significant.
Results
A total of 15 of 16 in vivo femur drillings from 12 domestic pigs were included in the present analysis. One drilling was excluded from further analysis, because the temperature monitoring device was damaged by the drilling device during the drilling procedure, which lead to a distortion of the measured values. The quality and size of all core specimens (n = 15) classified as sufficient by a pathologist. There were no complications during the drilling procedures, all of which were performed with a standardized drilling time of exactly 20 s.
Assessment of drill-related temperature changes within the porcine femur bone
An overview of the measured temperatures in the porcine femora is given in Table 1. The median temperature of the pigs was 39.7 °C at the start of the measurements and 39.9 °C at the end of the measurements (at 60 s), corresponding to a normal porcine body temperature, which is approximately 39 °C with a range of up to 40 °C [11]. During the drilling process, we observed varying temperature changes across our 15 measurements without any significant trend (see Fig. 3). The observed variations in the collected individual measurements are likely due to differences in the body temperature of the pigs, variability in bone properties or deviations of the drilling path at initial contact.
Figure 4 shows the corresponding boxplots for the range of temperature changes for all of our measurements. Each boxplot includes the minimum, the maximum, the sample median, and the first and third quartiles. A maximum change in temperature was observed in measurement number 8 with a maximum change of 1.2 °C. The median change in temperature (across all measurements) between the initial measurement and the temperature measured after drilling (at 20 s) was 0.1 °C and the median change in temperature between the initial and the final measurement after 60 s was also 0.1 °C. One of the pigs (measurements 4 and 5) had a lower body temperature of approximately 37 °C, which can be interpreted in the context of perioperative hypothermia due to an anaesthesia-related heat loss [12].
Across all measurements, no significant changes in temperature were found with a constant median temperature of 39.7 °C at the beginning of and after the drilling procedure (p = 0.84). Also, there was no significant difference between the end of the drilling procedure and the end of the temperature measurement after 60 s (p = 0.78) or between the first and last temperature measurement (p = 0.57). During the drilling process, the measured temperatures showed fluctuations of ± 0.5 °C compared to the initial measurement and started to fluctuate slightly more when the drilling was stopped after 20 s, but this trend was again not significant. Overall, we did not find any significant changes in the pigs’ temperature after bone drilling with a battery-powered drill device.
Histological examination of the core specimens
The obtained core specimens were subsequently seen by a pathologist (Fig. 5). After H&E staining, no signs of mechanical or thermal damage were found. The cortical bone as well as the central spongiosa and especially the cutting edge of the drilling device, where shearing energy, frictional effects and heat are most pronounced, appeared microscopically intact. Figure 5 shows histological sections from the core specimens including both the inner central spongiosa (A1–3) and the spongiosa near the cortical bone (B1–3) with magnifications of 100 × , 50 × and 25 × without any signs of drilling-related thermal damage such as thermal osteonecrosis.
Discussion
Currently, there is a scarcity of in vivo data on heat generation and potential bone damage in battery-powered drill-assisted bone biopsies. After in vivo examination of the femur in pigs, we found that battery-powered, drill-assisted bone biopsies with an active drilling time of 20 s over a total observation period of 60 s showed only a low heating effect of a median of 0.1 °C and did not result in any micro- or macroscopic mechanical or thermal damage.
The bone has a microstructure with several interfaces, e.g. between marrow spaces and mineralized matrix or between mineralized matrix and fibrous tissue [13]. Heat generation follows from plastic deformation and friction adjacent to the drilling device and the bone interface [14]. Depending on the increase in temperature and the exposure time, thermal injuries may occur with changes in bone tissue properties and protein denaturation, desiccation and dehydration, resulting in cell death (thermal necrosis) [5, 15,16,17,18]. Thermal osteonecrosis is thus a direct consequence of the combined effects of increased temperature and the duration of increased temperature [16, 19]. In a study on rabbits, Lundskog et al. observed that a temperature of 55 °C applied for a duration of 30 s caused irreversible damage of bone cells [20]. Eriksson et al. performed a vital microscopic rabbit study and showed that bone tissue loses its function when heated to 47 °C for 1 m [21]. As none of these previous studies was carried out in vivo on human bone tissue, the exact threshold temperature for the death of human bone tissue is still unknown. However, the majority of researchers assume that a temperature of over 47 °C for 1 min can be considered a threshold for thermal osteonecrosis [22].
During bone drilling, the temperature and the level of stress in the bone tissue depend on the drilling parameters, which include speed, feed rate, energy, cooling and depth, with the majority of experimental examinations recommending high speed [22]. Technical advances in bone biopsy systems led to the development of battery-powered devices [23, 24]. Lee et al. reported battery-powered bone drilling using the same device to be safe and effective for biopsy sampling of focal bone lesions [3]. In addition, they observed a reduction of patient pain, shorter intervention time and lower radiation dose as well as an improved user control [3]. Similar results were reported by Berenson et al., who found that bone marrow samples were secured more rapidly and that patients experienced less intermediate-term pain [25]. The results of our porcine in vivo study furthermore suggest that there is no relevant heat development in the bone during drill-assisted bone biopsy sampling and that there is no apparent thermal tissue damage. A surprising finding of our study was a decrease in temperature immediately after drilling in two pigs. What we can imagine is that the drilling device was placed and inserted particularly quickly and easily in these two animals and therefore still had almost room temperature, which could have potentially led to a short-term cooling of the temperature probe.
The diagnostic yield of a bone biopsy can be affected by a multitude of factors. These include the histopathological architecture of the bone lesion, location of the lesion and the presence of mechanical artefacts and thermal osteonecrosis [4]. One limitation of our study is that we investigated a single percutaneous drilling device with a single speed and direction of rotation and a narrow range of possible needle sizes of 10 to 11gauge for bone access and 12 to 13 gauge for biopsy. Another limitation of the present study is that the drilling system was only investigated in healthy porcine bones and not in bones with lytic, sclerotic or suspected infectious lesions. Therefore, the results of the present study may not be directly transferable to humans with lytic or sclerotic bone lesions. Besides, the present study only investigated the effects of an active drilling time of 20 s. In the clinical settings, longer drilling times may be necessary in some patients. For such cases, no conclusion regarding possible temperature development can be drawn from our results. A further shortcoming of the present study is that drilling-induced heat generation was investigated in the living pig femur instead of a human femur, which differ in cortical and trabecular thickness. On the other hand, pigs are considered to be a good non-human model for a better understanding of the human skeleton, with similarities in terms of bone biology and bone mineral density [26][26]. Finally, it is important to note that our study design differs from applications in orthopaedic surgery, where bone drilling procedures are mainly used for bone reconstruction surgery and to insert implants.
Conclusions
Overall, this preliminary study shows that battery-powered, drill-assisted harvesting of bone core specimens does not appear to cause mechanical or thermal damage, if the procedure is limited to an active drilling time of 20 s. As we only investigated normal bone tissue in pigs, the results may not be directly transferable to pathological bone lesions.
Data availability
All data are available within the manuscript and the supplementary material.
Change history
07 November 2021
Updated due to the missing funding information.
References
Ng CS, Salisbury JR, Darby AJ, Gishen P. Radiologically guided bone biopsy: results of 502 biopsies. Cardiovasc Intervent Radiol. 1998;21(2):122–8.
Cohen SC, Gore JM. Evaluation of a powered intraosseous device for bone marrow sampling. Anticancer Res. 2008;28(6B):3843–6.
Lee RK, Ng AW, Griffith JF. CT-guided bone biopsy with a battery-powered drill system: preliminary results. AJR Am J Roentgenol. 2013;201(5):1093–5.
Kihira S, Koo C, Lee A, Aggarwal A, Pawha P, Doshi A. Reduction of radiation dose and scanning time while preserving diagnostic yield: a comparison of battery-powered and manual bone biopsy systems. AJNR Am J Neuroradiol. 2020;41(3):387–92.
Lee J, Ozdoganlar OB, Rabin Y. An experimental investigation on thermal exposure during bone drilling. Med Eng Phys. 2012;34(10):1510–20.
Sui J, Wang C, Sugita N. Experimental study of temperature rise during bone drilling process. Med Eng Phys. 2020;78:64–73.
Vahldiek JL, Erxleben C, Bressem KK, Gemeinhardt O, Poch F, Hiebl B, et al. Multipolar RFA of the liver: influence of intrahepatic vessels on ablation zones and appropriateness of CECT in detecting ablation dimensions-results of an in-vivo porcine liver model. Clin Hemorheol Microcirc. 2018;70(4):467–76.
Xue J, Peng J, Yuan M, Wang A, Zhang L, Liu S, et al. NELL1 promotes high-quality bone regeneration in rat femoral distraction osteogenesis model. Bone. 2011;48(3):485–95.
Team RC. R: A language and environment for statistical computing. 2013.
Wickham H. Tidyverse: easily install and load the ‘tidyverse.’ R package version. 2017;1(1):2017.
Dent BT, Stevens KA, Clymer JW. Forced-air warming provides better control of body temperature in porcine surgical patients. Veterinary Sciences. 2016;3(3):22.
Diaz M, Becker DE. Thermoregulation: physiological and clinical considerations during sedation and general anesthesia. Anesth Prog. 2010;57(1):25–32 (quiz 33–24).
Aghvami M, Brunski JB, Serdar Tulu U, Chen CH, Helms JA. A thermal and biological analysis of bone drilling. J Biomech Eng. 2018; 140(10).
Staroveski T, Brezak D, Udiljak T. Drill wear monitoring in cortical bone drilling. Med Eng Phys. 2015;37(6):560–6.
Bachus KN, Rondina MT, Hutchinson DT. The effects of drilling force on cortical temperatures and their duration: an in vitro study. Med Eng Phys. 2000;22(10):685–91.
Palmisano AC, Tai BL, Belmont B, Irwin TA, Shih A, Holmes JR. Heat accumulation during sequential cortical bone drilling. J Orthop Res. 2016;34(3):463–70.
Udiljak T, Ciglar D, Skoric S. Investigation into bone drilling and thermal bone necrosis. Advances in Production Engineering & Management. 2007;2(3):103–12.
Augustin G, Zigman T, Davila S, Udilljak T, Staroveski T, Brezak D, et al. Cortical bone drilling and thermal osteonecrosis. Clin Biomech (Bristol, Avon). 2012;27(4):313–25.
Augustin G, Davila S, Udilljak T, Staroveski T, Brezak D, Babic S. Temperature changes during cortical bone drilling with a newly designed step drill and an internally cooled drill. Int Orthop. 2012;36(7):1449–56.
Lundskog J. Heat and bone tissue. An experimental investigation of the thermal properties of bone and threshold levels for thermal injury. Scand J Plast Reconstr Surg. 1972;9:1–80.
Eriksson AR, Albrektsson T. Temperature threshold levels for heat-induced bone tissue injury: a vital-microscopic study in the rabbit. J Prosthet Dent. 1983;50(1):101–7.
Pandey RK, Panda SS. Drilling of bone: a comprehensive review. J Clin Orthop Trauma. 2013;4(1):15–30.
Bucher CM, Lehmann T, Tichelli A, Tzankov A, Dirnhofer S, Passweg J, et al. Comparison of a powered bone marrow biopsy device with a manual system: results of a prospective randomised controlled trial. J Clin Pathol. 2013;66(1):24–8.
Miller LJ, Philbeck TE, Montez DF, Puga TA, Brodie KE, Cohen SC, et al. Powered bone marrow biopsy procedures produce larger core specimens, with less pain, in less time than with standard manual devices. Hematol Rep. 2011;3:1-e8.
Berenson JR, Yellin O, Blumenstein B, Bojanower D, Croopnick J, Aboulafia D, et al. Using a powered bone marrow biopsy system results in shorter procedures, causes less residual pain to adult patients, and yields larger specimens. Diagn Pathol. 2011;6(1):1–6.
Ben-Zvi Y, Reznikov N, Shahar R, Weiner S. 3D architecture of trabecular bone in the pig mandible and femur: inter-trabecular angle distributions. Frontiers in Materials. 2017;4:29.
Funding
Open Access funding enabled and organized by Projekt DEAL. This work was supported by a grant from the Deutsche Forschungsgemeinschaft, Ref.-No. RI1131/3–3. MRM is grateful for support from the Deutsche Forschungsgemeinschaft (DFG, SFB 1340/1 2018, 5943/31/41/91). LCA is grateful for her participation in the BIH Charité – Junior Clinician and Clinician Scientist Program funded by the Charité – Universitaetsmedizin Berlin and the Berlin Institute of Health.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare no competing interests.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Niehues, S.M., Elezkurtaj, S., Bresssem, K.K. et al. Evaluation of potential tissue heating during percutaneous drill-assisted bone sampling in an in vivo porcine study. Skeletal Radiol 51, 829–836 (2022). https://doi.org/10.1007/s00256-021-03890-w
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00256-021-03890-w