Abstract
Objective
This study evaluated the use of lacrimal fluid glucose concentrations as a minimally invasive, alternative sampling strategy for monitoring glucose concentrations in surgical/trauma ICU patients.
Design and setting
Prospective, paired sample study in an adult surgical/trauma ICU.
Patients
Patients receiving subcutaneous or intravenous insulin requiring routine capillary blood glucose measurements. Patients receiving ocular lubricants, artificial tears, or routinely administered ophthalmic medications and patients with facial injuries were excluded.
Interventions
Lacrimal fluid was collected using glass capillary tube placed near the cul-de-sac of the eye. Capillary blood glucose was determined using a bedside glucose meter as per routine ICU care.
Measurements and results
Lacrimal fluid glucose concentration was analyzed using high-performance liquid chromatography with pulse amperometric detection. Forty-four paired samples from five patients were analyzed. Pearson correlation between lacrimal fluid (µM) and blood glucose (mM) concentrations and the proportional change from baseline revealed no significant associations. Due to the very poor association, enrollment was discontinued after five patients.
Conclusions
Lacrimal fluid and blood glucose concentrations were poorly correlated, suggesting that the former is not a reliable alternative to blood glucose monitoring in surgical/trauma ICU patients requiring insulin therapy.
References
McCowen K, Malhotra A, Bistrian B (2001) Stress-Induced hyperglycemia. Crit Care Clin 17:107–124
Berghe G van den, Wouters P, Weekers F, Verwaest C, Bruyninckx F, Schetz M, Vlasselaers D, Ferdinande P, Lauwers P, Bouillon R (2001) Intensive insulin therapy in the critically ill patients. N Engl J Med 345:1359–1367
Golden S, Peart-Vigilance C, Kao W, Brancati F (1999) Perioperative Glycemic Control and the Risk of Infectious Complications in a Cohort of Adults With Diabetes. Diabetes Care 22:1408–1414
Furnary AP, Zerr KJ, Grunkemeier GL, Starr A (1999) Continuous intravenous insulin infusion reduces the incidence of deep sternal wound infection in diabetic patients after cardiac surgical procedures. Ann Thorac Surg 67:352–360–
Malmberg K, for the DIGAMI (Diabetes Mellitus Insulin Glucose Infusion in Acute Myocardial Infarction) Study Group (1997) Prospective randomised study of intensive insulin treatment on long term survival after acute myocardial infarction in patients with diabetes mellitus. BMJ 314:1512–1515
Tamada JA, Garg S, Jovanovic L, Pitzer KR, Fermi S, Potts RO (1999) Noninvasive glucose monitoring: comprehensive clinical results. JAMA 282:1839–1844
Wilson D, Potts RO, Ackerman NR, Fermi S (2000) Human acute skin irritation and sensitization potentials, in vivo: GlucoWatch Biographer. Diabetes 49:A129
Garg SK, Potts RO, Ackerman NR, Fermi SJ, Tamada JA, Chase HP (1999) Correlation of fingerstick blood glucose measurements with GlucoWatch biographer glucose results in young subjects with type 1 diabetes. Diabetes Care 22:1708–1714
Chico A, Vidal-Rios P, Subira M, Novials A (2003) The continuous glucose monitoring system is useful for detecting unrecognized hypoglycemias in patients with type 1 and type 2 diabetes but is not better than frequent capillary glucose measurements for improving metabolic control. Diabetes Care 26:1153–1157
Gasset A, Braverman L, Fleming M, Arky R, Alter B (1986) Tear glucose detection of hyperglycemia. Am J Ophthalmol 65:414–420
Gaur M, Sharma G, Kabra S, Sharma P, Nepalia L (1982) Tear glucose in uncontrolled and chemical diabetics. Indian J Ophthalmol 30:367–369
Chatterjee P, De S, Datta H, Chatterjee SM, Biswas M, Sarkar K, Mandal L (2003) Estimation of tear glucose level and its role as a prompt indicator of blood sugar level. J Indian Med Assoc 10:481–483
Chen R, Jin Z, Colon LA (1996) Analysis of tear fluid by CE/LIF: a noninvasive approach for glucose monitoring. J Capillary Electrophor 3:243–248
Farine S, Villard C, Moulin A, Mouren G, Puigserver A (1997) Comparative quantitative analysis of sucrose and related compounds using ion exchange and reversed phase chromatographic methods. Int J Biol Macromol 21:109–114
Lee Y (1996) Carbohydrate analyses with high-performance anion-exchange chromatography. J Chromatogr A 720:137–149
Sen D, Sarin G (1980) Tear glucose levels in normal people and in diabetic patients. Br J Ophthalmol 64:693–695
Das B, Sengupta S, Das B, Goswami N (1995) Tear glucose estimation—an alternative to blood glucose estimation. J Indian Med Assoc 93:127–128
Boyne MS, Silver DM, Kaplan J, Saudek CD (2003) Timing of changes in interstitial and venous blood glucose measured with a continuous subcutaneous glucose sensor. Diabetes 52:2790–2794
Stout P, Peled N, Erickson B, Hilgers M, Racchini J, Hoegh T (2001) Comparison of glucose levels in dermal interstitial fluid and finger capillary blood. Diabetes Technol Ther 3:81–90
Acknowledgements
We thank Joseph Beres for his hard work and dedication to this project.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
LeBlanc, J.M., Haas, C.E., Vicente, G. et al. Evaluation of lacrimal fluid as an alternative for monitoring glucose in critically ill patients. Intensive Care Med 31, 1442–1445 (2005). https://doi.org/10.1007/s00134-005-2747-5
Received:
Accepted:
Published:
Issue Date: