Prefraction displacement and intrafraction drift of the prostate due to perineal ultrasound probe pressure

Präfraktionelle Verschiebung und intrafraktionelle Drift der Prostata durch Druck perinealer Ultraschallköpfe

  • Original Article
  • Published:
Strahlentherapie und Onkologie Aims and scope Submit manuscript

A Data Descriptor to this article was published on 11 November 2019

Abstract

Objective

In image-guided EBRT of the prostate, transperineal ultrasound (US) probes exert pressure on the perineum both during planning and treatment. Through tissue deformation and relaxation, this causes target and risk organ displacement and drift. In this study, prefraction shift and intrafraction drift of the prostate are quantified during robotic transperineal 4DUS.

Methods

The position of the prostate was recorded for different positions of the probe before treatment in 10 patients (16 series of measurements). During treatment (15 patients, 273 fractions), intrafraction motion of the prostate was tracked (total of 27 h and 24 min) with the transperineal probe in place.

Results

Per 1 mm shift of the US probe in the cranial direction, a displacement of the prostate by 0.42 ± 0.09 mm in the cranial direction was detected. The relationship was found to be linear (R² = 0.97) and highly significant (p < 0.0001). After initial contact of the probe and the perineum (no pressure), a shift of the probe of about 5–10 mm was typically necessary to achieve good image quality, corresponding to a shift of the prostate of about 2–4 mm in the cranial direction. Tissue compression and prostate displacement were well visible. During treatment, the prostate drifted at an average rate of 0.075 mm/min in the cranial direction (p = 0.0014).

Conclusion

The pressure applied by a perineal US probe has a quantitatively similar impact on prostate displacement as transabdominal pressure. Shifts are predominantly in the cranial direction (typically 2–4 mm) with some component in the anterior direction (typically <1 mm). Slight probe pressure can improve image quality, but excessive probe pressure can distort the surrounding anatomy and potentially move risk organs closer to the high-dose area.

Zusammenfassung

Zielsetzung

In der bildgeführten Strahlentherapie der Prostata üben perineale Ultraschallköpfe während Planung und Behandlung Druck auf das Perineum aus. Durch Gewebedeformation verursacht dies Verschiebungen von Zielvolumen und Risikoorganen. In dieser Studie werden Verschiebungen vor und Relaxationen während der Behandlung unter transperinealem orts- und zeitaufgelöstem Ultraschall (US) quantifiziert.

Methoden

Vor der Behandlung (10 Patienten, 16 Messreihen) wurde die Lage der Prostata bei verschiedenen Schallkopfpositionen aufgezeichnet. Während der Behandlung (15 Patienten, 273 Fraktionen) mit anliegender perinealer Probe wurde die intrafraktionelle Bewegung der Prostata aufgezeichnet (insgesamt 27 h 24 min).

Ergebnisse

Pro 1 mm Verschiebung des Schallkopfs nach kranial verschob sich die Prostata um 0,42 ± 0,09 mm, ebenfalls in kranialer Richtung. Der Zusammenhang war linear (R² = 0,97) und hoch signifikant (p < 0,0001). Nach drucklosem Kontakt des US-Kopfs war für eine gute Bildqualität eine Verschiebung in das Perineum um typischerweise 5–10 mm notwendig, was einer Verschiebung der Prostata von etwa 2–4 mm in kranialer Richtung entspricht. Gewebedeformation und Prostataverschiebung waren deutlich sichtbar. Während der Behandlung driftete die Prostata mit einer mittleren Rate von 0,075 mm/min in kranialer Richtung (p = 0,0014).

Schlussfolgerung

Der Druck des perinealen Schallkopfs hat ähnlich großen Einfluss auf die Lage der Prostata, wie derjenige eines abdominellen. Verschiebungen geschehen hauptsächlich in kranialer Richtung (typisch 2–4 mm) mit einer geringen Komponente in ventraler Richtung (typisch <1 mm). Mäßiger Druck des Schallkopfs kann die Bildqualität verbessern, übergroßer Druck jedoch die umliegende Anatomie verformen und potentiell Risikoorgane in Regionen höherer Dosis verschieben.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Ballhausen H, Hieber S, Li M, Belka C, Reiner M (2014) Millimeter precision in ultrasound based patient positioning: experimental quantification of inherent technical limitations. Med Phys 41:081718

    Article  Google Scholar 

  2. Ballhausen H, Hieber S, Li M, Parodi K, Belka C, Reiner M (2015) Linearity of patient positioning detection : a phantom study of skin markers, cone beam computed tomography, and 3D ultrasound. Strahlenther Onkol 191:442–447

    Article  Google Scholar 

  3. Li M, Ballhausen H, Hegemann NS, Ganswindt U, Manapov F, Tritschler S et al (2015) A comparative assessment of prostate positioning guided by three-dimensional ultrasound and cone beam CT. Radiat Oncol 10:82

    Article  CAS  Google Scholar 

  4. Ballhausen H, Ballhausen BD, Lachaine M, Li M, Parodi K, Belka C et al (2015) Surface refraction of sound waves affects calibration of three-dimensional ultrasound. Radiat Oncol 10:119

    Article  Google Scholar 

  5. McGahan JP, Ryu J, Fogata M (2004) Ultrasound probe pressure as a source of error in prostate localization for external beam radiotherapy. Int J Radiat Oncol Biol Physics 60:788–793

    Article  Google Scholar 

  6. Dobler B, Mai S, Ross C, Wolff D, Wertz H, Lohr F et al (2006) Evaluation of possible prostate displacement induced by pressure applied during transabdominal ultrasound image acquisition. Strahlenther Onkol 182:240–246

    Article  Google Scholar 

  7. Fargier-Voiron M, Presles B, Pommier P, Rit S, Munoz A, Liebgott H et al (2014) Impact of probe pressure variability on prostate localization for ultrasound-based image-guided radiotherapy. Radiother Oncol 111:132–137

    Article  Google Scholar 

  8. Treece GM, Prager RW, Gee AH, Berman L (2002) Correction of probe pressure artifacts in freehand 3D ultrasound. Med Image Anal 6:199–214

    Article  CAS  Google Scholar 

  9. Treece GM, Gee AH, Prager RW (2005) RF and amplitude-based probe pressure correction for 3D ultrasound. Ultrasound Med Biol 31:493–503

    Article  Google Scholar 

  10. Harris EJ, Symonds-Taylor R, Treece GM, Gee AH, Prager RW, Brabants P et al (2009) Evaluation of a three-dimensional ultrasound localisation system incorporating probe pressure correction for use in partial breast irradiation. Br J Radiol 82:839–846

    Article  CAS  Google Scholar 

  11. Lachaine M, Falco T (2013) Intrafractional prostate motion management with the Clarity autoscan system. Med Phys Int 1:72–80

    Google Scholar 

  12. Wang KK, Vapiwala N, Deville C, Plastaras JP, Scheuermann R, Lin H et al (2012) A study to quantify the effectiveness of daily endorectal balloon for prostate intrafraction motion management. Int J Radiat Oncol Biol Physics 83:1055–1063

    Article  Google Scholar 

  13. Smeenk RJ, Louwe RJ, Langen KM, Shah AP, Kupelian PA, van Lin EN et al (2012) An endorectal balloon reduces intrafraction prostate motion during radiotherapy. Int J Radiat Oncol Biol Physics 83:661–669

    Article  Google Scholar 

Download references

Acknowledgements

We thank Andrea Beisel, Gabriela Danilkiewicz, Sandra Kohlhauser, and Anja Weber for their excellent technical assistance.

Funding

Funding for research with the Clarity system has been received from Elekta. Elekta was not involved in and had no influence on the study design, the collection, analysis, or interpretation of data, on the writing of the manuscript or the decision to submit the manuscript for publication.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Hendrik Ballhausen.

Ethics declarations

Conflict of interest

Elekta Germany supports research at the university hospital of Ludwig-Maximilians-Universität, chaired by Professor Belka. Elekta supported various congress presentations by C. Belka and M. Li. N.-S. Hegemann, F. Manapov, A. Kolberg, P.D. Thum, U. Ganswindt and H. Ballhausen declare that they have no competing interests.

Ethical standards

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. Informed consent was obtained from all individual participants included in the study.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Li, M., Hegemann, NS., Manapov, F. et al. Prefraction displacement and intrafraction drift of the prostate due to perineal ultrasound probe pressure. Strahlenther Onkol 193, 459–465 (2017). https://doi.org/10.1007/s00066-017-1105-1