The human temporomandibular joint (TMJ) is the diarthroidial articulation of the mandible to the temporal bone. The essential life functions of mastication, speech, airway support, and deglutition are supported by TMJ function and form. This places the TMJ complex under more cyclical loading and unloading than any other body joint over a lifetime [1].

End-stage disease is the very worst condition of an organ or disease state. At this point, the organ is barely functioning. Examples include end-stage kidney disease where the kidneys have shut down and the patient requires dialysis, or end-stage cardiac disease which means the heart is functioning very poorly and may need mechanical support or transplantation in order for the patient to survive.

Applying this term to joints, end-stage joint disease connotes a joint that is so negatively affected architecturally by disease or injury that it results in severe functional impairment for the patient.

As with all other joints, the TMJ is affected by all of the end-stage joint diseases resulting from developmental disorders, neoplasia, trauma, arthritic disease, failed prior joint surgery, or ankylosis, and require replacement. Therefore, to provide long-term safe and effective outcomes, any TMJ total joint replacement (TJR) device must be able to manage the anatomical, operational, and esthetic discrepancies presented, withstand the stresses and strains of functional loading, while exhibiting minimal wear-related concerns [2].

There are two categories of TMJ TJR devices approved by the Food and Drug Administration for use in the United States. First are stock or “off-the-shelf” devices, which the surgeon has to make fit at implantation (Biomet Microfixation TMJ Replacement System, Jacksonville, FL, USA; Nexus CMF, Salt Lake City, UT, USA). Second are custom or “patient-fitted” devices which are made to fit each specific case utilizing CAD/CAM technology (TMJ Concepts, Ventura, CA, USA; Nexus CMF, Salt Lake City, UT, USA) (Fig. 1).

Fig. 1
figure 1

a Stock metal-on-polyethylene device (Biomet Microfixation, Jacksonville, FL); b Stock metal-on-metal device (Nexus CMF, Salt Lake City, UT); c Custom metal-on-metal device (Nexus CMF, Salt Lake City, UT); d Patient-fitted metal-on-polyethylene TMJ Concepts device (TMJ Concepts, Ventura, CA—image credit—Hunterian Museum, Royal College of Surgeons of England, London, UK)

In the orthopedic literature, it has become evident that some THR failures are related to material breakdown due to wear and corrosion leading to metal ion release. The overall poor performance of such joints is the result of a local and/or system response to the combined effect of both wear and corrosion [38].

In a description of the general requirements for joint replacement devices, Katti emphasized the importance of device material biocompatibility, with no inflammatory or toxic response to wear beyond a tolerable level, the appropriate mechanical properties for the desired application, and lastly economically viable manufacturing and processing methods [9].

Implicit in these requirements is the importance of understanding wear and failure mechanisms of implanted devices. However, compared to orthopedic TJR devices, functional wear failure mechanisms for TMJ TJR implants have not been clearly defined. This may be because there are exponentially fewer TMJ TJR devices implanted annually than orthopedic TJR devices. What is understood is that not only must wear and corrosion influence TJR implant in vivo longevity, but also peri-implant tissue reactivity to these two phenomena.

Our research group has started initial translational investigations involving the analysis of failed retrieved TMJ TJR devices alloy microstructure compared to control, never implanted, TMJ TJR devices utilizing established orthopedic TJR device retrieval wear, and corrosion evaluation protocols.

Thirty-one failed TMJ TJR samples were obtained from two independent sources, a group of international TMJ surgeons and the retrieval collection at the TMJ Implant Registry and Repository located at the University of Minnesota. The inventory was comprised three failed retrieved TMJ TJR groups of devices: Group 1—Metal-on-metal (MoM): consisted of 19 failed retrieved metal (CoCrMo) condyle-on-metal (CoCrMo) fossa devices. Group 2—Polymethymethacrylate-on-metal (PoM): consisted of a total of 7 failed retrieved polymethylmethacrylate (PMMA) condyle-on-metal (CoCrMo) fossa devices. Group 3 consisted of 2 titanium-nitride (TiN)-coated condyle-on-fossa devices. In addition to the study failed retrieved TMJ TJR devices, a control Group 4, consisting of 3 never implanted MoM-type CoCrMo TMJ TJR devices, was included.

Microscopic images using the SmartScope automatic measuring system, white light interferometry (WLI), scanning electron microscopy (SEM), and Raman spectroscopy were conducted to characterize the surfaces of the device glenoid fossa (cup) and mandibular condyle (ball) components in each group where applicable.

Substantial surface damage was observed in the articulating areas between the condyle and the fossa components in all retrieval Groups. Damage included pitting corrosion, evidence of deposited corrosion products, specific wear patterns, hard phases, surface depressions, and bi-directional scratches, which are similar to the findings of metal-on-metal failed retrieved THR devices.

Electrochemical testing was performed on Group 1 and 3 retrieval samples and the Group 4 control. Electrochemical impedance spectroscopy (EIS) confirmed material properties as well as corrosion kinetics in vivo help mitigate corrosion as reflected by the Raman spectroscopy results.

In summary, this study demonstrated the potential role of wear and corrosion interactions on the failure of TMJ TJR devices. Further studies will include in vitro investigation of corrosion kinetics and the synergistic interactions of wear and corrosion and their specific roles in the underlying tribocorrosion mechanisms of TMJ TJR devices. The generation and distribution of corrosion products and wear particles due to the tribocorrosion processes will also be considered as it has several clinical implications and potentials sources of many side effects on the patients [10, 11].

This and future studies will guide future material choices and functional design improvements for TMJ TJR devices. Orthopedic TJR implant schemes may also be improved by understanding the degradation mechanism of TMJ TJR implants, as the materials employed in both TJR devices are similar.