Skip to main content
Log in

Characterization of sub-systems of a molecular biowire-based biosensor device

  • Original Paper
  • Published:
Microchimica Acta Aims and scope Submit manuscript

Abstract.

We describe the fabrication and characterization of components of a biosensor system that can be used for simultaneous screening of multiple pathogens in a sample. The two sub-systems are: (1) a mixed antibody immunosensor based on molecular biowires and (2) a multi-channel potentiostat for measuring conductance across the immunosensor. The immunosensor operates by converting binding events between antigen and antibody into a measurable electrical signal using polyaniline nanowires as labels and transducers. By mixing and patterning antibodies on a substrate, the response of the biosensor can detect the presence of either one of several pathogens present in the analyte through an equivalent change in conductance. The paper also presents a 42-channel potentiostat array which can be used for measuring the conductance change across the immunosensor. Each channel of the array uses a novel semi-synchronous ΣΔ modulator which enables detection down to femtoampere current ranges.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • InstitutionalAuthorNameFood and Drug Administration (2000) Bacteriological analytical manual EditionNumber8 Association of Analytical Chemists Arlington, VA

    Google Scholar 

  • InstitutionalAuthorNameIUPAC (1999) ArticleTitleElectrochemical biosensors: recommended definitions and classification Pure Appl Chem 71 2333 Occurrence Handle10.1351/pac199971122333

    Article  Google Scholar 

  • Z Muhammad-Tahir E C Alocilja (2003) ArticleTitleA conductometric biosensor for biosecurity Biosen Bioelectron 18 813 Occurrence Handle10.1016/S0956-5663(03)00020-4 Occurrence Handle1:CAS:528:DC%2BD3sXivV2kt7Y%3D

    Article  CAS  Google Scholar 

  • Z Muhammad-Tahir E C Alocilja (2003) ArticleTitleFabrication of a disposable biosensor for Escherichia coli o157:H7 detection IEEE sensor J 3 345 Occurrence Handle10.1109/JSEN.2003.815782 Occurrence Handle1:CAS:528:DC%2BD3sXnslKnt7g%3D

    Article  CAS  Google Scholar 

  • R Wiese Y Belosludtsev T Powdrill P Thompson M Hogan (2001) ArticleTitleSimultaneous multianalyte ELISA performed on a microarray platform Clini Chem 47 1451 Occurrence Handle1:CAS:528:DC%2BD3MXls1Squrc%3D

    CAS  Google Scholar 

  • C R Taitt J P Golden Y S Shubin L C Shriver-Lake K E Sapsford A Rasooly F S Ligler (2004) ArticleTitleA portable array biosensor for detecting multiple analytes in complex samples Microbial Ecol 47 175 Occurrence Handle10.1007/s00248-003-1011-1 Occurrence Handle1:CAS:528:DC%2BD2cXksFGntLY%3D

    Article  CAS  Google Scholar 

  • D Malamud H Bau S Niedbala P Corstjens (2005) ArticleTitlePoint detection of pathogens in oral samples Adv Dent Res 18 12 Occurrence Handle1:STN:280:DC%2BD2Mzks1altw%3D%3D Occurrence Handle10.1177/154407370501800104

    Article  CAS  Google Scholar 

  • Y Gao G Hu F Y H Lin P M Sherman D Li (2005) ArticleTitleAn electrokinetically-controlled immunoassay for simultaneous detection of multiple microbial antigens Biomed Microdevices 7 301 Occurrence Handle10.1007/s10544-005-6072-0 Occurrence Handle1:CAS:528:DC%2BD28XjtVygtw%3D%3D

    Article  CAS  Google Scholar 

  • K Murari M Stanacevic G Cauwenberghs N V Thakor (2005) ArticleTitleIntegrated potentiostat for neurotransmitter sensing IEEE Eng Med Biol Mag 24 23 Occurrence Handle10.1109/MEMB.2005.1549726

    Article  Google Scholar 

  • H S Narula J G Harris (2004) ArticleTitleVLSI potentiostat for amperometric measurements for electrolytic reaction ISCAS 1 457

    Google Scholar 

  • R D Beach R W Conlan M C Godwin F Moussy (2005) ArticleTitleTowards a miniature implantable in vivo telemetry monitoring system dynamically configurable as a potentiostat or galvanostat for two- and three-electrode biosensors IEEE T Instrum Meas 54 61 Occurrence Handle10.1109/TIM.2004.839757

    Article  Google Scholar 

  • A Gore S Chakrabartty S Pal E C Alocilja (2006) ArticleTitleA multichannel femtoampere-sensitivity potentiostat array for biosensing applications IEEE T Circ Syst 53 2357 Occurrence Handle10.1109/TCSI.2006.884432

    Article  Google Scholar 

  • P R Kinget A A Lazar L T Toth (2005) ArticleTitleOn the robustness of an analog VLSI implementation of a time encoding machine ISCAS 5 4221

    Google Scholar 

  • B Razavi (1995) Principles of data conversion system design Wiley New York 220

    Google Scholar 

  • B Linares-Barranco T Serrano-Gotarredona (2003) ArticleTitleOn the design and characterization of femtoampere current-mode circuits IEEE J Solid St Circ 38 1353 Occurrence Handle10.1109/JSSC.2003.814415

    Article  Google Scholar 

  • InstitutionalAuthorNameFDA-Food and Drug Administration (2000) Bacteriological analytical manual AOAC International Gaithersburg, MD

    Google Scholar 

  • Michigan State University (1998) Biosafety manual. Office of Radiation, Chemical, and Biological Safety

  • Y Liu S Chakrabartty E C Alocilja (2007) ArticleTitleFundamental building blocks for molecular bio-wire based forward-error correcting biosensors Nanotech 18 424017 Occurrence Handle10.1088/0957-4484/18/42/424017 Occurrence Handle1:CAS:528:DC%2BD2sXhsVartbfN

    Article  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Yang Liu.

Additional information

Correspondence: Yang Liu, Michigan State University, Department of Electrical and Computer Engineering, East Lansing, MI 48824, USA

Rights and permissions

Reprints and permissions

About this article

Cite this article

Liu, Y., Gore, A., Chakrabartty, S. et al. Characterization of sub-systems of a molecular biowire-based biosensor device. Microchim Acta 163, 49–56 (2008). https://doi.org/10.1007/s00604-008-0950-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00604-008-0950-0

Navigation