Ocular anatomic changes for different accommodative demands using swept-source optical coherence tomography: a pilot study

  • Cornea
  • Published:
Graefe's Archive for Clinical and Experimental Ophthalmology Aims and scope Submit manuscript

Abstract

Purpose

The purpose of our study was to assess the changes in ocular parameters for different accommodative demands using a new optical biometer based on swept-source optical coherence tomography (SS-OCT).

Methods

Seven subjects were included in this pilot study, and only one eye per participant was analyzed. Each eye was measured six times with the optical biometer IOLMaster 700 (Carl Zeiss Meditec, Jena, Germany). As this instrument is not able to change the vergence of the stimulus, to enable measurements at different accommodative states, a tilted first-surface mirror attached to the optical biometer was used to place the fixation stimulus at different vergences. Measurements were taken on the right eye of the subject while the left eye was looking through the mirror. Central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), axial length (AL), white-to-white (WTW) distance and keratometric readings were evaluated for three different accommodative states: 0.0 diopters (D), 1.5 D, and 3.0 D.

Results

No statistically significant differences were found for CCT, AL, WTW, K1 and K2 between the three accommodative states. As expected, changing the accommodative condition did not change CCT, AL, WTW, and keratometric outcomes. Nevertheless, statistically significant differences between the accommodative states were found for ACD and LT measurements. In addition, variations in ACD correlated linearly with variations in LT (R2 ≥ 0.99) when changing the vergence of the optotype.

Conclusion

A practical methodology to assess the changes in ocular parameters for different accommodative demands using the IOLMaster 700 based on SS-OCT has been described. Statistically significant changes that have been found that agree well with previous reports.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. Montés-Micó R, Carones F, Buttacchio A, Ferrer-Blasco T, Madrid-Costa D (2011) Comparison of immersion ultrasound, partial coherence interferometry, and low coherence reflectometry for ocular biometry in cataract patients. J Refract Surg 27:665–671

    Article  PubMed  Google Scholar 

  2. Srivannaboon S, Chirapapaisan C, Chonpimai P, Loket S (2015) Clinical comparison of a new swept-source optical coherence tomography-based optical biometer and a time-domain optical coherence tomography-based optical biometer. J Cataract Refract Surg 41:2224–2232

    Article  PubMed  Google Scholar 

  3. Kunert KS, Peter M, Blum M, Haigis W, Sekundo W, Schütze J, Büehren T (2016) Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography–based biometer versus partial coherence interferometry and optical low-coherence reflectometry. J Cataract Refract Surg 42:76–83

    Article  PubMed  Google Scholar 

  4. Akman A, Asena L, Güngör SG (2016) Evaluation and comparison of the new swept source OCT-based IOLMaster 700 with the IOLMaster 500. Br J Ophthalmol 100:1201–1205

  5. Lehman BM, Berntsen DA, Bailey MD, Zadnik K (2009) Validation of OCT-based crystalline lens thickness measurements in children. Optometry and vision science: official publication of the American Academy of Optometry 86:181

    Article  Google Scholar 

  6. Baikoff G, Lutun E, Ferraz C, Wei J (2004) Static and dynamic analysis of the anterior segment with optical coherence tomography. J Cataract Refract Surg 30:1843–1850

    Article  PubMed  Google Scholar 

  7. Tsorbatzoglou A, Nemeth G, Szell N, Biro Z, Berta A (2007) Anterior segment changes with age and during accommodation measured with partial coherence interferometry. J Cataract Refract Surg 33:1597–1601

    Article  PubMed  Google Scholar 

  8. Leng L, Yuan Y, Chen Q, Shen M, Ma Q, Lin B et al (2014) Biometry of anterior segment of human eye on both horizontal and vertical meridians during accommodation imaged with extended scan depth optical coherence tomography. PLoS One 9:e104775

    Article  PubMed  PubMed Central  Google Scholar 

  9. Richdale K, Bullimore MA, Zadnik K (2008) Lens thickness with age and accommodation by optical coherence tomography. Ophthalmic Physiol Opt 28:441–447

    Article  PubMed  PubMed Central  Google Scholar 

  10. McAlinden C, Khadka J, Pesudovs K (2011) Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmology. Ophthal Physiol Opt 31:330–338

    Article  Google Scholar 

  11. Montés-Micó R, Alió J, Muñoz G, Charman W (2004) Temporal changes in optical quality of air-tear film interface at anterior cornea after blink. Invest Ophthalmol Vis Sci 45:1752–1757

    Article  PubMed  Google Scholar 

  12. Tan WY (1982) Sampling distributions and robustness of t, F and variance-ratio in two samples and ANOVA models with respect to departure from normality. Communication in statistics – Theory and Methods 11:486–511

    Article  Google Scholar 

  13. Box GEP (1965) Some theorems on quadratic forms applied in the study of analysis of variance problems. II: effects of inequality of variance and of correlation between errors in the two-way classification. Ann Math Stat 25:484–498

    Article  Google Scholar 

  14. Charman WN (2008) The eye in focus: accommodation and presbyopia. Clin Exp Optom 91:207–225

    Article  PubMed  Google Scholar 

  15. Neri A, Ruggeri M, Protti A, Leaci R, Gandolfi SA, Macaluso C (2015) Dynamic imaging of accommodation by swept-source anterior segment optical coherence tomography. J Cataract Refract Surg 41:501–510

    Article  PubMed  PubMed Central  Google Scholar 

  16. Piñero DP (2015) Technologies for anatomical and geometric characterization of the corneal structure and anterior segment: a review. Semin Ophthalmol 30:161–170

    Article  PubMed  Google Scholar 

  17. Ramos JL, Li Y, Huang D (2009) Clinical and research applications of anterior segment optical coherence tomography—a review. Clin Experiment Ophthalmol 37:81–89

    Article  PubMed  Google Scholar 

  18. Bizheva K, Haines L, Mason E, MacLellan B, Tan B, Hileeto D, Sorbara L (2016) In vivo imaging and morphometry of the human pre-Descemet's layer and endothelium with ultrahigh-resolution optical coherence tomography. Invest Ophthalmol Vis Sci 57(1):2782–2787

    Article  PubMed  Google Scholar 

  19. Sisó-Fuertes I, Domínguez-Vicent A, del Águila-Carrasco A, Ferrer-Blasco T, Montés-Micó R (2015) Corneal changes with accommodation using dual Scheimpflug photography. J Cataract Refract Surg 41:981–989

    Article  PubMed  Google Scholar 

  20. Bayramlar H, Sadigov F, Yildirim A (2013) Effect of accommodation on corneal topography. Cornea 32:1251–1254

    Article  PubMed  Google Scholar 

  21. Woodman EC, Read SA, Collins MJ (2012) Axial length and choroidal thickness changes accompanying prolonged accommodation in myopes and emmetropes. Vis Res 72(1):34–41

    Article  PubMed  Google Scholar 

  22. Zhong J, Tao A, Xu Z, Jiang H, Shao Y, Zhang H, Liu C, Wang J (2014) Whole eye axial biometry during accommodation using ultra-long scan depth optical coherence tomography. Am J Ophthalmol 157:1064–1069

    Article  PubMed  PubMed Central  Google Scholar 

  23. Ostrin L, Kasthurirangan S, Win-Hall D, Glasser A (2006) Simultaneous measurements of refraction and A-scan biometry during accommodation in humans. Optom Vis Sci 83:657–665

    Article  PubMed  Google Scholar 

  24. Bolz M, Prinz A, Drexler W, Findl O (2007) Linear relationship of refractive and biometric lenticular changes during accommodation in emmetropic and myopic eyes. Br J Ophthalmol 91:360–365

    Article  PubMed  Google Scholar 

  25. Laughton DS, Sheppard AL, Davies LN (2016) A longitudinal study of accommodative changes in biometry during incipient presbyopia. Ophthalmic Physiol Opt 36:33–42

    Article  PubMed  Google Scholar 

  26. Ramasubramanian V, Glasser A (2015) Objective measurement of accommodative biometric changes using ultrasound biomicroscopy. J Cataract Refract Surg 41:511–526

    Article  PubMed  PubMed Central  Google Scholar 

  27. Ramasubramanian V, Glasser A (2015) Prediction of accommodative optical response in prepresbyopic subjects using ultrasound biomicroscopy. J Cataract Refract Surg 41:964–980

    Article  PubMed  PubMed Central  Google Scholar 

  28. Farouk MM, Naito T, Shinomiya K, Eguchi H, Sayed KM, Nagasawa T, Katome T, Mitamura Y (2015) Optical coherence tomography reveals new insights into the accommodation mechanism. J Ophthalmol 2015:510459

    Article  PubMed  PubMed Central  Google Scholar 

  29. Aloy MA, Adsuara JE, Cerdá-Durán P, Obergaulinger M, Esteve-Taboada JJ, Ferrer-Blasco T, Montés-Micó R (2017) PLOS One (accepted for publication)

Download references

Funding

The Spanish Government provided financial support in the form of research grant funding (Explora project, reference SAF2013–49284-EXP). The sponsor had no role in the design or conduct of this research.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to José J. Esteve-Taboada.

Ethics declarations

Conflict of interest

All authors certify that they have no affiliations with or involvement in any organization or entity with any financial interest (such as honoraria; educational grants; participation in speakers’ bureaus; membership, employment, consultancies, stock ownership, or other equity interest; and expert testimony or patent-licensing arrangements), or non-financial interest (such as personal or professional relationships, affiliations, knowledge or beliefs) in the subject matter or materials discussed in this manuscript.

Ethical approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.

Informed consent

Informed consent was obtained from all individual participants included in the study.

Electronic supplementary material

ESM 1

(DOCX 10 kb)

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Esteve-Taboada, J.J., Ferrer-Blasco, T., Aloy, M.A. et al. Ocular anatomic changes for different accommodative demands using swept-source optical coherence tomography: a pilot study. Graefes Arch Clin Exp Ophthalmol 255, 2399–2406 (2017). https://doi.org/10.1007/s00417-017-3801-z

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00417-017-3801-z

Keywords

Navigation