Abstract
A model study introduces the use of statistical signal processing to analyse the signals from an intracavitary probe. A complete derivation is given for the detection of one type of arrhythmogenic substrate, myocardial infarctions (MIs). Both the use of statistical signal processing and the detection of VT substrates, as opposed to activation maps, are unique. A quasi-stationary electromagnetic model with simplified geometry is presented. The model is used to simulate ventricular pacing in the presence of MI. The likelihood ratio is used for detection. A tabulation of the results from this model shows that an intracavitary probe can be used to detect MIs as small as 400 mm2 in 1 mV of noise with a detectability index of 0.495, where 0.5 indicates perfect detection. Sensitivity to noise can be reduced by analysing multiple heart beats. The results are only slightly affected by changing the probe from a cage frame design, which mechanically supports the electrodes on thin spokes, to a balloon design, which supports the electrodes on the surface, of an insulating balloon.
Similar content being viewed by others
References
Aha (1993): ‘1993 Heart and stroke facts statistics’. (American Heart Association, Dallas) pub. no. 55-0502
Blanchard, S. M., Walcott, G. P., Wharton, J. M., andIdeker, R. E. (1994): ‘Why is catheter ablation less sucessful than surgery for treating ventricular tachycardia that results from coronary artery disease?’,PACE,17, pp. 2315–2335
Burdic, W. S. (1984): ‘Underwater acoustic system analysis’, Prentice Hall New Jersey, USA)
Derfus, D. L., andPilkington, T. C. (1992a). ‘Assessing the effect of uncertainty in intracavitary electrode position on endocardial potential estimates’,IEEE Trans. Biomed. Eng.,39, 676–681
Derfus, D. L., Pilkington, T. C., Simpson, E. W., andIdeker, R. E. (1992): ‘A comparison of measured and calculated intracavitary potentials for electrical stimuli in the exposed dog heart’,IEEE Trans Biomed Eng,39, pp. 1192–1206
Fenoglio, J. J., Pham, T. D., Harken, A. H., Horowitz, L. N., Josephson, M. E., andWit, A. L. (1983): ‘Recurrent sustained ventricular tachycardia: structure and ultrastructure of subendocardial regions in which tachycardia originates’,Circulation,68(3), 518–533
Khoury, D. S., andRudy, Y. (1992): ‘A model study of volume conductor effects on endocardial and intracavitary potentials’,Circ Res,71, pp. 511–525
Khoury, D. S., Taccardi, B., Lux, R. L., Ershler, P. R., andRudy, Y. (1995): ‘Reconstruction of endocardial potentials and activation sequences from intracavitary probe measurements’,Circulation,91, 845–863
Kilpatrick, D., andJohnston, P. R. (1994): ‘Origin of the electrocardiogram’,IEEE Eng. Med. Biol. Mag.,13(4), pp. 479–486.
Malkin, R. A., Juhlin, S. P., Penzotti, J. E., Pilkington, T. C., andPlonsey, R. (1993): ‘Optimum detection of myocardial infarctions with an intracavitary probe’.Proceedings of the Comp. in Card. Conference., London, p. 5
Swetss, J. A. (1988): ‘Measuring the accuracy of diagnostic systems’,Science,240, pp. 1285–1293
Taccardi, B., Arisi, G., Macchi, E., Baruffi, S., andSpaggiari, S. (1987): ‘A new intracavitary probe for detecting the site of origin of ectopic ventricular beats during one cardiac cycle’,Circulation,75(1), pp. 272–281
Trees, H. L. V. (1968): ‘Detection, estimation, and modulation theory. (Wiley and Sons).
Author information
Authors and Affiliations
Corresponding author
Additional information
deceased
Rights and permissions
About this article
Cite this article
Malkin, R.A., Penzotti, J.E., Juhlin, S.P. et al. Statistical analysis of signals from an intracavitary probe in a diseased heart. Med. Biol. Eng. Comput. 35, 462–466 (1997). https://doi.org/10.1007/BF02525524
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02525524